Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K170886

    Validate with FDA (Live)

    Date Cleared
    2018-06-21

    (451 days)

    Product Code
    Regulation Number
    868.5630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K062263

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HOMED Mesh Nebulizer is an ultrasonic vibrating mesh nebulizer system designed to aerosolize physician-prescribed liquid medications for inhalation to a patient except for Pentamidine. The device may be used with patients 5 years and older in the home, hospital, and sub-acute care settings.

    Device Description

    The HOMED Mesh Nebulizer is a small, handheld, internally powered general purpose nebulizer which utilizes vibrating mesh technology to generate aerosol.

    AI/ML Overview

    The document provided outlines the K170886 510(k) premarket notification for the HOMED Mesh Nebulizer, comparing it to a predicate device, the Omron NE-U22 (K062263). The study described focuses on demonstrating substantial equivalence, primarily through non-clinical performance testing related to aerosol particle characterization, rather than human-in-the-loop clinical studies.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the HOMED Mesh Nebulizer are primarily established through comparative testing against a predicate device (Omron NE-U22). The goal is to demonstrate "substantial equivalence" in terms of performance and safety.

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/MetricAcceptance Criteria (Implied by Predicate Performance)HOMED Mesh Nebulizer Performance
    Particle Size (MMAD)Similar to Predicate (Omron NE-U22) Ranges
    Ventolin2.850-3.148 µm (Adult Flow)2.792-2.922 µm (Adult Flow)
    2.996-3.575 µm (Pediatric Flow)3.177-3.388 µm (Pediatric Flow)
    Budesonide2.686-2.761 µm (Adult Flow)2.532-2.577 µm (Adult Flow)
    2.778-2.997 µm (Pediatric Flow)2.739-2.850 µm (Pediatric Flow)
    Atrovent2.761-3.103 µm (Adult Flow)2.412-2.552 µm (Adult Flow)
    2.782-3.077 µm (Pediatric Flow)2.805-2.920 µm (Pediatric Flow)
    Geometric Std. Dev. (GSD)Similar to Predicate Ranges
    Ventolin2.814-3.011 (Adult Flow)2.818-2.866 (Adult Flow)
    2.533-2.736 (Pediatric Flow)2.450-2.767 (Pediatric Flow)
    Budesonide3.112-3.221 (Adult Flow)3.115-3.165 (Adult Flow)
    2.813-2.964 (Pediatric Flow)2.933-3.011 (Pediatric Flow)
    Atrovent2.824-3.070 (Adult Flow)2.746-2.845 (Adult Flow)
    2.778-3.019 (Pediatric Flow)2.580-2.746 (Pediatric Flow)
    Total Respirable Dose (0.5-5 µm) / %Similar to Predicate Ranges
    Ventolin62.1-64.9% (Adult Flow)64.8-67.2% (Adult Flow)
    55.7-64.5% (Pediatric Flow)55.5-62.6% (Pediatric Flow)
    Budesonide63.4-65.4% (Adult Flow)64.0-65.7% (Adult Flow)
    60.1-63.3% (Pediatric Flow)61.3-63.2% (Pediatric Flow)
    Atrovent61.3-67.0% (Adult Flow)67.3-69.1% (Adult Flow)
    54.1-73.6% (Pediatric Flow)65.3-69.8% (Pediatric Flow)
    Coarse Particle Dose (>4.7 µm) / %Similar to Predicate Ranges
    Ventolin33.1-37.2% (Adult Flow)30.4-32.2% (Adult Flow)
    33.9-43.7% (Pediatric Flow)37.2-39.3% (Pediatric Flow)
    Budesonide28.2-31.0% (Adult Flow)26.8-29.7% (Adult Flow)
    28.8-39.0% (Pediatric Flow)30.7-32.1% (Pediatric Flow)
    Atrovent12.6-38.6% (Adult Flow)23.1-25.9% (Adult Flow)
    6.0-57.6% (Pediatric Flow)26.6-30.5% (Pediatric Flow)
    Fine Particle Dose (<4.7 µm) / %Similar to Predicate Ranges
    Ventolin62.6-66.4% (Adult Flow)68.0-69.6% (Adult Flow)
    56.3-66.1% (Pediatric Flow)60.6-62.8% (Pediatric Flow)
    Budesonide69.0-87.5% (Adult Flow)70.3-73.2% (Adult Flow)
    61.0-71.1% (Pediatric Flow)67.9-69.3% (Pediatric Flow)
    Atrovent61.6-76.3% (Adult Flow)74.0-76.9% (Adult Flow)
    42.4-93.5% (Pediatric Flow)69.5-73.4% (Pediatric Flow)
    Ultra-fine Particle Dose (< 1.0 µm) / %Similar to Predicate Ranges
    Ventolin3.4-5.0% (Adult Flow)5.0-5.4% (Adult Flow)
    2.1-3.0% (Pediatric Flow)2.2-2.6% (Pediatric Flow)
    Budesonide7.8% (Adult Flow)8.9-9.2% (Adult Flow)
    6.7-7.4% (Pediatric Flow)7.8-8.7% (Pediatric Flow)
    Atrovent6.8-8.4% (Adult Flow)9.2-10.6% (Adult Flow)
    4.0-6.5% (Pediatric Flow)5.1-6.9% (Pediatric Flow)
    Total Delivered Dose (µg) / %Similar to Predicate Ranges
    Ventolin21.3-23.6% (Adult Flow)22.1-22.8% (Adult Flow)
    20.8-22.1% (Pediatric Flow)21.1-21.9% (Pediatric Flow)
    Budesonide17.5-19.6% (Adult Flow)18.0-18.6% (Adult Flow)
    15.9-16.8% (Pediatric Flow)16.1-16.8% (Pediatric Flow)
    Atrovent17.6-19.1% (Adult Flow)18.0-18.5% (Adult Flow)
    15.0-17.6% (Pediatric Flow)16.0-16.7% (Pediatric Flow)
    Shelf-life, Useful-life, Simulated Life CycleDemonstrated by testingMet specification criteria of 90 days for Medicine Cup
    CleaningDemonstrated by testingPerformed after each use via rinsing and boiling
    Electrical Safety, EMCMeets IEC 60601-1:2012, IEC 60601-1-2:2014, IEC 60601-1-11:2012Device met these requirements
    BiocompatibilityMeets ISO 10993-1 testsConcluded as biocompatible with no new issues
    Software Level of ConcernModerateModerate

