Found 2 results

510(k) Data Aggregation

    K Number
    K090720

    Validate with FDA (Live)

    Date Cleared
    2009-06-17

    (91 days)

    Product Code
    Regulation Number
    874.3302
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cochlear Baha BP100 sound processor is intended for use with the Baha auditory osseointegrated implant (for children aged 5 and older, or adults), or with the Baha Headband or Baha Softband (no age limitations), for the following patients and indications:

    • Patients who have a conductive or mixed hearing loss and can still benefit from sound amplification. The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) should be better than or equal to 45 dB HL.
    • Bilateral fitting of the BP100 is intended for patients who meet the above criterion in both ears, with bilaterally symmetric moderate to severe conductive or mixed hearing loss. Symmetrical bone-conduction thresholds are defined as less than a 10 dB average difference between ears (measured at 0.5, 1, 2, and 3 kHz), or less than a 15 dB difference at individual frequencies.
    • Patients who suffer from unilateral sensorineural deafness in one ear with normal hearing in the other ear (i.e. single-sided deafness or "SSD"). Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL.
    • Baha for SSD is also indicated for any patient who is indicated for an airconduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
    Device Description

    The Baha implant system works by combining a sound processor with an abutment and a small titanium implant placed in the skull behind the ear. The system is based on the process of "osseointegration" through which living tissue integrates with titanium in the implant. Thus, the titanium implant becomes one with the bone, allowing high-quality amplified and processed sound to be conducted via the skull bone directly to a cochlea with residual functionality. Baha sound processors can also be fitted as external (non-implanted) bone-conduction vibrators with either the Baha Headband or Baha Softband as an alternative to a conventional bone-conduction hearing aid.

    The Baha BP100 will be the newest addition to a current family of three marketed sound processors for use with the Baha auditory osseointegrated implant, or with the Baha Headband/Softband setup. In its initial release, it will provide a moderate gain ear-level sound processor that will serve as a premium-feature option to the currently marketed Baha Divino. At a later date, the BP100 platform may be used to provide a range of programmable products to meet end-user needs.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Cochlear Baha BP100, a bone conduction hearing aid. The submission focuses on demonstrating substantial equivalence to predicate devices rather than a de novo clinical study with specific acceptance criteria and performance metrics.

    Here's an analysis based on the provided text, addressing your questions:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state formal "acceptance criteria" in terms of specific performance metrics (e.g., sensitivity, specificity, accuracy) for a new clinical study. Instead, the primary "acceptance criterion" for this 510(k) submission is demonstrating substantial equivalence to existing, legally marketed predicate devices.

    The reported device performance is framed in terms of comparability to the predicate Baha Divino.

    Aspect of Performance/ComparisonAcceptance Criterion (Implicit)Reported Device Performance
    Gain/OutputTo be "comparable to" or "essentially equivalent (or better) than" the Baha Divino sound processor."Bench testing shows that the gain/output of the Baha BP100 are comparable to those for the marketed Baha Divino sound processor..." "Bench testing shows that the Baha BP100 is essentially equivalent (or better) than the Baha Divino in terms of gain/output..." "Bench testing supports the proposed Indications For Use, since gain/output are equivalent to the Baha Divino..."
    Intended UseTo have the "same Indications For Use" as the Baha Divino.The BP100 offers gain/output comparable to the Baha Divino, and "Thus, it will be marketed for use for conductive or mixed hearing loss patients who have average bone-conduction thresholds up to < 45 dB HL (across 0.5, 1, 2, and 3 kHz), the same Indications For Use as cleared for marketing of the Baha Divino." The stated Indications for Use in the submission for BP100 match those of the Baha Divino for the relevant patient population.
    Technological CharacteristicsTo have "substantially equivalent technology" to other marketed sound processors in the Baha family."It has substantially equivalent technology, and intended use, as the other three currently marketed sound processors for the Baha system (Baha Divino, Baha Intenso, Baha Cordelle II)."
    Software/Programming SystemThe programming software and interface should be "substantially equivalent to other programming software that has been used for fitting of hearing devices that have digital signal processing algorithms."The programming software and interface "...is substantially equivalent to other programming software that has been used for fitting of hearing devices that have digital signal processing algorithms." The fitting software is "...substantially equivalent to other programming software that has been used for fitting of hearing devices that have digital signal processing algorithms."
    Direct Bone Conduction (DBC) MeasurementThe intended use and process to accomplish DBC measurements should be "substantially equivalent to a procedure used by a previous hearing device cleared for marketing under the 510(k) process."The production of pure tones through the hearing device for standard audiometric measurements to be used in a prescription has the "same intended use, function, and technology as the predicate [K984547 for the LGOB test using the Resound digital 5000 series hearing devices]."
    Safety and Efficacy (Overall)To be considered safe and effective for the indicated patient population."Published data in the literature support the safety and efficacy of the Baha Divino for patients with average bone-conduction thresholds of < 45 dB HL (e.g. Kompis et al., 2007)." "...further clinical performance data are unnecessary to prove that the BP100 can successfully fit the same patient population..."

