510(k) Data Aggregation

    K Number
    K253150

    Validate with FDA (Live)

    Date Cleared
    2025-10-24

    (29 days)

    Product Code
    Regulation Number
    890.5650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K061866

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cryon-X Pro
    The private-label Cryon-X Pro device combines cold, heat, contrast, and compression therapy. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post-traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) is indicated.

    The device may optionally provide DVT therapy, intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.

    This device is intended to be used by or on the order of licensed healthcare professionals in rehabilitation facilities, outpatient clinics, athletic training settings, and home settings.

    Cryon-X One
    The private-label Cryon-X One device combines cold and compression therapy. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to be used by or on the order of licensed healthcare professionals in rehabilitation facilities/hospitals, outpatient clinics, athletic training settings, and home settings.

    Device Description

    This submission includes two models JKH-152 (private labelled as Cryon-X Pro) and JKH-151 (private labelled as Cryon-X One) as the subject devices. The private-label Cryon-X Pro has all the same features of combining cold, heat, contrast, and compression therapies as its original 510(k) cleared device in K223541, and Cryon-X One has the simplified features of cold and compression, which are the same as those of its original 510(k) cleared device in K223541. Each of the subject devices is an AC powered, software-controlled multimodality device, intended to be used by or on the order of licensed healthcare professionals in rehabilitation facilities/hospitals, outpatient clinics, athletic training settings, and home settings.

    The subject device is a prescriptive device, which is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy are indicated. It is optionally intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.

    The subject device and its accessories are clean and non-sterile. The device has a power switch, and then conducts the remaining operations on its touch screen, including the treatment temperature adjustment, air compression adjustment, and treatment time adjustment. The device works by circulating cooled or heated water and air through a treatment wrap that is placed on the treatment body area. The cooled or heated water circulates through the treatment wrap and provides cold or hot therapy, and the air compression inflates and deflates the treatment wrap to compress around the treatment body area. A connecting tube/hose is used to connect the device to the treatment wrap. The disposable or reusable wraps come with a variety of options to cover different body areas, including the universal, back, shoulder, ankle/foot, hip, knee, etc. To avoid any potential adverse skin reactions such as redness, irritation, and cod/hot injury, the sock/clothing should be worn by the patient prior to use.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K223541

    Validate with FDA (Live)

    Date Cleared
    2023-10-27

    (336 days)

    Product Code
    Regulation Number
    890.5650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K061866

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cold/Hot Compression combines cold, heat, contrast, and compression therapy. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) are indicated.

    Cold/Hot Compression also provide DVT therapy. It is intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.

    ColdHot Compression is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, athletic training settings, and home settings.

    Device Description

    The subject device is a prescriptive device, which combines cold/hot therapy and air compression. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) are indicated. It is intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.

    The subject device comes with the thermal-therapy wraps and the DVT wraps. The thermal-therapy wrap is nylon fabric on one side and velvet cloth on the other side; the DVT wrap is velvet on both sides. To avoid any potential adverse skin reactions such as redness, irritation, and cold/hot injury, the sock/clothing should be worn by the patient prior to use.

    The subject device includes a main device and an optional DVT device, and has a limited shelf life of 3 years, based on the charge retention characteristics of the device's battery.

    AI/ML Overview

    This document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed clinical study with acceptance criteria and performance metrics in the typical sense.

    Therefore, the information requested might not be fully available or explicitly stated in the provided text as it would be in a clinical study report.

    Based on the provided text, here's what can be extracted and inferred:

    1. Table of acceptance criteria and the reported device performance

    The concept of "acceptance criteria" here is primarily tied to demonstrating equivalence to the predicate device. The performance is assessed by comparing technical specifications.

    FeaturePredicate Device Performance / Range (Therm-X)Subject Device Performance / Range (Cold/Hot Compression)Equivalence Note
    Cold Therapy (Default)34°F, 45°F, 55°F50°FEquivalent (Note 1)
    Cold Therapy (Custom)34°F - 55°F41°F-55°FEquivalent (Note 1)
    Cold Therapy (Lowest)34°F or 40°F41°FEquivalent (Note 1)
    Heat Therapy (Default)105°F, 107°F, 110°F105°FEquivalent (Note 2)
    Heat Therapy (Custom)105°F - 110°F105°F-109°FEquivalent (Note 2)
    Heat Therapy (Highest)110°F109°FEquivalent (Note 2)
    Edema Pressure (Lowest)5 mm Hg or 20 mmHg15 mm HgEquivalent (Note 3)
    Edema Pressure (Highest)70 mmHg75 mm HgEquivalent (Note 3)
    DVT Pressure (Calf)50-70 mmHg55 mmHgEquivalent (Note 4)
    DVT Pressure (Foot)90 - 130 mmHg130 mmHgEquivalent (Note 4)
    Cycle Length (Longest)40 or 60 minutes60 minutesEquivalent (Note 5)
    Contrast Therapy (Heat)105°F105°FEquivalent (Note 6)
    Contrast Therapy (Cold)38°F49°FEquivalent (Note 6)
    Contrast Cycle (Heat)3-10 minutes10 minutesEquivalent (Note 7)
    Contrast Cycle (Cold)3-10 minutes20 minutesEquivalent (Note 7)
    Reservoir Fluid Capacity650 mL350 mLEquivalent (Note 8)

    Notes on Equivalence (as explained in the document):

