510(k) Data Aggregation

    K Number
    K060516

    Validate with FDA (Live)

    Date Cleared
    2006-05-15

    (77 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K043255, K052688

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intended Use for Ellipse I²PL is:

    Ellipse IPPL is intended for use in dermatology:

    • Hair removal (permanent hair reduction). .
    • Treatment of benign pigmented lesions (including, but not limited to solar . lentigines, ephilides, mottled pigmentation) and benign vascular lesions (including but not limited to diffuse redness, telangiectasias, port wine stains).
    • Treatment of inflammatory acne. .

    The Indications for Use for Ellipse I-PL are:

    ApplicationTreatment VariableFitzpatrick Skin Type
    123456
    Hair RemovalHR ApplicatorHR-S ApplicatorHair (Thin, Normal, Thick)
    Hair RemovalHR-D ApplicatorHair (Thin, Normal, Thick)
    Treatment of Benign Pigmented Lesionsand Benign Vascular LesionsPigmentation / Vessel Size
    Treatment of TelangiectasiasVessel size (Thin, medium, thick)
    Treatment of Port Wine StainsColor (Red, blue)
    Treatment of Individual Pigmented LesionsPigment Color
    Treatment of Inflammatory Acne
    Key: ✓ Allowed; ⊖ Not Allowed
    Device Description

    Ellipse I2PL is an Intense Pulsed Light (IPL) system used for long-term removal of unwanted hair and for treatment of sun-damaged skin, including uneven pigmentation, age spots, large pores, diffuse redness, and for the treatment of telangiectasias, port wine stains and inflammatory acne in the area of dermatology.

    The system consists of a console containing power unit and control electronics with control and display panel including software.

    Applicators/hand-pieces are connected to the system in order to generate light energy for treatment in the waveband 400 nm - 950 nm.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Danish Dermatologic Development A/S (DDD) Ellipse I2PL dermatologic IPL system. This document is a premarket notification for a medical device trying to demonstrate substantial equivalence to previously cleared devices, rather than a study designed to prove acceptance criteria through clinical trial data. Therefore, many of the requested elements about acceptance criteria and study design are not applicable or explicitly stated as would be found in a clinical study report.

    The primary "acceptance criteria" in a 510(k) submission is demonstrating substantial equivalence to a predicate device. The "study" proving this is essentially the comparison provided by the manufacturer.

    Here's an analysis based on the information provided:


    Table of Acceptance Criteria and Reported Device Performance

    As this is a 510(k) summary focused on substantial equivalence, there are no explicit quantitative "acceptance criteria" presented in the sense of performance thresholds (e.g., "device must achieve X accuracy"). Instead, the "acceptance criterion" is the demonstration of substantial equivalence to predicate devices, which implies similar safety and effectiveness for the stated Indications for Use. The "reported device performance" is implicitly that it performs as safely and effectively as the predicate devices.

    Acceptance Criteria (Implied by Substantial Equivalence Determination)Reported Device Performance
    Intended Use: Functionally identical to predicate devices.Met: Ellipse I2PL (B02) is stated to have identical modules, applications, and intended uses to Ellipse IPL (A01), and identical and substantially equivalent technology, modules, applications, and intended uses to Ellipse Flex PPT.
    Technological Characteristics: Similar to predicate devices, with no new issues of safety or effectiveness raised by differences.Met: Comparison matrix shows similar technology (IPL/broad spectrum), energy output, pulse duration, applicator spot size, charge time, and cooling method. Differences in energy spectrum (400-950nm for B02/Flex PPT vs. 555-950nm for A01) are noted but deemed not to raise new safety/effectiveness issues.
    Safety and Effectiveness: No significant differences in overall performance characteristics compared to predicate devices that would raise new questions of safety or effectiveness.Met: Danish Dermatologic Development A/S (DDD) concludes that no significant differences exist and the added applications do not raise new issues of safety and effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not applicable. This document is a 510(k) summary for substantial equivalence, not a clinical study report. It does not contain data from an external test set with a defined sample size for performance evaluation.
    • Data Provenance: Not applicable. The "data" here refers to the specifications and intended uses of the device and its predicates, as presented by the manufacturer, rather than clinical data from a patient population.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not applicable. No external test set requiring expert-established ground truth is mentioned in this submission. The "ground truth" for demonstrating substantial equivalence rests on comparing the device's technical specifications and intended uses against already cleared predicate devices.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. There was no test set requiring ground truth adjudication. The FDA's review process for 510(k) submissions involves regulatory experts evaluating the manufacturer's claims and comparison data.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is an Intense Pulsed Light (IPL) system, not an AI diagnostic or assistance device. Therefore, no MRMC study or AI-related effectiveness metrics are relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical medical device (IPL system), not a standalone algorithm.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: For a 510(k) clearance based on substantial equivalence, the "ground truth" is established by the previously cleared predicate devices and their established safety and effectiveness profiles. The manufacturer's submission effectively argues that the new device's characteristics align sufficiently with these predicates such that it poses no new questions of safety or effectiveness. There is no specific "expert consensus, pathology, or outcomes data" presented as ground truth for this specific device's performance beyond what was established for its predicates.

    8. The Sample Size for the Training Set

    • Not applicable. This device does not involve machine learning or an algorithm that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. This device does not involve machine learning or an algorithm that requires a training set with established ground truth.