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: No specific sample size for a "test set" or a new clinical study involving human subjects for the Baha BP100 is reported. The primary evidence presented is bench testing comparing the BP100 to predicate devices.
    • Data Provenance: The document refers to "Published data in the literature" (Kompis et al., 2007) to support the safety and efficacy of the Baha Divino (a predicate device). This suggests the provenance is retrospective clinical literature on a different device, not new data generated specifically for the BP100. Bench testing data would be generated from the device itself.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable as no new clinical study with a "test set" requiring expert-established ground truth was conducted for the Baha BP100 for this submission. The submission relies on bench testing and previously published clinical data on predicate devices.

    4. Adjudication Method for the Test Set:

    This information is not applicable for the same reason as above. No new clinical test set requiring adjudication was performed.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not reported. The submission explicitly states that "further clinical performance data are unnecessary" given the bench testing equivalence to the Baha Divino and its existing clinical literature.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    The device is a hearing aid, which requires a human user (patient). The "algorithm" here refers to the digital signal processing within the hearing aid. The document implicitly refers to standalone device performance in terms of its gain/output characteristics as measured by bench testing, independently of human interaction beyond fitting. However, this is not an "algorithm-only" study in the typical sense of AI diagnostics, but rather engineering performance of the device's internal processing.

    7. The Type of Ground Truth Used:

    For the demonstration of substantial equivalence, the "ground truth" implicitly relies on:

    • Bench Testing Data: Direct measurements of the BP100's gain/output characteristics, compared to similar measurements for the Baha Divino.
    • Predicate Device Clinical Performance: The established safety and efficacy of the Baha Divino (supported by published literature like Kompis et al., 2007) served as the "ground truth" for clinical performance for the patient population, which the BP100 is then argued to serve equally well due to its comparable technical performance.

    8. The Sample Size for the Training Set:

    This information is not applicable. The Baha BP100 is a medical device, not an AI/ML algorithm that is "trained" in the typical sense. Its digital signal processing algorithms are designed and programmed, not learned from a dataset.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as there is no "training set" for the device's inherent functionality described in the conventional AI/ML context. The device's design and programming are based on established audiological principles and engineering practices for hearing aids.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082108

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-11-14

    (112 days)

    Product Code
    Regulation Number
    874.3302
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K042017

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OBC is intended for the following patients and indications;

    • Patients with conductive or mixed hearing losses, who can still benefit from amplification of the sound. The pure tone average (PTA) bone conduction (BC) threshold for the indicated ear should be better than 45 dB HL (measured at 0.5, 1, 2 and 3 kHz).
    • Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5. 1. 2 and 4 kHz, or less than 15 dB at individual frequencies.
    • Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear, who for some reason will not or cannot use an AC CROS. The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should then be better than 20 dB HL (measured at 0.5, 1, 2 and 3 kHz).
    Device Description

