Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K170423

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-10-25

    (254 days)

    Product Code
    Regulation Number
    880.6300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K071355

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DentureID Microchip is intended to enable access to secure patient identification and device information when used with complete dentures, partial dentures and other removable oral appliances.

    Device Description

    DentureID Microchip enables access to secure patent identification and device information when used with removable oral appliances. The microchip is permanently embedded into a denture and is a digital link to owner contact information and information about the denture in the event that a repair or replacement is needed. All of the information is controlled on the secure DentureID.com website by the denture owner and dental professional. The DentureID Microchip can be read by any NFC compatible Android Smart Phone (after downloading the DentureID.com App from Google Play) and holding the phone against the microchip. The information on the DentureID.com website may be modified at any time by the patient or dental professional by entering a username and password. The information that is on the website will appear on the smart phone when reading the DentureID Microchip. DentureID Microchips are classified as RFID ISO 14443 which use NFC (near-field-communication). DentureID Microchips are designed to be read by ISO 14443 NFC-compliant smart phones with DentureID App installed. DentureID Microchips are encased in a medical-grade epoxy resin. The size is 1.5mm X 6 mm in diameter. They are inserted into the buccal flange of a denture and completely covered with self-cure repair resin. The DentureID Microchip does not directly contact the patient.

    AI/ML Overview

    This document describes the non-clinical performance testing of the DentureID Microchip.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test DescriptionStandardAcceptance Criteria (Implied by Result)Reported Device Performance
    Biocompatibility
    CytotoxicityISO 10993-5:2009Non-cytotoxic (score '0')The test article scored '0' at 24, 48, and 72 ± 4 hours and is considered non-cytotoxic under the conditions of this test.
    IrritationISO 10993-10:2010Mean test and control scores of extract dermal observations less than 1.0The differences in the mean test and control scores of the extract dermal observations were less than 1.0, indicating that the requirements of the ISO Intracutaneous Reactivity Test have been met by the test article.
    SensitizationISO 10993-10:2010No sensitization response greater than '0' for test article extractsNone of the negative control animals challenged with the control vehicles were observed with a sensitization response greater than '0'. None of the animals challenged with the test article extracts were observed with a sensitization response greater than '0'. The normal saline extract of the test material had a sensitization response of '0' under valid test conditions. The sesame oil extract of the test material had a sensitization response of '0' under valid test conditions. Under the conditions of this protocol, the test article did not elicit a sensitization response.
    Chemical Characterization of Materials (Leachate)Proprietary standards of Nelson Laboratories and ChemTech Ford LaboratoriesPotential toxicity from leaching of harmful chemicals is not significantThe incremental substances found in the study are presented and compared to toxicity standards for the respective materials. As a result of the study, we feel that the potential toxicity from leaching of harmful chemicals from the DentureID RFID tag is not significant.
    Information Security ProceduresNot specifiedPassedPassed
    Software ValidationNot specifiedPassedPassed
    Migration Testing of Implanted TransponderNot specifiedPassedPassed
    Performance Testing of Implanted TransponderNot specifiedPassedPassed
    Electromagnetic Compatibility (EMC)
    Radiated EmissionsEN 55011:2009PassPass
    Electrostatic Discharge ImmunityEN 61000-4-2:2008PassPass
    Radiated Electromagnetic Field ImmunityEN 61000-4-3:2010PassPass
    Magnetic Field ImmunityEN 61000-4-8:2009PassPass
    Data Integrity (during EMC)Not specifiedNo loss or corruption of data, latency, or throughputNo loss or corruption of the data, latency or through-put, which was coordinated with the electromagnetic compatibility (EMC) performance of the microchip, scanner and wireless data link.
    Magnetic Resonance Imaging Compatibility (MRI)
    Magnetic Field Interactions at 3-TeslaASTM F2052-15No additional risk or hazard to a patient in the 3-Tesla MRI environment or less with regard to torque (qualitative measurement of 0)The qualitatively measured torque at 3-Tesla for the DentureID was 0, no torque. As such, this device will not present an additional risk or hazard to a patient in the 3-Tesla MRI environment or less with regard to torque.
    MRI-related heating, 1.5-Tesla and 3-TeslaASTM F2182-11aMaximum temperature rise matches background (e.g., 1.5°C at 1.5T, 1.9°C at 3T)1.5-Tesla demonstrated a maximum of 1.5°C temperature rise and the 3-T system demonstrated a 1.9°C rise. Both of these temperature rises matched the maximum background temperature rise. In conclusion, MRI 1.5 and 3 do not induce significant heating to DentureID Microchip.
    Artifacts at 3-TeslaASTM F2119-07 (Reapproved 2013)Localized signal voids corresponding to device size and shape; maximum artifact size around 10-mm relative to device size.The artifacts that appeared on the MR images were shown as localized signal voids (i.e., signal loss) that corresponded to the size and shape of this device. The gradient echo pulse sequence produced larger artifacts than the T1-weighted, spin echo pulse sequence for the device. The maximum artifact size (i.e., as seen on the gradient echo pulse sequence) extends approximately 10-mm relative to the size and shape of this device.
    Effects of MRI at 1.5-Tesla and 3-Tesla on FunctionNot specified (Internal investigation protocol)100% pre and post-exposure performanceDentureID Microchips performed 100% pre and post exposures.
    Simulated Wear (Denture Cleaning)Not specified (Internal protocol based on 5-year simulated wear with cleaning medium/brush)No impact on device performance; clear self-cure dental acrylic not affectedThe clear self-cure dental acrylic placed over the DentureID Microchip was not affected by the scrubbing action. Therefore, cleaning dentures will not impact the performance of DentureID Microchip.