    The acceptance criteria for the nebulizer's performance are primarily based on demonstrating comparable performance to the predicate device across various aerosol characteristics (MMAD, GSD, different particle doses, total delivered dose) for different medications and flow rates. This is explicitly stated in "Summary of Non-Clinical Performance Testing" and demonstrated by Tables 3 and 4 where the proposed device's performance values consistently fall within or are very similar to the ranges reported for the predicate device, or are directly compared to show "very similar" results.

    Study Details:

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document does not specify exact sample sizes for the particle characterization tests (e.g., how many nebulizer units were tested, or how many runs were performed for each medication/flow rate). It presents 95% confidence intervals for the results, implying repeat measurements were taken.
    • Data Provenance: The data appears to be from non-clinical laboratory testing ("We performed a number of non-clinical tests") conducted by the manufacturer, Shenzhen Homed Medical Device Co,.Ltd., presumably in China (the manufacturer's location is in Shenzhen, PRC). The data is prospective as it's generated specifically for this premarket submission to demonstrate equivalence.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts:

    • This information is not applicable in this context. The "ground truth" for nebulizer performance in this 510(k) submission is established by objective physical measurements using standardized cascade impaction techniques and comparison to the predicate device's measured performance. There are no human experts "establishing ground truth" in the way a radiologist might for an AI image analysis system.