    The OBC bone anchored hearing aid system consists of an external sound processor unit and an implant with a skin penetrating abutment. The implant with the abutment is surgically anchored in the bone behind the ear. Vibrations generated by the sound processor are transmitted directly through the skull bone to the cochlea as bone conduction sound. The sound processor has a coupling so that it can be easily connected to and disconnected from the abutment by the user. The sound processor can alternatively be connected to headband accessories, to function as a conventional bone conductor.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Oticon Medical OBC Bone Anchored Hearing Aid System, seeking substantial equivalence to a predicate device, the BAHA Divino. The provided text primarily focuses on demonstrating substantial equivalence rather than presenting an extensive study with specific acceptance criteria and performance metrics typically found for novel device claims.

    However, based on the information provided, we can infer the "acceptance criteria" relate to the functional and performance characteristics that need to be "substantially equivalent" to an existing, legally marketed device. The "study" here is essentially the comparison and verification process detailed in the submission.

    Here's a breakdown of the requested information:

    1. Table of acceptance criteria and the reported device performance

    Since this is a substantial equivalence claim, the "acceptance criteria" are implicitly matching or performing equivalently to the predicate device, BAHA Divino. The "reported device performance" is the claim of equivalence.

    Acceptance Criteria (Inferred from Predicate Device BAHA Divino)Reported Device Performance (OBC Bone Anchored Hearing Aid System)
    Design: Bone conduction sound processor connected to an implant surgically placed in the bone behind the deaf earBone conduction sound processor connected to an implant surgically placed in the bone behind the deaf ear (Yes, substantially equivalent)
    Intended Use: Improvement of hearing for patients with conductive and mixed hearing losses, bilateral fitting and single sided deafnessImprovement of hearing for patients with conductive and mixed hearing losses, bilateral fitting and single sided deafness (Yes, substantially equivalent)
    Material: Implant: Titanium; Sound processor coupling: PEEKImplant: Titanium; Sound processor coupling: PEEK (Yes, substantially equivalent)
    Power Requirement: Zinc-air batteryZinc-air battery (Yes, substantially equivalent)
    Max Gain: 33dB33 dB (Yes, substantially equivalent)
    Frequency Response: 125 Hz - 8 kHz125 Hz - 8kHz (Yes, substantially equivalent)
    Sound Processing: DigitalDigital (Yes, substantially equivalent)
    Safety and Efficacy: No new issues of safety and efficacy compared to predicateNo new issues of safety and efficacy compared to predicate (Concluded based on laboratory testing and comparison)

    2. Sample size used for the test set and the data provenance

    The document states: "Laboratory testing of the OBC system was conducted to determine device functionality and conformance to design input requirements." However, it does not specify a sample size for a test set (e.g., number of patients or devices tested). It also does not mention data provenance in terms of country of origin or whether it was retrospective or prospective, as the testing described appears to be internal laboratory testing for functional equivalence.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This submission is for substantiating equivalence based on technical specifications and functionality testing against a predicate device, not typical clinical performance evaluation requiring expert ground truth for a diagnostic or therapeutic outcome.

    4. Adjudication method for the test set

    Not applicable. There's no mention of an adjudication process as there isn't a clinical trial or performance study involving human interpretation of results requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a hearing aid system, not an AI-powered diagnostic tool, and the submission does not describe an MRMC study or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a bone-anchored hearing aid, which requires human interaction (wearing and using it) and is not an algorithm-only standalone system. The "standalone" performance here refers to the device's functional characteristics as described in the comparison table.

    7. The type of ground truth used

    The "ground truth" in this context is the established functional and performance characteristics of the predicate device, BAHA Divino. The OBC system's performance was compared against these established characteristics using laboratory testing. The submission confirms that the "intended use, function and procedure" of the OBC are substantially equivalent to the BAHA Divino.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device where a "training set" would be used in the conventional sense. The "training" for this device would refer to its design and engineering process, not data-driven model training.

    9. How the ground truth for the training set was established

    Not applicable. As above, there is no "training set" in the context of AI/ML device development.