    • Note 1 (Cold Therapy): The subject device's cold therapy temperatures are within or similar to the range of the predicate/reference devices, and are adjustable by professionals. Differences are considered insignificant and do not raise new safety/effectiveness concerns.
    • Note 2 (Heat Therapy): Similar to cold therapy, the heat therapy temperatures are within or similar to the predicate/reference devices and adjustable. Differences are insignificant.
    • Note 3 (Edema Pressure): The subject device's lowest pressure is within the predicate's range and similar to the reference. The highest pressure is slightly higher but still adjustable by professionals. Differences are insignificant.
    • Note 4 (DVT Pressure): The subject device's DVT pressures for calf and foot fall within or are slightly higher than the predicate/reference devices. Differences are insignificant.
    • Note 5 (Cycle Length): The subject device's longest cycle length is similar to or longer than the predicate's, though shorter than the reference. Cycle length is adjustable. Differences are insignificant.
    • Note 6 (Contrast Therapy Temp): The subject device's contrast therapy temperatures are either similar to the predicate or identical to the reference. No new safety/effectiveness issues.
    • Note 7 (Contrast Cycle Length): The contrast therapy cycle length is slightly different from the predicate but identical to the reference. No new safety/effectiveness issues.
    • Note 8 (Reservoir Fluid Capacity): The subject device's capacity is smaller than the predicate's but larger than the reference's. This difference does not raise new safety/effectiveness issues.

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "Clinical testing was not needed in support of this 510(k) application."

    Therefore, there is no test set in the sense of patient data from a clinical study. The "testing" involved non-clinical bench tests and comparison of specifications to predicate devices. The data provenance is from internal JKH Health Co., Ltd. testing and comparison to publicly available predicate and reference device information.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As no clinical testing was performed, there was no "test set" requiring expert ground truth establishment for a patient-based diagnostic or therapeutic outcome.

    4. Adjudication method for the test set

    Not applicable. There was no clinical test set requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical medical device (Cold/Hot Compression unit), not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-only device.

    7. The type of ground truth used

    The "ground truth" for the non-clinical tests would be the measured physical parameters of the device (e.g., actual temperature output, actual pressure levels, cycle times) against the device's design specifications and the specifications of the predicate devices. For safety standards, the ground truth is adherence to the requirements of the standards (e.g., ANSI AAMI ES60601-1, IEC 60601-1-2).

    8. The sample size for the training set

    Not applicable. There is no mention of a "training set" as this is not an AI/machine learning device. The design was likely based on engineering principles and comparison to existing devices.

    9. How the ground truth for the training set was established

    Not applicable. No training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162481

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-02-16

    (163 days)

    Product Code
    Regulation Number
    870.5800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K061866, K133274

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RF1400 Active Compression Wrap is intended to be a portable system, prescribed by healthcare professionals, to treat the following conditions by stimulating blood flow in the legs:

    · Aid in the prevention of DVT (deep vein thrombosis);

    • Enhance blood circulation;

    · Diminish post-operative pain and swelling;

    • · Reduce wound healing time;
      · Aid in the treatment and healing of: stasis dermatits, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency, chronic lymphedema, and reduction of edema in the lower limbs;

    · As a prophylaxis for DVT by persons expecting to be stationary for long periods of time.

    · Reduction of edema associated with soft tissue injuries, such as burns, postoperative or post-immobilization edema, or ligament sprains.

    The device can be used in the home or clinical setting. The device is intended for use in an adult patient population.

    Device Description

    The RF1400 Active Compression Wrap is a lightweight, portable, rechargeable battery powered, prescriptive device that helps stimulate blood flow in the lower limb through the use of intermittent sequential compression. The wrap contains nickel titanium, martensite phase change wires, using a battery-powered microprocessor to "excite" and "relax" the wires resulting in compression. The battery and microprocessor components are protectively housed in a plastic controller case that is permanently attached to the wrap. A single, touch control button interface and a RGB LED light indicator provide the user interface, and there is a port for connecting the battery charger plug. The wrap is available in a wide range of sizes XS, S, M, and L, to accommodate varying anatomy sizes.

    The wrap is divided into three discrete zones which are externally applied to the limb. After one zone is fully activated for a period of time and turns off, then the next zone is activated. This cycle continues until all three zones have activated and turned off. Then the sequence is repeated after a short delay. This cycle repeats until the unit is turned off. The wrap may be used on one or both legs. When used on both legs, the wraps operately. The wrap is supplied with a rechargeable battery, which can be charged when not in use.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the "RF1400 Active Compression Wrap." This device is intended to stimulate blood flow in the legs through intermittent sequential compression. However, the document does not contain information about a study that proves the device meets specific acceptance criteria based on clinical outcomes or a comparison against human performance.

    The provided text focuses on:

    • Regulatory clearance (510(k) summary): It primarily describes the device, its intended use, contraindications, and why it is considered "substantially equivalent" to predicate devices based on technological characteristics and bench testing.
    • Nonclinical testing: It mentions electrical safety, EMC, software validation, environmental/shipping, life cycle, and performance testing (biocompatibility, surface temperature, equivalent pressure delivery, cycle time).

    Therefore, I cannot provide a detailed answer to your request, specifically regarding:

    1. A table of acceptance criteria and reported device performance related to clinical outcomes.
    2. Sample sizes for test or training sets for a clinical study.
    3. Number of experts or their qualifications for ground truth establishment.
    4. Adjudication methods for a clinical test set.
    5. MRMC comparative effectiveness study or human reader improvement with AI assistance.
    6. Standalone algorithm performance.
    7. Type of ground truth used (expert consensus, pathology, outcomes data).
    8. Sample size for the training set (if it were an AI/ML device).
    9. How ground truth for the training set was established.

    This document pertains to a mechanical compression device, not typically an AI/ML device, and the substantial equivalence determination for such devices often relies on engineering specifications and bench testing demonstrating equivalent function to existing predicate devices, rather than large-scale clinical outcome studies in the same way an AI diagnostic device might require.

    In summary, the provided document does not contain the information requested about acceptance criteria and a study proving the device meets them in the context of clinical performance, particularly as it relates to AI/ML device evaluation.