    Study Information:

    The provided document describes non-clinical performance testing to demonstrate the substantial equivalence of the DentureID Microchip to its predicate devices. It does not detail a clinical study involving human readers or a training set in the context of typical AI device evaluation.

    Here's a breakdown of the available information:

    2. Sample Sizes and Data Provenance for Test Set:

    • Cytotoxicity, Irritation, Sensitization: These tests involve biological samples (e.g., cell cultures, animals). The document doesn't specify the exact number of samples/animals used but refers to the standards (ISO 10993-5, ISO 10993-10) which define such sample sizes. The data provenance is implied to be laboratory testing in facilities adhering to these standards, likely in the US given the FDA submission. These are retrospective tests conducted on device materials.
    • Chemical Characterization: Not specified, but involved proprietary standards of Nelson Laboratories and ChemTech Ford Laboratories. Retrospective laboratory testing.
    • EMC Testing: Not specified for the number of devices tested, but it's laboratory testing of the DentureID Microchips. Retrospective.
    • MRI Compatibility: Not specified for the number of devices tested, but it's laboratory testing of DentureID Microchips to ASTM standards. Retrospective.
    • Simulated Wear: Not specified for the number of DentureID Microchips installed in dentures, but it's a non-clinical performance test over a 5-year simulation. Retrospective.

    3. Number of Experts and their Qualifications for Ground Truth:

    • This information is not applicable in the context of this non-clinical performance testing. The "ground truth" for these tests is established by adhering to recognized international standards (ISO, ASTM, EN) and laboratory protocols, with results interpreted by qualified laboratory personnel (e.g., toxicologists, engineers) who execute these specific tests.

    4. Adjudication Method for the Test Set:

    • Not applicable. These are objective, quantitative and qualitative laboratory tests against defined scientific and engineering standards, not subjective interpretations requiring adjudication by experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was NOT done. This document focuses on the technical and safety performance of the device itself (hardware and its interaction with a smartphone app), not on the diagnostic or interpretative ability of human readers or AI in a clinical setting. Therefore, there is no effect size of how much human readers improve with AI vs. without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance):

    • Yes, in essence, standalone performance testing was done for the core function of the microchip. The various tests (Biocompatibility, EMC, MRI Compatibility, Simulated Wear, Migration Testing, Performance Testing of Implanted Transponder) evaluate the physical and functional characteristics of the DentureID Microchip itself, independent of a human interpreting its output. The "algorithm" here is the RFID chip's ability to respond with its serial number and the smartphone app's ability to retrieve information from a database. "Performance Testing of Implanted Transponder" is a direct measure of this standalone function.

