Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K152339

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-06-24

    (310 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    Artsana InsuPen® Insulin Pen Needles, K051783, BD Pen Needles, K051899, BD 32G x 4mm Pen Needle, K100005

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unifine Pentips range of pen needles are intended for use with pen injector devices for the subcutaneous injection of drugs, including insulin.

    Device Description

    Unifine® Pentips® and Unifine® Pentips® Plus are sterile, non-toxic, non-pyrogenic, single use pen needles for the subcutaneous injection of drugs from pen injectors. The pen needles are used by consumers, caregivers, and healthcare professionals. They are available in a variety of lengths (4mm, 5mm, 6mm, 8mm and12mm) and gauges (29G, 31G and 32G).

    The pen needle assembly consists of a double-ended cannula, a needle hub, a needle shield, primary container and sterility seal that covers the assembly inside the primary container.

    The hub has internal threads, which allows the pen needle to be screwed onto a pen injector device. The Non Patient (NP) end of the cannula (inside the hub) penetrates the rubber septum of the cartridge in the pen injector device. The Patient end and NP end of the cannula are lubricated using a silicone based lubricant to aid in injection and penetration into the rubber septum. The inner needle shield is injection molded and assembled over the Patient end of the needle to protect the needle from damage and accidental needle-sticks. The pen needle assembly is then inserted into an iniection molded outer container and sealed with a protective peel-away label to ensure a sterility barrier and tamper evidence. The outer container can be used to remove the hub with cannula from the pen injector device after the injection is completed. The peel-away label is pre-printed with information, which includes the lot number, expiry date, and size (length) and gauge of the needle. The individual needle assemblies are then packaged in bags and/or cartons, and placed into shippers with appropriate labeling. The shipper cases are then placed on pallets for sterilization.

    Unifine Pentips Plus pen needles differ from Unifine Pentips pen needles because the primary container consists of two 'chamber contains a new, unused pen needle and the second chamber is empty. The empty chamber serves as a built-in pen needle remover, and may prevent accidental needlesticks. Users can insert their used needle into the chamber, then unscrew from the injection pen. The primary container is designed to hold the used pen needle until it can be disposed of in a suitable sharps container. The actual pen needle is identical between Unifine Pentips and Unifine Pentips Plus in both the physical appearance and functionality.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for medical devices, specifically pen needles. It is not a study proving an AI/ML medical device meets acceptance criteria. Therefore, most of the requested information regarding AI/ML device performance, ground truth establishment, sample sizes for training/test sets, expert adjudication, and MRMC studies, cannot be extracted from this document.

    This document focuses on demonstrating substantial equivalence of the Unifine® Pentips® and Unifine® Pentips® Plus pen needles to legally marketed predicate devices, primarily K973899 (Unifine® Pentips®). This is a regulatory pathway for Class II medical devices in the US, where the manufacturer needs to show that their new device is as safe and effective as an already cleared device.

    Here's what can be extracted based on the provided text, and where the requested information is not applicable or not present:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for these pen needles are primarily defined by adherence to ISO standards and internal performance tests. The document states that the devices meet these standards/criteria.

    Test / CharacteristicAcceptance Criteria (Standard / Description)Reported Device Performance
    Dimensions for needlesISO 11608-2:2012 Section 4.2.2Meets standard
    Needle pointsISO 11608-2:2012 Section 4.5Meets standard
    Freedom from defectsISO 11608-2:2012 Section 4.6Meets standard
    LubricationISO 11608-2:2012 Section 4.7Meets standard
    Dislocation of measuring point at patient endISO 11608-2:2012 Section 4.8Meets standard
    Bond between hub and needle tubeISO 11608-2:2012 Section 4.4Meets standard
    Ease of assembly/disassemblyISO 11608-2:2012 Section 4.10Meets standard
    SterilizationISO 11608-2:2012 Section 4.11, ISO 11137-2:2007 (UK), ISO 11135-1:2006 (Italy)Meets standard
    Flow rateISO 11608-2:2012 Section 4.3Meets standard
    Needle dose accuracyISO 11608-2:2012 Section 4.9, 11.4.2Meets standard
    Needle hub torque removalISO 11608-2:2012 Section 4.9, 11.4.3Meets standard
    Force required to remove pen needle from primary containerEstablished internal acceptance criteria based on device specificationsMeets acceptance criteria
    Force required to fit pen needle into primary container (UP)Established internal acceptance criteria based on device specificationsMeets acceptance criteria
    Force to remove needle shield from pen needleEstablished internal acceptance criteria based on device specificationsMeets acceptance criteria

    Note: The document states "The devices comply with the acceptance criteria established based on the specifications of the devices" for internal tests, but does not provide the specific numerical criteria for these tests.


