510(k) Data Aggregation

    K Number
    K101805

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-04-08

    (284 days)

    Product Code
    Regulation Number
    870.3470
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    Reference Devices :

    K061727 Gore PRECLUDE® Vessel Guard

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HydroFix Vaso Shield™ is indicated for use as a cover for vessels following anterior vertebral surgery.

    Device Description

    The MiMedx HydroFix™ Vaso Shield is a flexible sheet of polyvinyl alcohol (PVA) material provided in various dimensions. It is a slippery, clear pliable, conformable, nanoporous provided in valled over that minimizes tissue contact between the vessels and the spine. There are no holes or perforations. There are no markings on either side of the sheet, raised (embossed) or printed. The sheet is provided sterile and hydrated.

    AI/ML Overview

    Acceptance Criteria and Study for HydroFix™ Vaso Shield (K101805)

    The provided document describes the 510(k) premarket notification for the HydroFix™ Vaso Shield. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than establishing novel safety and effectiveness criteria through a new study with predefined acceptance criteria. Instead, the "acceptance criteria" are implied by the demonstration of comparable performance to existing, legally marketed devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Technological Characteristics and Functionality comparable to predicate devices.The device is a flexible sheet of polyvinyl alcohol (PVA). It is described as slippery, clear, pliable, conformable, and nanoporous, designed to minimize tissue contact between vessels and the spine. It has no holes or perforations, no markings, and is provided sterile and hydrated. This aligns with the basic function of a "vessel guard."
    Similar Indications for Use as predicate devices.Indicated as a cover for vessels during anterior vertebral surgery. This is directly comparable to the stated intended use of the Paradis Vaso Shield™ (K090022 and K093551) and Gore PRECLUDE® Vessel Guard (K061727).
    Demonstrated reduction in risk of potential vessel damage in vivo.A sheep model (animal study) was conducted to demonstrate the reduction in the risk of potential vessel damage during anterior vertebral revision surgery. The implanted devices (HydroFix™ Vaso Shield) allowed a separation of the soft tissue adjacent to the study site from the study site. This provides evidence for the device's protective function.
    Safety profile comparable to predicate devices.While not explicitly detailed, the 510(k) process inherently requires a demonstration that the new device does not introduce new or different questions of safety or effectiveness compared to predicate devices. The absence of adverse findings in the animal study and the comparison to established predicate devices serve as the basis for this.
    Materials comparable to predicate devices or acceptable for medical use.The device is made of polyvinyl alcohol (PVA). The overall assessment of substantial equivalence suggests this material is acceptable for the intended purpose, aligning with the materials of predicate devices (polypropylene, polyethylene terephthalate, or polytetrafluoroethylene for the general class, though specific predicate materials are not detailed for comparison).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document mentions a "Large Animal Implantation A sheep model (animal study)." It does not specify the exact number of sheep used in the study.
    • Data Provenance: The study was an in vivo animal study conducted on sheep. The country of origin for the data is not specified. It is a prospective study as an experiment was conducted to generate the data for the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not provide details on the number or qualifications of experts used to establish ground truth for the animal study. Veterinary surgeons or researchers would likely have been involved in the implantation and assessment of the devices in the sheep model.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the animal study data. The results would likely have been analyzed and interpreted by the research team involved.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. An MRMC comparative effectiveness study was not performed. This type of study is typically used for diagnostic devices involving human interpretation of images or other data, which is not applicable to this physical implantable device.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No. This is not
    applicable. The HydroFix™ Vaso Shield is a physical implantable device, not an algorithm or software. Its performance is assessed through its physical interaction with tissues in vivo.

    7. Type of Ground Truth Used

    The ground truth for the in vivo testing was established through direct observation and assessment of the implanted devices in the animal model. This would likely involve:

    • Direct observation/histology: Assessing the physical separation of soft tissue, the presence or absence of vessel damage, and tissue responses at the implant site.
    • Pathology: Likely the most definitive ground truth, involving histological examination of tissues around the implant to confirm device presence, tissue separation, and absence of adverse reactions or damage.

