Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K141943

    Validate with FDA (Live)

    Date Cleared
    2015-06-12

    (330 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K103345, K050359

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Breathe Technologies Life2000 Ventilation System is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult patients who require the following types of ventilatory support: - Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask). - Assist/Control mode of ventilation. The ventilator is suitable for use in home and institutional settings.

    Device Description

    The Breathe Technologies Life2000 Ventilation System is a portable, battery powered, critical care ventilator. The Ventilator administers the physician-prescribed volume to the patient via the attached Breathe Technologies Patient Universal Connector which connects into the patient's tracheostomy tube, endotracheal tube, or any off the shelf non-invasive mask. It can also be used with the Breath Technologies NIOV Pillows Interface, a type of nasal mask. The ventilator is small and light enough to be worn on a patient's belt, or slung over their shoulder. It is connected to a separate, third party, gas supply. The Ventilator is intended for Institutional or Home Use. It is not intended for use during emergency/medical transportation.

    AI/ML Overview

    The Breathe Technologies Life2000 Ventilation System is a portable, battery-powered critical care ventilator for adult patients requiring continuous or intermittent ventilatory support.

    Here's an analysis of its acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance
    Feature/ParameterPredicate Device (LTV-1200, K060647) Acceptance CriteriaApplication Device (Life2000 Ventilation System) PerformanceEvaluation / Difference Status
    Indications for UseContinuous/intermittent ventilatory support, adult & pediatric (>5kg), invasive/non-invasive, A/C, SIMV, CPAP, NPPV modes, institutional, home & transport settingsContinuous/intermittent ventilatory support, adult patients only, invasive/non-invasive, Assist/Control mode, institutional & home settingsDifference: Application device's indications are a subset of the predicate (adults only, fewer modes, no transport). This is considered acceptable as it's a narrower use case.
    Product Classification CodeCBKCBKIdentical
    CFR Citation21 CFR 868.589521 CFR 868.5895Identical
    Principal OperatorTrained personnel under direction of a physicianTrained personnel under direction of a physicianIdentical
    Environment of UseInstitution, Home, and TransportInstitution & HomeDifference: Application device not validated for transport. This is an accepted difference as it narrows the intended use.
    Patient InterfaceInvasive (ET tube) or non-invasive (mask)Invasive (ET tube) or non-invasive (mask)Identical
    Power SourceBattery Powered, run while chargingBattery Powered, run while chargingIdentical
    Operational ModesVolume Control, Volume Assist/Control, Volume Assist, Pressure Control, Pressure Support, SIMV, CPAP, NPPVVolume Control, Volume Assist/Control, Volume AssistDifference: Application device modes are a subset of the predicate. This is an accepted difference as fewer documented modes are presented.
    Active Exhalation Valve?Yes, MechanicalYes, Mechanical/PneumaticClinically Equivalent
    Design DesignationPortable Critical CarePortable Critical CareIdentical
    Size WxLxH (in)3 x 10 x 123.2 x 7.7 x 1.0Difference: Application device is much smaller. Considered an improvement potentially making it easier for patients to handle, with no new safety/effectiveness issues raised.
    Weight13.4 lbs1.1 lbsDifference: Application device is much lighter. Considered an improvement potentially making it easier for patients to handle, with no new safety/effectiveness issues raised.
    Volume Setting Range50 - 2000 ml/breath50 - 750 ml/breathDifference: Application device range is a subset of the predicate. This is an accepted difference as it's a narrower range.
    Resultant Tidal Volume50 - 2000 ml/breath50 - Up to 2000 ml/breath due to venturi effectIdentical
    PEEP Setting0 – 20 cmH2O0 - 10 cmH2ODifference: Application device range is a subset of the predicate. This is an accepted difference as it's a narrower range.
    PIP Alarms & MonitoringYesYesIdentical
    Adjustable Inspiration Time0.3 - 9.9 seconds0.15 to 3 secondsDifference: I-time of application device is a subset of predicate. Only clinically relevant times are used. This is an accepted difference.
    Supply GasOxygen, AirOxygen, AirIdentical
    Method of supply gas pressurizationInternal turbine for Air, Compressed Source for O2Compressed source for Air, Compressed source for O2Clinically Equivalent
    Sterilized?Ventilator: No, Patient Circuit: NoVentilator: No, Patient Circuit: NoIdentical
    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • The document does not specify sample sizes for specific performance tests (e.g., comparative waveform testing, human factors).
      • The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective, beyond general statements of compliance with international and US standards.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • The document identifies general "qualified, trained personnel under the direction of a physician" as the intended operators for the device. However, it does not specify the number or qualifications of experts used to establish ground truth or conduct specific performance assessments for the testing mentioned (e.g., comparative waveform testing, human factors).
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • The document does not specify any adjudication methods for the test sets.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • No MRMC comparative effectiveness study is mentioned, as this device is a mechanical ventilator, not an AI-assisted diagnostic tool. The "comparative effectiveness" studies mentioned are related to device performance compared to a predicate, not human reader performance.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