    Regarding the study that proves the device meets acceptance criteria (for an AI/ML context):

    This document describes a regulatory submission for a physical medical device (pen needles), not an AI/ML software device. Therefore, the concepts of "test set," "training set," "ground truth," "experts for ground truth," "adjudication," and "MRMC studies" as they apply to AI/ML performance evaluation are not relevant here. The "study" described is a series of non-clinical performance (bench) tests and biocompatibility tests to demonstrate substantial equivalence to existing devices.

    Here's why the specific questions are not applicable or cannot be answered from this document:

    2. Sample sized used for the test set and the data provenance:

    • Not Applicable in AI/ML context: This is a physical device. "Test set" refers to samples tested during bench testing.
    • Sample Size: The document does not specify the number of pen needles tested for each performance test. It only states that the tests were performed and results "meet standard" or "meet acceptance criteria."
    • Data Provenance: Not applicable in the AI/ML sense. Data is from laboratory bench testing of manufactured pen needles. The manufacturing locations mentioned are the UK (Owen Mumford) and Italy (Artsana S.p.A.).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable: There is no "ground truth" in the AI/ML sense (e.g., expert-annotated images) for a physical device like a pen needle. The "ground truth" is defined by the objective physical and mechanical properties measured against established ISO standards and internal specifications. Bench testing is typically performed by trained technicians.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable: No expert adjudication process as would be used for clinical image interpretation or AI model validation. Bench tests have objective, measurable outcomes.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable: This type of study (MRMC, human readers, AI assistance) is for clinical decision support or diagnostic AI devices. This document is for a mechanical device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable: This is not an algorithm or software device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not Applicable in AI/ML context: The "ground truth" for these pen needles is defined by objective physical and mechanical measurements against established ISO standards (e.g., dimensions, flow rate, bond strength) and internal engineering specifications. Biocompatibility is assessed through standardized biological tests.

    8. The sample size for the training set:

    • Not Applicable: This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established:

    • Not Applicable: As above, no training set or AI/ML ground truth. The "ground truth" for manufacturing and quality control of pen needles is based on engineering specifications and adherence to international standards for medical devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K121632

    Validate with FDA (Live)

    Device Name
    COMFORT EZ
    Date Cleared
    2012-08-10

    (67 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K080904,K100005,K051899

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These disposable sterile insulin pen needles are intended for subcutaneous injection of insulin in the treatment of diabetes.

    Device Description

    The Pen Needle consists of a sterile cap, needle cap needle hub, which can be fixed with needle and protected by blister paper. The sterile cap functions to maintain the sterility of the needle because sterile cap covers the needle hub and needle cap with blister paper sealed on the opening hole of sterile cap. The needle hub can be connected to the pen. The needle cap cover is intended to provide physical protection to the needle tube. They are supplied with a sterile fluid path, (EO), nontoxic, and non pyrogenic, for single use only, disposable. The devices operate on the principles of common piston syringes.

    AI/ML Overview

    This document describes a 510(k) submission for the Comfort EZ Pen Needle, a Class II medical device. The submission primarily focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a study of AI-based diagnostic performance. Therefore, many of the requested elements for describing a study proving a device meets acceptance criteria related to AI or diagnostic accuracy are not applicable.

    Based on the provided text, the device in question is a Comfort EZ Pen Needle, which is a medical device for drug delivery (insulin) and not a diagnostic AI system. Therefore, the typical acceptance criteria and study designs relevant to AI/diagnostic performance (such as sensitivity, specificity, MRMC studies, ground truth establishment by experts, etc.) are not applicable to this document.

    The "study" described in this 510(k) summary is a set of bench tests and demonstrations of substantial equivalence to existing predicate devices, as is standard for many medical devices, particularly those for which equivalence can be shown based on design and performance characteristics rather than complex diagnostic interpretation.

    Here's an attempt to address the points based on the provided text, noting where the requested information is not relevant to a pen needle:

    1. A table of acceptance criteria and the reported device performance:

    Since this is a non-AI, non-diagnostic device, the acceptance criteria are based on physical and functional properties, and the performance is reported against these.

