510(k) Data Aggregation

    K Number
    K121938

    Validate with FDA (Live)

    Date Cleared
    2012-08-01

    (30 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    2 - 999
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K022544, K050108, K011703

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dornier's Medilas H RFID Laser Fibers are intended to be used as an accessory for the Dornier Medilas H Holmium Laser ("Laser"). This laser is intended for cutting, vaporization, ablation, and coaqulation of soft tissue in conjunction with endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, gastroscopes, and colonoscopes), or for open surgery for contact or non-contact surgery with or without a handpiece for use in incision/excision, vaporization, ablation and coagulation of soft tissue. The Laser is indicated for use in medicine and surgery, in the following medical specialties:

    Arthroscopy, Urology, Lithotripsy, Pulmonology, Gastroenterology, Gynecology, ENT, General Surgery

    Device Description

    The Medilas H RFID Laser Fiber is a modification to Dornier's currently cleared Dornier Medilas H Laser Fiber K022544.

    AI/ML Overview

    The provided text, K121938, describes Dornier's Medilas H RFID Laser Fibers, which is a modification of an already cleared device. This submission is a Special 510(k) notice, indicating that the device has the same intended use and similar indications, principles of operation, and technological characteristics as its predicate devices, with only minor differences that do not raise new questions of safety or effectiveness.

    Therefore, the submission primarily focuses on demonstrating that the modified device performs as safely and effectively as the predicate devices. This type of submission does not typically involve the extensive clinical studies with human subjects, ground truth establishment, or multi-reader, multi-case studies that would be associated with entirely new medical devices or software algorithms.

    Based on the provided document, here is the information requested, with "N/A" for criteria not addressed in this type of submission:


    1. Table of Acceptance Criteria and Reported Device Performance:

    The document broadly states that "Performance data demonstrates that the Dornier Medilas H RFID Laser Fibers are as safe and effective as the listed predicates." Specific quantitative acceptance criteria and detailed performance metrics are not typically required or presented in this type of 510(k) summary for accessory modifications. The "performance data" likely refers to bench testing to ensure the fiber's physical and functional properties (e.g., power transmission, durability, RFID functionality) are comparable to the predicate.

    Acceptance CriteriaReported Device Performance
    Safety equivalent to predicate devicesPerformance data demonstrates safety equivalent to predicate devices.
    Effectiveness equivalent to predicate devicesPerformance data demonstrates effectiveness equivalent to predicate devices.
    RFID functionality (implied)The RFID functionality, as a new feature, would have been tested to ensure it operates as intended (e.g., communication with the laser system), but specific metrics are not detailed in this summary.
    Physical and functional equivalence to predicate laser fibers (implied)Likely demonstrated through bench testing, but specific metrics are not detailed in this summary.

    2. Sample size used for the test set and the data provenance: N/A (No human subject test set is described for this type of submission.)

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A (Ground truth established by experts is not applicable to this type of device modification submission.)

    4. Adjudication method for the test set: N/A (No human subject test set or adjudication process is described.)

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: N/A (This is not an AI device, and no MRMC study was conducted.)

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: N/A (This is a physical medical device accessory, not an algorithm.)

    7. The type of ground truth used: N/A (Not applicable. The equivalence is based on engineering design, bench testing, and comparison to existing predicate devices, not clinical ground truth in the context of diagnostic accuracy.)

    8. The sample size for the training set: N/A (Not applicable, as this is a physical device accessory, not a machine learning algorithm.)

    9. How the ground truth for the training set was established: N/A (Not applicable.)

    Ask a Question

    Ask a specific question about this device

    K Number
    K120926

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-07-27

    (122 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K972272, K011207, K050108, K011703

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EndoBeam™ Holmium Laser Fibers are indicated for a variety of surgical uses including open, laparoscopic or endoscopic ablation, incision, excision, vaporization, and coagulation of soft and cartilaginous tissue and in surgical procedures involving vaporization, ablation and fragmentation of calculi.

    The delivery system may be used in surgical specialty or procedures for which compatible Holmium and Nd:YAG lasers have received regulatory clearance. Refer to the applicable laser system user manual for complete information regarding applications, contraindications, precautions and warnings when using this fiber.