    4. Adjudication Method for the Test Set:

    • This is not applicable. As the "ground truth" is based on objective physical measurements, there is no need for expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No, an MRMC study was NOT done. This submission is for a medical device (nebulizer) that delivers medication, not an AI or imaging diagnostic device that involves human interpretation of cases. The study focuses on the physical performance characteristics of the nebulizer, not on improving human reader performance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This question is partially applicable given the software component ("Software driven - Yes - firmware"). The software verification and validation testing ("Software Verification and Validation Testing") was conducted. The specific performance of the software (firmware) is related to its control of the vibrating mesh module (e.g., maintaining 117 kHz frequency, nebulization rate, power consumption) and LED indicators. While an explicit "standalone performance study" in the context of diagnostic accuracy isn't mentioned, the software V&V and the overall device performance (particle characterization) inherently demonstrate the algorithm's correct function in controlling the nebulization process. The "Level of Concern" for the software is "moderate," indicating that software failure could lead to a delay in appropriate medical care.

    7. The Type of Ground Truth Used:

    • The ground truth for the device's technical performance (e.g., particle size, delivered dose) is based on objective, standardized laboratory measurements using a cascade impactor, following established protocols (e.g., "Particle Characterization per Cascade Impactor").
    • For other aspects like electrical safety and biocompatibility, the ground truth is established by compliance with recognized consensus standards (e.g., IEC 60601-1, ISO 10993-1).

    8. The Sample Size for the Training Set:

    • This is not applicable. This document describes a 510(k) submission for a physical medical device (nebulizer), not an AI/Machine Learning model that requires a "training set" in the conventional sense. The "firmware" is mentioned, implying some level of software control, but it's not an adaptive learning algorithm.

    9. How the Ground Truth for the Training Set was Established:

    • This is not applicable for the same reason as point 8. There is no training set for an AI model. For the device's development and manufacturing, the "ground truth" for its design and desired performance would have been established through engineering specifications, previous research, and predicate device analysis.
    Ask a Question

    Ask a specific question about this device

    K Number
    K133105

    Validate with FDA (Live)

    Date Cleared
    2014-06-24

    (267 days)

    Product Code
    Regulation Number
    868.5630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K062263

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Breeze Nebulizer (NBR-101) is a vibrating mesh nebulizer system designed to aerosolize liquid medications for inhalation by the patient. The device may be used with pediatric (ages 2 years old and above) and adult patients in the home. It is not intended for use with Pentamidine.

    Device Description

    Breeze Nebulizer (NBR-101) is a vibrating mesh nebulizer that delivers aerosolized medication to the lower respiratory tract by using a vibrating mesh to create aerosol and provide fine particles to the patient's lungs. It is similar to the predicate device, the FDA-cleared Model NE-U22 Micro Air Vibrating Mesh Nebulizer, cleared under 510(k) K062263. They are identical in purpose, function, core technology and method of operation.

    Breeze nebulizer (NBR-101) is a portable size, curvaceous body design that is convenient to hold, and ability to detect the amount of medications available and to turn off automatically. The open button is made of soft materials and gives off an ice blue light, coupled with an overall elegant white exterior.

    Breeze Nebulizer (NBR-101) is battery powered, 4 "AAA" and the dimensions is 58(W) X 145(H) X 70(D). The medication container capacity is 8ml maximum and the residual volume is approximately 0.1ml.

    AI/ML Overview

    This document describes the 510(k) premarket notification for the Breeze Nebulizer (NBR-101), specifically focusing on its substantial equivalence to a predicate device rather than a study demonstrating it meets specific clinical performance acceptance criteria. As such, information regarding AI, ground truth, sample sizes for training/test sets, or expert involvement in establishing ground truth is not applicable.

    The core of the submission relies on non-clinical testing to demonstrate safety and effectiveness.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document lists various standards and regulations applied for different testing items. The reported performance is generally stated as meeting the requirements of these standards. Specific numerical acceptance criteria or detailed performance metrics are not provided in this summary.

