510(k) Data Aggregation

    K Number
    K160703

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-06-10

    (88 days)

    Product Code
    Regulation Number
    882.5805
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    DEN070003, K061053, K083538, K130233, K133408

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NeuroStar TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.

    Device Description

    The NeuroStar TMS Therapy System is a computerized, electromechanical medical device that produces and delivers non-invasive, magnetic stimulation using brief duration (185 usec nominal) rapidly alternating, or pulsed, magnetic fields to induce electrical currents directed at spatially discrete regions of the cerebral cortex. This method of cortical stimulation by application of brief magnetic pulses to the head is known as Transcranial Magnetic Stimulation or TMS. NeuroStar TMS Therapy is a non-invasive tool for the stimulation of cortical neurons for the treatment of adult patients with Major Depressive Disorder (MDD) who have failed to receive satisfactory improvement from prior antidepressant medication as described under Intended Use. The NeuroStar System is used for patient treatment by prescription only under the supervision of a licensed physician. It can be used in both inpatient and outpatient settings including physician's offices and clinics, and hospitals.

    The NeuroStar TMS Therapy System is an integrated system consisting of a combination of hardware, software, accessories and consumable supplies. It includes a Mobile Console which houses the electronics, includes a software controlled graphical user interface, and gantry that supports the Treatment Coil. The ferromagnetic Treatment Coil delivers NeuroStar TMS Therapy®. The Head Support System provides accurate positioning of the Treatment Coil using a laser-guided alignment system. The SenStar® Connect is a non-sterile, multi-use consumable which is applied to the Treatment Coil, provides contact sensing to monitor contact of the treatment coil with the patient's head throughout a treatment session, quality control by monitoring the magnetic field level prior to patient treatment and provides surface field cancellation to reduce stimulation of the scalp. The TMS TrakStar™ practice data management system consists of a stand-alone computer and data management software that facilitates recording and retrieval of patient and treatment information and communication of data among multiple NeuroStar TMS Systems.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the NeuroStar TMS Therapy System, focusing on a labeling modification to allow for a shorter treatment time by varying inter-train intervals. This is not a typical device performance study with acceptance criteria for diagnostic accuracy, but rather an assessment of safety and effectiveness equivalence with a predicate device under modified usage parameters.

    Therefore, a table of acceptance criteria and reported device performance in the conventional sense (e.g., sensitivity, specificity) is not directly applicable. Instead, the study aims to demonstrate that a change in inter-train interval does not negatively impact the safety and efficacy of the device, which was already established for the predicate.

    Here's an analysis based on the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a 510(k) for a labeling change based on equivalence, there aren't explicit numeric acceptance criteria like those for diagnostic devices (e.g., minimum sensitivity of 90%). The acceptance criteria essentially revolve around demonstrating non-inferiority in safety and efficacy with the modified inter-train interval compared to the established performance of the predicate device.

    Acceptance Criterion (Implicit)Reported Device Performance (as demonstrated by the study)
    Safety: The risk of inadvertent seizure and the pattern of common adverse events with the modified inter-train interval must not systematically vary from established risks."This analysis showed that there is no evidence to indicate that the risk of inadvertent seizure or the pattern of common adverse events reported for TMS varies in a systematic manner across the range of inter-train intervals used within the reported treatment parameter sets." (Based on 61 studies, 67 active treatment arms, 2836 subjects)
    Efficacy: TMS efficacy with the modified inter-train interval must not be negatively impacted."This analysis showed that variations across the range of inter-train intervals used within the reported treatment parameter sets do not impact TMS efficacy." (Based on 44 studies, 50 active treatment arms where complete information on TMS treatment parameters and clinical outcome for primary efficacy measure was available. This also included four peer-reviewed meta-analyses.)
    Substantial Equivalence: The modified device (with labeling change) remains substantially equivalent to the predicate."The NeuroStar TMS Therapy System that is the subject of this premarket 510(k) notification is the same (substantially equivalent) device cleared by the FDA under DEN070003/K061053, K083538, K130233 and K133408." "New clinical data from a literature review... demonstrate that the revised inter-train interval range does not impact the safety and efficacy..."
    Physiological Parameters: The change remains within existing product specifications."This change remains within the existing product specification for inter-train interval (range 10-60 seconds for > 1 pulse per second)."

    2. Sample Size Used for the Test Set and Data Provenance

    The "test set" here refers to the data used for the clinical performance assessment.

    • Sample Size:

      • Effectiveness (Efficacy) Analysis: 44 studies (total of 50 active treatment arms) with complete information identified the TMS treatment parameter set and clinical outcome for the primary efficacy measure.
      • Safety Analysis: 61 studies (comprising 67 active treatment arms) among 2836 subjects provided adverse event data.
      • Within these, 11 clinical studies utilized the NeuroStar TMS Therapy System or progenitor iron-core devices, encompassing 1069 subjects, of whom 770 subjects received active TMS. The remaining studies used reference TMS devices.
      • The overall database search identified 79 studies and research outcomes on 3359 subjects, among whom 2162 subjects were exposed to active TMS treatment.
    • Data Provenance: The studies were identified via a "database search" to identify "human studies evaluating TMS and variations in inter-train interval." This indicates the data is likely retrospective, drawing from published clinical trials and literature. The country of origin of the data is not specified but would presumably be diverse given the nature of a database search for TMS studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not provided in the document. The study relies on clinical outcomes and reported adverse events from historical clinical trials, rather than forming a new "test set" with a panel of experts. The "ground truth" for efficacy would be the reported clinical outcomes (e.g., reduction in depression symptoms) in the included studies, and for safety, the reported adverse events. This ground truth was established by the original researchers and clinicians involved in those 79 studies.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. Given that the study is a retrospective review of existing literature and clinical trial data, it's unlikely that a new, external adjudication method was applied to the outcomes of those historical studies. The outcomes were presumably adjudicated by the methods employed within each individual study.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is a treatment system (Transcranial Magnetic Stimulator) and not an AI-powered diagnostic device that involves human "readers" interpreting medical cases. Therefore, the concept of human readers improving with or without AI assistance is not applicable here.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone (algorithm only) performance study was not done. The NeuroStar TMS Therapy System is a physical medical device for treatment, not an algorithm, and it operates with human supervision ("under the supervision of a licensed physician").

