510(k) Data Aggregation
(314 days)
MFUSE™ is indicated for temporary external use for controlling moderate to severe bleeding.
MFUSE™ is a sterile, single-use, disposable device designed to stop bleeding by creating an barrier over the wound. The device will be used by healthcare professionals to temporarily stop moderate to severe bleeding. The device is comprised of polyacrylic acid (PAA) in a chitosan matrix. The PAA / chitosan structure is mixed with mineral oil to form a paste that is cream to light yellow in color. MFUSE™ is provided in a 5gm (mL) quantity syringe. The syringe is packaged in a foil pouch and is terminally sterilized with gamma sterilization.
N/A
Ask a specific question about this device
(31 days)
Prescription, Rx UltraClot™ is intended as a hemostatic dressing for emergency external use and temporary wound treatment to achieve hemostasis of moderate to severe bleeding.
Over the Counter, OTC
UltraClot™ OTC is intended as a topical hemostatic dressing for the local management of bleeding from minor cuts, lacerations, and abrasions.
UltraClot™ is a hemostatic agent comprising a clay-based powder contained in a dissolving pouch with a non-stick gauze pad backing that is placed on a moderate to severe wound and held in place until hemostasis is achieved. The device is packaged in a vacuum sealed foil bag and is provided sterile.
The provided text is a 510(k) summary for the UltraClot™ and UltraClot™ OTC devices. These documents are typically for medical devices seeking FDA clearance based on substantial equivalence to a predicate device, rather than requiring extensive clinical trials to prove efficacy against defined acceptance criteria in the same way a new drug or high-risk medical device might.
The key information regarding "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of this 510(k) is related to demonstrating substantial equivalence to a predicate device through non-clinical testing, particularly animal models and biocompatibility.
Here's an analysis of the provided text in relation to your questions, noting that many of the requested details for AI/software studies are not applicable to this type of device submission:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" in a quantitative, tabular format as one might find for a software-based diagnostic AI. Instead, the "acceptance criteria" for this 510(k) are implicitly met by demonstrating substantial equivalence to the predicate device, Quikclot®, in terms of mechanism, safety, and efficacy.
| Acceptance Criteria (Implied for Substantial Equivalence) | Reported Device Performance (UltraClot™) |
|---|---|
| Similar adsorptive mechanism | Has a similar adsorptive mechanism to Quikclot®. |
| Equivalent speed of hemostasis | Found to be at least substantially equivalent to Quikclot® in terms of speed of hemostasis. |
| Equivalent efficacy of hemostasis | Found to be at least substantially equivalent to Quikclot® in terms of efficacy. |
| Absence of exothermia | Demonstrates absence of exothermia. |
| Reduced blood loss | Demonstrates reduced blood loss. |
| Biocompatibility (safety for intended use) | Biocompatibility testing demonstrated UltraClot™ to be safe for its intended use (mutagenicity, cytotoxicity, intracutaneous studies). |
| Similar indications for use | Intended as a hemostatic dressing for emergency external use and temporary wound treatment to achieve hemostasis of moderate to severe bleeding (similar to predicate). |
| OTC version labeling conformance | Labeling modified in conformance with 21CFR 801 Subpart C, consistent with other legally marketed OTC wound management devices. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: The document mentions "animal models" for testing speed and efficacy. However, it does not specify the number of animals (sample size) used in these models.
- Data Provenance: Not explicitly stated, but clinical studies are generally prospective. Animal studies are experimental. The document does not specify the country of origin. This is a non-clinical study.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable to the type of device and study described. The evaluation in animal models would typically be based on physiological measurements (e.g., clotting time, blood loss) rather than expert interpretation of images or clinical assessments requiring a panel of human experts to establish "ground truth."
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is not a study involving human readers or adjudicators of a test set in the context of diagnostic AI.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a physical hemostatic device, not an AI or software-based diagnostic tool. No MRMC study was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical hemostatic device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" in the animal model studies would be objective physiological measures of hemostasis, such as:
- Time to hemostasis
- Volume of blood loss
- Presence/absence of exothermic reaction
These are direct measurements of the device's physical and physiological effect in a controlled animal setting, rather than subjective expert consensus or pathology in the context of diagnostic performance.
8. The sample size for the training set
Not applicable. This is not a machine learning or AI-based device that requires a training set.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
For Prescription Use: MedTrade Products CELOX Hemostatic Granules on Sheet is indicated for temporary external use to control moderate to severe bleeding.
For OTC (Over The Counter) Use: MedTrade Products CELOX Hemostatic Granules on Sheet is indicated for temporary external use to control bleeding of lacerations, minor cuts and abrasions.
The MedTrade Products CELOX Hemostatic Granules on Sheet is nearly identical to the legally marketed MedTrade Products CELOX Hemostatic Granules (K061079, cleared April 20, 2006) in product design composition and processing in that the same CELOX Hemostatic Granules are mechanically heat bonded on to a viscose sheet. The Sheet provides a controlled and accurate delivery System to ensure that the CELOX Hemostatic Granules are applied directly onto the wound site as quickly and efficiently as possible. On contact with blood CELOX Hemostatic Granules that are heat bonded on to the surface of the viscose sheet cause hemostasis.