    Testing ItemStandard and Regulations AppliedReported Device Performance
    Electromagnetic Compatibility & Electrical SafetyIEC 60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performanceMeets requirements
    IEC 60601-1-2 Edition 3:2007-03, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests. (General)Meets requirements
    BiocompatibilityISO 10993-5:2009(E) Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity.Meets requirements
    ISO 10993-10:2002/Amd.1:2006(E) Biological evaluation of medical devices --- Part 10: Tests for irritation and delayed-type hypersensitivity (7.4 Maximization test for delayed hypersensitivity)Meets requirements
    ISO 10993-3:2003/(R)2009, Biological Evaluation Of Medical Devices - Part 3: Tests For Genotoxicity, Carcinogenicity, And Reproductive Toxicity. (Biocompatibility)Meets requirements
    ISO 10993-6:2007/(R)2010, Biological Evaluation Of Medical Devices -- Part 6: Tests For Local Effects After Implantation. (Biocompatibility)Meets requirements
    ISO 10993-11:2009, Biological Evaluation Of Medical Devices -- Part 11: Tests For Systemic Toxicity. (Biocompatibility)Meets requirements
    ISO 10993-12:2012, Biological Evaluation Of Medical Devices -- Part 12: Sample Preparation And Reference Materials. (Biocompatibility)Meets requirements
    UsabilityIEC 60601-1-6:2006 Medical electrical equipment -Part 1-6: General requirements for safety -Collateral Standard: Usability.Meets requirements
    PerformanceEN 13544-1:2007 - Respiratory therapy equipment - Part 1: Nebulizing systems and their componentsMeets requirements
    EN 13544-2:2002+A1 - Respiratory therapy equipment - Part 2: Tubing and connectorsMeets requirements

    The document also provides a comparative table of characteristics between the proposed device and the predicate device, highlighting their similarities and differences. While this is not "acceptance criteria" in the traditional sense, it demonstrates how the device's characteristics are compared against a legally marketed device.

    ItemProposed Device (Breeze Nebulizer NBR-101)Predicate Device (Model Micro Air Vibrating Mesh Nebulizer NE-U22)
    ClassificationClass IIClass II
    Product CodeCAFCAF
    Intended UseVibrating mesh nebulizer system to aerosolize liquid medications for inhalation by pediatric (ages 2 and above) and adult patients in the home. Not for Pentamidine.Ultrasonic (vibrating mesh) nebulizer system to aerosolize liquid medications for inhalation by pediatric and adult patients in the home, hospital, and sub-acute care settings. Not for Pentamidine.
    TechnologyVibrating meshVibrating mesh
    Environment of UseHomeHome, Hospital, Sub-acute Institutions
    Patient PopulationPediatric (ages 2 and above), adultPediatric and adult
    Nebulizer components cleanableYesYes
    Software drivenYesNo
    Vibrating Capacity107kHz180kHz
    ButtonON/OFF SwitchON/OFF Switch
    Reservoir size8.0ml7.0ml
    Nebulization Rate0.2~0.4 ml/min0.25-0.9 ml/min
    Operation condition3℃ ~40℃; Max 70% RH0℃ ~ 45℃; 30% - 85% RH
    Storage condition-10℃~80℃; Max 70% RH-25°C ~ 70°C; 10% - 90% RH
    Power sourceFour AAA batteriesTwo AA batteries; AC adapter AC 120V (60Hz/DC3V)
    Power consumption1.5W1.5W
    Power indicatorLEDLED
    Dimensions58mm(W) x 70mm(D) x 145mm(H)38mm(W) x 51mm(D) x 104mm(H)
    WeightNot specified in table123.6 g (without batteries)

    2. Sample size used for the test set and the data provenance

    The document does not specify a "test set" in the context of a clinical performance study. The studies mentioned are non-clinical (bench tests, safety tests), and the sample size for these engineering and regulatory compliance tests is not provided in this summary. The data provenance is from digiO2 International Co., Ltd. in New Taipei City, Taiwan, performing internal non-clinical testing. These are assumed to be prospective tests conducted for the purpose of the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a submission for a medical device that relies on non-clinical engineering and regulatory compliance testing rather than clinical data or expert-derived ground truth.

    4. Adjudication method for the test set

    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-powered diagnostic tool, and no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an algorithm.

