Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K172314

    Validate with FDA (Live)

    Date Cleared
    2017-09-20

    (50 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K033594, K162005, K052983

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ampleye Scleral RGP Lens (roflufocon D, roflufocon E, hexafocon A, paflufocon D) for daily wear is indicated for use for the management of irregular astigmatism, corneal degeneration or dystrophy caused by keratoconus, keratoglobus, pellucid marginal degeneration (PMD), post corneal trauma/scaring, post keratoplasty, post K-Pro, post RK, post LASIK, Salzmann's nodular degeneration. Cogan's dystrophy, granular corneal dystrophy, lattice corneal dystrophy or Reis-Bucklers dystrophy.

    The Ampleye Scleral RGP Lens (roflufocon D, roflufocon E, hexafocon A, paflufocon D) is also indicated for therapeutic management of ocular surface disease from dry eye including ocular pemphigoid, Stevens-Johnson syndrome, symblepharon formation, graft vs host disease, persistent epithelial defect, exposure keratitis, neurotrophic keratopathy( herpes zoster, familial dysautonomia), Sjogern's syndrome, filamentary keratitis, limbal stem cell deficiency, atopy, ectodermal dysplasia. When prescribed for therapeutic use for irregular astigmatism or ocular surface diseases, the Ampleye Scleral RGP Lens may also provide correction of refractive error including myopia, presbyopia and regular astigmatism.

    Eye care practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

    Device Description

    The Ampleye Scleral RGP Lens (roflufocon D, roflufocon E, hexafocon A, paflufocon D) for daily wear is a large diameter rigid gas permeable contact lens design that vaults over the cornea and rests on the conjunctiva overlying the sclera. The Ampleye Scleral RGP Lens is lathe cut from one of the following hydrophobic, FDA Group #3 fluoro-silicone acrylate materials: roflufocon D, roflufocon E, hexafocon A, or paflufocon D.

    AI/ML Overview

    The provided document is a 510(k) summary for a contact lens device, the Ampleye Scleral RGP Lens. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than undergoing a full clinical study with acceptance criteria in the way a novel therapeutic or diagnostic device would.

    Therefore, the information typically requested in your prompt regarding acceptance criteria and detailed study results for device performance (e.g., accuracy, sensitivity, specificity, statistical significance with effect size) is not present for this contact lens device in this particular document.

    Here's how to interpret the document in the context of your request:

    1. Table of Acceptance Criteria and Reported Device Performance:

    • No specific acceptance criteria for a "device performance" study (e.g., accuracy metrics) are provided.
    • Instead, the "performance" demonstrated is the ability to manufacture lenses to established finished product specifications within ANSI Z80.20-2016 tolerance and bioburden levels within acceptance criteria. These are manufacturing and safety standards, not clinical performance metrics in the sense of predictive or diagnostic accuracy.
    Acceptance Criteria (Manufacturing/Safety)Reported Device Performance (Ampleye Scleral RGP Lens)
    Lenses manufactured within ANSI Z80.20-2016 tolerances for various parameters (Base Curve, Center Thickness, Diameter, Spherical Power, Cylindrical Power, Cylindrical Axis, Multifocal Power).All lenses were manufactured to established finished product specifications within the ANSI Z80.20-2016 tolerance.
    Bioburden: < 100 CFU per lensColony Forming Units (CFU) per lens was within the established acceptance criteria of less than 100 CFU per lens.

    2. Sample size used for the test set and the data provenance:

    • Test Set for Manufacturing Verification: The document states "Bench Testing-manufacturing verification testing was conducted to demonstrate the ability of Art Optical Contact Lens, Inc. to manufacture lenses, on a repeatable basis, from supplied lens blanks to a variety of prescribed parameters." It does not specify the sample size for this bench testing.
    • Test Set for Bioburden Testing: "Bioburden testing conducted on rigid gas permeable lenses manufactured at Art Optical Contact Lens, Inc." The sample size for bioburden testing is not specified.
    • Data Provenance: The testing was "conducted at Art Optical Contact Lens, Inc.", implying the data is generated internally by the manufacturer. It is "retrospective" in the sense that it's performed on manufactured lenses to meet a standard, but not a prospective clinical trial.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This document describes manufacturing and safety testing, not a study where expert-established ground truth is used for performance evaluation (e.g., for an AI algorithm).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a study requiring expert adjudication of results.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was done. This is a contact lens device, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This device does not involve an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For Manufacturing Verification: The "ground truth" or reference standard would be the ANSI Z80.20-2016 tolerance specifications.
    • For Bioburden Testing: The "ground truth" or reference standard would be the established acceptance criteria of < 100 CFU per lens, likely derived from microbiological standards.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established:

    • Not applicable. There is no "training set" for ground truth establishment.

    Summary of the document's approach to "proving" substantial equivalence:

    The document primarily relies on:

    • Reference to previously cleared 510(k)s (K033594, K162005, and K052983) for the safety and effectiveness of the contact lens materials (roflufocon D, E, hexafocon A, paflufocon D). This means the materials themselves have already undergone review and met safety/effectiveness standards.
    • Bench testing to confirm manufacturing capabilities to specific ANSI standards.
    • Bioburden testing to ensure manufacturing hygiene and product safety.
    • Comparison to predicate devices (BostonSight PD Prosthetic Device, K161461, and Custom Stable RGP Scleral Contact Lens, K170335) across various characteristics like materials, production method, classification, intended use, and physical properties to demonstrate that the new device does "not raise different questions of safety and effectiveness."

    This approach is typical for contact lens 510(k) submissions, where the focus is on established material safety and equivalence of design/manufacturing to existing cleared devices, rather than a direct clinical performance study for accuracy or similar metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171950

    Validate with FDA (Live)

    Date Cleared
    2017-08-16

    (48 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K033594, K071043, K071266

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Visionary Optics Scleral Contact Lens for daily wear is indicated for use for the management of multiple ocular conditions, such as, degenerations that leads to an irregular corneal shape (e.g. keratoconus, keratoglobus, pellucid marginal degeneration, Salzmann's Nodular Degeneration), dystrophies (e.g. Cogan's dystrophy, Granular Corneal Dystrophy, Lattice Corneal Dystrophy,), post-surgery (e.g. corneal transplant, LASIK, radial keratotomy), and corneal scarring from infection or trauma.

    The Visionary Optics Scleral Contact Lens for daily wear is also indicated for therapeutic management of ocular surface disease including dry eye (e.g. ocular manifestations of Graftversus-Host disease. Sjogren's syndrome, dry eve syndrome), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical and thermal burns, radiation, filamentary keratitis), epidermal ocular disorders, disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex. Herpes zoster. Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for distorted cornea or ocular surface disease, the Visionary Optics Scleral Contact Lens may incidentally provide correction of refractive error in persons with myopia, hyperopia, astigmatism or presbyopia.

    Eye care practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

    Device Description

    The Visionary Optics Scleral Contact Lens (roflufocon D, roflufocon E, hexafocon A, hexafocon B) for daily wear is a large diameter rigid gas permeable contact lens design that vaults over the cornea and rests on the conjunctiva overlying the sclera. The Visionary Optics Scleral Contact Lens is lathe cut from one of the following hydrophobic, FDA Group #3 fluorosilicone acrylate materials: roflufocon D, roflufocon E, hexafocon A, or hexafocon B.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device, the Visionary Optics Scleral Contact Lens. This type of submission focuses on demonstrating "substantial equivalence" to existing legally marketed devices, rather than conducting new clinical trials to prove safety and effectiveness from scratch. Therefore, the information typically requested about acceptance criteria and detailed study designs for a new device's performance may not be fully present in this document.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the traditional sense for a new clinical study. Instead, it relies on demonstrating that the physical and material properties of the Visionary Optics Scleral Contact Lens are within established specifications and comparable to predicate devices. The "reported device performance" is largely inferred from meeting these specifications and the substantial equivalence claim.