This is a 510(k) premarket notification for the Medtrade Products CELOX Hemostatic Granules on Sheet. It does not contain information about the acceptance criteria or a study that proves the device meets those criteria in the format requested.
The document is a letter from the FDA to Medtrade Products Ltd. confirming the substantial equivalence of their device to legally marketed predicate devices. It outlines the regulatory classification, product codes, and general controls provisions of the Act. It also specifies the Indications for Use for both prescription and over-the-counter use of the device.
To address your request, specifically point 1-9, the document does not provide the details required, as it is a regulatory clearance letter and not a detailed study report. It states that the device is "nearly identical" to a previously cleared device (K061079) and that "the indications for use are similar to the HemCon ChitoFlex Surgical (K071519) and Scion Cardio-Vascular Clo-Sur Plus PAD (K032986)." This implies that the substantial equivalence was based on similarity to previously cleared devices rather than new performance studies detailed in this document.
Ask a specific question about this device
(55 days)
WoundStat™ wound dressing for OTC is indicated for temporary external use to control bleeding from minor cuts and abrasions of the skin.
WoundStat™ is a clay-based, granufar hemostatic agent that is poured on a moderate to severe wound and held in place until it adheres to the wound and hemostasis is achieved. The device is packaged in a foil package and is provided sterile.
This document is a 510(k) premarket notification for WoundStat™, a wound dressing intended for over-the-counter (OTC) use to control bleeding from minor cuts and abrasions. It's an administrative update to a previous 510(k) K071936 which originally cleared WoundStat™ for prescription (Rx) use.
Here's an analysis of the acceptance criteria and supporting study information:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly define specific performance acceptance criteria (e.g., stopping bleeding within X seconds for Y% of cases) for the OTC version of WoundStat™. Instead, the core of the submission relies on demonstrating substantial equivalence to its own previously cleared Rx version and other legally marketed OTC hemostatic agents.
The reported device performance, therefore, is primarily that:
- Mechanism of action is unchanged from the Rx version.
- It is demonstrated to be as safe and effective as the WoundStat™ for Rx use (K071936).
- It is similar to other legally marketed OTC wound management devices in terms of intended use and effectiveness for temporary external use to control bleeding from minor cuts and abrasions.
| Acceptance Criteria (Explicitly Stated in Document) | Reported Device Performance |
|---|---|
| Safety & Effectiveness (as compared to Rx version) | Deemed "as safe and effective" as WoundStat™ for Rx use (K071963). |
| Meeting OTC labeling requirements | Labeling revised "in conformance with 21 CFR 801 Subpart C" and "modeled after legally marketed OTC wound management devices". |
| Similar mechanism of action | Mechanism of action is "unchanged" from the Rx version. |
2. Sample Sizes Used for the Test Set and Data Provenance:
The document does not describe a new clinical study or test set specifically for this OTC 510(k) submission. Instead, it leverages the previous clearance of WoundStat™ for Rx use (K071936). The basis for "as safe and effective" likely stems from data or analyses provided in the K071936 submission. Without access to K071936, the specifics of its test set sample size and data provenance cannot be determined from this document.
3. Number of Experts Used to Establish Ground Truth and Qualifications:
This document does not mention the use of experts to establish a ground truth for a new test set for the OTC clearance. The assessment is based on comparison to existing cleared devices and the previous Rx clearance. Any expert involvement would have been part of the K071936 submission or the development of the predicate devices.
4. Adjudication Method:
Not applicable, as no new specific test set or clinical study requiring adjudication is described in this document for the OTC clearance.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
Not applicable. This document is for a medical device (wound dressing), not an AI/imaging diagnostic tool. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant.
6. Standalone (Algorithm Only) Performance Study:
Not applicable. This is a physical wound dressing, not an algorithm or software. Its performance is inherent to its physical and chemical properties when applied.
7. Type of Ground Truth Used:
For this OTC submission, the "ground truth" implicitly relies on:
- Previous FDA clearance of its own Rx version (K071936): This implies that the Rx version was found safe and effective based on its own ground truth (e.g., animal studies, clinical data on hemostasis).
- Established safety and effectiveness of predicate OTC devices: The comparison to other legally marketed OTC hemostatic agents (QuikClot Sport™, CELOX, BleedArrest™, BloodStop) suggests that the "ground truth" for these devices' performance in controlling minor bleeding is generally accepted by the FDA for OTC applications.
8. Sample Size for the Training Set:
Not applicable. This is a physical device, not a machine learning model, so there is no "training set."
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as there is no training set for a physical device.
Summary of Device Meeting Acceptance Criteria:
The device, WoundStat™ for OTC use, meets the "acceptance criteria" by demonstrating substantial equivalence to its own previously cleared prescription version (K071936) and other legally marketed OTC hemostatic agents. The study proving this essentially refers back to the data and rationale provided in the K071936 submission, along with a revised labeling conforming to OTC regulations (21 CFR 801 Subpart C). The core logic is that if the Rx version was safe and effective for its broader indications, and the mechanism of action remains the same, then the OTC version with more limited indications (minor cuts and abrasions) and appropriate labeling is also safe and effective. No new clinical trials or performance studies are described in this specific 510(k) update for the OTC clearance.
Ask a specific question about this device