    7. The type of ground truth used

    For the non-clinical tests, the "ground truth" implicitly refers to the specified performance parameters and safety limits defined by the referenced international standards (e.g., IEC 60601-1, ISO 10993, EN 13544-1). Compliance with these standards serves as the benchmark.

    8. The sample size for the training set

    Not applicable. There is no mention of a "training set" as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132247

    Validate with FDA (Live)

    Date Cleared
    2014-01-08

    (173 days)

    Product Code
    Regulation Number
    868.5630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K062263

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ultrasonic mesh nebulizer model NE105 is an ultrasonic (vibrating mesh) nebulizer system designed to aerosolize medications for inhalation by the patient. The device may be used by adult and pediatric patients at the discretion of their physician at home, hospital &sub-acute care setting.

    Device Description

    The Ultrasonic Mebulizer NE105 is similar to the predicate device, the FDA-cleared Model NE-U22 Micro Air Vibrating Mesh Nebulizer, cleared under 510(k) K062263. Similarities are that both, the NE105 and the predicate NE-U22, are identical in purpose, function, core technology and method of operation. They are single-patient use, reusable electronic nebulizers, using micro-perforated vibrating mesh technology to aerosolize liquid medications. They have the same 'Indication for Use' stating that they are for inhalation therapy by adults and pediatric patients in a home, hospital & sub-acute care setting. Both devices are hand-held and portable. Power input for both devices is provided by either AA batteries or an AC/DC adapter. The model NE105 incorporates several design differences in comparison with the predicate NE-U22, mainly body shape, dimension/weight, a larger capacity reservoir, power source and software controlled electronics.

    AI/ML Overview

    The provided 510(k) summary for the Ultrasonic Mesh Nebulizer NE105 focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with acceptance criteria and performance data in a typical clinical trial format.

    Here's a breakdown of the requested information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a substantial equivalence submission for a nebulizer, the "acceptance criteria" are primarily based on demonstrating performance similar to the predicate device and compliance with relevant safety and performance standards. There isn't a direct table of clinical performance metrics like sensitivity/specificity. Instead, the performance is assessed through non-clinical tests.

    Acceptance Criteria (Demonstrated Equivalence/Compliance)Reported Device Performance
    EMC and Electrical Safety (Compliance with IEC 60601-1, IEC 60601-1-2)Verification by testing laboratories; certified with Declaration of Conformity. Device conforms to all applicable requirements in its intended operational environment.
    Biocompatibility (Compliance with ISO 10993-1 for body-contacting components)Testing (Cytotoxicity, Sensitization, Subchronic Toxicity, Genotoxicity, Implantation) confirmed biocompatibility of materials.
    Software Performance (Performs within specifications, safe for intended use)Test results concluded software performs within specifications and is safe. Permanent hazard analysis implemented in development process; clear software architecture; test protocol sufficiently verifies main functional operation.
    Cleaning Effectiveness (Sufficiency of cleaning instructions)Validation of manual cleaning method confirmed nebulizer can be cleaned as per instructions.
    Simulated Lifetime Performance (No change in performance, effectiveness, safety after simulated lifetime)"Simulated Lifetime Study" by renowned laboratory and in-house 1000-cycle life test concluded no change in performance, effectiveness, and safety when cleaned as per instructions.
    Aerosol Characterization (Particle Size Distribution) (Results in line with predicate device)Performed by Piper Medical Lab. All tests confirmed results were in line with the predicate device.
    Aerosol Characterization (Total Medication Dose Delivered) (Results in line with predicate device)Performed by Piper Medical Lab. All tests confirmed results were in line with the predicate device.
    Inter-Sample and Intra-Sample Variability (No measurable inter-sample variance)Tests showed no indication of any measurable inter-sample variance.
    Overall Performance Equivalence to PredicateModel NE105 performed equally well compared with the predicate device across all aerosol characterization tests.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify a "sample size" in the context of human subjects or clinical data, as it explicitly states, "The NE105 did not conduct, nor rely upon, clinical tests to determine substantial equivalence." The testing described is primarily laboratory-based and engineering/design validation. For simulated lifetime testing, "an in-house 1000-cycle life test was performed," which could be considered a sample size for that specific test.
    • Data Provenance: Not applicable as clinical data was not used for this submission. The tests were performed by various "testing laboratories" and "Piper Medical Lab," implying engineering/testing facilities. The submitter is from China.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This submission relies on objective engineering and laboratory test results against established standards and comparison to a predicate device, not subjective expert assessment of clinical data to establish a "ground truth."