    7. Type of Ground Truth Used:

    • The ground truth for this non-clinical testing is based on:
      • Standardized test methods and protocols: Defined by international standards (ISO, ASTM, EN) and proprietary laboratory standards.
      • Objective measurements and observations: E.g., cytotoxicity scores, temperature rises, torque measurements, pass/fail criteria for EMC.
      • Chemical analysis: For leachate testing.

    8. Sample Size for the Training Set:

    • Not applicable. This device (DentureID Microchip) is a passive RFID transponder and its associated database/app. It does not employ machine learning or AI models that require a "training set" in the conventional sense for image analysis or diagnostic tasks. Its function is to communicate a serial number, which then links to pre-entered data on a website.

    9. How Ground Truth for Training Set Was Established:

    • Not applicable, as there is no training set for an AI/ML model for this device. The information stored on the DentureID.com website, which is retrieved by the microchip's serial number, is entered by a dental lab or dentist. This user-provided information serves as the "ground truth" for the data it's designed to return, but it's not a training set for an algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K100551

    Validate with FDA (Live)

    Date Cleared
    2010-08-12

    (167 days)

    Product Code
    Regulation Number
    880.2740
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K071355, K062642, K060076

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ORLocate™ system is indicated for use in recording and counting the number of RFID-tagged surgical sponges, laparotomy sponges, towels and other tagged items used during surgical procedures in which counting is required. In addition, the product is indicated for providing a non-invasive means of detecting retained RFID-tagged surgical sponges, towels and other tagged items within a surgical site, as an adjunctive detection method to current surgical counting systems and methods.

    Device Description

    Haldor ORLocate™ system is an RFID system providing a solution that enables the enumeration of sponges and surgical manual instruments, utilizing passive tags for keeping track of the items during surgery and to identify counting problems. In addition, the system provides a non-invasive means of locating retained surgical items within a surgical site. The submission consists of the ORLocate™ system which includes: cart and antennas. Additionaly the submission includes accessories which are: associated single use surgical sponges, gauzes, pads and surgical towels each fitted with a uniquely coded RFID tag and uniquely coded RFID tag used for surgical instruments. The RF frequency the system uses is 13.56 MHz according to ISO 15693. The system supplies also a semi-automatic application to help in counting untagged items, the count information is first entered manually and the calculations are automatic.

    AI/ML Overview

    The ORLocate™ System is an RFID-based system for counting and detecting surgical items. The provided document, a 510(k) Summary, details the device's technical characteristics, intended use, and non-clinical performance data to demonstrate substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" with numerical thresholds for performance. However, based on the non-clinical performance data section, the criteria can be inferred from the tests performed and the conclusion that the device functions as intended and is as safe and effective as predicate devices.

    Acceptance Criteria (Inferred)Reported Device Performance
    Biocompatibility of tagged itemsDemonstrated
    Permanent attachment of tags to sponges and instrumentsDemonstrated
    Software functions as intendedValidated, properly counting sponges in body fluids
    Safety equivalent to predicate devicesTest results demonstrate this
    Effectiveness equivalent to predicate devicesTest results demonstrate this
    Proper functioning in body fluidsSoftware properly counted sponges in body fluids
    Counting accuracyTesting performed, deemed "as safe and effective"
    System interference with OR devicesTesting performed
    ORLocate sponge X-ray detectionTesting performed
    ORLocate Tag pull testTesting performed
    Electromagnetic compatibility (IEC 60601-1-2:2007)Testing performed
    Electrical safety (IEC 60601-1:1988 + A1:1991 + A2:1995 and EN 60601-1:1990 + A1:1993 + A2:1995 + A3:1996)Testing performed

    2. Sample Size Used for the Test Set and Data Provenance

    The document states "Non-clinical testing included demonstrating performance of system and tagged items in laboratory tests." However, it does not specify the sample sizes used for any of the tests (e.g., how many sponges were tested for counting accuracy, how many instruments for tag pull test).

    The data provenance is described as "laboratory tests," implying controlled settings rather than real-world clinical data. The document does not mention the country of origin of the data explicitly, but the 510(k) owner is based in Israel, suggesting the testing likely occurred there or in collaboration with international labs. The study is retrospective in the sense that the testing was performed and then reported for the 510(k submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide information on the number of experts used or their qualifications for establishing ground truth for the test set. Given the nature of the non-clinical tests (e.g., biocompatibility, tag attachment, software counting accuracy), ground truth would likely be established through objective measurements and validated procedures rather than solely expert consensus.

