Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K111542

    Validate with FDA (Live)

    Date Cleared
    2011-10-20

    (140 days)

    Product Code
    Regulation Number
    870.2100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K052989

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CardioQ-EDM cardiac function and fluid status monitoring system is designed to provide clinicians with real-time information about a patient's left ventricular blood flow and key hemodynamic parameters. The CardioQ-EDM's beat-to-beat data on cardiovascular status can be used by the managing clinician to evaluate and optimize hemodynamic performance in anesthetized, sedated or conscious sedated patients in the operating room, intensive care unit, emergency room or ward.

    Device Description

    The CardioQ-EDM system employs esophageal Doppler techniques using 4 MHz continuous thoracic aorta, displaying this data as a maximum velocity curve, a descriting there is derived measurements. Thus, real-time information about verouty opesant in particular left ventricular flow, is displayed continuously. This is the same as for the predicate CardioQ K031706. CardioQ-EDM is an enhancement of the CardioQ K031706; obsolete The components have been replaced and an improved/modernized user interface created, by utilizing the full capabilities of the new components.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Deltex Medical CardioQ-EDM Esophageal Doppler Monitor, aiming to demonstrate substantial equivalence to a predicate device (Deltex Medical CardioQ - K031706). Since it's a 510(k) submission, the "acceptance criteria" are implied by comparing the new device's performance to the legally marketed predicate device, rather than explicit numerical thresholds typically seen in AI/ML performance studies. The study described is a comparative bench study.

    Here's the breakdown of the information requested based on the provided text, recognizing that it's a medical device submission focused on substantial equivalence rather than AI/ML performance:


    1. Table of Acceptance Criteria and Reported Device Performance

    For a 510(k) submission, acceptance criteria are generally to demonstrate that the new device (CardioQ-EDM) performs as safely and effectively as the predicate device (CardioQ K031706). The "reported device performance" is the conclusion of the bench testing that substantial equivalence was achieved. Specific numerical acceptance criteria are not explicitly stated for individual device parameters, but rather the overall conclusion of comparable performance across all listed characteristics.

    Feature/ParameterAcceptance Criteria (Implied - substantial equivalence to predicate)Reported Device Performance (CardioQ-EDM)
    Indications for UseSame as predicate (with minor textural updates)Same as predicate
    Patient PopulationSame as predicateSame as predicate
    Patient StatusSame as predicateSame as predicate
    Insertion RouteOral or NasalOral or Nasal (predicate was Oral, Nasal added later)
    ContraindicationsSame as predicateSame as predicate
    System DesignSame as predicateSame as predicate
    System ComponentsFunctionally equivalent/improved as predicateUpdated components, modernized user interface
    Mode of OperationSame as predicateSame as predicate
    Ultrasonic Clutter RejectionSame as predicateSame as predicate
    Spectral DisplaySame as predicateSame as predicate
    Velocity Spectrum Display Time Range (x-axis)Performance comparable to predicateFull screen: 4.3 seconds (Predicate: 3.6 sec) Split screen: 1.4 seconds (Predicate: 1.4 sec)
    Velocity Display Scales (y-axis)Performance comparable to predicate50, 100, 200 cm/s, 250 cm/s (Predicate: 50, 100, 200 cm/s)
    Doppler Audio ConfirmationSame as predicateSame as predicate
    DisplayPerformance comparable to predicateColor 10.4" TFT LCD screen (800 x 600 pixels) (Predicate: 640 x 480 pixels)
    Ranges of Directly Measured ParametersComparable to or improved from predicatePeak Velocity (PV) 10 - 250 cm/s (Predicate: 10 - 220 cm/s) Other parameters same range
    Ranges of Calculated ParametersSame as predicateSame as predicate
    Operating ModesSame as predicateSame as predicate
    Controls & User InterfacePerformance and functionality comparable to predicateSimilar controls with updated implementation
    Parameters DisplayedPerformance comparable to predicate, with potential for improvement/increase of displayed parametersEight of listed parameters displayed (Predicate: Six)
    Parameters Update RateSame as predicateSame as predicate
    Trend HistoryPerformance comparable to or improved from predicateUp to 240 hours/Unlimited (Predicate: Up to 48 hours)
    Trend Temporal ResolutionSame as predicateSame as predicate
    Acoustic OutputCompliance with NEMA UD 2Compliant with NEMA UD 2
    Electrical Safety & EMCCompliance with IEC 60601 seriesCompliant with IEC 60601 series
    Overall PerformanceSubstantially equivalent to predicateSubstantially equivalent performance achieved

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: This was a bench testing study, therefore there is no "patient test set" in the traditional sense. The comparison was between the physical device (CardioQ-EDM) and the predicate device (CardioQ K031706) within a laboratory/bench environment. The text does not specify a "sample size" of individual tests or measurements, but rather refers to "Comparative bench testing of the CardioQ-EDM... using a CardioQ..."
    • Data Provenance: The study was conducted by Deltex Medical Limited, based in the United Kingdom. The data is from bench testing, implicitly a prospective assessment of the new device's engineering and performance characteristics against the predicate's known specifications and performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. For this type of device (a diagnostic/monitoring tool) and study (bench testing for substantial equivalence), "expert ground truth" as understood in clinical evaluation or AI/ML performance is not relevant. The "ground truth" for the bench testing would be the engineering specifications and validated performance characteristics of the predicate device, against which the new device's measurements and functionalities were compared. This would be established by experts in device design, testing, and regulatory compliance, but not in the sense of clinical annotation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • None. This concept is not applicable to bench testing where objective measurements and engineering specifications are compared. Discrepancies would be resolved through engineering re-evaluation or re-testing, not through a consensus-based adjudication of clinical observations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC study was NOT done. This submission is for a medical device (Esophageal Doppler Monitor), not an AI/ML diagnostic tool. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not relevant. The device measures physiological parameters, and its performance is assessed against directly measured physical quantities or established specifications.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable in the context of AI/ML algorithms. The CardioQ-EDM is a standalone medical device that provides direct measurements and calculated parameters. Its "standalone performance" refers to its ability to accurately measure and compute these physiological metrics as designed, compared to the predicate device. It's not an AI algorithm that processes data to generate a finding that a human then interprets or acts upon.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • The "ground truth" for the bench testing was primarily the engineering specifications and established performance characteristics of the predicate device (CardioQ K031706), along with compliance with recognized consensus standards (IEC 60601 series for electrical safety/EMC, NEMA UD 2 for acoustic output). The goal was to show that the new device met or exceeded these, demonstrating substantial equivalence.

    8. The sample size for the training set

    • Not applicable. This submission is for a medical device that does not use machine learning algorithms, thus there is no "training set."

    9. How the ground truth for the training set was established

    • Not applicable. As there is no training set, there is no ground truth established for it.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1