    K Number
    K083561

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-02-11

    (71 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K043195

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neoss ProActive Implant is for single-stage and two-stage surgical procedures and cement or screw retained restorations. The Neoss ProActive Implant is intended for immediate placement and function on single tooth and /or multiple tooth applications recognizing sufficient bone stability and appropriate occlusal loading, to restore chewing function. Multiple tooth applications may be splinted with a bar.

    Device Description

    The Neoss ProActive is an modified version of the Neoss Bimodal Implant (K043195), Identical to K043195, the ProActive Implant is a threaded, internal abutment connection, root-form titanium dental implant. The ProActive assortment consists of a number of implants with a diameter of Ø3,5 to Ø5,5 mm and lengths between 7,0 - 19.0 mm having the same internal abutment dimension independent of implant diameter. The internal connection being equipped with interlocking elements for an insertion tool and the non-rotational locking of the abutment. Supplied sterile.

    AI/ML Overview

    The provided 510(k) summary for the Neoss ProActive Implant does not contain the detailed information requested regarding acceptance criteria and a study proving device performance as typically understood for AI/ML-based medical devices.

    This submission is for a dental implant, which is a physical device, not a software or AI/ML diagnostic tool. Therefore, the traditional metrics of sensitivity, specificity, accuracy, and detailed ground truth generation processes that would be relevant for devices making diagnostic claims are not applicable or present in this type of submission.

    Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (Neoss Bimodal Implant K043195) due to a modification in the surface finish process. The performance is assessed through comparisons of physical characteristics and biological responses in animal models, rather than a clinical study with a human test set and expert adjudication.

    Here's a breakdown of what is and is not available based on the provided text:

    1. Table of acceptance criteria and reported device performance:

    Acceptance CriteriaReported Device Performance
    Implied/General:
    Compliance to device specificationsAll functions verified to operate as designed.
    Safety and Effectiveness equivalent to predicate device (K043195)Demonstrated through similar intended use, operating principle, basic device design, physical properties, and materials.
    Increased strength of osseointegration when compared to machined surfaces (animal testing)Exhibit an increase in the strength of osseointegration
    Increased bone to implant contact when compared to grit-blasted surfaces (animal testing)Exhibit an increased bone to implant contact

    Note: These are based on biological responses in animal testing and general performance verification, not metrics like sensitivity or specificity.

    2. Sample size used for the test set and data provenance:

    • The document mentions "animal testing" but does not specify the sample size, type of animals, or country of origin for these tests. This is a pre-clinical, non-human study.
    • The study is retrospective in the sense that it's comparing a modified device to a previous one, but the specific animal tests themselves would be prospective experiments.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable as the "ground truth" for a physical dental implant's performance would be established through mechanical testing, material characterization, and biological response in animal models, not expert clinical interpretation of images or other data.

    4. Adjudication method for the test set:

    • Not applicable for this type of device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a physical dental implant, not an AI/ML diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical dental implant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The closest concept to "ground truth" here would be the biological outcomes observed in animal models (e.g., strength of osseointegration, bone-to-implant contact) and engineering/material properties verified during performance testing.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device that requires a training set. The "development" of the device involves engineering design and pre-clinical testing.

    9. How the ground truth for the training set was established:

    • Not applicable.

    In summary: The provided 510(k) pertains to a hardware modification (surface finish) of an existing dental implant. The safety and effectiveness data provided are based on pre-clinical animal testing and engineering verification, not clinical studies with human participants or the detailed performance metrics typically associated with AI/ML-based diagnostic devices.