    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method for the test set. Given that the non-clinical tests are largely objective performance evaluations (e.g., measuring count accuracy, pull force, EMC compliance), an adjudication method in the context of expert review is unlikely to be relevant.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document focuses on non-clinical performance data to establish substantial equivalence, not on human reader performance with or without AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    Yes, the non-clinical tests described are essentially standalone performance evaluations of the algorithm and device. The "counting accuracy test" and validation that the "software functioned as intended under simulated use, properly counting sponges in body fluids" demonstrate the algorithm's performance without direct human-in-the-loop assistance for the core counting mechanism. The system is designed to perform automatic counts and provide alerts if items are missing, which is a standalone algorithmic function.

    7. The Type of Ground Truth Used

    The ground truth used for the non-clinical tests would have been established through:

    • Objective measurement/validation: For counting accuracy, the actual number of sponges present would be the ground truth. For tag pull tests, the measured force would be compared against a standard.
    • Established standards: Compliance tests like IEC 60601-1-2:2007 and IEC 60601-1:1988 + A1:1991 + A2:1995 refer to external ground truths established by international standards bodies.
    • Simulated environment: For the software, simulated body fluids were used to test performance, implying a controlled and known environment against which the device's output was compared.

    8. The Sample Size for the Training Set

    The document does not mention a training set sample size. The system described is an RFID detection and counting system, which typically relies on pre-programmed logic for tag identification and counting, rather than a machine learning model that requires a "training set" in the conventional sense. If there are any adaptive or learning components, they are not detailed in this summary.

    9. How the Ground Truth for the Training Set Was Established

    Since a training set is not mentioned for machine learning purposes, the method for establishing its ground truth is not applicable/not provided. The system's operational parameters (e.g., RFID tag protocols, counting logic) would be established through engineering design and validation, not model training.

    Ask a Question

    Ask a specific question about this device

    K Number
    K073180

    Validate with FDA (Live)

    Date Cleared
    2007-11-19

    (6 days)

    Product Code
    Regulation Number
    880.2740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K071355, K062642

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ClearCount Medical Solutions SmartSponge™ PLUS System is indicated for use in counting and recording the number of RFID-tagged surgical sponges, laparatomy sponges and towels used during surgical procedures, as well as for providing a non-invasive means of locating retained RFID-tagged surgical sponges, towels, and other tagged items within a surgical site.

    Device Description

    The SmartSponge™ PLUS System includes surgical sponges, laparatomy pads and surgical towels, each of which contains a unique radio frequency identification (RFID) tag permanently attached to the gauze or fabric. The tags allow the sponges and towels to be individually recognized by an RFID reader.

    The SmartBucket is a specially designed cart containing a microcontroller unit with specialized software designed for mobile data collection. Integrated RFID technology allows capture of the information coded on the unique RFID tag on the sponges, pads and towels. The microcontroller unit counts the initial number of sponges introduced into a surgical case, and using the custom software program, reports the total sponges discarded at the end of the procedure, and compares that number to the original. By providing a count of the items entered into surgery, and a count of those discarded and removed permanently from the surgical field, personnel can be alerted to sponges that may still remain in the surgical field prior to closing the patient.

    A Detection Wand is an additional antenna that is tethered by a cable to the SmartBucket. It is powered and controlled by the SmartBucket. The antenna functions as an additional RFID antenna to the system, functioning in an identical manner to the internal SmartBucket antennas. By using a keypad the user may select activate the Detection Wand antenna. When in Detection Wand mode, the system uses the Wand antenna to recognize RFID-tagged items that may be inside the surgical site.

    A Detection Mat is a disposable or reusable element with multiple RFID tags embedded inside, along with several passive printed circuit traces. Like the RFID-tagged sponges, the Detection Mat tags contain unique identifying numbers and are distinguishable by the system software. The Detection Mat is placed on the operating room table before the patient is brought into the room and is covered by the standard sheets or drapes used in surgery, thus not making contact with the patient. The RFID tags in the Matt provide feedback to the user that the Detection Wand is being held close enough to the patient to ensure proper reading. The tags in the Detection Mat also ensure that the Detection Wand scan has covered the appropriate areas of the patient. The passive circuit traces help to enhance the readability of the RFID tags in the Detection Mat.

