Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K241314

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-16

    (98 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    KLS Mini Osteosynthesis System (K943347): The KLS Mini Osteosynthesis System is indicated for 1) Fractures, 3) Reconstruction procedures of the craniomaxillofial skeletal system.
    KLS Chin Plate System (K943348): The KLS Chin Plate System is indicated for 1) Fractures, 3) Reconstruction procedures of the craniomaxillofacial skeletal system.
    KLS-Martin Micro Osteosynthesis System (1.0MM) (K944561): The KLS-Martin Micro Osteosynthesis System (1.0MM) is used in oral-maxillo-cranio-facial surgery to stabilize fractures. The bone segments are attached to the plate with screws to prevent movement of the segments.
    KLS-Martin Micro Osteosynthesis System (1.5MM) (K944565): The KLS-Martin Micro Osteosynthesis System (1.5MM) is used in oral-maxillo-cranio-facial surgery to stabilize fractures. The bone segments are attached to the plate with screws to prevent movement of the segments.
    KLS Martin Centre-Drive Drill-Free Screw (K971297): The KLS Martin Centre-Drive Drill-Free Screws are in rigid internal fixation of the oral-maxillo-cranio-facial bones. The bone screws are used to anchor plates which are contoured to fit the bone fragments. The addition of the self drilling feature is the only difference between the submitted device and the predicate device referenced.
    KLS-Martin Temporary Condylar Implant (K990667): The KLS-Martin Temporary Condylar Implant is only intended for temporary reconstruction of the mandibular condyle in patients who have undergone resective procedures to remove malignant or benign the removal of the mandibular condyle. This device is not for permanent implantation, for patients with TMF or treatment of temporomandibular joint disease (TMD).
    KLS-Martin Mandibular/Reconstruction System II (K032442): The KLS-Martin Mandibular/Reconstruction System II is intended for use in the stabilization of mandibular fractures and mandibular reconstruction.
    KLS-Martin Ortho Anchorage System (K033483): The KLS-Martin Ortho Anchorage System is intended to be surgically placed in the mouth for use an an anchor for orthodontic procedures.
    KLS-Martin Ortho Anchorage System (Plates) (K040891): The KLS-Martin Ortho Anchorage System (Plates) are implants intended to be surgically placed in the mouth for use as an anchor for orthodontic procedures in patients.
    KLS Martin Rigid Fixation - Sterile (K060177): The KLS Martin Rigid Fixation - Sterile is in sterile packaging, osteosynthesis products with the following indications for use:
    K051236: The RESORB-X® SF Sonotrode is only intended for use for insertion of the RESORB-X® SF pins.
    K032442: The KLS Martin Mandibular/Reconstruction System II is intended for use in the stabilization of mandibular fractures and mandibular reconstruction.
    K971297: The KLS Martin Centre-Drive Drill-Free screws are in rigid internal fixation of the oral-maxillo-cranio-facial bones. The bone screws are used to anchor plates which are contoured to fit the bony surface and stabilize the bone fragments. The addition of the self drilling feature is the only difference between the predicate device reference
    K944565: The KLS-Martin Micro Osteosynthesis System is used in oral-maxillo-cranio-facial surgery to stabilize fractured bone segments. The bone segments are attached to the plate with screws to prevent movement of the segments.
    K944561: The KLS-Martin Micro Osteosynthesis System is used in oral-maxillo-cranio-facial surgery to stabilize fractured bone segments. The bone segments are attached to the plate with screws to prevent movement of the segments.
    KLS Martin Drill-Free MMF Screw (K042573): The KLS Martin Drill-Free MMF Screws is intended for use in maxilonandibular fixation of fractures of the maxilla, mandible, or both.
    Drill Free MMF Screw (K083432): The Drill Free MMF Screw is intended for use in maxillomandibular fixation of fractures of the maxilla, mandible, or both.
    KLS Martin L1 MMF System (K173320): The KLS Martin L1 MMF System is intended for temporary stabilization of maxillary fractures. It is designed to maintain proper occlusion during intraoperative bone healing (app. 6-8 weeks). It is indicated for the temporary treatment of maxillomandibular fixation (MMF) in adults or adolescents who have permanent teeth present (ages 12 and older).

