    Reference Devices :

    K071250, K101505

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CA, an auxiliary lens adapter, when used in conjunction with RTVue, is indicated for in vivo imaging and measurement of the cornea and the other ocular structures of the anterior segment of the eve, including pachymetry and corneal power.

    Device Description

    The already cleared (K071250) RTVue CA Module (CAM) is an instrument based on Fourier-Domain Optical Coherence Tomography (OCT) for in vivo imaging and measurement of the cornea and other ocular structures of the anterior segment of the eye with US FDA 510(k) clearance (K071250) in 2007. The RTVue CAM has been used in clinical practice for imaging the cornea, measuring corneal thickness, and visualizing the anterior segment angle. The RTVue CAM device uses the same Optical Coherence Tomography (OCT) technology that was previously cleared by FDA (K101505). The CAM adapter gives the user an option to use the RTVue device as previously approved for retina scans, or to use it for cornea and anterior eye scans. Aside from the CAM auxiliary attachment, the RTVue is virtually unchanged for the CAM use except the CAM software module provides for menu selections in the graphical user interface, which are selected by the operator to label corresponding corneal landmarks instead of those of the retina. The system scans a beam into patient's eye and uses a low coherence interferometer to measure the reflectivity of the ocular tissue. The cross-sectional ocular tissue structure is composed of sequence of A-scans. The RTVue has a traditional patient and instrumentinterface like most ophthalmic devices. The device is mounted on a motorized patient table. The patient will rest their head on the forehead and chin rest. The operator uses joystick to align the device to patient's eye. The computer has a graphic user interface for acquiring, and displaying the acquired image. The RTVue image acquisition speed and image resolution remain the same when used in conjunction with CAM.

    AI/ML Overview

    The provided document describes the RTVue CAM with Corneal Power Measurement device and its performance data to support substantial equivalence. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance CriteriaReported Device Performance
    Calibration AccuracyAccuracy of ±0.25D for corneal power measurement.Implicitly met by the bench test results showing error < 0.25D.
    Acceptance TestSystem must pass an acceptance test after calibration, using a spherical target with a different radius of curvature from the calibration target.Stated that the system must pass this test before the feature is enabled. No specific quantitative performance data provided for this test in the document.
    Weekly Validation TestThe acceptable difference between the validation test result (using the calibration target) and the initial calibration value is set to ±0.25D. If exceeded, the software prevents corneal power measurement.Stated that this test is required, and the software enforces the ±0.25D limit. No specific quantitative performance data provided for this test in the document.
    Bench Test Accuracy (Post-Calibration)Difference between the mean measured curvature and the ground truth for each ceramic ball should be less than 0.25D.Ceramic Ball 1: Error = 0.0655D < 0.25D Ceramic Ball 2: Error = 0.1187D < 0.25D Ceramic Ball 3: Error = 0.0471D < 0.25D Ceramic Ball 4: Error = 0.1308D < 0.25D

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Bench Test: Four ceramic balls of different known diameters were used. For each ball, 3 measurements (scans) were performed.
    • Data Provenance: The data is from a "bench test" conducted by Optovue. The country of origin is not explicitly stated, but the manufacturer is based in Fremont, CA, USA. The data is retrospective in the context of the 510(k) submission, as it was collected prior to market approval.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This section describes a bench test using ceramic balls with known physical properties. Therefore, no human experts were involved in establishing the ground truth for this test set. The ground truth (true power and diameter of the ceramic balls) was determined by the physical characteristics of the manufactured ceramic balls.

    4. Adjudication Method for the Test Set

    • No adjudication method was used, as the test set consisted of inanimate ceramic balls with known ground truth values. The measurements were compared directly against these known values.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    • No MRMC comparative effectiveness study involving human readers and AI assistance was conducted or mentioned in the provided document. The device is a measurement tool, not an AI diagnostic aid for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    • Yes, a standalone performance test (bench test) of the RTVue CAM with Corneal Power Measurement was performed to verify its accuracy in measuring corneal power. This test involved the device measuring known ceramic balls.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • For the bench test, the ground truth was physical measurements/specifications of the ceramic balls (known diameter, from which a "truth value" for power was calculated using a specific refractive index).