    Acceptance CriteriaReported Device Performance
    BiocompatibilityMet (implied by "device is as safe, as effective, and performs in a substantially equivalent manner to the previous predicate devices")
    Compatibility with ISO "Type A" pensMet ("They are compatible with ISO "Type A" standard pens.")
    Mechanical testingMet (implied by "Bench tests were performed... The tests demonstrated that the device is as safe, as effective, and performs in a substantially equivalent manner...")
    Sterility testing (including EO residues)Met ("These syringes are EO sterilized. The sterility assurance level is 10^-6.")
    Non-toxic and non-pyrogenicMet ("supplied with a sterile fluid path, (EO), nontoxic, and non pyrogenic")
    Single-use onlyDesigned accordingly ("for single use only, disposable")
    Substantial Equivalence to Predicate DevicesDemonstrated and accepted by FDA (K121632 approved based on equivalence to Feel Fine (K080904), K100005, and K051899 BD Pen Needles)
    Labeling ComparabilityEvaluated and found comparable to predicate (BD) labeling.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated for each bench test. Bench tests typically involve a specific number of units to ensure statistical significance for the particular test being conducted (e.g., a certain number of needles for mechanical force testing, or samples for sterility testing). The document does not provide these specifics.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). This is standard for internal product development and testing of a device rather than a clinical study with patient data. The tests would be performed in a controlled laboratory environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This device does not involve diagnostic interpretation or "ground truth" established by human experts in the way an AI diagnostic device would. Bench tests follow established laboratory protocols and standards (e.g., ISO standards for pen compatibility, AAMI standards for sterility). Compliance with these standards serves as the "ground truth" for device performance.

    4. Adjudication method for the test set:

    • Not Applicable. There's no human interpretation or subjective assessment that would require adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI or diagnostic device, so an MRMC study is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is not an algorithm or AI device. The "performance" described is the physical and functional performance of the pen needle itself.

    7. The type of ground truth used:

    • The "ground truth" for this device's performance is established by international standards (e.g., ISO "Type A" for pen compatibility), manufacturing specifications, and industry best practices for medical device safety and efficacy (e.g., biocompatibility standards, sterility assurance levels). It's not based on expert consensus, pathology, or outcomes data in the way a diagnostic device's ground truth would be.

    8. The sample size for the training set:

    • Not Applicable. This device does not involve machine learning or a "training set."

    9. How the ground truth for the training set was established:

    • Not Applicable. There is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K093109

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-06-18

    (260 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K062500, K090111, K050106, K051899

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NovoTwist® needles are intended for use with pen injector devices for the subcutaneous injection of insulin, liraglutide and somatropin.

    Device Description

    NovoTwist® needle is designed for single use in conjunction with Novo Nordisk injection delivery devices built with bayonet coupling. Prior to giving an injection, the protective tab is removed from the outer needle cap of the single-use disposable needle. With the disposable needle remaining in the outer needle cap, it is then pressed onto the injection delivery device with a bayonet coupling and twisted a quarter of a revolution until tightness and a confirmatory click is heard. Then the outer and inner caps of the needle are removed. Injection is performed by following the procedure described in the User Manual provided with the pen injection device and instructions from your health care professional. After the injection, the needle is removed from the skin. The needle is detached from the injection device and disposed of in accordance with national/local regulations. For each subsequent injection, another disposable needle must be used. Function checks for the delivery device can be performed with the NovoTwist® needles by using the needle cap. The NovoTwist® needle is used in exactly the same manner as the NovoFine® and the BD Pen needle except it is built with an improved attachment mechanism. NovoTwist® needle is attached to the delivery device via a bayonet thread. This allows the user to press down and twist the needle a quarter of a revolution instead of tightening through several revolutions as with the NovoFine® needle, BD pen needle, and many other commercially marketed hypodermic needles that operate on the same principles. This improved mechanism is intended to reduce the attachment time. In addition, users can hear a confirmatory click when the NovoTwist® needle is attached correctly to the device. This user friendly feature is designed to ensure the needle is attached to the device correctly before beginning the injection.

    AI/ML Overview

    NovoTwist® Needle Acceptance Criteria and Study Details

    1. Acceptance Criteria and Reported Device Performance

    The provided document describes a 510(k) summary for the NovoTwist® needle, claiming substantial equivalence to existing predicate devices. It references compliance with several ISO standards and outlines specific non-clinical tests. However, it does not explicitly state specific pass/fail acceptance criteria (e.g., "bubble leak rate must be less than X%") nor does it present quantitative results for how the device performed against these criteria. Instead, it generally states that the tests support the device meeting "specified requirements" and being "safely used."

