510(k) Data Aggregation

    K Number
    K200061

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-07-09

    (178 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    Multi-Point Thermal Sensor (K060390)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HEATfx Microwave Ablation System and Accessories are intended for the ablation (coagulation) of soft tissue in percutaneous, open surgical, and laparoscopic surgical settings.

    The HEATfx Microwave Ablation System is not intended for use in cardiac procedures.

    Device Description

    The HEATfx Microwave Ablation System is a mobile system intended to deliver microwave energy for thermal tissue ablation using a minimally invasive procedure. The HEATfx Microwave Ablation System is comprised of a console with optional cart, microwave ablation needles with cooling water manifold, optional thermal sensor, and related accessories.

    The HEATfx console is computer-controlled with a touch screen user interface that allows the user to control and monitor the procedure. The HEATfx System utilizes three 2450 MHz, 145W generators to deliver microwave energy to up to three HEATfx Needles.

    The HEATfx Microwave Ablation Needles are inserted into the target lesion under ultrasound or CT image guidance during laparoscopic procedures, or using direct visualization during surgical procedures, and operated using the console. The needles incorporate an antenna that transmits the microwave energy uniformly around the needle shaft. When connected to the HEATfx console, the console's internal pump and a disposable water manifold circulate sterile water through the needle for needle cooling and facilitating a uniform ablation zone.

    The optional Multi-Point Thermal Sensor allows the user to monitor tissue temperatures near critical structures during an ablation procedure.

    The system must be used with an irrigation baq of sterile water (not provided with the system).

    AI/ML Overview

    The provided text describes the HEATfx Microwave Ablation System's 510(k) submission to the FDA. It does not contain information about the acceptance criteria or a study proving the device meets those criteria in the context of an AI/machine learning device.

    The document focuses on the substantial equivalence of the HEATfx Microwave Ablation System to a predicate device (Emprint Ablation System) based on technological characteristics, performance data (system, mechanical, electrical, software, needle, thermal sensor, ex-vivo, in-vivo, biocompatibility, usability, packaging, sterility, pyrogenicity testing), and intended use.

    Therefore, for the specific questions related to AI/ML device acceptance criteria and studies (such as sample sizes for test/training sets, expert ground truth establishment, MRMC studies, or standalone algorithm performance), the provided text does not contain the necessary information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141485

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-07-02

    (27 days)

    Product Code
    Regulation Number
    878.4350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K060390, K021261, K113860, K110946

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Galil Medical Cryoablation Systems are intended for cryoablative destruction of tissue during surgical procedures. The cryoablation systems is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery, ENT, gynecology, proctology, and urology. The system is designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts.

    Galil Medical Cryoablation Systems have the following specific indications:
    • Urology Ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia (BPH)
    • Oncology Ablation of cancerous or malignant tissue and benign tumors, and palliative intervention
    • Dermatology Ablation or freezing of skin cancers and other cutaneous disorders
    Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, actinic and seborrheic keratosis, cavernous hemangiomas, peri-anal condylomata, and palliation of tumors of the skin
    • Gynecology Ablation of malignant neoplasia or benign dysplasia of the female genitalia
    • General surgery Palliation of tumors of the rectum, hemorrhoids, anal fissures, pilonidal cysts, recurrent cancerous lesions, ablation of breast fibroadenomas
    • ENT Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth
    • Thoracic surgery Ablation of arrhythmic cardiac tissue cancerous lesions
    • Proctology Ablation of benign or malignant growths of the anus or rectum, and hemorrhoids

    Device Description

    Galil Medical's predicate IceSphere needles and Galil's proposed IceSphere needles are sterile, single use, disposable devices used in conjunction with Gali! Medical's cleared cryoablation systems (SeedNet, SeedNet Gold; Presice; Visual-ICE) for cryoablative destruction of tissue during surgical procedures. The needles have a sharp cutting tip, a 1.5mm (17G) shaft, a color-coded handle, gas tubing, and a connector. Galil's predicate and proposed iceSphere needles are available in 2 handle configurations: straight and 90°. The needles contain shaft markings to aid the physician in needle placement.

    Galil is requesting clearance of the proposed vacuum insulated IceSphere needles. The vacuum insulated modification represents a one-for-one component exchange within the internal non-patient contacting lumen of the cryoablation needle shaft.

    AI/ML Overview

    The provided text is a 510(k) summary for the Galil Medical IceSphere Cryoablation Needle. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain information regarding traditional acceptance criteria, a test set with ground truth, or details of a study with human readers or standalone AI performance.

    Instead, the performance data discussed pertains to engineering verification testing to demonstrate that the modified device (with vacuum insulation) meets design specifications compared to the predicate non-insulated device.

    Therefore, many of the requested fields cannot be filled based on the provided document.

    Here's a breakdown of what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    The document states: "The insulated IceSphere successfully passed all the design properties testing in accordance with the established acceptance criteria compared to the non-insulated IceSphere predicate. Test results demonstrated that the proposed insulated IceSphere needles meet defined specifications and do not raise any new safety or effectiveness issues as compared to the predicate."

    This implies that the acceptance criteria were defined as the modified device performing equivalently to the predicate device in design properties testing. Specific numerical acceptance criteria or performance metrics are not provided.

    Acceptance Criteria (Implied)Reported Device Performance
    Equivalent performance to predicate non-insulated IceSphere Cryoablation Needles in design properties testing.The insulated IceSphere successfully passed all design properties testing in accordance with established acceptance criteria, demonstrating it meets defined specifications and raises no new safety or effectiveness issues compared to the predicate.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This document describes engineering verification tests on the device itself, not a clinical study on patient data. Therefore, typical "test set" and "data provenance" as applied to AI/diagnostic devices are not applicable and not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not a diagnostic device or a study requiring expert readers for ground truth. This information is not applicable and not provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not a diagnostic device or a study requiring adjudication. This information is not applicable and not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not a diagnostic device or a study involving human readers or AI assistance. This information is not applicable and not provided.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not an algorithm or AI device. This information is not applicable and not provided.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Given this is an engineering verification of a physical medical device (cryoablation needle), the "ground truth" would be objective engineering measurements and comparisons to design specifications and the predicate device's measured performance. No clinical "ground truth" like pathology or outcomes data is relevant for the reported testing. This information is not applicable in the traditional sense of diagnostic AI.

    8. The sample size for the training set

    This document does not describe the development of an AI algorithm with a training set. This information is not applicable and not provided.

    9. How the ground truth for the training set was established

    This document does not describe the development of an AI algorithm with a training set. This information is not applicable and not provided.