510(k) Data Aggregation

    K Number
    K071849

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-12-04

    (518 days)

    Product Code
    Regulation Number
    888.3045
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K040419, K041168

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AlloFuse is indicated for orthopedic applications as filler for gaps or voids that are not intrinsic to the stability of the bony structure. AlloFuse® is indicated to be packed gently into bony gaps in the skeletal system as a bone graft extender (extremities, spine, and pelvis) and as bony void filler of the extremities and pelvis. These defects may be surgically created or from the result of traumatic injury to the bone.

    Device Description

    AlloFuse is derived from selected donated human bone tissue that has been processed into particles. The bone particles are subsequently demineralized using a hydrochloric acid process. The demineralized bone matrix (DBM) is combined with an inert reverse phase carrier and formulated into a gel or putty-like consistency.

    The carrier is a solution of polyethylene oxide polypropylene oxide block copolymer dissolved in water exhibiting reverse phase characteristics (i.e. an increase in viscosity as temperature increases).

    AI/ML Overview

    Here's an analysis of the provided text, outlining the acceptance criteria and the studies mentioned for the AlloFuse Gel and Putty device:

    Executive Summary:

    The provided document describes the 510(k) submission for AlloFuse Gel and Putty, which claims substantial equivalence to the predicate device, DynaGraft® II. The primary "acceptance criteria" is demonstrating substantial equivalence due to the proposed device being the same in design, materials, and function as the predicate device. This is supported by:

    1. Viral Inactivation Validation: Testing of the DBM processing method to ensure viral inactivation.
    2. Osteoinductive Potential: An in vivo assay in athymic rats confirms osteoinductive potential for every lot.
    3. Product Performance Testing: Reference to studies on the predicate device (DynaGraft® II) in rabbit and sheep models (radiographic and histological) and human clinical studies for spinal fusions.

    1. Table of Acceptance Criteria and Reported Device Performance

    It's important to note that the document does not explicitly present a table of numerical "acceptance criteria" and direct performance metrics for the AlloFuse device itself in the way one might see for a diagnostic AI model. Instead, the acceptance is based on demonstrating substantial equivalence to a predicate device. The performance data presented primarily pertains to the predicate device or the processing methods.

    Acceptance Criterion (Implicit)Reported Device Performance
    Premarket Requirement: Demonstrated Substantial Equivalence to Predicate Device (DynaGraft® II)The proposed device is stated to be "the same device in design, materials of construction and function" as the previously cleared DynaGraft® II. AlloSource also holds an exclusive license to manufacture the predicate device. This is the primary claim for substantial equivalence.
    Material Safety (Viral Inactivation)The DBM processing method (for AlloFuse) was evaluated for its viral inactivation potential, demonstrating "significant viral inactivation potential for a wide range of potential human viruses" using a panel of viruses.
    Biological Activity (Osteoinductive Potential)AlloFuse has shown "osteoinductive potential in athymic rats." "Every lot of final product is tested via an in vivo assay to ensure osteoinductive potential."
    Clinical Performance / EfficacyNot directly tested on AlloFuse in human trials for this submission. Instead, the submission relies on: - Product Performance Testing (Predicate): DBM in RPM carrier (DynaGraft® II formulation) evaluated in rabbit and sheep models by radiographic and histological methods. - Clinical Studies (Predicate): Clinical studies using DynaGraft® II DBM Putty and Gel for spinal fusions "demonstrating acceptable outcomes."
    Sterility & Donor Tissue QualityAlloFuse is provided sterile and for single patient use. Donor bone meets AATB requirements.

    2. Sample Size Used for the Test Set and Data Provenance

    Due to the nature of this 510(k) submission, direct "test sets" for clinical performance of the new device (AlloFuse) are not reported. The submission relies on data from the predicate device and specific lab tests for the new device:

    • Viral Inactivation Validation:
      • Sample Size: Not specified for the "select panel of viruses," but assumed to be multiple viruses representing various types.
      • Provenance: Lab-based (in-vitro) testing of the DBM processing method.
    • Osteoinductive Potential:
      • Sample Size: Assumed to be adequate for "an in vivo assay to ensure osteoinductive potential of the final product," performed for every lot of final product.
      • Provenance: Athymic rat model (animal study), conducted by the manufacturer.
    • Product Performance Testing (Predicate Device - DynaGraft® II):
      • Sample Size: Not specified for the rabbit and sheep models.
      • Provenance: Animal models (rabbit, sheep), nature (retrospective/prospective) not specified but typically prospective for such studies.
    • Clinical Studies (Predicate Device - DynaGraft® II):
      • Sample Size: Not specified.
      • Provenance: Human clinical studies, typically prospective. Country of origin not specified, but likely US-based given the FDA submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Given this is a 510(k) for a device claiming substantial equivalence rather than a new drug or complex diagnostic, explicit "ground truth" established by a panel of human experts for a specific test set (like an AI study) is not detailed.

    • Viral Inactivation: Ground truth would be based on established virology protocols and detection methods, likely performed by qualified microbiologists/virologists in a laboratory setting.
    • Osteoinductive Potential: Ground truth is established by histological and potentially radiographic evaluation of new bone formation in the athymic rat model, assessed by experts in pathology and histology.
    • Product Performance Testing (Predicate - Animal Models): Ground truth was established by radiographic and histological methods, interpreted by relevant veterinary or medical experts (e.g., veterinary radiologists, pathologists).
    • Clinical Studies (Predicate - Human): Ground truth (acceptable outcomes in spinal fusions) would have been established by clinicians (e.g., orthopedic surgeons, radiologists) based on follow-up examinations, imaging, and patient outcomes.

