510(k) Data Aggregation

    K Number
    K063567

    Validate with FDA (Live)

    Device Name
    EZ-MIO STERNAL
    Date Cleared
    2007-01-26

    (59 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K052195, K970380

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sternal IO access is indicated for adult patients when rapid fluid or pharmacological resuscitation is required in emergencies

    Device Description

    The EZ-MIO Sternal is very similar to the EZ-MIO, manual driver, previously cleared under 510(k) K052195. The two minor differences between the devices are in the driver and addition of a collar to the needle set. The predicate EZ-MIO consists of a proprietary pentagon shaft permanently attached to an ergonomically designed reusable driver. The EZ-MIO Sternal is single use and has a smaller driver permanently attached to the needle set. The second difference is the EZ-MIO Sternal has a permanently attached collar added to the same needle set used by the EZ-MIO. The collar is used to control depth of penetration. Both products utilize the same needle set and are designed to allow the user to manually insert a needle set consisting of a stylet and catheter through the cortex of a bone to a desired depth within the bone marrow to facilitate the infusion of desired fluids. After insertion of the needle set, both systems function identically by allowing the user to remove a stylet from the catheter by turning the stylet hub counter clockwise leaving a standard Luer lock catheter securely seated in the bone. The catheter Luer lock permits attachment of standard syringes and IV tubing for administration of drugs and fluids (not supplied). The needle set size used in the sternum is the identical needle set used by the predicate EZ-MIO (15G X 25mm). The EZ-MIO system is approved for use in the proximal tibia under 510(k) K052195. This submission requests an indication for use of the EZ-MIO Sternal in the sternum location already established by the predicate device Pyng F.A.S.T.1 (K970380).

    AI/ML Overview

    This Premarket Notification [510(k)] for the Vidacare EZ-MIO Sternal device does not contain the information requested in your prompt regarding acceptance criteria and a study proving the device meets those criteria.

    This document is a 510(k) summary submitted to the FDA for market clearance. It primarily focuses on demonstrating substantial equivalence to previously cleared devices. It does not include:

    • Acceptance criteria: Specific performance metrics with defined thresholds.
    • Study details: Information about a clinical or other performance study designed to demonstrate the device meets acceptance criteria.
    • Sample sizes (test, training): There's no mention of a test set or training set.
    • Data provenance, ground truth establishment, expert qualifications, adjudication methods: These are all related to performance studies, which are not described here.
    • MRMC comparative effectiveness study or standalone algorithm performance: These concepts are relevant to AI/algorithm-driven devices, which the EZ-MIO Sternal (an intraosseous infusion system) is not.

    The document states: "The EZ-MIO Sternal has the exact same technology as the EZ-MIO cleared under 510(k) K052195. There have been no major changes to the design or components and therefore the comparison of technological characteristics such as target population, needle design, technique, sterility, biocompatibility and where it is used is identical."

    This indicates that the submission relies on the established safety and effectiveness of the predicate devices rather than new performance studies specifically for the EZ-MIO Sternal. The primary purpose of this 510(k) is to extend the indications for use of an existing device concept to a new anatomical location (sternum) based on equivalence to another device already cleared for that location (PYNG Medical Corp. F.A.S.T. 1).

    Ask a Question

    Ask a specific question about this device

    K Number
    K062956

    Validate with FDA (Live)

    Date Cleared
    2006-12-05

    (67 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K052195, K032885

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EZ-MIO manual driver and EZ-IO power driver provide intraosseous access in the distal tibia of adults as an alternative to IV access during emergencies.

    Device Description

    The EZ-MIO manual driver, previously cleared under 510(k) K052195, consists of a proprietary pentagon shaft permanently attached to an ergonomically designed handle. The manual driver is designed to allow the user to manually insert a needle set consisting of a stylet and catheter through the cortex of the bone to a desired depth within the bone marrow to facilitate the infusion of desired fluids. After insertion of the needle set, the manual driver is detached from the needle set, leaving the stylet and cannula firmly seated in the bone. The stylet is then separated and removed from the catheter by turning the stylet hub counter clockwise leaving a standard Luer lock catheter securely seated in the bone. The catheter Luer lock permits attachment of standard syringes and IV tubing for administration of drugs and fluids (not supplied). The size needle that can be used in the distal tibia utilizing the MIO manual driver is identical to the predicate 15q X 25mm. The EZ-MIO system is approved for use in the proximal tibia under 510(k) K052195. This submission extends the indication for use of the manual device to include the distal tibia in adults utilizing the same technique and device previously cleared for the proximal tibia via 510(k) K052195.