      • This is not applicable to a mechanical ventilator device. The device's performance is inherently standalone in its function, but its operation always involves a human user (qualified personnel under a physician's direction).
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • For the "Summary of Supporting Data," the ground truth for device performance is established through compliance with recognized standards such as ISO 10993-1, FDA Draft Reviewer Guide for Ventilators, ASTM F1246-91, IEC 60601-1, IEC 60601-1-2, and ISO 80601-2-12. This implies engineering and clinical performance metrics defined by these standards serve as the "ground truth" for the device's functional and safety capabilities.
      • Specific tests mentioned like "Ventilator Cleaning validation," "Risk characterization of VOC's," "Static Analysis Tool verification of software," "Comparative Waveform Testing," and "Human Factors and Usability Testing" would have their own defined ground truths based on established protocols for those types of evaluations (e.g., cleanliness standards, toxicology limits, software quality metrics, physiological waveform accuracy, human-computer interaction success rates).
    7. The sample size for the training set

      • This is not applicable for a mechanical ventilator. Training sets are typically associated with machine learning or AI models, which are not described as integral to this device's function.
    8. How the ground truth for the training set was established

      • This is not applicable for a mechanical ventilator.
    Ask a Question

    Ask a specific question about this device

    K Number
    K132013

    Validate with FDA (Live)

    Device Name
    SWIFT AIR
    Manufacturer
    Date Cleared
    2013-10-21

    (112 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K050359,K090244

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Swift™ Air channels airflow noninvasively to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bilevel device.

    The Swift Air is:

    • to be used by patients (> 66 lb/30 kg) for whom positive airway pressure has been prescribed
    • intended for single-patient re-use in the home environment and multipatient re-use in the hospital/institutional environment.
    Device Description

    The Swift™ Air provides an interface such that air flow from a positive pressure source is directed to the patient's nasal nares. The mask is held in place with adjustable headgear that straps the mask to the face.

    Swift™ Air is a prescription device supplied non-sterile.

