510(k) Data Aggregation

    K Number
    K250286

    Validate with FDA (Live)

    Date Cleared
    2025-07-03

    (153 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K051864

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rapid2 Magnetic Stimulators (Magstim Rapid2, Magstim Super Rapid2, Magstim Super Rapid2 Plus1) are intended to stimulate peripheral nerves for relief of chronic intractable, post-traumatic and post-surgical pain for patients 18 years or older.

    Device Description

    The Magstim Rapid2, Magstim Super Rapid2, Magstim Super Rapid2 Plus1 (herein collectively referred to as "Rapid2 Magnetic Stimulators") are computerized, electromechanical medical devices that provide brief and focused magnetic pulses in order to non-invasively stimulate peripheral nerves for relief of chronic intractable, post-traumatic and post-surgical pain for patients 18 years or older. The subject device is intended to be used in hospitals and clinics such as pain management clinics.

    Rapid2 Magnetic Stimulators are integrated systems consisting of a combination of hardware, software, and accessories. Rapid2 Magnetic Stimulators are offered in multiple configurations:

    • Rapid2
    • Super Rapid2
    • Super Rapid2 Plus1

    All three configurations have identical intended use/indications for use, common specifications, equivalent performance characteristics and equivalent composition to each other. Specifically, Rapid2 and Super Rapid2 have received prior clearance under K051864 for Peripheral Nerve Stimulation (Product Code: GWF, Regulation 21 CFR 882.1870). All Rapid2 Magnetic Stimulators are made up of components that have received prior clearance under K051864 (e.g., the 3190-00, 3192-00 and 3193-00 coils) and components which have received prior clearance under K051864 but have received modifications due to aspects like obsolescence (Mainframe, Power Supply etc.).

    All Rapid2 Magnetic Stimulators are composed from the following main components:

    • Stimulating Unit & Power Supply
    • User Interface
    • Stimulating Coil
    • System and Stimulating Coil Cart and Holding Arm

    Rapid2 Magnetic Stimulators include temperature monitoring via two independent temperature sensors to ensure surfaces of the coils do not reach unacceptable levels. The cut-off is set to act at 40°C at which point the system will automatically be disabled. Over-temperature conditions are also communicated on the User Interface (UI) via a temperature gauge and alarm system. Rapid2 Magnetic Stimulators also includes the 3910-00 air-cooled coil to further mitigate any temperature conditions. The 3910-00 air-cooled coil comes with all 3 configurations (Rapid2, Super Rapid2 and Super Rapid2 Plus1) as standard.

    AI/ML Overview

    The provided document is an FDA 510(k) Clearance Letter for the Rapid2 Magnetic Stimulators. It does NOT contain information about a study proving the device meets acceptance criteria related to its performance in pain relief. Instead, it focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing of technical characteristics, safety standards compliance, and physical properties.

    The document does not describe an AI/ML-driven device, nor does it present results from a clinical study with patients or human readers using AI. The acceptance criteria and performance metrics described are related to physical and electrical characteristics of the magnetic stimulator, not diagnostic accuracy or clinical effectiveness in a traditional sense of "performance" as one might expect for an AI diagnostic device.

    Therefore, many of the requested items (e.g., sample size for test/training sets, data provenance, number of experts for ground truth, MRMC study, standalone performance, ground truth type) cannot be answered based on the provided text, as they pertain to clinical or AI/ML performance evaluation, which is not the subject of this 510(k) summary.

    However, I can extract information related to the technical acceptance criteria and how they align with the device's measured performance as described in the summary:


    Acceptance Criteria and Device Performance (Based on Technical and Safety Equivalence)

    The acceptance criteria for the Rapid2 Magnetic Stimulators are primarily based on demonstrating substantial equivalence to the predicate device (MagVenture Pain Therapy) in terms of technical characteristics, safety, and effectiveness for the stated indications for use. The "performance" reported is adherence to these characteristics and safety standards.