    ParameterAcceptance Criteria (Implicit from Tolerances & Predicates)Reported Device Performance (Visionary Optics Scleral Contact Lens)
    Bench Testing (Manufacturing Verification)All lenses manufactured to established finished product specifications within ANSI Z80.20 tolerance.All lenses were manufactured to established finished product specifications within the ANSI Z80.20 tolerance.
    Bioburden TestingLess than 100 CFU per lens.Colony forming units (CFU) per lens was within the established acceptance criteria of less than 100 CFU per lens.
    Physical Parameters (Example)
    Base Curve± 0.2mmRange: 5.5mm to 25.00mm, Tolerance: ± 0.2mm (Implicitly met)
    Center Thickness± 0.1mmRange: 0.10mm to 3.00mm, Tolerance: ± 0.1mm (Implicitly met)
    Diameter± 0.20mmRange: 12.00mm to 26.00mm, Tolerance: ± 0.20mm (Implicitly met)
    Spherical PowerVarious (e.g., ± 0.12 (0 to = 5D))Range: -35.00 D to +35.00 D, Tolerances as specified (Implicitly met)
    Cylindrical PowerVarious (e.g., ± 0.25 (0 to = 2D))Range: +10.00 D to -10.00 D, Tolerances as specified (Implicitly met)
    Cylindrical Axis± 5°Range: 1° to 180°, Tolerance: ± 5° (Implicitly met)
    Bifocal Add± 0.25DRange: +.12 D to +6.00 D, Tolerance: ± 0.25D (Implicitly met)
    Material Properties (Example - Roflufocon D)
    Refractive Index- (Compared to predicate)1.4333
    Light Transmission (clear)>97% (Compared to predicate)>97%
    Water Content<1% (Compared to predicate)<1%
    Oxygen Permeability (Dk)- (Compared to predicate)$100 \times 10^{-11} \frac{cm^2}{sec} \frac{ml O_2}{ml \times mm Hg @ 35°C}$
    UV Light Blocking>98% UVB, >95% UVA (Compared to predicate)>98% UVB, >95% UVA

    Note: The acceptance criteria for many of the physical and material properties are implicit in the comparison to the predicate devices and the fact that the new device states these values. The document asserts that the new device "demonstrate no significant differences from the predicate devices-supporting the substantial equivalence claim."

    2. Sample Size for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not specify a distinct "test set" sample size for a clinical trial involving human subjects. The non-clinical studies mention "All lenses" for manufacturing verification and "rigid gas permeable lenses manufactured at Visionary Optics LLC" for bioburden testing, implying a representative sample from their production. For clinical evidence, it relies on "a series of clinical case reports" and references previous 510(k) clearances (K033594, K071043, and K071266) for the materials used.
    • Data Provenance:
      • Non-Clinical (Bench & Bioburden): From Visionary Optics LLC's own manufacturing and testing facility. This is prospective for the purpose of this submission (i.e., new tests performed for this device).
      • Clinical: Primarily by reference to previous 510(k) submissions for the lens materials. The "series of clinical case reports" would typically be retrospective, gathering outcomes from actual use of the device. No country of origin is explicitly stated for these case reports or the referenced 510(k) clinical data, but for FDA submissions, it's generally assumed to be U.S. or internationally recognized standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable or provided in the context of this 510(k) submission. This is not a study requiring expert-derived ground truth as would be found in a diagnostic AI study. The "ground truth" for the non-clinical tests would be the measured physical properties and bioburden counts. For the clinical case reports, the "ground truth" would be the patient outcomes as documented by the eye care practitioners who prescribed the lenses.

    4. Adjudication Method for the Test Set

    Not applicable. There is no expert adjudication process described for the non-clinical tests or the clinical case reports in this submission.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This type of study is common for diagnostic imaging devices involving human readers. For contact lenses, the evaluation focuses on physical parameters, material safety, and clinical performance (patient outcomes), not on the interpretation of images by multiple readers. The submission relies on "substantial equivalence" to predicate devices and existing data for the materials.

    6. Standalone Performance (Algorithm Only)

    Not applicable. This device is a physical contact lens, not an algorithm or AI. Therefore, the concept of "standalone (algorithm only)" performance does not apply.

