Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K120160

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-05-14

    (116 days)

    Product Code
    Regulation Number
    876.5130
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K051593,K030956

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ROCAMED RocaUS Platinum is intended to be a conduit for passage of endoscopes and other urological devices for the purpose of performing ureteroscopy procedures. The dual working lumen dilator with luer lock connections allows the user to insert guidewires and fluids.

    Device Description

    The ROCAMED RocaUS Platinum is designed to create a conduit for urological procedural instruments. The device consists of two components: a flexible, coil reinforced sheath and a semirigid dual lumen dilator catheter with tapered distal tip. Both components are radiopague and have hydrophilic coating. This device is sold in two sizes, 10/12 and 12/14 FR.

    AI/ML Overview

    The provided documentation is a 510(k) summary for the ROCAMED RocaUS Platinum, a ureteral access sheath. This type of submission focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific clinical performance acceptance criteria in the same way a new AI/ML-driven diagnostic device might.

    Therefore, many of the requested categories (e.g., sample size for test set, number of experts for ground truth, MRMC study, training set information) are not applicable or not present in this document because it is not a clinical study report for a diagnostic or treatment outcome device. The "performance data" referred to in this document relates to physical and functional testing, not clinical performance measures like accuracy or sensitivity.

    Here's an attempt to fill in the table and address the questions based on the provided text, indicating where information is not applicable (N/A) or not provided (NP):


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a 510(k) for a medical device (ureteral access sheath) demonstrating substantial equivalence, the "acceptance criteria" are primarily met through comparison to predicate devices across technological characteristics and intended use, supported by physical and functional testing. There are no explicit performance metrics like sensitivity, specificity, or accuracy derived from a clinical study with a test set that would have numerical acceptance criteria.

    Acceptance Criteria CategoryReported Device Performance (ROCAMED RocaUS Platinum)
    Intended UseConduit for passage of endoscopes and other urological devices for performing ureteroscopy procedures. Dual working lumen dilator allows guidewire and fluid insertion.
    Technological Characteristics- Disposable, sterile, X-ray opaque, coil reinforced, atraumatic tip, tapered dilator, radiopaque marks, hydrophilic coating, injection of contrast media, proximal end funnel. - 2 lumens - Dilator Material: LDPE+BaSO4 - Sheath Material: Pebax-SST-PTFE - Fr Size: 10/12, 12/14 - Length: 35 cm - Guidewire compatibility: 0.032", 0.035"
    SafetyImplied by substantial equivalence to legally marketed predicate devices and successful physical/functional testing. Changes in dilator material (LDPE instead of PTFE) stated to "ease insertion into the patient and reduce trauma."
    Performance (Functional)Results of physical and functional testing support substantial equivalence. (Specific reported performance values for these tests are not detailed in this summary).
    MaterialsMaterials chosen are considered substantially equivalent or improved (e.g., LDPE for dilator) compared to predicate devices, based on physical and functional testing.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not applicable. This document refers to physical and functional testing, not a clinical test set of patient data. The specifics of samples used for these engineering tests (e.g., number of devices tested for tensile strength or burst pressure) are not provided in this summary.
    • Data Provenance: Not applicable for clinical test data. The physical and functional testing would have been conducted by Promepla SAM.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Ground truth as typically understood for diagnostic device performance (e.g., pathology, expert consensus on images) is not relevant for this type of medical device submission, which relies on physical and mechanical properties and comparison to predicates.

    4. Adjudication method for the test set

    Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-assisted diagnostic tool and therefore did not undergo an MRMC study related to AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable in the context of clinical "ground truth." The "ground truth" for this device's performance would be against engineering specifications and predicate device characteristics, verified through physical and functional testing.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML algorithm.

    9. How the ground truth for the training set was established

    Not applicable. This device is not an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1