    AI/ML Overview

    Here's an analysis of the provided text regarding the ClearCount Medical SmartSponge™ PLUS System, focusing on acceptance criteria and supporting study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Counting Functionality: Accurately count and record the initial number of RFID-tagged surgical items and the number discarded post-procedure.The SmartBucket (part of the system) "counts the initial number of sponges introduced into a surgical case, and using the custom software program, reports the total sponges discarded at the end of the procedure, and compares that number to the original." This function is implicitly stated to work as intended. "The customized software program uses the scanned information to count the number of items used at the beginning of a surgical procedure, and then again before surgical closure." No specific numerical accuracy rates are provided for the counting function in this document.
    Detection/Localization Functionality: Non-invasive means of locating retained RFID-tagged surgical items within a surgical site."Non-Clinical testing included simulated use in patient models that represented worst case biological situations... and in all cases the ClearCount SmartSponge™ PLUS System performed as intended." "The validated software functioned as intended under simulated use, properly locating all tags."
    RFID Tag Readability: Read tags through blood, bodily fluids, and tissue."The scanner can read the tag through blood and other bodily fluids and tissue." (This is a design claim rather than a measured performance metric from the study specifically.)
    Biocompatibility of Transponder Tags: Tags are safe for contact within the surgical environment."Biocompatibility of the transponder tag was illustrated and is comparable to the commercially available predicates."
    Electrical Safety Standards: Compliance with IEC 60601-1."The system has also been designed to meet the following electrical safety standards and electromagnetic compatibility standards: IEC 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Safety"
    Electromagnetic Compatibility Standards: Compliance with IEC 60601-1-2."The system has also been designed to meet the following electrical safety standards and electromagnetic compatibility standards: IEC 60601-1-2 (Second Edition, 2001) Medical Electrical Equipment - Part 1: General Requirements for Safety: Electromagnetic Compatibility - Requirements and Tests"

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "simulated use in patient models that represented worst case biological situations." However, no specific sample size (number of simulated cases, number of sponges, or number of tests) for the test set is provided.

    The data provenance is from non-clinical testing, using simulated patient models. The document does not specify the country of origin, but given the FDA submission, it's presumably conducted under U.S. regulatory standards or by a manufacturer seeking to market in the U.S. It explicitly states "Non-Clinical testing," confirming it's not a prospective or retrospective study involving actual patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The "ground truth" for the simulated use in patient models would likely have been established by the study designers or engineers setting up the scenarios of "worst case biological situations" and knowing the expected location and quantity of tagged items. There is no mention of independent experts establishing this ground truth.

    4. Adjudication Method for the Test Set

    This information is not provided. Without details on who assessed the device's performance in the "simulated use," an adjudication method cannot be determined.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. The device's primary function is to automate sponge counting and aid in detection, not to assist human readers in interpreting medical images in the traditional sense of an MRMC study. The "Detection Wand" is a tool for locating tagged items, not for improving human interpretation of visual data.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a form of standalone performance was implicitly evaluated for the detection component. The statement "The validated software functioned as intended under simulated use, properly locating all tags" suggests the algorithm's ability to locate tags independently within the simulated environment. While a human operates the Detection Wand, the core detection by the RFID system and its interpretation by the software is a standalone algorithm function. The counting function is also purely algorithmic.

    7. The Type of Ground Truth Used

    The ground truth used was based on pre-defined scenarios within simulated patient models where the exact quantity and location of RFID-tagged items would be known to the experimenters. This is an engineered or experimental ground truth, not derived from expert consensus, pathology, or outcomes data from real patients.

    8. The Sample Size for the Training Set

    The document does not provide information on the sample size for a training set. The device appears to rely on established RFID technology and programmed logic, rather than a machine learning model that would typically require a training set in the conventional sense. If there was any "training" (e.g., for system calibration), its details are not mentioned.

    9. How the Ground Truth for the Training Set was Established

    Since no training set is mentioned or implied for a typical machine learning model, this information is not applicable / not provided. The functionality seems to be based on direct sensing and programming, not a learned model from data.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1