    AI/ML Overview

    This document describes the premarket notification for the ResMed Swift™ Air nasal mask. The acceptance criteria and supporting studies are based on demonstrating substantial equivalence to predicate devices, rather than establishing de novo performance metrics.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryPredicate Device(s)Acceptance CriteriaReported Device Performance (Swift™ Air)
    Intended UseMirage FX (K102746)Identical intended use as the predicate device, for use with Positive Air Pressure therapy equipment and the same patient population.Identical intended uses: Channels airflow noninvasively to a patient from a PAP device (CPAP or bilevel), for patients > 66 lb/30 kg, single-patient re-use in home, multi-patient re-use in hospital/institutional environment.
    BiocompatibilitySwift™ FX (K090244)All materials used in the construction must be deemed as safe as those of the predicate devices, in accordance with FDA Guidance #G95-1 and ISO 10993-1. Specific tests for heated humidified gas pathway (ISO 10993-3, -5, -6, -10) and skin contact (ISO 10993-5, -10).All materials used in the construction of the new mask are deemed as safe as those of the predicate devices. Appropriate biological tests (ISO 10993-3 Genotoxicity, ISO 10993-5 Cytotoxicity, ISO 10993-6 Implantation, ISO 10993-10 Sensitization for heated gas pathway; ISO 10993-5 Cytotoxicity, ISO 10993-10 Sensitization and Irritation for skin contact) were conducted and passed.
    CO2 Performance (Venting)Ultra Mirage II (K050359)Adequate venting to flush out expired CO2, satisfying predefined pass/fail criteria and being substantially equivalent to the predicate device.CO2 performance of the new device (both vent types) was tested, including physical and functional dead-space measurements. The device satisfied all predefined pass/fail criteria and was shown to be substantially equivalent to the predicate Ultra Mirage II.
    Pressure-Flow Characteristics & Flow ImpedanceSwift™ FX (K090244)Identical pressure-flow characteristics and flow impedance to the predicate device, operating on the same ResMed flow generator settings.Pressure-flow and through impedance bench test results of the new mask were substantially equivalent to the predicate Swift FX.
    Mechanical Integrity & Performance(General Safety)Withstand simulated normal use and reasonable abuse scenarios, and the effects of storage temperature, humidity, and transportation shock & vibration.Mechanical integrity and performance were tested to simulated normal use and reasonable abuse scenarios. The device was tested to demonstrate it can withstand storage temperature, humidity, and transportation shock & vibration. Device development complies with ISO 14971:2007 (risk management).
    Cleaning and Reuse(General Safety)Device can be safely reused (single-patient home, multi-patient hospital) following validated disinfection protocols, and function as intended after specified cleaning/disinfection cycles.Validation of cleaning and reuse was completed. After 20 cycles of cleaning/disinfection in accordance with described methods, the device was shown to function as intended. The device satisfied pass/fail criteria and was shown to be substantially equivalent to the predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    The review document does not specify a "test set" in the context of clinical data for performance evaluation of the Swift™ Air mask. The evaluation relied heavily on bench testing and comparison to predicate devices to demonstrate substantial equivalence.

    • Clinical Data: "Use of vented nasal masks with CPAP or Bilevel therapy is proven technology and is well accepted by the medical community. Bench testing is sufficient to demonstrate substantial equivalence to the predicate devices." This indicates no new clinical studies with a specific patient "test set" were conducted for the Swift™ Air for this submission.
    • Bench Testing: The sample sizes for the various bench tests (CO2 performance, pressure-flow, mechanical integrity, cleaning/reuse) are not explicitly stated in the provided text. The data provenance is presumed to be internal laboratory testing conducted by ResMed.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. As noted above, the evaluation was based on bench testing and comparison to predicate devices, not on expert-adjudicated clinical patient data or specific "ground truth" established by experts for a test set.

    4. Adjudication Method for the Test Set

    Not applicable. There was no clinical "test set" requiring expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. An MRMC study was not performed. This device is a medical accessory, and its effectiveness is determined by its physical and functional characteristics in delivering CPAP/Bilevel therapy, rather than diagnostic interpretation requiring human readers.

    6. Standalone Performance Study (Algorithm only without Human-in-the-Loop Performance)

    Yes, implicitly. The performance data presented (CO2, pressure-flow, mechanical integrity, cleaning/reuse) represents the "standalone" performance of the device itself, under bench test conditions. This is the equivalent of "algorithm only" in the context of a physical medical device.

    7. Type of Ground Truth Used

    The "ground truth" for the Swift™ Air device in this submission was established by:

    • Predicate Device Performance: The primary "ground truth" was the established performance and safety profiles of the predicate devices (Mirage FX, Swift™ FX, Ultra Mirage II) that have already been cleared by the FDA.
    • Industry and Regulatory Standards: Adherence to standards such as FDA Guidance #G95-1, ISO 10993-1 (Biocompatibility), ISO 14971:2007 (Risk Management), and ISO 5356-1:2004 (Conical Connectors) served as additional "ground truth" for safety and performance requirements.
    • Predefined Pass/Fail Criteria: For the bench tests (e.g., CO2 performance, mechanical integrity), "predefined pass/fail criteria" were used as the immediate ground truth, though these criteria would have been derived from relevant standards and predicate performance.

