    Intended Use

    Intended Use:
    The Stryker Navigation System - OrthoMap® 3D 1.1 Module is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery. The system is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified. The system offers the following features and functions:
    • Import of multi-modality image data
    • Surgical planning, such as image manipulation and visualization and procedural planning
    • Surgical navigation, such as manual and automatic registration and precise positioning of instruments and implants
    Indications:
    The system should be operated only by trained personnel such as surgeons and clinic staff.
    The system can assist surgeons in the following, but not limited to, orthopedic surgical procedures:
    • Orthopedic oncology procedures
    • Cam type FAI (femoroacetabular impingement) surgery
    • ACL (anterior cruciate ligament) reconstruction

    Device Description

    The Stryker Navigation System - OrthoMap® 3D 1.1 Module is part of the product series of the Stryker Navigation System. It is based on a wireless optical tracking localization device. The system comprises of hardware and software for surgical planning and computer assisted surgery. It supports orthopedic oncology procedures, cam type FAI (femoroacetabular impingement) surgery procedures and ACL (anterior cruciate ligament) reconstruction procedures.

    AI/ML Overview

    The provided text describes the Stryker Navigation System - OrthoMap® 3D 1.1 Module, a planning and intraoperative guidance system for computer-assisted orthopedic surgery.

    However, the document does not explicitly state specific acceptance criteria in a quantitative or tabular format, nor does it provide a full study report with detailed numerical performance metrics for the device against such criteria. Instead, it offers a summary of verification and validation activities.

    Based on the information provided, here's an analysis of the device's performance and the supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table of quantitative acceptance criteria. It broadly states that "the specified acceptance criteria for system and clinical accuracy can be reached" but does not detail what these criteria were.

    Acceptance Criteria (Quantitative)Reported Device Performance
    Not explicitly stated in the document. The document mentions that the data from cadaveric studies "revealed that the specified acceptance criteria for system and clinical accuracy can be reached," but the specific numerical criteria are not provided."the specified acceptance criteria for system and clinical accuracy can be reached for cam type FAI and ACL treatments."
    Safety and Effectiveness: "No new types of issues of safety and effectiveness are introduced by using this device.""The combination of the results of the bench testing, the cadaveric verification study, and the clinical confirmative trials demonstrate the safety and effectiveness of the system when used for the cam type FAI and ACL procedures."

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "cadaveric verification studies" and "clinical confirmative trials."

    • Cadaveric Verification Studies (Test Set): Performed on "cadaveric" subjects. The document does not specify the exact sample size (number of cadavers or procedures). The provenance is not explicitly stated as a country, but "Stryker Leibinger GmbH & Co. KG" is based in Germany. The studies were likely prospective in nature for verification.
    • Clinical Confirmative Trials (Test Set): The system "has been used clinically for cam type FAI treatments." The sample size and location (country) are not specified. These trials were likely prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number of experts or their qualifications used to establish ground truth for the test sets (cadaveric or clinical). It mentions the system "should be operated only by trained personnel such as surgeons and clinic staff," implying expert involvement during clinical use, but not explicitly in establishing ground truth for validation.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method (e.g., 2+1, 3+1, none) for establishing ground truth in the cadaveric or clinical test sets.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, the document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was performed. The focus is on the device's performance, not specifically on comparing human readers with and without AI assistance. The device is a "guidance system" for surgery, implying human-in-the-loop, but not in a diagnostic reading context.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The device is described as a "planning and intraoperative guidance system" and assists surgeons. This implies a human-in-the-loop system, not a standalone algorithm without human intervention. The performance studies focus on the overall system (hardware and software) in conjunction with clinical use.

    7. The Type of Ground Truth Used

    • For the cadaveric verification studies, the ground truth for "system and clinical accuracy" would likely be established through precise measurements using gold-standard techniques (e.g., coordinate measuring machines, optical tracking, direct anatomical measurements) compared against the system's output.
    • For the clinical confirmative trials, the ground truth would typically be based on intraoperative assessments by experienced surgeons, post-operative imaging, and potentially patient outcomes, confirming the safe and effective use of the system.
      However, the specific methods are not detailed in the summary.

    8. The Sample Size for the Training Set

    The document does not specify any sample size for a training set. This device is a navigation system for surgery, not an AI/ML diagnostic algorithm that typically undergoes a distinct "training" phase with labeled data in the sense of deep learning or machine learning models. Its algorithms are likely based on established geometric and image processing principles, refined and verified through the described bench and cadaveric testing.

    9. How the Ground Truth for the Training Set was Established

    As there's no mention of a distinct "training set" in the context of an AI/ML model, this question is not applicable based on the provided information. The software is part of a "Navigation System" that performs planning and guidance based on image data and anatomical structures. Its "ground truth" for development would stem from engineering principles, mathematical models, and knowledge of anatomy and surgical procedures, validated through the performance testing described.