    Acceptance Criteria Category/CharacteristicSubject Device Performance (Rapid2 Magnetic Stimulators)Predicate Device Performance (MagVenture Pain Therapy)Evaluation / Proof of Meeting Criteria
    Indications for UseStimulate peripheral nerves for relief of chronic intractable, post-traumatic and post-surgical pain for patients 18 years or older.Stimulate peripheral nerves for relief of chronic intractable, post-traumatic and post-surgical pain for patients 18 years or older.Identical. Meets criteria by having the same intended use.
    Anatomical SitesAny area, such as hand, arm, waist, buttock, thigh, calf, back and lower back etc.Any area, such as hand, arm, waist, buttock, thigh, calf, back and lower back etc.Identical.
    Treatment FacilitiesHospitals & ClinicsHospitals & ClinicsIdentical.
    Treatment Time13 minutes per session (800 seconds)13 minutes per session (800 seconds)Identical.
    Pulse FrequencyRapid2: 0.1 – 50 Hz (pps); Super Rapid2 and Super Rapid2 Plus1: 0.1 – 100 Hz (pps)MagPro R30 & MagPro R30 with MagOption: 0.1 – 30 Hz (pps); MagPro X100 & MagPro X100 with MagOption: 0.1 – 100 Hz (pps)Similar range. Subject device's range covers or extends slightly beyond predicate, but the recommended protocol (0.5Hz) is well within both.
    Pulse Amplitude0 – 100%0 – 100%Identical.
    On-cycle duty period2-800 Seconds (0.5 Hz and up to 400 pulses)2-800 Seconds (0.5 Hz and up to 400 pulses)Identical.
    Off-cycle rest periodN/AN/AIdentical.
    Maximum Repetition RateRapid2: 50Hz; Super Rapid2: 100Hz; Super Rapid2 Plus1: 100HzMagPro R30 & MagPro R30 with MagOption: 30 pulses per second; MagPro X100 & MagPro X100 with MagOption: 100 pulses per secondUpper limit identical compared to predicate. Substantial equivalence demonstrated despite differences in how maximum output is achieved (explained in "SE Note 1").
    Pulse WidthBiphasic (300-425 µs)Biphasic (280-320 µs)Similar range. Differences deemed not to raise new safety/effectiveness questions due to compensating factors (see "SE Note 2").
    Pulse ModeStandardStandardSame.
    Temperature ControlAutomatic disable at 40°C; includes air-cooled coil; UI communication of over-temperature.Automatic disable at 43°C.Comparable/Better. Subject device has a lower cutoff and additional cooling/reporting features.
    Peak Magnetic Field at Coil Surface1.0-1.5T1.15-2.6TSubstantially equivalent. Subject device is a subset of the predicate's range. Differences explained in "SE Note 2" as not raising new safety/effectiveness concerns.
    Peak Magnetic Field Gradient (dB/dt) at 20mm from Coil Center9-12kT/s9-24kT/sSubstantially equivalent. Subject device is a subset of the predicate's range. Differences explained in "SE Note 2" as not raising new safety/effectiveness concerns.
    WaveformBiphasic, Biphasic BurstBiphasic, Monophasic, Biphasic Burst, Halfsine (combinations vary by predicate configuration)Substantially equivalent. Subject device's waveform is within the range available in the predicate.
    Software/Firmware ControlYesYesIdentical. Verified per IEC 62304.
    Power Supply TypePower Supply via dedicated power supply modules each using a separate input mains line cord.Power Supply via Isolation Transformer.Similar.
    Power Consumption230/240V Systems – 3000VA peak per input; 115V Systems – 2300VA peak per inputMaximum 2700VASimilar.
    User InterfaceLCD Capacitive TouchscreenLED DisplaySimilar. A difference in display technology, but performs the same function.
    Housing Material ConstructionStimulator: PUR, Stainless/Galvanized Steel; Coils: PC, PURStimulator: Aluminum, Aluzinc; Coils: PVC, ABS, PA, POMSimilar. Different specific materials but serve the same function.
    Applied Parts (Coils)Various, including previously cleared (K051864, K080499, K130403) and new coils (4150-00, 4170-00, 4189-00, 4190-00, 4510-00).Various, all previously cleared.Substantially equivalent coil range. New coils are evaluated for safety and function to be equivalent (see "SE Note 2").
    Applied Part AreaButterfly Coils: 152mm – 191mm; Circular Coils: 124.5mmButterfly Coils: 150mm; Circular Coils: 110-126mm; Special Coils: 160x80 mmSubstantially equivalent. (see "SE Note 2").
    SterilizationNon-sterile when used.Non-sterile when used.Identical.
    Electrical SafetyComplies with IEC 60601-1 Ed. 3.2Complies with IEC 60601-1 Ed. 3.1Meets/Exceeds. Complies with a newer edition of the standard.
    Mechanical SafetyComplies with IEC 60601-1 Ed. 3.2Complies with IEC 60601-1 Ed. 3.1Meets/Exceeds. Complies with a newer edition of the standard.
    Thermal SafetyComplies with IEC 60601-1 Ed. 3.2Complies with IEC 60601-1 Ed. 3.1Meets/Exceeds. Complies with a newer edition of the standard.
    Radiation SafetyNo radiation generated.No radiation generated.Same.
    BiocompatibilityComplies with ISO 10993-1, -5, -10; materials tested for Skin Irritation, Cytotoxicity, Skin Sensitization.Complies with ISO 10993.Same/Exceeds. Detailed compliance with relevant parts of the standard.
    Standards ComplianceISO 13485 (company); IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 62366-1 (device).EN ISO 13485 (company).Same/Equivalent. Device-specific standards compliance indicated.
    AcousticsTested per 60601-1 type testing and in-house, demonstrating substantially equivalent acoustic output. Labeling requires earplugs (30dB noise reduction).(Not specified beyond "similar")Comparable. Demonstrated and mitigated with user instructions.
    E-Field Decay & Linearity of OutputPerformance data showed "very similar" E-Field decay, linearity, and electric/magnetic field spatial distributions.(Not explicitly detailed, but implied to be baseline for comparison)Comparable. Results demonstrate equivalent effects at 0-2cm from coil surface.