    Device Description

    KLS Mini Osteosynthesis System (K943347): The KLS Mini Osteosynthesis System consists of titanium non-locking plates ranging in thickness from 0.6mm - 2.5mm and titanium screws ranging in diameter from 1.5mm - 2.3mm.
    KLS Chin Plate System (K943348): The KLS Chin Plate System consists of titanium plates ranging in thickness of 0.6mm and titanium screws ranging in diameter from 1.5mm - 2.3mm.
    KLS-Martin Micro Osteosynthesis System (1.0mm) (K944561): The KLS-Martin Micro Osteosynthesis System is designed to aid in the alignment and stabilization of the skeletal system after a facial fracture or surgery. The bone plates, bone plates, bone screws and accessories of various shapes and sizes for use in oral-maxillo-cranio-facial surgery. The bone plates are manufactured from CP Titanium and range in thickness from 0.3mm - 0.6mm. The bone screws are manufactured from Titanium Alloy and range in diameter from 1.0mm - 1.2mm.
    KLS-Martin Micro Osteosynthesis System (1.5mm) (K944565): The KLS-Martin Micro Osteosynthesis System is designed to aid in the alignment and stabilization of the skeletal system after a facial fracture or surgery. The bone plates and screws of various shapes and sizes for use in oralmaxillo-cranio-facial surgery. The bone plates are manufactured from CP Titanium and range in thickness from 0.3mm - 0.6mm. The bone screws are manufactured from Titanium Alloy and range in diameter from 1.5mm - 1.8mm.
    KLS Martin Centre-Drive Drill-Free Screw (K971297): The KLS Martin Centre-Drive Drill-Free Screws are designed to eliminate the need for pre-drilled pilot holes. They are self-tapping with one step insertion. They are intended for use in rigid internal fixation of the oral-maxillo-cranio-facial bones. The bone screws are used to anchor plates where are contoured to fit the bone fragments. The bone fragments. The bone screws are manufactured from Titanium Alloy and range in diameter from 1.0mm - 2.0mm.
    KLS-Martin Temporary Condylar Implant (K990667): The KLS-Martin Temporary Condylar Implant is a solid condylar head which attaches with fastening screws to a KLS-Martin Fracture/ Reconstruction Plate. The implant is available for left and right placement. The KLS-Martin Temporary Condy intended for temporary reconstruction of the mandibular condyle in patients who have undergone resective procedures to benign tumors requiring the removal of the mandibular condyle. This device is not for permanent implantation, for patients with TMJ or traumatic injuries, or for treatment of temporomandibular joint disease (TMD).
    KLS-Martin Mandibular/Reconstruction System II (K032442): The KLS-Martin Mandibular/Reconstruction System II includes several different designs of titanium plates and screws intended for use in the stabilization and fixation of mandibular fractures and reconstruction. The plates are manufactured from either CP Titanium or Titanium Alloy and range in thickness from 1.0mm - 3.0mm. The screws are manufactured from either CP Titanium Alloy and range in diameter from 2.0mm - 3.2mm.
    KLS-Martin Ortho Anchorage System (K033483): The KLS-Martin Ortho Anchorage System consists of a titanium screw designed to aid in dental movement by providing a rigid skeletal fixation point. The screw is intended to be surgically placed in the mouth for orthodontic procedures. The screws are manufactured from either CP Titanium or Titanium Alloy.
    KLS-Martin Ortho Anchorage System (Plates) (K040891): The KLS-Martin Ortho Anchorage System (Plates) consists of titanium non-locking plates to aid in dental movement by providing a rigid skeletal fixation point. The plates are manufactured from either CP Titanium Alloy and are fixated with titanium screws and are utilized as an anchor for orthodontic procedures in the palatal, maxilla or mandible region.
    KLS-Martin Drill-Free MMF Screw (K042573): The KLS-Martin Drill-Free MMF Screw provides temporary occlusal and fracture stabilization. These screws may be applied prior to or after exposure of the fracture. The KLS-Martin Drill-Free MMF Screw is in maxillomandibular fixation to provide stabilization of fractures of the maxilla, or both. The screws are manufactured from either CP Titanium Alloy and are provided in 2.0mm diameter with lengths ranging from 8mm - 12mm.
    KLS Martin Rigid Fixation - Sterile (K060177): The KLS Martin Rigid Fixation - Sterile includes titanium plates of various shapes and thickness, titanium screws of various length and diameter, stainless steel twist drills of various length and stainless steel sonotrode tips that are provided in sterile packaging. The KLS Martin Rigid Fixation - Sterile is intended to provide KLS Martin's previously cleared osteosynthesis products in sterile packaging.
    Drill Free MMF Screw (K083432): The Drill Free MMF Screw provides temporary occlusal and fracture stabilization. These screws may be applied prior to or after exposure of the fracture. The Drill Free MMF Screw is in maxillomandibular fixation to provide stabilization of fractures of the maxilla, mandible, or both. The screws are manufactured from Stainless Steel and are provided in 2.0mm diameter with lengths ranging from 8mm - 12mm.
    KLS Martin L1 MMF System (K173320): The KLS Martin L1 MMF System is a bone-borne maxillomandibular fixation (MMF) system consisting of metalic archbars with sliding locking plates that attach to the dental arches with screws. The system is intended to provide temporary stabilization of mandibular and maxillary fractures as well as maintain properative bone fixation and postoperative bone healing (app. 6-8 weeks). The patient is brought into occlusion by wiring around the archbar wire hooks. The L1 MMF system plates are manufactured from CP Titanium (ASTM F67), are available in either a 7-hole siding plate configuration with two different lengths, and are 0.5mm in plate thickness. The L1 MMF system sliding locking plates are fixated with either 2.0 x 6 mm selfdrilling locking screws manufactured from Ti-6Al-4V (ASTM F136). Implants are available both sterile. The system also includes the necessary instruments to facilitate placement of the implants.