    4. Adjudication Method for the Test Set

    No explicit adjudication method (e.g., 2+1, 3+1) is mentioned, as this is not an AI diagnostic study relying on human reader consensus. Outcomes from animal and human studies would be assessed via standard clinical or laboratory practices, which often involve multiple reviewers but not necessarily a formal adjudication process as seen in AI studies.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This 510(k) is for a bone void filler device, not an AI diagnostic tool. Therefore, there's no discussion of human readers improving with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This submission is for a medical device (bone void filler), not an algorithm or AI system.

    7. The Type of Ground Truth Used

    • Viral Inactivation Validation: Established lab-based measurements of viral reduction.
    • Osteoinductive Potential: Histological evidence of new bone formation in an in vivo athymic rat model.
    • Product Performance Testing (Predicate): Radiographic and histological findings in animal models.
    • Clinical Studies (Predicate): Clinical outcomes (e.g., successful spinal fusion) and patient follow-up data.

    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" in the context of an AI algorithm for this device. The development of the manufacturing processes and material formulation would be based on general scientific knowledge and prior product development, not a labeled dataset for training an algorithm.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no AI training set, there is no ground truth established for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072265

    Validate with FDA (Live)

    Date Cleared
    2008-01-09

    (147 days)

    Product Code
    Regulation Number
    888.3045
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K060794, K050690, K071206, K040262, K041168

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PROGENIX™ DBM Putty is intended for use as a bone graft substitute in bony voids or gaps of the skeletal system not intrinsic to the stability of the bony structure (i.e. spine, pelvis and extremities). The voids or gaps may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. PROGENIX™ DBM Putty provides a bone void filler that is resorbed/remodeled and is replaced by host bone during the healing process. When used in the extremeties or pelvis, the device is used by itself. When used in the spine, the device must be mixed with autograft bone and used as a bone graft extender.

    Device Description

    PROGENIX™ DBM Putty contains human demineralized bone matrix (DBM) in a biocompatible carrier. The carrier is a mixture of bovine collagen with a natural polysaccharide (sodium alginate). The components are mixed in phosphate buffered saline to achieve a flowable or moldable consistency. PROGENIX™ DBM Putty is a single use product intended for use as a bone graft substitute, bone graft extender, and bone void filler in bony voids or gaps of the skeletal system (i.e. spine, pelvis and extremities) not intrinsic to the stability of the bony structure. Additionally, this product is not designed to impart any mechanical strength to the surgical site. PROGENIX™ DBM Putty is provided in ready-to-use malleable forms that may be molded or manipulated by the surgeon into various shapes. This product has been shown to be osteoconductive as well as osteoinductive in an athymic rat assay, allowing for bony ingrowth across the graft site while resorbing at a rate consistent with bony healing.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Medtronic Sofamor Danek PROGENIX™ DBM Putty

    This document describes the acceptance criteria and the study used to demonstrate the performance of the PROGENIX™ DBM Putty, as outlined in the provided 510(k) summary (K072265).

    1. Table of Acceptance Criteria and Reported Device Performance

    The primary acceptance criteria for the expanded indication of PROGENIX™ DBM Putty (use as a bone graft extender in spinal fusion procedures) were based on demonstrating its effectiveness in producing spinal fusion.

    Acceptance CriterionReported Device Performance (as demonstrated in the rabbit study)
    Spinal Fusion Effectiveness
    Radiographic fusionEffectiveness in producing spinal fusion by radiographic criteria
    Manual palpation for fusionEffectiveness in producing spinal fusion by manual palpation
    OsteoconductivityDemonstrated
    OsteoinductivityDemonstrated (in an athymic rat assay)

    2. Study Design and Data Provenance

    The study that demonstrated the expanded indication for PROGENIX™ DBM Putty was a rabbit study.

    • Sample Size: Not explicitly stated in the provided text, but the study was conducted on a sufficient number of rabbits to "suggest that PROGENIX™ DBM Putty is effective."
    • Data Provenance: The study was conducted in a pre-clinical animal model (rabbit). No country of origin is specified for the study location. The study is prospective as it evaluates the device's performance in a controlled environment.

    3. Number of Experts and Qualifications for Ground Truth

    The provided text does not specify the number of experts used or their qualifications for establishing the ground truth of the rabbit study. However, the assessment of spinal fusion (radiographic and manual palpation) would typically involve trained veterinary radiologists and/or surgeons.

    4. Adjudication Method

    The adjudication method used for the rabbit study is not explicitly stated. However, given the nature of radiographic assessment and manual palpation, it is likely that the evaluation was performed by one or more trained individuals.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was performed as this is a pre-clinical animal study for a bone graft substitute, not a diagnostic imaging device.

    6. Standalone Performance (Algorithm Only)

    Not applicable. This device is a bone graft substitute, not an algorithm or software. Its performance is assessed through biological outcomes in vivo.

    7. Type of Ground Truth Used

    The ground truth used in the rabbit study was based on direct observation and objective assessments of fusion in an animal model:

    • Radiographic criteria: Imaging evidence of spinal fusion.
    • Manual palpation criteria: Physical assessment of spinal fusion.

    8. Sample Size for the Training Set

    Not applicable. This is a biological device, not an AI/ML algorithm that requires a training set. The "study" refers to the pre-clinical validation for the expanded indication.

    9. How Ground Truth for Training Set Was Established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device