    8. Sample Size for the Training Set

    Not applicable. There is no mention of a "training set" in the context of this device. This is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How Ground Truth for the Training Set Was Established

    Not applicable. As there was no training set, there was no ground truth for a training set to be established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103434

    Validate with FDA (Live)

    Device Name
    3B FLEX-LITE
    Manufacturer
    Date Cleared
    2011-10-13

    (324 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K110008, K063036, K050359

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 3B Flex-Lite™ channels airflow noninvasively to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or a bilevel system. The 3B Flex-Lite™ is:

    • to be used by adult patients (>66lb / 30 kg); (1)
    • to be used for single-patient reuse in the home environment and multipatient (2) reuse in the hospital/institutional environment.
    • (3) to be used in the following environments: home, hospital, and sub-acute institutions.
    • intended for prescription use. (4)
    Device Description

    The 3B Flex-Lite™ is a mask interface, of the nasal pillow variety, that directs airflow from a positive pressure device to the patient's nose. The mask is held in place with adjustable headgear that straps the mask to the face.

    The 3B Flex-Lite™ is safe when used under the conditions and purposes intended as indicated in the labeling provided with the product.

    The 3B Flex-Lite™ is a prescription device supplied non-sterile.

    AI/ML Overview

    The provided document describes the 3B Flex-Lite Nasal Mask, a device intended to channel airflow noninvasively to a patient from a positive airway pressure (PAP) device. The document focuses on demonstrating substantial equivalence to predicate devices rather than establishing novel performance criteria. Therefore, the information provided does not directly outline "acceptance criteria" as distinct performance thresholds that the device must meet for a specific clinical task. Instead, the "acceptance criteria" are implicitly met by demonstrating substantial equivalence to legally marketed predicate devices through comparable technological characteristics and performance.

    Here's an analysis based on the provided text, focusing on how the device meets the implicit acceptance criteria of substantial equivalence:

    1. Table of Acceptance Criteria and Reported Device Performance

    As explicit acceptance criteria with numerical targets are not stated, this table will frame the "acceptance criteria" as comparability to predicate devices for key performance and safety characteristics.

    Acceptance Criteria (Implicit for Substantial Equivalence)Predicate Device Performance (Nasal-Aire II / Swift FX)Reported Device Performance (3B Flex-Lite™)Conformance
    Intended UseSame as proposed deviceChannels airflow noninvasively to a patient from a PAP device. Used by adult patients (>66lb / 30 kg); for single-patient reuse in home/multi-patient reuse in hospital/institutional environment; in home, hospital, and sub-acute institutions; for prescription use.Conforms
    Environment of UseHospitals, sub-acute institutions, sleep laboratories and home.SameConforms
    Patient PopulationAdultSameConforms
    ContraindicationsNoneNoneConforms
    Single patient, multi-useYesYesConforms
    ComponentsNasal interface (Nasal-Aire II); Nasal interface with pillows and headgear (Swift FX)Nasal interface with pillows and headgearConforms (similar to Swift FX, evolving from Nasal-Aire II)
    Dead SpaceInterface - 24-28 ml / Pillows 2-4 mlInterface 10.5ml / Pillows 3.5 mlConforms (similar or improved, within acceptable range for safety)
    Fixed leak portYesYesConforms
    Deliverable pressure range3-18 cmH2O (Nasal-Aire II); 4-20 cmH2O (Swift FX)4-20 cmH2OConforms (similar to Swift FX, encompassing broader range)
    Nasal Interface MaterialSiliconeSilicone (Biocompatibility per ISO 10993-1)Conforms
    Headgear MaterialNone (Nasal-Aire II); Breath-O-Prene™ (Swift FX)Breath-O-Prene™Conforms (similar to Swift FX)
    Performance Bench Testing (Flow vs. Leak Pressure)PerformedPerformedConforms
    Performance Bench Testing (Dead Space)PerformedPerformedConforms
    Performance Bench Testing (CO2 Rebreathing)PerformedPerformedConforms
    Safety and Effectiveness ConclusionEstablished for predicate devicesSubstantially equivalent to predicate devices. Does not raise new questions of safety and effectiveness; at least as safe and effective.Conforms (as determined by FDA review)

