Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K113272

    Validate with FDA (Live)

    Date Cleared
    2012-02-29

    (117 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    K102775, K070513, K100011

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INOmax DS delivery system delivers INOmax® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

    The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.

    The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

    The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

    The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

    Device Description

    The INOmax DSIR uses a "dual-channel" design to ensure the safe delivery of INOmax. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOmax delivery independent of monitoring but also allows the monitoring system to shutdown INOmax delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm.

    AI/ML Overview

    The provided document describes the INOmax DSIR, a nitric oxide delivery system, and its compatibility with additional respiratory care devices. The submission focuses on non-clinical testing to demonstrate substantial equivalence, rather than a study involving clinical outcomes or diagnostic accuracy. Therefore, information related to observer performance studies (e.g., MRMC studies, standalone performance), ground truth establishment for diagnostic tasks, expert qualifications, and adjudication methods is not applicable to this submission.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Requirements)Reported Device Performance (Testing Conclusion)
    O2 dilutionPerformed within published specifications when used with each of the selected respiratory care devices.
    Effect on respiratory care devicePerformed within published specifications when used with each of the selected respiratory care devices.
    INOmax DSIR delivery accuracyPerformed within published specifications when used with each of the selected respiratory care devices.
    NO2 generationPerformed within published specifications when used with each of the selected respiratory care devices.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes non-clinical testing involving the INOmax DSIR and three specific respiratory care devices:

    The testing involved using six INOmax DSIR settings: [0 (baseline), 1, 5, 20, 40] ppm (the sixth value is cut off but implied to be another concentration in ppm). The document states, "The three respiratory care devices were set up and calibrated according to the manufacturer's recommendations, and tested using the settings established for each respiratory care device test." This suggests a systematic testing approach across different settings for each device. However, a specific numerical "sample size" in terms of number of patient cases or repeated measurements for statistical analysis is not detailed in the provided text.

    The data provenance is prospective non-clinical testing conducted by INO Therapeutics/Ikaria, likely at their facilities, to evaluate compatibility and performance. There is no indication of country of origin of patient data as no patient data was used.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This was a non-clinical device compatibility and performance study, not a diagnostic accuracy study requiring expert-established ground truth. The "ground truth" was the expected performance according to published specifications and manufacturer recommendations for the devices.

    4. Adjudication Method for the Test Set

    Not applicable. This was a non-clinical device compatibility and performance study, not a diagnostic accuracy study requiring adjudication of expert interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. No MRMC study was conducted, as this submission concerns the hardware and software performance of a nitric oxide delivery system and its compatibility with other respiratory devices, not a diagnostic algorithm involving human readers or AI.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. The INOmax DSIR is a medical device for delivering and monitoring nitric oxide, not a standalone diagnostic algorithm. Its performance was tested as a standalone system and in conjunction with other respiratory care devices.

    7. The Type of Ground Truth Used

    The "ground truth" for this non-clinical testing was based on:

    • Manufacturer's specifications: The INOmax DSIR was expected to "perform within published specifications."
    • Manufacturer's recommendations: The respiratory care devices were set up and calibrated "according to the manufacturer's recommendations."
    • Expected compatibility: The overall aim was to conclude that the INOmax DSIR and the three respiratory care devices are compatible.

    8. The Sample Size for the Training Set

    Not applicable. This device is a hardware and software system for medical gas delivery and monitoring, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there was no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1