Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K090629

    Validate with FDA (Live)

    Date Cleared
    2009-06-04

    (87 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K062829

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Easy Step / DIACHEX PRO Blood Glucose Monitoring System is intended for use in the quantitatively measurement of glucose (sugar) in fresh capillary whole blood from the fingertip and the alternative sites: the palm and the forearm. The Easy Step / DIACHEX* PRO Blood Glucose Test Strips are for testing outside the body (in vitro . diagnostic use). The Easy Step / DIACHEX* PRO Blood Glucose Monitoring System is intended for use at home (over the counter [OTC]) by persons with diabetes, or in clinical setting by healthcare professionals as an aid in monitoring the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus. The alternative site testing in the systems can be used only during steady-state blood glucose conditions. It is not intended for neonatal testing.

    Device Description

    The Easy Step / DIACHEX* PRO Blood Glucose Monitoring System designed by Tyson Bioresearch Inc., an amperometric biosensor, is adopted for its ease of use, its ability to process accurate results utilizing only a small volume of blood, and its quick response time. Easy Step / DIACHEX* PRO provide a convenient and safe monitoring system for diabetes health care professionals, hospitals and most importantly, people with diabetes.

    The Easy Step / DIACHEX* PRO Blood Glucose Monitoring System is intended for use in the quantitatively measurement of glucose (sugar) in fresh capillary whole blood from the fingertip and the alternative sites: the palm and the forearm. The Easy Step / DIACHEX* PRO Blood Glucose Test Strips are for testing outside the body (in vitro diagnostic use). When the edge of the test strip is touched to a drop of blood, the test strip draws the blood into the sample chamber and the glucose reading is displayed on the meter after 5 seconds. The test measures glucose from 20 mg/dL (1.1mmol/L) to 600 mg/dL (33.3 mmol/L). The test strip is calibrated to display the equivalent of plasma glucose values to allow the comparison of results with laboratory methods.

    AI/ML Overview

    This submission pertains to the Easy Step / DIACHEX* PRO Blood Glucose Monitoring System.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the device are based on the ISO 15197 standard for clinical accuracy.

    Acceptance Criteria (ISO 15197)Reported Device Performance (Easy Step)Reported Device Performance (DIACHEX* PRO)Reported Device Performance (DIACHEX* INFINITY - Predicate)
    95% of individual glucose results falling within ±15 mg/dL at glucose concentration for samples <75 mg/dL
    95% of individual glucose results falling within ±20% at glucose concentrations ≥75 mg/dL (when compared to a laboratory reference)Finger: 98.4% (127/129) Palm: 100.0% (113/113) Forearm: 98.2% (111/113)Finger: 99.2% (128/129) Palm: 98.2% (111/113) Forearm: 96.5% (109/113)Finger: 100% (129/129) Palm: 98.2% (111/113) Forearm: 97.3% (110/113)

    Additionally, the following
    performance characteristics were evaluated:

    • Within-Day Precision:
      • Easy Step: CV% ranged from 2.08% to 5.80% across 5 glucose levels. Bias% ranged from -3.1% to 2.8%.
      • DIACHEX PRO:* CV% ranged from 2.17% to 5.08% across 5 glucose levels. Bias% ranged from -2.0% to 2.1%.
      • Predicate Device (DIACHEX INFINITY):* CV% ranged from 5.45% to 2.45% (two data points missing for the highest and lowest glucose levels). Bias% ranged from -3.4% to 2.6%.
    • Linearity (20 to 600 mg/dL):
      • Easy Step: Slope 1.01, Intercept -1.11, R$^2$ 0.9997, sy.x 5.45
      • DIACHEX PRO:* Slope 1.00, Intercept 0.41, R$^2$ 0.9996, sy.x 6.66
      • Predicate Device (DIACHEX INFINITY):* Slope 1.01, Intercept -0.23, R$^2$ 0.9998, sy.x 6.77
    • Clinical Accuracy (Fingertip vs. YSI):
      • Easy Step: N=129, Slope 1.03, Intercept 2.18, R$^2$ 0.9887
      • DIACHEX PRO:* N=129, Slope 1.01, Intercept 3.10, R$^2$ 0.9890
      • Predicate Device (DIACHEX INFINITY):* N=129, Slope 1.02, Intercept 1.44, R$^2$ 0.9871
    • Alternate Site Testing (Palm vs. YSI, Forearm vs. YSI, Palm vs. Finger, Forearm vs. Finger): Results consistently showed slopes close to 1, intercepts close to 0, and high R$^2$ values (>0.96) for all proposed and predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Accuracy (Primary Test Set):
      • Sample Size: 129 participants for fingertip testing. 113 participants for alternate site testing (palm and forearm).
      • Data Provenance: The study was an "in-house study" performed by technicians, implying it was conducted by the manufacturer or a contracted lab. The document does not specify the country of origin of the data beyond the manufacturer being in Taiwan. The study appears to be prospective, as it involved collecting samples from participants specifically for this evaluation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The ground truth for the test set was established using a YSI glucose analyzer (likely a YSI 2300 STAT Plus Glucose & Lactate Analyzer or similar laboratory reference method). There is no mention of experts (e.g., radiologists, pathologists) being used to establish the ground truth for this type of glucose monitoring device. The YSI system itself serves as the reference standard.

    4. Adjudication Method for the Test Set

    Not applicable. For blood glucose monitoring systems, the ground truth is typically a direct measurement from a highly accurate laboratory reference instrument (like the YSI analyzer), rather than a subjective assessment requiring expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not presented. This type of study is usually conducted for imaging devices where human readers interpret results, and the AI's impact on human performance is evaluated. For blood glucose meters, the performance is assessed against a laboratory reference.

    6. Standalone Performance Study

    Yes, a standalone (algorithm only without human-in-the-loop performance) study was primarily conducted. The "Performance Characteristic Summary" section details the device's accuracy (precision, linearity, clinical accuracy) directly compared to the YSI reference method. The "clinical accuracy" study involved technicians taking readings, but the core evaluation is of the device's measurement against a golden standard.

    7. Type of Ground Truth Used

    The type of ground truth used was laboratory reference method data, specifically results obtained from a YSI glucose analyzer. This is considered the "gold standard" for glucose measurement in many clinical and research settings.

    8. Sample Size for the Training Set

    The document does not provide a specific sample size for a "training set." The submission describes performance characteristics of the finalized device. For IVD devices like blood glucose meters, the development process involves internal calibration and validation steps, but a distinct "training set" in the context of machine learning is not typically disclosed in 510(k) summaries, nor is it always applicable in the same way it would be for complex AI algorithms in image analysis.

    9. How the Ground Truth for the Training Set Was Established

    As mentioned above, information on a distinct "training set" is not provided. For the initial calibration and development of such a device, similar laboratory reference methods (e.g., YSI) would be used to establish the correct response for different glucose concentrations. The chemical formula and enzyme used in the test strip are identical to a previously cleared device (K062829), suggesting that the fundamental detection mechanism, which would have been rigorously characterized, remains unchanged.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1