    Regarding the specific questions that cannot be answered from the provided text:

    1. Sample sizes used for the test set and the data provenance: Not applicable or provided. The "test set" here refers to non-clinical testing of device characteristics, not a clinical study on patients or data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable or provided. Ground truth in this context would implicitly be engineering specifications, laboratory measurements, and standard compliance testing, not expert clinical assessment of patient data.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable or provided. No adjudication process detailed for establishing technical specifications.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device, nor is it a diagnostic device being evaluated for reader performance.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. Not an AI/ML algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for this device's performance evaluation is based on engineering specifications, direct physical measurements (e.g., of magnetic fields, temperatures, electrical properties), and compliance with international safety and performance standards (e.g., IEC 60601 series, ISO 10993).
    7. The sample size for the training set: Not applicable or provided. This device does not use a "training set" in the machine learning sense.
    8. How the ground truth for the training set was established: Not applicable or provided.

    In summary, the provided document details the 510(k) clearance process for a non-AI/ML magnetic stimulator for pain relief. The "acceptance criteria" and "performance" are framed around demonstrating substantial technical and safety equivalence to a legally marketed predicate device, rather than clinical efficacy data from patient studies or AI algorithm performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN170078

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-08-17

    (322 days)

    Product Code
    Regulation Number
    882.5802
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K122288, K992911, K051864, K162935

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Brainsway Deep Transcranial Magnetic Stimulation System is intended to be used as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder.

    Device Description

    The Brainsway DTMS System enables direct non-invasive activation of brain structures. TMS is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold, and is directed in an appropriate orientation relative to the brain's neuronal pathways, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure.

    The Brainsway DTMS System is composed of the following five main components:

    • Electromagnetic Coil (HAC-Coil)
    • TMS Neurostimulator
    • Cooling System
    • Positioning Device
    • Personal Head Cap
    AI/ML Overview

    The provided text describes a clinical study to establish the safety and effectiveness of the Brainsway Deep Transcranial Magnetic Stimulation (DTMS) System for the treatment of Obsessive-Compulsive Disorder (OCD).

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for effectiveness appear to be based on the primary effectiveness endpoint: a statistically significant reduction in Yale-Brown Obsessive Compulsive Scale (YBOCS) scores compared to sham treatment. Additionally, clinically meaningful reductions in OCD severity are highlighted.