    Therefore, the table below reflects what is implied as acceptance for specific tests and the general statement of performance:

    Acceptance Criterion (Implied)Reported Device Performance
    Compliance with ISO 11608-2:2000 (Pen-injectors for medical use - Part 2: Needles - Requirements and test methods)"NovoTwist® 30G and 32G Tip needles have been subjected to many non-clinical tests throughout the development by Novo Nordisk R&D, and these tests comply with ISO 11608-2:2000..."
    Performance and functionality evaluation (prototype tests)"several prototype tests with NovoTwist® needles were conducted to evaluate the performance and functionality of the needles." (Implied successful evaluation)
    Bubble leak test for sterility (per ISO 11607-1:2006)Conducted as part of final design verification test. "...bubble leak test to ensure sterility..." (Implied successful compliance with standard)
    Freedom from defects (per ISO 7864:1993)Conducted as part of final design verification test for: defect needle point, patient needle point, silicone on patient-end needle, bond between hub and needle tube, visual control of quality of sealing. "...Freedom from defects in accordance with ISO 7864:1993..." (Implied successful compliance with standard)
    Biocompatibility (per ISO 10993-1)Reference made to prior successful biocompatibility tests for NovoFine needles (predicate device), as the cannula and silicone oil (components in contact with human tissue) are identical. "...we reference the biocompatibility tests to those previously conducted for the NovoFine needles in accordance with ISO 10993-1... which have been cleared by the Agency." (Implied successful compliance through substantial equivalence of materials)
    Shelf life"shelf life test has been performed on final released NovoTwist® 30G and 32G Tip needles in accordance with ISO 11607-1:2006..." (Implied successful determination of shelf life according to standard)
    Handling test and Usability test"handling test and usability test are performed..." (Implied successful performance indicating safe and effective use; specifics not detailed.)
    Overall Design Verification"The final design verification test consists of 34 mechanical and visual tests... In summary, the final design verification test supports that Novo Twist needles meet the specified requirements and that the needles can be safely used in various real-life situations of handling and use according to Instructions for Use." (Implied all 34 tests met their respective acceptance criteria.)

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size for Test Set: The document states that the final design verification test involved "34 mechanical and visual tests" but does not specify the exact number of NovoTwist® needles tested for each of these tests. It only mentions that "The needle samples tested are manufactured, sterilized, packaged and labelled on validated equipment and QA-released in accordance with cGMP..."
    • Data Provenance: The document implies the tests were conducted by "Novo Nordisk R&D" and that "The needle samples tested are manufactured...in accordance with cGMP for Medical Devices (ISO 13485:2003 Medical devices - Quality management systems Requirements for regulatory purposes)." This suggests the data is prospective and likely originated from internal company testing, presumably in facilities compliant with quality management systems. The country of origin for the data is not explicitly stated, but Novo Nordisk is a Danish multinational company.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable as the described tests are non-clinical (mechanical, visual, sterility, biocompatibility, shelf-life, handling, and usability) and do not involve human assessment of medical images or patient outcomes that would require expert ground truth establishment in the traditional sense (e.g., by radiologists). The "ground truth" for these tests would be defined by the technical specifications and standards (e.g., ISO 11608-2) the device is being tested against.

    4. Adjudication Method for the Test Set

    This information is not applicable. The tests described are objective, physical, or chemical measurements/inspections against predefined specifications or standards. They do not involve subjective assessments requiring an adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical tests are required." The submission relies on non-clinical testing and substantial equivalence to predicate devices.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable. The device, NovoTwist® needle, is a physical medical device (hypodermic needle), not a software algorithm or AI-powered system where a "standalone" algorithm-only performance would be relevant.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical tests is based on established international standards and internal specifications. For example:

    • ISO 11608-2:2000 (Pen-injectors for medical use - Part 2: Needles - Requirements and test methods)
    • ISO 11607-1:2006 (Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems)
    • ISO 7864:1993 (Sterile hypodermic needles for single use)
    • ISO 10993-1 (Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management system)
    • Internal design specifications for mechanical, visual, and functional performance.

    8. The Sample Size for the Training Set

    This information is not applicable. As a physical medical device (hypodermic needle), there is no "training set" in the context of machine learning or artificial intelligence. The development process involves prototype testing and design verification, not algorithm training.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1