    8. Sample Size for the Training Set

    No training set is mentioned or applicable. This device's performance was evaluated through an in vivo study, not through a machine learning or algorithm-based approach that requires training data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K100984

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-07-07

    (90 days)

    Product Code
    Regulation Number
    870.3470
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K062161, K960532, K061727

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    XYLOS™ Vessel Guard is indicated as a cover for vessels during anterior vertebral surgery.

    Device Description

    XYLOS™ Vessel Guard is a flexible, non-resorbable, implantable sheet composed of microbialderived cellulose. The device is presented ready-to-use in a sterile double-pouched package. It is intended for one time use.

    AI/ML Overview

    The provided K100984 510(k) summary for the XYLOS™ Vessel Guard focuses on demonstrating substantial equivalence to predicate devices through material similarity and biomechanical testing, rather than clinical performance based on a study with acceptance criteria often associated with AI/software devices. Therefore, a direct response to some of the specific questions is not possible from the provided text, as they pertain to aspects that are not relevant to this type of medical device submission.

    However, I can extract the relevant information and indicate where details regarding AI-specific criteria are not applicable.

    Here's the breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/TestAcceptance CriteriaReported Device Performance
    BiocompatibilityChemical CompositionIdentical to MTA™ Protective Sheet (K090778)Biologically and chemically identical to MTA™ Protective Sheet (K090778)
    Biocompatibility PropertiesEquivalent to MTA™ Protective Sheet (K090778)Equivalent biocompatibility properties of the MTA™ Protective Sheet
    Biomechanical PerformanceTensile StrengthEquivalent to Replication Medical Vessel Guard (K082782) and its predicatesDemonstrated biomechanical equivalence to Replication Medical Vessel Guard (K082782) and its predicates (PRECLUDE® IMA Sleeve, K960532; PRECLUDE® Vessel Guard, K061727)
    Suture Pull-out TestingEquivalent to Replication Medical Vessel Guard (K082782) and its predicatesDemonstrated biomechanical equivalence to Replication Medical Vessel Guard (K082782) and its predicates

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not applicable/not provided in the context of this 510(k) submission. The device is a physical implantable sheet, and its substantial equivalence was demonstrated through benchtop biomechanical testing and material comparison, not through a clinical study with a "test set" of patient data or clinical outcomes.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not applicable/not provided. Ground truth, in the context of AI or diagnostic studies, refers to a definitive diagnosis or outcome, usually established by experts. For this device, "ground truth" would relate to the established material properties and biomechanical performance standards of the predicate devices. These are typically assessed by engineers and materials scientists in laboratory settings.

    4. Adjudication Method for the Test Set

    This information is not applicable/not provided. Adjudication methods (like 2+1 or 3+1) are used to resolve discrepancies in expert opinions when establishing ground truth in clinical or image-based studies. Since this submission relies on benchtop testing and material comparisons, such adjudication is not relevant.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices or AI algorithms that assist human readers in interpreting data or images. The XYLOS™ Vessel Guard is an implantable surgical aid, and its effectiveness is demonstrated through its physical properties and biological compatibility, not through an assessment of human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone performance study was not done. This concept applies to AI algorithms that operate independently. The XYLOS™ Vessel Guard is a passive physical device.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is based on:

    • Established Material Composition: The chemical and biological identity to the predicate MTA™ Protective Sheet.
    • Established Biomechanical Performance Standards: The tensile strength and suture pull-out performance as demonstrated by the predicate devices (Replication Medical Vessel Guard, PRECLUDE® IMA Sleeve, PRECLUDE® Vessel Guard).

    Essentially, the ground truth for this submission is the well-characterized properties and performance of the legally marketed predicate devices.

    8. The Sample Size for the Training Set

    This information is not applicable/not provided. The concept of a "training set" applies to machine learning or AI models, where data is used to teach the algorithm. This device is a physical product, not an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable/not provided for the same reasons as #8.