    4. Adjudication Method for the Test Set

    Not applicable. There's no clinical test set requiring adjudication in this submission.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device for drug delivery (nebulizer), not an AI-powered diagnostic or interpretive device. Therefore, no MRMC or AI-assisted studies were performed or are relevant.

    6. If a Standalone (i.e. algorithm only, without human-in-the-loop performance) was done

    Not applicable. This device does not involve an algorithm for interpretation or diagnosis; it is a physical device for drug delivery.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established through:

    • Compliance with recognized standards: IEC 60601-1, IEC 60601-1-2, ISO 10993-1.
    • Benchmarking against a predicate device: Specifically for aerosol characterization (particle size distribution and total medication dose delivered).
    • Design validation tests: Ensuring the device meets its own specified design requirements (e.g., software functionality, cleaning effectiveness, simulated lifetime).

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI device that requires a "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122060

    Validate with FDA (Live)

    Date Cleared
    2013-03-21

    (251 days)

    Product Code
    Regulation Number
    868.5630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K062263

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Clever TD-7001 Nebulizer is an ultrasonic (vibrating mesh) nebulizer system designed to aerosolize physician-prescribed solution for inhalation by the patient, except for Pentamidine. The Clever TD-7001 Nebulizer is for use with children and adult patients in the homecare environment.

    Device Description

    Clever TD-7001 Nebulizer is an ultrasonic (vibrating mesh) nebulizer system designed to aerosolize physician-prescribed solution for inhalation by the patient. It is for use with children and adult patients in the homecare environment. Clever TD-7001 Nebulizer consists of a Nebulizer unit, a medication cap and a Jand Medication cup are single medication cup. The Nebulizer unit, me patient re-usable. The Nebulizer unit trol circuitry and is powered by using 2 AA alkaline batteries organ A onnects to electrical outlets. The patient interface comprises a mouthpiê medication cap contains the metal alloy mesh. When the device is turne ultrasonic piezoelectric element vibrate to cause the mesh vibrate. ation solution tin the reservoir of medication cup is aerosolized by the vibrating the aerosols can be inhaled through the mouthpiece or mask.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Clever TD-7001 Nebulizer, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Aerosol Characterization Testing (Particle Size Distribution)
    Equivalent performance to predicate device (K062263) for 3 drugs (Ipratropium bromide, Ventolin, Cromolyn sodium) with no significant difference (p > 0.035).The test "has shown Clever TD-7001 Nebulizer consistent in repeatabil for each three classes of drug, and demonstrated equivalent performance? ate device K062263 that no significant difference (p>0.035) in the Clever TD-7001 Nebulizerst as the ame performance characteristics as the predicate device and meets its product specification as well."
    Airpath Testing (Gas Sample Analysis)
    Device does not emit potential toxic gases (carbon monoxide, carbon dioxide, ozone, volatile organic compounds)."Gas sample analysis of Clever TD-7001 Nebulizer has performed and shown the device does not emit potential toxic gases that may cause harmful influences to human, including carbon monoxide, carbon dioxide, ozone, or volatile organic compounds (VOCs)."
    Output of particulate matter conforms to EPA PM2.5 standard."The output of particulate matter conformed to EPA requirements of the PM2.5 standard."
    Materials (Biocompatibility)
    Meet ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation), ISO 10993-10 (sensitization), ISO 10993-6 (implantation), and ISO 10993-3 (genotoxicity) for gas path contact materials."Biocompatibility evaluations of cytotoxicity (ISO 10993-5), irritation (ISO 10993-10), sensitization (ISO 10993-10), implantation (ISO 10993-6) and genotoxicity (ISO 10993-3) tests have been conducted for the gas path contact materials." (Implies meeting the standards.)
    Safety and EMC (Electrical Safety and Electromagnetic Compatibility)
    Meet EN 60601-1:1988+A1:1991+A2:1995 and EN 60601-1-1-2:2001+A1:2006, CISPR 11:2003."The electromagnetic compatibility and electric safety of the proposed device are tested to meet the following standards: - EN 60601-1:1988+ A1:1991+ A2:1995 - EN 60601-1-1-2:2001+A1:2006, CISPR 11: 2003" (Implies meeting the standards.)