    7. Type of Ground Truth Used

    • Non-Clinical (Bench Testing): Established engineering specifications and measurements (e.g., dimensions, powers).
    • Non-Clinical (Bioburden Testing): Microbiological culture results.
    • Clinical (Relevant to Referenced 510(k)s and Case Reports): Patient physiological responses, visual acuity, comfort, and other clinical outcomes observed by eye care practitioners.

    8. Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm that requires a "training set." The materials and manufacturing processes are established based on validated engineering and material science principles.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set for an algorithm is involved.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170335

    Validate with FDA (Live)

    Date Cleared
    2017-03-24

    (50 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K033594

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Custom Stable(TM) Rigid Gas Permeable Scleral Contact Lenses for daily wear are indicated for use for the management of multiple ocular conditions, such as, degenerations that lead to an irregular corneal shape (e.g. keratoconus. keratoglobus, pellucid marginal degeneration, Salzmann's Nodular Degeneration), dystrophy, granular corneal dystrophy, Lattice Corneal Dystrophy), post-surgery (e.g. corneal transplant, LASIK, radial keratotomy), and corneal scarring. The lens may also be prescribed for the management of ocular surface diseases (e.g. dry eye syndrome, Keratoconjunctivitis Sicca (Graft vs Host Disease, Sjogren's syndrome, Filamentary Keratitis), limbal stem cell deficiency, epidermal ocular disorders, neurotrophic keratitis, and corneal exposure/lagophthalmos). When prescribed for therapeutic use, the Custom Stable RGP Scleral Lenses is also indicated for correction of refractive error in persons with myopia, hyperopia or presbyopia.

    Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

    Device Description

    The Custom Stable "11 Rigid Gas Permeable Scleral Contact Lens for daily wear is a large diameter rigid gas permeable contact lens design that vaults over the cornea and rests on the conjunctiva overlying the sclera. The Custom Stable "" Rigid Gas Permeable Scleral Contact Lens is lathe cut from one of the following hydrophobic, fluoro-silicone acrylate materials:

    • · roflufocon D (supplied by Contamac Ltd.)
    • · roflufocon E (supplied by Contamac Ltd.)
    AI/ML Overview

    The device described is the Custom Stable™ Rigid Gas Permeable Scleral Contact Lens. This document assesses its substantial equivalence to predicate devices rather than providing a study where the device meets acceptance criteria directly. However, based on the provided information, we can infer acceptance criteria from the performed tests and the comparison to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Criteria/TestAcceptance CriteriaReported Device Performance
    Bench Testing (Manufacturing Verification)All lenses manufactured to established finished product specifications within ANSI Z80.20 tolerance.All lenses were manufactured to established finished product specifications within the ANSI Z80.20 tolerance.
    Bioburden TestingColony forming units (CFU) per lens less than 100.Colony forming units (CFU) per lens was less than 1.
    Substantial Equivalence (General)No significant differences from predicate devices in terms of safety and effectiveness. Does not raise different questions of safety and effectiveness.Testing demonstrates no significant differences from the predicate devices, supporting the substantial equivalence claim. No new questions of safety and effectiveness are raised.
    Components/Materials/FormulationEquivalent to predicate devices (roflufocon D & E contact lens materials, and potentially others as per prednisone).Uses roflufocon D & E contact lens materials, matching predicate devices.
    FDA Classification & Group NumberClass II, Group #3 Fluoro Silicone Acrylate.Class II, Group #3 Fluoro Silicone Acrylate.
    Lathe Cut Manufacturing ProcessUtilizes lathe-cut manufacturing.Utilizes lathe-cut manufacturing process.
    Large Diameter (Scleral) DesignHas a large diameter scleral design.Has a large diameter (scleral) design.
    Actions and Intended UseActions and intended use are equivalent to predicate devices.Actions and intended use are equivalent to predicate devices, extending indications based on clinical case reports.
    Therapeutic Indications for UseTherapeutic indications for use are equivalent or justified compared to predicate devices.Therapeutic indications for use include conditions like keratoconus, dystrophies, post-surgery, and ocular surface diseases, supported by clinical case reports and compared to predicates.
    Physical Properties(Implicit/Explicit from predicate and material specifications): Refractive Index, Light Transmission, Water Content, Dynamic Contact Angle, Specific Gravity, Modulus, Shore D Hardness, Oxygen Permeability (Dk), UV Light Blocking.Roflufocon D: Refractive Index 1.4333, Light Transmission >97% (clear), >90% (tinted), Water Content <1%, Dynamic Contact Angle 3°, Specific Gravity 1.166, Modulus 697 MPa, Shore D Hardness 75, Dk 100, UV Blocking >98% UVB, >95% UVA. Roflufocon E: Refractive Index 1.4332, Light Transmission >97% (clear), >90% (tinted), Water Content <1%, Dynamic Contact Angle 6°, Specific Gravity 1.155, Modulus 77 MPa, Shore D Hardness 77, Dk 125, UV Blocking >98% UVB, >95% UVA.
    Lens Parameters(Implicit/Explicit from predicate and manufacturing capabilities): Chord Diameter, Center Thickness, Base Curve, Spherical Powers.Chord Diameter: 14.8 mm to 17.8 mm, Center Thickness: 0.20 mm to 0.40 mm, Base Curve: 6.6 mm to 11.0 mm, Spherical Powers: -30.00 D to +30.00 D (0.125 D steps).