    Device Description

    The EndoBeam™ Holmium Laser Fibers are free-beam delivery devices that transmit laser energy in a forward direction. The devices are either 2.5 meters (8.2 ft.) or 3.0 meters (9.8 ft) in length and are terminated with a laser specific connector on the proximal end. These delivery systems are capable of delivering Ho:YAG (2140nm) and Nd:YAG (1064nm). The devices are either single use or reusable and are supplied Ethylene Oxide (EtO) sterilized.

    The Bard Holmium series of laser fibers are fiber optic laser energy delivery devices consisting of an SMA 905 connector, strain relief, and a silica core/silica clad fiber jacketed with ethylene tetrafluoroethylene (ETFE). The fibers are further equipped with a polished, flat output tip. These fibers may be used in a variety of laser-based surgical cases as an integral part of laser systems.

    The line of reusable laser fibers will feature a protective sheath outside of the jacketed fiber and extending from the strain relief to the midpoint of the fiber length.

    AI/ML Overview

    This document describes the 510(k) premarket notification for the "EndoBeam™ Holmium Laser Fibers" (K120924/K120926). The information provided focuses on demonstrating substantial equivalence to predicate devices, rather than an AI/ML-based device requiring detailed performance metrics against a defined ground truth derived from expert consensus or pathology. Therefore, many of the requested fields are not applicable in this context.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The submission focuses on demonstrating substantial equivalence by showing that the EndoBeam™ Holmium Laser Fibers have the same technological characteristics and intended use as predicate devices, and that nonclinical performance testing confirms their safety and effectiveness. The acceptance criteria are implicitly met by successful completion of these nonclinical tests and by showing equivalence to legally marketed devices.

    Acceptance Criteria (Implied)Reported Device Performance
    Technological Characteristics EquivalenceThe EndoBeam™ Holmium Laser Fibers have the same technological characteristics as the predicate devices: silica core/cladding, hard-fluoropolymer coating, and ETFE fiber jacket.
    Intended Use EquivalenceThe EndoBeam™ Holmium Laser Fibers have the same intended use as the predicate devices, indicated for a variety of surgical uses including ablation, incision, vaporization, and coagulation of soft/cartilaginous tissue and fragmentation of calculi.
    Nonclinical Functional Performance (Safety & Effectiveness)Nonclinical functional performance testing included visual, functional, and tensile strength testing carried out per internal test methods and IEC 61754-22.
    BiocompatibilityNonclinical biocompatibility testing was conducted in accordance with ISO 10993-1:2009 and FDA Bluebook Memorandum G95-1.
    Substantial Equivalence to Predicate DevicesNonclinical test data demonstrate that the device is safe and effective and is substantially equivalent to the legally marketed predicate devices (Laser Peripherals Holmium Bare Fibers (K972272), Laser Peripherals Reusable Holmium Fiber (K011207), SureFlex™ Laser Fiber (K050108), SlimLineEZ™ Fiber Delivery Device (K011703)).

    Regarding the other requested information:

    This 510(k) submission is for a physical medical device (holmium laser fibers) and does not involve AI or algorithms that would require a test set, ground truth established by experts, or training data in the typical sense of AI/ML performance evaluation. Therefore, the following points are not applicable to this submission.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable: This filing does not describe a software or AI/ML device that uses a test set of data for performance evaluation. The "performance data" refers to nonclinical functional and biocompatibility testing of the physical device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable: There is no "ground truth" establishment by experts in the context of an AI/ML system for this type of device submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable: No test set or expert adjudication is described for this device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable: This is a physical laser fiber, not an AI-assisted diagnostic or treatment system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable: This is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable: Ground truth for AI/ML performance is not relevant here. The "ground truth" for this device's safety and effectiveness is established through adherence to recognized standards (e.g., ISO 10993-1:2009, IEC 61754-22) and by demonstrating equivalence to predicate devices that have prior regulatory clearance and established safety records.

    8. The sample size for the training set

    • Not Applicable: No training set is involved.

    9. How the ground truth for the training set was established

    • Not Applicable: No training set with ground truth is involved.