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the specific sample sizes for the test sets in the performance characteristic studies. It mentions "three drugs (Ipratropium bromide, Ventolin, and Cromolyn sodium)" were used for the aerosol characterization testing, implying at least three distinct test cases.

    The data provenance is not specified in terms of country of origin or whether it was retrospective or prospective. It is, however, implied to be generated specifically for the submission of this device, suggesting prospective testing conducted by the manufacturer, TaiDoc Technology Corporation (Taiwan).


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided in the given text. For device performance testing of a nebulizer, "ground truth" would typically refer to objective measurements obtained from controlled laboratory conditions, rather than expert interpretation of data.


    4. Adjudication Method for the Test Set

    This information is not applicable/provided in the context of the described performance testing. The studies are objective measurements (e.g., particle size distribution, gas analysis) rather than subjective assessments requiring adjudication.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not conducted. This type of study is typically relevant for AI-powered diagnostic imaging devices, not for a medical device like a nebulizer.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not applicable as the device is a nebulizer and does not involve an algorithm or AI. The performance tests described (aerosol characterization, airpath, materials, safety/EMC) are inherently "standalone" in the sense that they measure the device's physical and functional properties directly.


    7. The Type of Ground Truth Used

    The ground truth used for these studies is based on:

    • Objective Measurement Standards:
      • For Aerosol Characterization: Comparison to the performance of a legally marketed predicate device (K062263) using established methods like Cascade Impactor, with statistical significance testing (p > 0.035).
      • For Airpath Testing: Compliance with EPA requirements (PM2.5 standard) and general toxic gas emission limits.
      • For Materials: Adherence to international biocompatibility standards (ISO 10993 series).
      • For Safety and EMC: Compliance with international electrical safety and electromagnetic compatibility standards (EN 60601 series, CISPR 11).

    8. The Sample Size for the Training Set

    This information is not applicable as the device is not an AI/ML algorithm requiring a training set.


    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as the device is not an AI/ML algorithm requiring a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102454

    Validate with FDA (Live)

    Date Cleared
    2011-10-25

    (424 days)

    Product Code
    Regulation Number
    868.5630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K062263, K870027, K072019

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The I-neb AAD system with TIM is a nebulizer system designed to aerosolize liquid medication for inhalation by the patient in the home care, nursing home, sub-acute institution, or hospital environment.

    The I-neb Insight AAD System is monitoring software that provides feedback to the patient recording treatment events, including treatment times and compliance data which is also available to the clinician. It is an accessory to an accessory for use with the I-neb AAD system.

    The I-neb AAD system with TIM is intended for patients 3 years and older who can coordinate breathing.

    Device Description

    The I-neb AAD System is an ultrasonic (vibrating mesh) nebulizer system designed to aerosolize liquid medication (except pentamidine) for inhalation by the patient in the home care, nursing home, sub-acute institution, or hospital environment.

    The I-neb Insight AAD System is an accessory for use with the I-neb AAD system and controls monitoring software that provides feedback to the patient recording treatment events, including treatment times and compliance data which is also available to the clinician.

    The I-neb AAD system with TIM incorporates a modification of the predicate I-neb AAD system with (tidal breathing mode (TBM)), K042991. While the I-neb Insight incorporates updated software to accommodate the added TIM feature, it is a modification of the predicate I-neb Insight AAD System, K052941.