    2. Sample Size Used for the Test Set and Data Provenance:

    • Bench Testing: The document does not specify a separate "test set" sample size for bench testing. It indicates "All lenses were manufactured..." suggesting testing was conducted on a representative sample or a substantial portion of manufactured lenses to verify production capabilities. The provenance of this data is internal to Valley Contax, Inc. (manufacturer), and is prospective in relation to the manufacturing verification.
    • Bioburden Testing: The sample size is not explicitly stated, but bioburden testing was conducted on "rigid gas permeable lenses manufactured at Valley Contax, Inc." The data provenance is internal to Valley Contax, Inc. and is prospective from the manufacturing process.
    • Clinical Studies: The document refers to "a series of clinical case reports" for performance demonstration. The specific sample size for these case reports is not given, nor is the country of origin. These are likely retrospective case reports from practitioners fitting the lenses.
    • Reference to Predicate Devices: For the bulk of safety and effectiveness, the device relies on existing data from predicate devices (K033594 - Optimum GP and K161461 - BostonSight PD). The sample sizes and provenance for those original approvals are not detailed in this document.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This document describes a 510(k) submission, which primarily focuses on demonstrating substantial equivalence to legally marketed predicate devices, not on a de novo clinical trial with extensive ground truth establishment by external experts for a new device's performance.

    • For the internal bench and bioburden testing, the "ground truth" is defined by established specifications (ANSI Z80.20 tolerance for manufacturing, less than 100 CFU for bioburden), which are likely set by industry standards and potentially internal quality assurance experts. No specific number or qualifications of "experts" are provided.
    • For the "series of clinical case reports," the implicit "experts" are the eye care practitioners who prescribed and managed the patients. Their qualifications are not specified but would typically be optometrists or ophthalmologists.
    • For the safety and effectiveness data derived from predicate devices, the "ground truth" and expert involvement would have been established during the approval process of those predicate devices, but this information is not included in the current document.

    4. Adjudication Method for the Test Set:

    • Bench Testing & Bioburden Testing: No explicit adjudication method is mentioned beyond compliance with pre-defined acceptance criteria (e.g., ANSI standards, CFU limits). This suggests direct measurement and comparison to thresholds rather than a multi-reader review process.
    • Clinical Case Reports: Given they are "case reports," it's unlikely a formal adjudication method (like 2+1, 3+1 consensus) was used. Case reports typically document individual patient outcomes as observed and reported by a single practitioner or clinical team.
    • Predicate Device Data: No information on adjudication methods for the predicate device studies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:

    No MRMC comparative effectiveness study was mentioned, nor is there any indication of AI involvement in this device. This is a contact lens, not an AI-powered diagnostic or assistive technology.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    No standalone (algorithm only) performance was evaluated, as this device is a physical contact lens and does not involve an algorithm.

