510(k) Data Aggregation

    K Number
    K092913

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-10-16

    (24 days)

    Product Code
    Regulation Number
    870.1750
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K040207

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EP-4 Cardiac Stimulator is indicated for use during clinical cardiac electrophysiology procedures. The EP-4 Cardiac Stimulator is intended to be used for diagnostic electrical stimulation of the heart for the purpose of refractory measurements, initiation and termination of tachyarrhythmias, and measurements of electrical conduction.

    Device Description

    The EP-4 Cardiac Stimulator consists of a touch-screen computer, an AC-powered stimulation module, which provides electrical stimulation to the heart during electrophysiology case studies. The EP-4 consists of a touch-screen computer that controls a two- or four-channel AC-powered stimulation module, and a keyboard. The stimulator is capable of performing single channel, simultaneous, or sequential stimulation at programmed output settings using existing/predicate programmed protocols. The device is independent and self-contained, with the exception of marker outputs and an external sync input connector. It can be configured as a stand-alone system or in conjunction with the EP-WorkMate™ Recording System. It may also be used with other EP recording systems.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the EP-4™ Cardiac Stimulator, which is an external programmable pacemaker pulse generator. The purpose of this submission is to demonstrate substantial equivalence to a predicate device, the EP-4™ Clinical Stimulator (K040207).

    The document does not contain acceptance criteria in terms of numerical performance metrics (e.g., sensitivity, specificity, accuracy, or specific physiological parameters with defined thresholds) or a formal study designed to "prove" the device meets such criteria through a statistical analysis against a ground truth.

    Instead, the submission focuses on demonstrating substantial equivalence to a predicate device. This is a regulatory pathway where a new device is shown to be as safe and effective as a legally marketed device that is not subject to premarket approval (PMA). The "study" described is primarily bench testing to confirm that design changes did not adversely affect safety or efficacy.

    Here's a breakdown of the requested information based on the provided text, highlighting what is present and what is not:


    1. A table of acceptance criteria and the reported device performance

      The document does not provide a table of acceptance criteria with numerical performance metrics. The "acceptance criteria" in the context of this 510(k) submission are implicitly tied to demonstrating that the EP-4 Cardiac Stimulator is "substantially equivalent" to its predicate device (EP-4 Clinical Stimulator).

      The reported "performance" is described qualitatively:

      • "Bench testing was performed to confirm that the changes met design requirements and did not adversely affect the safety or efficacy of the product."
      • "Where design differences exist between the proposed device and the predicate device, performance testing demonstrated that these differences do not adversely affect the device's safety and effectiveness."
      • The device has the "same intended use and fundamental scientific technology" and "all technological characteristics... are substantially equivalent" to the predicate.

      No specific numerical performance values (e.g., stimulation accuracy within a certain mV or ms, battery life, etc.) that would typically be associated with explicit acceptance criteria are provided in this regulatory summary.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      Not applicable. The document describes "bench testing," which refers to laboratory testing of the device itself, not a clinical study involving patients or a dataset of patient data. Therefore, there is no "test set" in the context of patient data, nor is there information on data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      Not applicable. As this was bench testing for substantial equivalence, clinical ground truth establishment in the traditional sense (e.g., by medical experts for diagnostic accuracy) was not the focus. The "ground truth" for bench testing would be engineering specifications and predicate device performance, verified by engineers and quality control personnel. No information on specific expert qualifications for this type of verification is provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      Not applicable, as there is no specific "test set" of patient data requiring clinical adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      No. The EP-4 Cardiac Stimulator is a medical device for electrical stimulation, not an AI-assisted diagnostic tool. Therefore, an MRMC study and AI-related effectiveness are not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      Not applicable. This device is a hardware system with embedded software for cardiac stimulation, not an algorithm being tested in isolation. The device's functionality is inherently "standalone" in operation once programmed, but this is distinct from testing an AI algorithm's performance without human interaction.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      For bench testing, the "ground truth" would be established against the design requirements, engineering specifications, and performance characteristics of the predicate device. The goal was to confirm that the changes "met design requirements" and did not "adversely affect the safety or efficacy" when compared to the predicate. This is verified through a combination of standard engineering tests, electrical measurements, and functional checks.

    8. The sample size for the training set

      Not applicable. The device's software is menu-driven and firmware-based, controlling stimulation functions. It's not described as a machine learning system that requires a "training set" in the AI sense. The software's development followed "St. Jude Medical's Quality System requirements" and "design controls requirements documented in 21 CFR 820.30," which dictate rigorous testing and verification, but not "training data" in the AI context.