    2. Sample Size Used for the Test Set and Data Provenance

    The document refers to "Performance bench testing of both the 3B Flex-Lite™ and the Nasal Aire II were conducted" and explicitly mentions a "Piper Medical" test report dated March 23, 2011, annexed at Appendix A. However, the specific sample size for these bench tests (e.g., number of units tested) and the country of origin of the data are not detailed in the provided text. The testing appears to be retrospective in the sense that it evaluates the manufactured device against established performance benchmarks and comparisons to existing devices, rather than a prospective clinical trial with human subjects.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The "ground truth" for the bench tests would typically be defined by engineering specifications, regulatory standards, and the performance characteristics of the predicate devices. The individuals conducting and reporting the tests (e.g., Piper Medical) would be engineers or technicians with expertise in medical device testing, but their specific qualifications or the number involved are not stated.

    4. Adjudication Method for the Test Set

    This information is not provided. Given that the tests are bench tests, an adjudication method in the clinical sense (e.g., 2+1 physician consensus) is not directly applicable. The "adjudication" would be based on comparison of test results to predefined criteria or to the predicate device's measured performance.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not applicable to this device. The 3B Flex-Lite™ Nasal Mask is a physical medical device (mask interface), not an AI-powered diagnostic or assistive technology for human readers. Therefore, there is no discussion of human reader improvement with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    A standalone performance study of an algorithm is not applicable to this device, as it is a physical device, not an algorithm or AI system. The performance tests ("Non-clinical Tests") evaluated the physical mask's characteristics.

    7. The Type of Ground Truth Used

    The ground truth used for evaluating the 3B Flex-Lite™ appears to be based on:

    • Engineering specifications and regulatory standards for medical devices (implicitly, concerning safety, materials, and performance parameters like dead space, CO2 rebreathing, flow vs. leak pressure).
    • Performance characteristics of legally marketed predicate devices (InnoMed Nasal-Aire II and Resmed Swift FX) as established through their own testing and marketing. The goal was to demonstrate that the new device is "substantially equivalent" in terms of performance to these predicates.

    8. The Sample Size for the Training Set

    This is not applicable. The 3B Flex-Lite™ Nasal Mask is a physical medical device, not a machine learning model, and therefore does not have a "training set" in the context of AI.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090490

    Validate with FDA (Live)

    Device Name
    MIRAGE ECHO
    Manufacturer
    Date Cleared
    2009-05-06

    (70 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K071808, K072940, K081321, K050359

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mirage Echo channels airflow noninvasively to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bilevel system. The Mirage Echo is: to be used by adult patients (> 66 lb/30 kg) for whom positive airway pressure has been prescribed intended for single-patient re-use in the home environment and multipatient re-use in the hospital/institutional environment.

    Device Description

    The Mirage Echo provides an interface such that airflow from a positive pressure source is directed to the patient's nose. The mask is held in place with adjustable headgear that straps the mask to the face. Mirage Echo is safe when used under the conditions and purposes intended as indicated in the labeling provided with the product. Mirage Echo is a prescription device supplied non-sterile.

    AI/ML Overview

    Here's an analysis of the Mirage Echo Traditional 510k, based on the provided text, focusing on acceptance criteria and supporting studies.

    Based on the provided information, the 510(k) submission for the Mirage™ Echo nasal mask describes a comparative study against predicate devices (Mirage Micro and Ultra Mirage II Mask) to demonstrate substantial equivalence, rather than a study with explicit, quantitative acceptance criteria for device performance as one might see for an AI algorithm.

    The core of the "study" is a set of comparisons and assertions of substantial equivalence, relying on bench testing and the existing clinical acceptance of similar technology.

    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a traditional 510(k) for a medical device (nasal mask), the "acceptance criteria" are primarily established implicitly by demonstrating substantial equivalence to predicate devices, focusing on safety, performance characteristics, and intended use. The performance is reported as "substantially equivalent" to the predicates.