    7. The Type of Ground Truth Used:

    • Bench Testing: Ground truth is based on engineering and manufacturing specifications (e.g., ANSI Z80.20 tolerance for lens parameters).
    • Bioburden Testing: Ground truth is based on microbiological safety standards (CFU count).
    • Clinical Case Reports: Ground truth is based on observed clinical outcomes and practitioner assessments of patient response to the lens (e.g., comfort, visual acuity, ocular health).
    • Predicate Device Data: Ground truth would have been established during the original predicate approvals, likely through clinical trials and performance testing against objective measures and clinical endpoints.

    8. The Sample Size for the Training Set:

    This is a physical medical device (contact lens), not an AI/machine learning algorithm. Therefore, there is no "training set" in the context of data used for algorithm development.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161461

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-07-25

    (60 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K033594, K022128, K071266

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Boston Sight PD Prosthetic Device for daily wear is indicated for the management of a distorted corneal surface that:

    1. precludes satisfactory spectacle lens correction
    2. demonstrates significant improved rigid contact lens corrected vision
    3. is incapable of wearing traditional corneal lenses because of the inability to achieve adequate lens centration/stability and/or tolerance to physical contact with a lens
      Causes of corneal distortion include corneal degeneration (e.g. keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration), corneal dystrophy (e.g. lattice dystrophy, Reis-Bucklers dystrophy), and scarring from surgery (e.g. corneal transplant, LASIK, radial keratotomy), infection, or trauma.
      The BostonSight PD Prosthetic Device for daily wear is also indicated for therapeutic use in eyes with ocular surface disease from dry eye (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, limbal stem cell deficiency (e.g.Stevens-Johnson syndrome, chemical and thermal burns, radiation), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for distorted cornea or ocular surface disease, the BostonSight PD Prosthetic Device may incidentally provide correction of refractive error.
      The BostonSight PD Prosthetic Device may be cleaned and disinfected using a chemical (not heat) care system.
    Device Description

    The BostonSight PD Prosthetic Device is a daily wear, prosthetic device for the ocular surface lathe cut from one of the following fluoro-silicone acrylate materials:

    • · roflufocon D (supplied by Contamac Ltd.)
    • · roflufocon E (supplied by Contamac Ltd.)
    • · oprifocon A (supplied by Bausch & Lomb, Inc.)
    • · hexafocon B (supplied by Bausch & Lomb, Inc.)
      The BostonSight PD Prosthetic Device is designed to vault over the cornea and rest on the conjunctiva overlying the sclera, resulting in a tear reservoir between the back surface of the prosthetic device and the corneal surface. The tear reservoir masks optical distortions from an irregular corneal surface, and in combination with the device itself, protects the ocular surface from an adverse external environment, including but not limited to dysfunctional evelids and margins. The design parameters are customized to allow for tear exchange underneath the device.
      The BostonSight PD Prosthetic Device may be shipped "dry" or "wet" in a polypropylene contact lens case. The primary container for shipping the BostonSight PD Prosthetic Device is the Bausch & Lomb Frequent Replacement Contact Lens Case, with clearance under 510(k) K896685. When shipped "wet". The BostonSight PD Prosthetic Device manufactured from material supplied by Bausch & Lomb, inc. may be packaged and shipped in Boston Advance Comfort Formula Conditioning Solution (K974466) or Boston SIMPLUS Multi-Action solution (K024289). The BostonSight PD Prosthetic Device manufactured from material supplied by Contamac, Ltd. may be packaged and shipped "wet" in in the OPTIMUM by Lobob Cleaning and Disinfecting Storage solution, with clearance under 510(k) K014162.
    AI/ML Overview

    This document is a 510(k) Summary of Safety and Effectiveness for the BostonSight PD Prosthetic Device, a rigid gas permeable contact lens. As such, it does not contain a study or data proving the device meets acceptance criteria. Instead, it asserts substantial equivalence to predicate devices based on device characteristics, indications for use, and a reliance on prior clearances for materials and clinical performance of similar devices.