    9. How the ground truth for the training set was established

      Not applicable, as there is no "training set" for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092810

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-10-09

    (28 days)

    Product Code
    Regulation Number
    870.1425
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K040207

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EP-WorkMate™ System is indicated for use during clinical electrophysiology procedures.

    The EP-WorkMate® system with an EP-4 stimulator is intended to be used for diagnostic electrical stimulation of the heart for the purpose of refractory measurements, initiation and termination of tachy-arrhythmias, measurements of electrical conduction, and arrhythmia mapping.

    Device Description

    The EP-WorkMate system is a computer-based electrophysiological recording and monitoring system that is used to capture, display, store, and retrieve surface and intracardiac electrical signals during electrophysiology studies. It consists of a computer. two 21" high-resolution monitors, a multi-channel signal amplifier and filtering system (signal conditioning unit), a catheter junction box, and a laser printer. The system may also be configured with an integrated EP-4 clinical stimulator and touch-screen computer monitor (cleared in K040207).

    The EP-WorkMate is connected to electrophysiology catheters that are guided into various locations within the heart, and to surface electrocardiogram (ECG) cables. Intracardiac and ECG signals are then acquired from electrodes on the indwelling catheters and ECG leads, and transmitted to the amplifier, which amplifies and conditions the signals before they are received by the EP-WorkMate computer for measurement and display.

    During the procedure, cardiac signals are acquired and an automated software waveform detector (trigger) performs online recognition of cardiac activation on preselected leads. Temporal interval measurements are computed on multiple channels on a beat-by-beat basis and dynamically displayed on the real-time display. Menu-driven software is utilized for data acquisition and analysis, interval posting, and instant data retrieval with waveform markers and intervals displayed.

    Signals are also presented on a review monitor for measurement and analysis. Continuous capture of the digitized signals can be invoked, and the user can also retrieve and display earlier passages of the current study without interruption of the realtime display. The system can also acquire, display and record data from other interfaced devices in use during the procedure, such as imaging devices and ablation generators,

    AI/ML Overview

    This 510(k) summary does not contain information typically found in a study proving a device meets acceptance criteria, especially for AI/ML-driven devices. This is a submission for an electrophysiology recording and monitoring system, which as described, appears to be a hardware and software system for acquiring, displaying, and storing physiological signals, rather than an AI/ML diagnostic tool.

    The document states that "Bench testing was performed to confirm that the changes met design requirements and did not affect the safety or effectiveness of the product." This indicates an engineering validation approach rather than a clinical study with acceptance criteria for diagnostic performance.

    Given the provided text, I cannot extract answers for many of your questions, as they pertain to clinical studies, AI performance metrics, and expert adjudication, which are not detailed in this submission.

    Here's what I can infer from the provided text, addressing your questions where possible:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a quantitative, performance-based manner typical for diagnostic device efficacy (e.g., sensitivity, specificity, accuracy). Instead, it states that "Bench testing was performed to confirm that the changes met design requirements and did not affect the safety or effectiveness of the product." This implies acceptance related to meeting engineering specifications and maintaining safety and effectiveness attributes similar to the predicate device.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    Not applicable or not provided. The bench testing performed would likely involve hardware and software validation testing, not a clinical test set with patient data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable or not provided. Ground truth establishment by experts is typically relevant for diagnostic AI/ML devices or clinical studies with human interpretation, which is not described here.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable or not provided.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is not described as an AI assistance tool for human readers in a diagnostic context.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The description is of a system that acquires and displays signals for human interpretation, not an algorithm providing a standalone diagnostic output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable or not provided. For "bench testing," ground truth would likely be based on known signal inputs or predefined engineering specifications rather than clinical outcomes or expert consensus.

    8. The sample size for the training set

    Not applicable. The device is not described as an AI/ML model requiring a training set in the conventional sense.

    9. How the ground truth for the training set was established

    Not applicable.


    Summary of what is known about the "study":

    • Type of Study: Bench testing.
    • Purpose: To confirm that changes in the device met design requirements and did not affect the safety or effectiveness of the product.
    • Methodology: The development was performed in accordance with St. Jude Medical's Quality System requirements and in compliance with Quality System Regulation design control requirements (21 CFR 820.30). A Declaration of Conformity with Design Controls was provided.
    • Conclusion: Performance testing demonstrated that any operational and performance differences from the predicate device do not adversely affect the device's safety and effectiveness, leading to a substantial equivalence determination.