    Acceptance Criterion (Effectiveness)Reported Device Performance
    Primary Endpoint: Statistically significant difference in change from baseline YBOCS scores to 6 weeks (post-randomization) between DTMS and Sham groups in the mITT analysis set.Met: In the mITT analysis set (N=94), the DTMS group showed an adjusted mean YBOCS score reduction of 6.0 points (95% CI: [4.0;8.1]), while the Sham group showed a reduction of 3.3 points (95% CI: [1.2;5.3]). The difference between the slopes of 2.8 points was statistically significant (p-value: 0.0127), indicating superior effectiveness of DTMS. The ITT analysis set (N=99) did not achieve statistical significance (p-value: 0.0988) due to a higher reduction in YBOCS in the Sham group (4.1 points) compared to the mITT Sham group (3.3 points), but a clinically meaningful reduction (6.0 points) was still observed in the DTMS group.
    Secondary Endpoints: - Response rate (≥30% YBOCS reduction from baseline) at 6 weeks. - Partial Response rate (≥20% YBOCS reduction from baseline) at 6 weeks. - Improvement in CGI-I ("Improved" category: moderately improved to very much improved) at 6 weeks. - Improvement in CGI-S ("Improved" category) at 6 weeks. - Sustained YBOCS reduction at 10 weeks.Met (with qualifications): - Response Rate: DTMS: 38.1% (16/42), Sham: 11.1% (5/45) in mITT. - Partial Response Rate: DTMS: 54.8% (23/42), Sham: 26.7% (12/45) in mITT. - CGI-I "Improved": DTMS: 49% (20/41), Sham: 21% (9/43) in mITT. - CGI-S "Improved": DTMS: 61% (25/41), Sham: 33% (14/43) in mITT. - YBOCS at 10 Weeks: Adjusted mean YBOCS reduction of 6.5 points (95% CI: [4.3;8.7]) in the DTMS group, indicating stability. (Note: Formal statistical testing was stopped at the first secondary endpoint, so p-values for other secondary endpoints were unadjusted for multiplicity, but results show favorable trends for DTMS).
    Safety: Device is safe and tolerable, with adverse events being minor, reversible, and typical for TMS devices, and no significant differences in vital signs, physical/neurological exams, or cognitive status between treatment groups. Absence of serious related adverse events.Met: - AEs reported by 73% (DTMS) vs. 69% (Sham). - Most frequent AE was headache (37.5% DTMS, 35.3% Sham). - Other pain/discomfort (administration site, jaw, facial, muscle, neck, dizziness) were mild/moderate and resolved. - No reports of hypoacusis (hearing loss). - No notable differences in vital signs, physical/neurological exams between groups. - One SAE reported, assessed as not device-related (suicidal thoughts preceding study).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Clinical Study Population):
      • Total enrolled: 100 subjects.
      • ITT Analysis Set (received at least one treatment): 99 subjects (48 DTMS, 51 Sham).
      • mITT Analysis Set (met eligibility criteria): 94 subjects (47 DTMS, 47 Sham). This was the principal data analysis set for efficacy.
      • Per-Protocol Analysis Set (without major protocol deviations): 93 subjects (46 DTMS, 47 Sham).
    • Data Provenance: Prospective, randomized, double-blinded, multi-site, sham-controlled clinical trial. The study was conducted at 11 study sites in the United States (9 sites), Israel (1 site), and Canada (1 site).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • The ground truth for the effectiveness of the treatment was established through clinical assessments using standardized scales like the YBOCS, Sheehan Disability Scale (SDS), and Clinical Global Impression - Severity (CGI-S) and Improvement (CGI-I).
    • These assessments were performed by independent raters. While the exact number of independent raters or their specific qualifications (e.g., years of experience as psychiatrists or psychologists) is not explicitly stated, the study design mentions that "the independent rater" was blinded to the treatment administered. This implies that trained clinical professionals, likely psychiatrists or psychologists, administered these critical rating scales.

    4. Adjudication Method for the Test Set

    • The study design involved a double-blinded approach where both study personnel (including the operator and independent rater) and study subjects were blinded to whether active or sham treatment was administered.
    • The primary and secondary outcome measures (YBOCS, SDS, CGI-S, CGI-I) are standardized clinical scales that are typically administered by trained personnel. There is no explicit mention of an "adjudication method" like 2+1 or 3+1 for discordance between multiple experts for individual case ground truth, as these scales are designed for direct clinical assessment. The "ground truth" here is the patient's symptom severity and improvement as measured by these scales.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. This study directly evaluated the effectiveness of a medical device (DTMS system) for treating OCD in patients, not an AI system assisting human readers. Therefore, the concept of "how much human readers improve with AI vs. without AI assistance" is not applicable to this study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • No, a standalone (algorithm only) performance study was not done. The study evaluated a medical device that is administered to patients by trained operators, with outcomes assessed by independent clinical raters. This is a human-in-the-loop system in the context of patient treatment and assessment, not an AI algorithm acting autonomously.

    7. The Type of Ground Truth Used

    • The ground truth used for determining treatment effectiveness was based on expert clinical assessments and standardized patient-reported outcomes. Specifically:
      • Yale-Brown Obsessive Compulsive Scale (YBOCS): A clinician-administered scale that measures the severity of OCD symptoms.
      • Sheehan Disability Scale (SDS): A patient-rated scale that assesses functional impairment.
      • Clinical Global Impression - Severity (CGI-S) and Improvement (CGI-I): Clinician-rated scales that measure global severity and improvement of illness.
      • Response/Remission Definitions: Pre-defined thresholds of YBOCS score reduction or absolute score for clinical response and remission.
    • For safety, ground truth was assessed through adverse event reporting, vital signs, physical/neurological examinations, and cognitive tests.

    8. The Sample Size for the Training Set

    • This was a clinical trial evaluating a medical device, not a machine learning model. Therefore, there was no "training set" in the context of machine learning model development. The sample size used for the clinical study (94-99 subjects in the analysis sets) was for direct evaluation of the device's efficacy and safety.

    9. How the Ground Truth for the Training Set Was Established

    • As there was no machine learning "training set," this question is not applicable. The clinical trial directly provided the data (including outcome measures like YBOCS scores) used to assess the device's performance.