    Therefore, most of the requested information regarding acceptance criteria, sample sizes, expert involvement, and statistical analyses cannot be extracted from this document. However, I can provide the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a quantitative manner for performance. Instead, it relies on substantial equivalence to predicate devices, implying that if the new device is sufficiently similar to the legally marketed predicate devices, it is considered safe and effective. The table below summarizes the comparison to predicate devices, which implicitly defines the "performance" as being comparable in key design and material aspects.

    CharacteristicBostonSight PD Prosthetic Device (Subject Device)BostonSight IC Corneal & Scleral Lens (K153066) (Predicate Device)Boston® Scleral (itafluorofocon B) RGP Contact Lens (P860022/S40) (Predicate Device)
    Indication for UseTherapeutic use in eyes with ocular surface disease from dry eye, limbal stem cell deficiency, disorders of the skin, neurotrophic keratitis, and corneal exposure. May incidentally correct refractive error.Correction of refractive error in aphakic and non-aphakic persons. For management of irregular corneal conditions (keratoconus, pellucid marginal degeneration, post-keratoplasty or LASIK).Significantly reduced vision due to distorted corneal surface: 1. precludes satisfactory spectacle lens correction 2. demonstrates significantly improved RGP contact lens corrected vision 3. incapable of wearing traditional corneal lenses Also for ocular surface disorders benefiting from physical protection and saline bath.
    Device ClassificationClass II, Lenses, Rigid Gas Permeable, Daily Wear, HQDClass II, Lenses, Rigid Gas Permeable, Daily Wear, HQDClass II, Lenses, Rigid Gas Permeable, Daily Wear, HQD
    Production MethodLathe-cutLathe-cutLathe-cut
    FDA Group #Group # 3 Fluoro Silicone AcrylateGroup # 3 Fluoro Silicone AcrylateGroup # 3 Fluoro Silicone Acrylate
    Water Content<1%<1%<1%
    UV Absorber/BlockerYESYESYES
    Materialsroflufocon D, roflufocon E, oprifocon A, hexafocon B(Implied similar types, as it's the primary predicate for materials, but not explicitly listed in this table.)itafiuorofocon B
    Physical Properties(Specific Dk, refractive index, etc. are listed for each material on page 6, considered comparable to predicates.)(Not detailed in this summary table, but assessed in K153066 which is referenced)(Not detailed in this summary table, but assessed in P860022/S40 which is referenced)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This document does not describe a new study with a test set. Instead, it relies on prior clearances for predicate devices and materials. Therefore, no information on sample size for a test set or data provenance is available here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no new study with a test set requiring ground truth establishment is described in this submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a contact lens, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable, as no new study with a test set requiring ground truth establishment is described in this submission. The "ground truth" for contact lens performance and safety is typically established through clinical trials submitted for PMA devices or through extensive post-market experience for predicate devices.

    8. The sample size for the training set

    Not applicable, as no new clinical study with a training set is described.

    9. How the ground truth for the training set was established

    Not applicable.

    Summary based on the provided document:

    The BostonSight PD Prosthetic Device gained 510(k) clearance by asserting substantial equivalence to existing legally marketed predicate devices. The "study" proving it meets acceptance criteria is primarily an analysis of its characteristics (materials, design, intended use) compared to previously cleared devices.

    Specifically, the submission cross-references:

    • Non-clinical studies and part of the clinical assessment: BostonSight IC Corneal & Scleral Lens (K153066).
    • Therapeutic indications for use and device design: Boston® Scleral (itafluorofocon B) RGP Contact Lens (P860022/S40).
    • Clinical studies for specific materials:
      • OPTIMUM GP (roflufocon D, roflufocon E) Daily Wear Contact Lenses (K033594)
      • Boston EQUALENS II (oprifocon A) Rigid Gas Permeable Contact Lenses (K022128)
      • Boston XO2 (hexafocon B) Daily Wear Contact Lens (K071266)

    The document specifies that the new device does not raise different questions of safety and effectiveness than the identified predicate devices. Therefore, direct "acceptance criteria" and "device performance" in the context of a new, standalone study (as is often the case for AI/diagnostic devices) are not presented.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1