    K094022

    Validate with FDA (Live)

    Device Name
    THE WATER LILY
    Date Cleared
    2010-06-11

    (163 days)

    Product Code
    Regulation Number
    876.5220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K050992

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Water Lily The is indicated for colon cleansing when medically indicated, such as before radiological or endoscopic examination.

    Device Description

    The Water Lily 10 is a professional colon hydrotherapy system for use when medically indicated. It introduces filtered water at a comfortable temperature into the large intestine, thus cleansing the colon of its contents. Water temperature is controlled by means of a mixing valve; flow is controlled by one switch operated by the user. The system is manually sanitized prior to each use with a suitable broad spectrum disinfectant (not included). The system includes disposable tubing and a sterile, disposable rectal nozzle intended for single use only (ConMed Corporation K050992).

    AI/ML Overview

    The provided document describes the K094022 510(k) submission for the Water Lily™ Colon Hydrotherapy System. This device is a colonic irrigation system and its acceptance criteria and the study proving it meets them are detailed primarily through a comparison to a predicate device, the Angel of Water™ Colon Hydrotherapy System (K003720), and non-clinical testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    This device is cleared based on demonstrating substantial equivalence to a predicate device. The acceptance criteria are implicitly met by showing similar technical characteristics and performance to the predicate. The "reported device performance" below refers to the results of non-clinical testing for specific safety parameters.

    Acceptance Criteria (Inferred from Predicate Equivalence and Safety Tests)Reported Device Performance (Water Lily™)
    Intended UseColon cleansing when medically indicated, such as before radiological or endoscopic examination.
    Target PopulationSame as predicate
    DesignSame principles of operation as predicate
    PerformanceSimilar to predicate
    Materials UsedSimilar to predicate (e.g., single-use sterile disposable nozzle, ConMed K050992)
    Where UsedSimilar to predicate
    Patient ControlYes
    Temperature Range99°-103° F
    Temperature cannot rise above 103° F (safety mechanism)Yes. When water temperature reaches 104°F, an audible alarm sounds, and a solenoid valve closes water flow to the client. The therapist then cools the system to 104°F and resets the unit.
    Water flow controlled by mixing valveYes
    Filtration: UltravioletYes
    Filtration: Carbon FiltrationYes
    Pressure inside patient's colon will not exceed 2 psi (safety mechanism)Formula for PSI: 2.5 Ft drop = 1.075 psi. Confirmed by 0-3 psi gage.
    Electrical Safety - UV filtration systemApproved by UL and CSA
    Electrical Safety - Controller-thermocouple and solenoid valveApproved by UL and CSA

    2. Sample size used for the test set and the data provenance

    The document does not detail a "test set" in the traditional sense involving human subjects or clinical data for evaluating the device's performance against the acceptance criteria. Instead, the evaluation is based on:

    • Non-clinical bench testing: This was used to verify specific safety parameters (pressure, temperature shutdown, electrical safety). There is no "sample size" indicated for these tests, as they are typically performed on a limited number of manufactured units to confirm design specifications.
    • Comparison to a predicate device: The Water Lily™ was compared to the Angel of Water™ (K003720) based on shared technical characteristics. This is a design and functional comparison, not a data-driven clinical comparison.
    • Data provenance: Not applicable in the context of clinical trials as no clinical trials were performed. The testing described is bench testing.

    Therefore, there is no explicit sample size for a "test set" or data provenance information (country of origin, retrospective/prospective) because the submission focuses on substantial equivalence through design and non-clinical engineering verification rather than clinical studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device's clearance is based on substantial equivalence to a predicate device and non-clinical bench testing. There was no "ground truth" derived from expert consensus for a test set as would be found in a clinical study for diagnostic accuracy or efficacy.

    4. Adjudication method for the test set

    Not applicable. As no clinical "test set" was described requiring expert assessment or ground truth establishment, no adjudication method was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done

    No. The document does not mention any clinical studies, including MRMC comparative effectiveness studies, demonstrating how human readers (or users in this context) improve with or without AI assistance. This device is a physical medical device, not an AI-driven diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This device is a colon hydrotherapy system and does not involve any algorithms or AI, thus rendering "standalone performance" (in the context of AI) irrelevant.

    7. The type of ground truth used

    Not applicable in the sense of clinical "ground truth." For the non-clinical tests, the "ground truth" equates to engineering specifications and safety standards (e.g., pressure not exceeding 2 psi, temperature shutdown at 104°F, UL/CSA electrical safety approvals). The primary ground for clearance is substantial equivalence to the legally marketed predicate device.

    8. The sample size for the training set

    Not applicable. This device does not involve machine learning or AI, so there is no concept of a "training set."

    9. How the ground truth for the training set was established

    Not applicable, as no training set was used.