    The EZ-IO powered driver, previously cleared under 510(k) K032885 for adult use, consists of a reusable battery powered driver connected to a single use disposable intraosseous (IQ) needle assembly. Upon activation, the drill supplies power to the needle set in order to penetrate through the cortex of the bone to a desired depth within the bone marrow. After insertion of the needle set, the power driver is detached from the needle set, leaving the stylet and cannula firmly seated in the bone. The stylet is then separated and removed from the catheter by turning the stylet hub counter clockwise leaving a standard Luer lock catheter securely seated in the bone. The catheter Luer lock permits attachment of standard syringes and IV tubing for administration of drugs and fluids (not supplied). The size needle that can be used with the EZ-IO powered driver is identical to the predicate 15g X 25mm. The EZ-IO powered intraosseous system is cleared for use in the proximal tibia under 510(k) K032885. This submission extends the indication for use of the powered device to include the distal tibia in adults utilizing the same technique and device previously cleared for the proximal tibia via 510(k) K032885.

    AI/ML Overview

    This document is a 510(k) summary for the EZ-MIO and EZ-IO Intraosseous Infusion Systems, seeking to extend their indications for use to include the distal tibia in adults. The submission primarily relies on equivalence to previously cleared devices (K052195 and K032885) and does not describe a new study with specific acceptance criteria or performance metrics directly comparable to a table.

    Therefore, many of the requested details about acceptance criteria, specific device performance, sample sizes, expert qualifications, and ground truth establishment for a new study validating the extended indication are not present in this summary. The submission argues that since there are "no changes to the design or components," the devices are substantially equivalent to their predicates for the new anatomical site based on existing technology.

    However, based on the provided text, here's what can be extracted and inferred:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a quantitative manner or present a new study with specific performance metrics for the distal tibia insertion that would traditionally be reported in such a table. The argument for expanded indication is based on the devices being identical to their cleared predicates and the technique being similar.

    Inferred Acceptance Criteria:

    • Substantial Equivalence: The primary "acceptance criterion" for this 510(k) is demonstrating substantial equivalence to the predicate devices (EZ-MIO, K052195; EZ-IO, K032885) for the new indication (distal tibia in adults). This implies that the device performs similarly in terms of safety and effectiveness as demonstrated by the predicate devices for their cleared indications.
    • No Design Changes: The device must have no changes to its design or components compared to the cleared predicate devices.
    • Same Technique: The insertion technique for the distal tibia must be the same as the technique cleared for the proximal tibia.

    Reported Device Performance (Inferred from Substantial Equivalence):
    Since no new performance data for the distal tibia is presented, the "reported device performance" is essentially that the EZ-MIO and EZ-IO systems, when used in the distal tibia with the specified 15g x 25mm needle, are expected to perform comparably to their performance in the proximal tibia, which was deemed safe and effective in their original 510(k) clearances. This includes:

    • Successfully facilitating the infusion of desired fluids.
    • Firmly seating the stylet and cannula in the bone.
    • Allowing separation and removal of the stylet, leaving a standard Luer lock catheter.

    2. Sample size used for the test set and the data provenance

    The document does not describe a new "test set" or a new clinical study to evaluate performance in the distal tibia. The submission is based on the technological characteristics being identical to the predicate devices. Therefore, details like sample size, country of origin, or retrospective/prospective nature are not applicable to this specific submission for an expanded indication. The "data provenance" would refer to the original studies conducted for the predicate devices, which are not detailed here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as no new test set or ground truth establishment for an expanded indication study is described in this summary.

    4. Adjudication method for the test set

    Not applicable, as no new test set requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device for intraosseous infusion, not an AI-assisted diagnostic or imaging device, and no MRMC study is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is a medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For this particular 510(k) expanding the indication, the "ground truth" is implied to be the established safety and effectiveness of the predicate devices for their original indications, and the argument that the same principles apply due to identical device characteristics and technique. No new empirical "ground truth" for the distal tibia is presented.

    8. The sample size for the training set

    Not applicable, as no new training set or new study is described.

    9. How the ground truth for the training set was established

    Not applicable, as no new training set or new study is described. The relevant "ground truth" in this submission pertains to the established performance of the predicate devices.