Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K112760

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-11-22

    (61 days)

    Product Code
    Regulation Number
    880.6860
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    K062297, K083097, K073244, K09051, K091174, K070769, K070765, K071087, K090371

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Amsco® V-PRO™ maX Low Temperature Sterilization System, with VAPROXTM HC Sterilant, is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in healthcare facilities. The three pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for processing medical devices sensitive to heat and moisture.

    The Amsco V-PRO maX Low Temperature Sterilization System's Lumen Cycle, cleared under K062297, can sterilize:

    • Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • Medical devices, including rigid endoscopes, with a single stainless steel lumen with:
      • an inside diameter of 1 mm or larger and a length of 125 mm or shorter
      • an insider diameter of 2 mm or larger and a length of 250 mm or shorter
      • an inside diameter of 3 mm or larger and a length of 400 mm or shorter
        The validation testing for all lumen sizes was conducted using a maximum of twenty (20) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs.

    The Amsco V-PRO maX Low Temperature Sterilization System's Non Lumen Cycle, cleared under K083097, can sterilize:
    Non-lumened instruments including non-lumened rigid endoscopes and non-lumened instruments with stainless steel diffusion-restricted areas such as the hinged portion of forceps or scissors.
    The validation studies were conducted using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs.

    The Amsco V-PRO maX Low Temperature Sterilization System's Flexible Cycle, the subject of this submission, can sterilize single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of two load configurations:

    1. Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load.
      The flexible endoscopes may contain either:
    • a single lumen with an inside diameter of 1 mm or larger and a length of 1050 mm or shorter
    • or two lumens with:
      • one lumen having an inside diameter of 1 mm or larger and a length of 998 mm or shorter
      • and the other lumen having an inside diameter of 1 mm or larger and a length of 850 mm or shorter
        The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope).
    1. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments including instruments with diffusion-restricted areas such as the hinged portion of forceps or scissors.
      The flexible endoscope can contain either:
    • a single lumen with an inside diameter of 1 mm or larger and a length of 1050 mm or shorter
    • or two lumens with:
      • one lumen having an inside diameter of 1 mm or larger and a length of 998 mm or shorter
      • and the other lumen having an inside diameter of 1 mm or larger and a length of 850 mm or shorter
        The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (if not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total load weight of 24.0 lbs.
    Device Description

    The Amsco V-PRO maX Low Temperature Sterilization System is a new sterilizer model to be added to the Amsco V-PRO family of sterilizers. The V-PRO product line currently consists of the Amsco V-PRO 1 and Amsco V-PRO 1 Plus Sterilizers.

    The V-PRO maX Sterilizer has three pre-programmed cycles: the Lumen Cycle (K062297), the Non Lumen Cycle (K083097) and the Flexible Cycle (K102330). The V-PRO maX Low Temperature Sterilization System is intended for terminal sterilization of cleaned, rinsed, dried and packaged reusable surgical instruments used in healthcare facilities.

    The V-PRO maX Sterilizer uses VAPROX HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). Its three preprogrammed cycles all utilize a conditioning phase, a sterilization phase and an aeration phase. The packaged sterilized devices are ready for use at the completion of the cycle; no cool down or aeration period is required.

    The Flexible Cycle has been designed to sterilize surgical flexible endoscopes (such as those used in ENT, Urology and Surgical care) and bronchoscopes. The flexible endoscopes can be single or dual channeled and they may be processed in a flexible endoscope-only load or in combination with non-lumened medical instruments. The cycle is approximately 35 minutes long.

    The Verify V24 SCBI (K073244 and K09051), Verify VH2O2 Process Indicator (K091174), V-PRO Sterilization Trays (K070769), and Vis-U-All Tyvek pouches (K070765, K071087, and K090371) have been validated for use in the V-PRO maX Sterilizer.

    AI/ML Overview

    The provided 510(k) summary (K112760) describes a modification to the Amsco V-PRO maX Low Temperature Sterilization System, involving minor software changes to enhance usability. The study submitted for this modification is a Software Validation.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Device ModificationTestingAcceptance CriteriaResults
    Software ModificationsSoftware ValidationSoftware shall be appropriately verified and validated.PASS

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the software validation test set (e.g., number of test cases run, or specific inputs). It only indicates that "Software Validation" was performed.

    Regarding data provenance, the document does not specify country of origin for the validation data. The nature of the study (software validation) suggests it would be conducted in a controlled environment, likely by the manufacturer (STERIS Corporation, based in Mentor, Ohio, USA). The study is retrospective in the sense that the software was already developed and then validated.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not provide details on the number of experts, their qualifications, or their role in establishing ground truth for the software validation. For a software validation, ground truth would typically be defined by the software requirements and specifications, rather than expert interpretation of data.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the software validation. For software validation, discrepancies (bugs) are typically identified and fixed, not adjudicated by independent parties.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically used for diagnostic devices involving human interpretation of images or other data. The Amsco V-PRO maX is a sterilizer, and its effectiveness is determined by its ability to reliably sterilize medical devices, not by human interpretation of its output in a clinical setting. Therefore, an MRMC study with human readers and AI assistance is not applicable to this device.

    6. Standalone (Algorithm Only Without Human-in-the Loop Performance) Study

    Yes, a standalone study was performed. The "Software Validation" is an assessment of the algorithm (software)'s performance against its specifications and intended function without human intervention in its operation other than initiating the process and evaluating the outcome of the software's execution.

    7. Type of Ground Truth Used

    For the "Software Validation," the ground truth would be defined by the software requirements and specifications. The software's performance is validated against these pre-defined functional and non-functional requirements to ensure it operates as intended.

    8. Sample Size for the Training Set

    The document does not mention a "training set" because this is a software validation for a sterilizer. The software for a sterilizer is typically developed based on engineering principles and verified through testing against specifications, rather than being "trained" on a dataset like a machine learning algorithm.

    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of a training set, the establishment of ground truth for a training set is not applicable to this submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1