    8. The Sample Size for the Training Set

    • The document does not provide information about a specific "training set" or its sample size. The device is a measurement instrument, and the upgrade is a software change for calculating corneal power from acquired images. It mentions a "calibration" process using a single ceramic ball with a prespecified radius of curvature. This calibration can be considered part of setting up the system, rather than a typical machine learning "training set."

    9. How the Ground Truth for the Training Set was Established

    • As noted above, no traditional "training set" for a machine learning algorithm is described. The "ground truth" for the device's calibration involves using a single ceramic ball with a known curvature (diameter 15.875mm, equivalent power 47.37D using index 1.376). This known physical property of the calibration sphere serves as the ground truth for the calibration process.
    Ask a Question

    Ask a specific question about this device

    K Number
    K101505

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-09-15

    (106 days)

    Product Code
    Regulation Number
    886.1570
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K071250,K033123

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RTVue with Normative Database is an optical coherence tomography system indicated for the in vivo imaging and measurement of the retina, retinal nerve fiber layer, and optic disc as an aid in the diagnosis and management of retinal disease. The RTVue with Normative Database is also a quantitative tool for the comparison of retinal nerve fiber layer, and optic disc measurements in the human eye to a database of known normal subjects. It is intended for use as a diagnostic device to aid in the detection and management of ocular diseases.

    Device Description

    The RTVue is a computer controlled ophthalmic imaging and measurement system that employs optical coherence tomography to image and measure the posterior segment of the eye. The device is currently cleared for in vivo imaging and measurement of the various retinal layers (K062552). Imaging and measurements include but are not limited to the internal limiting membrane (ILM), the retinal nerve fiber layer (RNFL), the ganglion cell complex (GCC), the retinal pigment epithelium (RPE), the outer retinal thickness, the total retinal thickness and optic disk structures including the cup and neuroretinal rim as an aid in the diagnosis and management of retinal disease. The measurements for the ILM and RPE are height measurements relative to the RPE reference plane. The RNFL. GCC. the outer retinal thickness and total retinal thickness measurements where RNFL is the thickness of the RNFL layer, the GCC is the thickness from the ILM to the inner plexiform layer (IPL), the outer retinal thickness is the thickness from the IPL to the RPE, and total retinal thickness is the thickness from the ILM to the RPE. The current submission is for a software modification to add a normative database. With the additional normative database (NDB), the RTVue can compare the measured data from the GCC, the RNFL, the full retinal thickness, optic disc cup and optic disc rim measurements to the normative database. The RTVue with Normative Database provides a statistical reference of the GCC, the RNFL, the full retinal thickness, optic disc cup and optic disc rim measurements to a database of known normal subjects. The RTVue with Normative Database will provide the analysis information to be used as a clinical reference to aid in the diagnosis and management of ocular diseases.

    AI/ML Overview

    1. Acceptance Criteria and Reported Device Performance:

    The document describes a precision study to evaluate the RTVue with Normative Database. The acceptance criteria are implicitly demonstrated by presenting the repeatability and reproducibility standard deviations for various measurements. The "Performance Data" section outlines these results.
    The acceptance criteria are not explicitly stated, however based on the precision study results for repeatability and reproducibility, the device performance is within acceptable limits for a medical device.