    Acceptance Criteria (Implied from Substantial Equivalence to Predicates)Reported Device Performance (Mirage Echo)
    Safety: Materials deemed safe (ref: ISO 10993-1).All components are fabricated using materials deemed safe (ref: ISO 10993-1).
    Intended Use: Channels airflow noninvasively from a PAP device to adult patients (> 66 lb/30 kg) for single-patient home reuse and multipatient hospital/institutional reuse.Same intended use as predicates. Channels airflow noninvasively to a patient from a PAP device (CPAP/bilevel system), for adult patients (> 66 lb/30 kg), for single-patient home reuse and multipatient hospital/institutional reuse.
    Technological Characteristics:
    - Provides a seal via silicone interface.Provides a seal via silicone interface.
    - Offered in various sizes for adequate fit.Offered in various sizes to ensure adequate fit.
    - Incorporates vent holes for continuous air leak to flush CO2 and minimize rebreathing without interfering with performance.Incorporates vent holes to provide continuous air leak to flush out and minimize the amount of CO2 rebreathed. The design ensures vents do not interfere with intended performance.
    - Connects to conventional air delivery hose via standard conical connectors (ISO 5356-1:2004).Connects to a conventional air delivery hose via standard conical connectors (ref: ISO 5356-1:2004).
    - Constructed using molded plastic and silicone components.Constructed using molded plastic and silicone components.
    - Operates on the same ResMed flow generator settings.Designed to operate on the same ResMed flow generator settings.
    - Substantially equivalent pressure-flow characteristics and flow impedance.Pressure-flow characteristics and flow impedance of both the new device and the predicate device are substantially equivalent.
    - Can be reused in home and hospital/institution environment.Can be reused in the home and hospital/institution environment.
    - CO2 performance is substantially equivalent.The CO2 performance of the new device and the predicate device are substantially equivalent.
    Effectiveness: At least as safe and effective as the predicate devices and does not raise new questions of safety and effectiveness.Asserted to be at least as safe and effective and does not raise new questions of safety and effectiveness. "Use of vented nasal masks with CPAP or Bilevel therapy is Clinical Data proven technology and is well accepted by the medical community." "Bench testing is sufficient to demonstrate safety and efficacy of the Mirage ECHO, as was the case with the predicate device."

    2. Sample Size for the Test Set and Data Provenance

    The provided text does not mention a "test set" in the context of clinical data or patient-specific evaluation. The performance data refers to bench testing. Therefore, information regarding:

    • Sample size for the test set
    • Data provenance (e.g., country of origin, retrospective/prospective)
      is not applicable as no clinical test set data is described. The submission relies on "bench testing" and comparison of technical specifications.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable. Since no clinical test set data is described, there's no mention of experts establishing a ground truth for such a set. The "ground truth" for this type of submission is typically derived from established engineering principles, international standards (e.g., ISO), and the performance of legally marketed predicate devices.

    4. Adjudication Method for the Test Set

    This information is not applicable. No clinical test set data is described that would require an adjudication method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret results, potentially with AI assistance. The Mirage Echo is a medical device (nasal mask) for therapy delivery, not a diagnostic tool requiring interpretation.

    6. Standalone (Algorithm Only) Performance Study

    A standalone performance study was not done, nor would it be relevant for this type of device. The Mirage Echo is a physical medical device, not an AI algorithm.

    7. Type of Ground Truth Used

    The "ground truth" in this context is established through bench testing results, adherence to international standards (e.g., ISO 10993-1 for biocompatibility, ISO 5356-1:2004 for connectors), engineering specifications, and the established performance and safety profile of legally marketed predicate devices. There is no mention of expert consensus, pathology, or outcomes data being used to establish ground truth for this submission, as it focuses on demonstrating substantial equivalence in physical and functional characteristics.

    8. Sample Size for the Training Set

    This information is not applicable. The Mirage Echo is a physical medical device, not an AI algorithm that undergoes a "training set" process.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable. As it's not an AI algorithm, there is no "training set" or ground truth for such a set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K061099

    Validate with FDA (Live)

    Date Cleared
    2006-08-11

    (114 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K031883, K050359

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BreatheX Nasal Interface System is intended for use as accessory device to CPAP machines in the treatment of obstructive sleep apnea (OSA).