    AI/ML Overview

    The provided text describes a 510(k) summary for the I-neb AAD Systems with TIM and I-neb Insight AAD System. The primary method for demonstrating substantial equivalence is through comparative bench testing against predicate devices. The document does not describe a study in the context of AI/ML device performance, as it is a medical device from 2011, long before the widespread use of AI in medical imaging. Therefore, many of the questions related to AI device performance are not applicable.

    Here's an analysis based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are implicitly tied to demonstrating substantial equivalence to existing predicate devices (K062263 – Omron U-22, K870027 - Salter 8900 nebulzier, K072019 - Activaero - AKITA2 APIXNEB, K042991 - I-neb AAD, K052491 - I-neb Insight). The "acceptance criteria" are not explicitly stated as numerical thresholds but rather as the proposed device exhibiting comparable performance characteristics to the predicates in specific areas.

    Acceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance (I-neb AAD with TIM & I-neb Insight AAD)
    Particle Characterization (MMAD, GSD, Respirable Fraction %) comparable to predicates."Comparison of particle characterization testing included the evaluation of MMAD, GSD, Respirable Fraction (%) the predicates and the proposed device were found to be substantially equivalent."
    Comparative Dosing (gravimetric dose, Filter dose, Treatment time) equivalent to predicates."Comparative Dose for the I-neb AAD system in TBM vs. TIM mode and to the predicates Omron U22 (K062263) and Salter 8900 (K870027) was performed. Parameters measured and compared included gravimetric dose, Filter dose and Treatment time. Results the predicates and the proposed device were substantially equivalent."
    Indications for Use for general purpose nebulization (I-neb AAD with TIM) and monitoring accessory (I-neb Insight AAD).I-neb AAD with TIM: "General Purpose use." "a nebulizer system designed to aerosolize liquid medication for inhalation by the patient in the home care, nursing home, sub-acute institution, or hospital environment." I-neb Insight AAD: "monitoring software that provides feedback to the patient recording treatment events, including treatment times and compliance data which is also available to the clinician."
    Technology: Vibrating mesh nebulizer and breath-triggered nebulization."Identical vibrating mesh nebulizer technology to predicates." "Identical breath triggered nebulization technology to predicate."
    Materials: Gas and fluid pathway materials identical to predicate I-neb AAD."The materials in the gas and fluid pathway are identical to predicate device - K042991 -- I-neb AAD System."
    Environment of Use: Home care, nursing home, sub-acute institution, or hospital environment."Identical to predicate."
    Patient Population: 3 years and older who can coordinate breathing."Identical to predicates."
    I-neb Insight Technology/Indications: Monitoring software and feedback."Identical technology to predicate - K052491 - I-neb Insight AAD system." "The I-neb Insight is monitoring software that provides feedback to the patient recording treatment events, including treatment times and compliance data which is also available to the clinician."

    Study Details:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • The document mentions "3 drugs" were used for particle characterization.
      • The testing was "performed at flow rates of 15 lpm and 30 lpm."
      • The sample sizes for "comparative dosing" are not explicitly stated, but it was done between the I-neb AAD system in TBM vs. TIM mode and compared to Omron U22 and Salter 8900.
      • This was bench testing, not human clinical data. Therefore, the concept of "country of origin for data" or "retrospective/prospective" human data does not apply here.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not applicable. This submission relies on bench testing, not expert-adjudicated ground truth from human data for AI/ML performance.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not applicable. This was bench testing, not a human consensus or adjudication study.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • No. This is a 2011 submission for a nebulizer, not an AI/ML diagnostic tool. An MRMC study was not conducted, nor would it be relevant for this device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Not applicable. This device is a nebulizer system, not an AI algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • The "ground truth" for this submission are the performance specifications and characteristics of the predicate devices, against which the proposed device's bench test results (e.g., MMAD, GSD, gravimetric dose) were compared to demonstrate substantial equivalence.
    7. The sample size for the training set

      • Not applicable. This device does not involve an AI/ML algorithm requiring a training set.
    8. How the ground truth for the training set was established

      • Not applicable. This device does not involve an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1