    Table of Performance Results

    Scan Type & MeasurementPatient GroupRepeatability SD (Min, Max)Reproducibility SD (Min, Max)
    EMM5 Scan Results
    Full Retina FoveaNormal Patients2.57 (1.66, 2.57)17.72 (1.66, 17.72)
    Retina Patients2.96 (1.83, 3.15)13.78 (1.83, 13.78)
    Glaucoma Patients3.01 (1.95, 3.01)3.01 (2.05, 5.80)
    Full Retina PeripheralNormal Patients1.96 (1.66, 2.57)2.01 (1.66, 17.72)
    Retina Patients2.69 (1.83, 3.15)2.69 (1.83, 13.78)
    Glaucoma Patients2.35 (1.95, 3.01)2.66 (2.05, 5.80)
    ONH Scan Results
    Average RNFLNormal Patients1.76 (1.76, 6.18)2.11 (2.11, 9.78)
    Glaucoma Patients1.56 (1.56, 5.86)7.87 (3.92, 16.33)
    RNFL Avg (2 Hemis, 4 Quads, 8 Sectors)Normal Patients3.92 (1.76, 6.18)5.40 (2.11, 9.78)
    Glaucoma Patients4.48 (1.56, 5.86)7.84 (3.92, 16.33)
    Cup AreaNormal Patients0.02 (0.02, 0.03)0.02 (0.02, 0.03)
    Glaucoma Patients0.07 (0.07, 0.21)0.07 (0.07, 0.21)
    Rim AreaNormal Patients0.03 (0.02, 0.03)0.03 (0.02, 0.03)
    Glaucoma Patients0.18 (0.07, 0.21)0.18 (0.07, 0.21)
    Vertical CD RatioNormal Patients0.01 (0.01, 0.04)0.01 (0.01, 0.04)
    Glaucoma Patients0.05 (0.05, 0.08)0.05 (0.05, 0.08)
    GCC Scan Results
    Inner Retina AverageNormal Patients1.86 (1.86, 2.07)7.56 (7.32, 7.92)
    Glaucoma Patients1.18 (1.18, 1.59)2.55 (1.59, 3.36)
    GCC FLVNormal Patients0.37 (0.37, 0.37)0.46 (0.46, 0.46)
    Glaucoma Patients0.88 (0.88, 0.88)1.04 (1.04, 1.04)
    GCC GLVNormal Patients0.83 (0.83, 0.83)2.44 (2.44, 2.44)
    Glaucoma Patients1.03 (1.03, 1.03)1.71 (1.71, 1.71)

    2. Sample Size for Test Set and Data Provenance:

    • Sample Size for Test Set: A total of 36 subjects were used in the precision study. This included 4 normal subjects, 4 retina patients, and 4 glaucoma patients for each of the 3 RTVue devices used (12 subjects per device).
    • Data Provenance: The data was collected at two clinical sites. The document implies a prospective nature as subjects were "enrolled in the study," and "data was carefully reviewed for completeness and quality." The country of origin is not explicitly stated but can be inferred as the United States given the FDA submission.

    3. Number of Experts and Qualifications for Ground Truth:

    • Number of Experts: The ground truth for subject classification (normal, glaucoma, retina pathology) was established by the "Principal Investigator (PI) at each clinical site." The exact number of PIs is not specified but there were two clinical sites.
    • Qualifications of Experts: The PIs diagnosed glaucoma and various retina pathologies. Their specific qualifications (e.g., "Radiologist with 10 years of experience") are not detailed beyond being a "Principal Investigator" responsible for diagnosis.

    4. Adjudication Method:

    The document mentions that all data was "carefully reviewed for completeness and quality in two levels," including a review of the CRF by the PI against inclusion/exclusion criteria, and a review of individual scans for image quality. There is no explicit mention of an adjudication method (like 2+1 or 3+1 consensus) for establishing the diagnostic ground truth for patient classification. The PI at each site made the diagnosis.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study is described. The study focuses on the precision (repeatability and reproducibility) of the RTVue device itself, not on the improvement of human readers with AI assistance. The additional normative database (NDB) functions as a reference tool for comparison rather than a direct AI-assisted diagnostic aid for human readers in the context of this study.

    6. Standalone Performance:

    Yes, a standalone performance study was done. The precision study evaluates the RTVue device (algorithm only, in the sense of its measurement capabilities and normative database comparison) without human interpretation as the primary outcome. The results presented are the repeatability and reproducibility of the device's measurements.

    7. Type of Ground Truth Used:

    The ground truth used for classifying subjects into normal, glaucoma, or retina pathology groups was expert consensus (diagnosis by the Principal Investigator).

    8. Sample Size for Training Set:

    The "Normative Databases" section states that normative data was collected from 480 individuals (over 640 eyes) for the database. This acts as the "training set" for the normative database comparison feature of the device.

    9. How Ground Truth for Training Set was Established:

    The ground truth for the normative database was established by enrolling "known healthy eyes" from these 480 individuals. This implies that subjects were deemed healthy based on a medical evaluation, likely conducted by qualified medical professionals at the 11 clinical sites following an IRB approved protocol. The criteria for "healthy eyes" would have been part of this protocol, but are not explicitly detailed in the document.