    The BreatheX Nasal Interface System is for use on adult spontaneously breathing (non-ventilator dependant) patients at home or in the sleep clinic.

    The BreatheX Nasal Interface System is an accessory intended for use with devices that deliver Continuous Positive Airway Pressure (CPAP) and bi-level positive airway pressure in treating adult patients.

    Device Description

    The BreatheX Nasal Interface System has been designed for use as an accessory with commonly marketed CPAP units. The device consists of a plastic manifold with integral exhaust port, nasal mask cushion, a set of nasal pillows, adjustable headgear straps, and an interface tube with a 22mm fitting.

    The Nasal Interface System allows a patient to use either the nasal pillows or nasal mask cushion interchangeably. The lightweight construction of the device is designed to maximize patient comfort.

    The device is labeled for single patient use.

    AI/ML Overview

    The Hoffman Laboratories BreatheX™ Nasal Interface System is a CPAP device accessory. The provided document does not detail specific quantitative acceptance criteria or a formal study proving the device met such criteria in the way one might expect for a diagnostic or therapeutic medical device involving clinical outcomes.

    Instead, the submission focuses on demonstrating substantial equivalence to already legally marketed predicate devices through non-clinical testing and comparison of technological characteristics.

    Here's an breakdown of the information requested, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific Criteria (Implicitly based on predicate devices and standards)Reported Device Performance
    Environmental PerformanceCompliance with standards for temperature, humidity, shock, vibration, moisture ingress."The BreatheX Nasal Interface System passed all of the tests."
    Static PerformanceComparable dead space to predicate devices."The device met specified requirements and was comparable to the applicable specifications of the predicate devices."
    Dynamic PerformanceComparable pressure-flow characteristics to predicate devices."The device met specified requirements and was comparable to the applicable specifications of the predicate devices."
    Sound Output LevelComparable sound output level to predicate devices."The device met specified requirements and was comparable to the applicable specifications of the predicate devices."
    BiocompatibilityMaterials comply with biocompatibility requirements for intended use."The materials utilized in the device comply with biocompatibility requirements appropriate for the intended use."
    Risk & Hazard AnalysisResidual risks acceptable for intended use."Risk and Hazard analyses were performed, and the results of the analyses demonstrated that the residual risks were acceptable for the intended use."
    Technological EquivalenceSimilar technological characteristics as predicate devices (e.g., rigid plastic manifold, soft silicone rubber seal, head straps, integral exhaust port)."The BreatheX Nasal Interface device utilizes similar technological characteristics as the predicate devices... Like the predicate devices, the manifold... has an integral exhaust port..."

    2. Sample sized used for the test set and the data provenance

    • Sample Size: Not specified quantitatively. The "tests" mentioned are likely engineering and bench tests, not involving human subjects. For example, "dead space" or "pressure-flow characteristics" would be measured on a single or a small number of device units.
    • Data Provenance: The tests were conducted internally by Hoffman Laboratories, LLC. "Laboratory and standards compliance tests" are mentioned, suggesting controlled lab environments. No country of origin for data is stated beyond the company's location in Chatsworth, CA, USA. The data is retrospective in the sense that it was collected prior to the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not applicable and not provided. The "ground truth" for this type of device (a CPAP accessory) in a 510(k) submission is typically based on engineering specifications, industry standards, and comparison to predicate devices, rather than expert interpretation of clinical data.

    4. Adjudication method for the test set

    • Not applicable and not provided. As mentioned above, this is not a study requiring adjudication of expert opinions.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC study was conducted. This device is not an AI-powered diagnostic or therapeutic tool that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No standalone algorithm performance study was done. This device does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for the non-clinical tests was based on engineering specifications, industry standards (e.g., for environmental testing), and predefined performance characteristics that were considered acceptable or comparable to the predicate devices. For example, the "specified requirements" for dead space and pressure-flow characteristics served as the ground truth against which the device's performance was measured. Biocompatibility requirements per relevant standards also acted as ground truth.

    8. The sample size for the training set

    • Not applicable. This device is not an AI or machine learning system, so there is no training set.

    9. How the ground truth for the training set was established

    • Not applicable, as there is no training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1