    7. The Type of Ground Truth Used

    The ground truth used in this study (a literature review) is based on:

    • Clinical Outcomes/Efficacy Data: Reported improvement in Major Depressive Disorder (MDD) symptoms from the primary efficacy measures in the analyzed clinical trials.
    • Adverse Event Data: Verbatim adverse event terms reported in the reference studies, including the incidence of seizure.

    This is essentially outcomes data and reported clinical findings from a large body of pre-existing scientific literature.

    8. The Sample Size for the Training Set

    This information is not applicable. The study described is a review of existing clinical data to support a labeling change for an already cleared device, not an AI or algorithm development study that would involve a distinct "training set" in the context of machine learning. The device itself was developed and cleared based on its own clinical trials in the past.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130233

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-04-30

    (90 days)

    Product Code
    Regulation Number
    882.5805
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K061053,K083538

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NeuroStar TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from one prior antidepressant medication at or above the minimal effective dose and duration in the current episode.

    Device Description

    The NeuroStar TMS Therapy System is a computerized, electromechanical medical device that produces and delivers non-invasive, magnetic stimulation using brief duration (~200 µsec) rapidly alternating, or pulsed, magnetic fields to induce electrical currents directed at spatially discrete regions of the cerebral cortex. This method of cortical stimulation by application of brief magnetic pulses to the head is known as Transcranial Magnetic Stimulation or TMS.
    The NeuroStar TMS System is a non-invasive tool for the stimulation of cortical neurons for the treatment of adult patients with Major Depressive Disorder (MDD) who have failed to receive satisfactory improvement from prior antidepressant medication as described under Intended Use. The NeuroStar TMS System is used for patient treatment by prescription only under the supervision of a licensed physician. It can be used in both inpatient and outpatient settings including physician's offices and clinics, and hospitals.
    The NeuroStar TMS System is an integrated system consisting of a combination of hardware, software, accessories and consumable supplies. It includes a Mobile Console which houses the electronics, includes a software controlled graphical user interface, and gantry that supports the Treatment Coil. The ferromagnetic Treatment Coil delivers the TMS TherapyTM. The Head Support System provides accurate positioning of the Treatment Coil using a laser-guided alignment system. A singleuse device, the SenStar® Treatment Link, which is applied to the Treatment Coil, provides contact sensing to monitor contact of the treatment coil with the patient's head throughout a treatment session, quality control by monitoring the magnetic field level prior to patient treatment, surface field cancellation to reduce stimulation of the scalp, and acts as a hygiene barrier from patient to patient. The TMS TrakStar (previously known as Practice Data Management System - PDMS) consists of a stand-alone computer and data management software that facilitates recording and retrieval of patient and treatment information and communication of data among multiple NeuroStar TMS Systems.

    AI/ML Overview

    This 510(k) summary (K130233) describes a labeling change to the NeuroStar TMS Therapy System, not a new device requiring performance data. Therefore, most of the requested information regarding acceptance criteria and a study proving device performance is not applicable to this submission.

    Here's a breakdown of the relevant information provided and why other sections are not applicable:

    1. A table of acceptance criteria and the reported device performance
      Not applicable. This 510(k) is for a labeling change to an already cleared device (K061053/K083538). No new performance data or acceptance criteria were required for this specific submission.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
      Not applicable. No new test set data was generated or required for this labeling change.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
      Not applicable. No ground truth establishment was needed for a labeling change.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
      Not applicable. No test set requiring adjudication was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
      Not applicable. This device is a transcranial magnetic stimulator for treatment, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
      Not applicable. This device is a physical medical device, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
      Not applicable.

    8. The sample size for the training set
      Not applicable. This 510(k) is not for an AI/machine learning device requiring a training set.

    9. How the ground truth for the training set was established
      Not applicable.

    Summary of the K130233 Submission:

    • Device: NeuroStar® TMS Therapy System
    • Submission Type: 510(k) for a labeling change.
    • Predicate Device: NeuroStar TMS Therapy® System [K061053/K083538]
    • Reason for 510(k): To update the prescribing statement in the labeling to comply with the FDA's Class II Special Controls Guidance for Repetitive Transcranial Magnetic Stimulation (rTMS) Devices (26 July 2011).
    • Conclusion: The FDA determined the device with the proposed labeling change is substantially equivalent to the predicate device, as no other changes were made to the device or its intended use. The substantial equivalence was based on similarities in indications for use, principles of operation, design for delivery of TMS, materials, shelf life, and packaging.