    AI/ML Overview

    The document describes the KLS Martin Oral-Max Implants - MR Conditional, a bundled submission of various osteosynthesis systems and screws intended for use in craniomaxillofacial surgery. The purpose of this submission is to support the conditional safety and labeling modification of these devices in the magnetic resonance (MR) environment.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are implied by the non-clinical tests conducted to support MR Conditional safety, aligning with relevant ASTM standards and FDA guidance. The reported device performance is that the devices can be safely scanned under specified conditions.

    Acceptance Criteria (from ASTM/FDA Guidance)Reported Device Performance (Summary from Submission)
    Magnetically induced displacement force within acceptable limits (ASTM F2052-21)Not explicitly quantified but implied as acceptable for MR Conditional labeling.
    Magnetically induced torque within acceptable limits (ASTM F2213-17)Not explicitly quantified but implied as acceptable for MR Conditional labeling.
    Image artifacts within acceptable limits (ASTM F2119-07, R2013)Not explicitly quantified but implied as acceptable for MR Conditional labeling.
    RF-induced heating (ASTM F2182-19e2) resulting in a temperature rise below 6 ℃Achieved under specified scanning conditions: 1.5 T/64 MHz and 3 T/128 MHz at a whole-body averaged specific absorption rate (wbSAR) of 2 W/kg or head SAR of 3.2 W/kg for an hour-long scanning session.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The testing involved computational modeling and simulation (CM&S). The "test set" in this context refers to the simulated scenarios and device configurations.

    • Sample Size for Test Set: "the entire portfolio of KLS Martin maxillofacial implants" was simulated. This implies that all devices grouped under "KLS Martin Oral-Max Implants - MR Conditional" were included in the simulations. The document also mentions "various in-vivo device positions and landmarks," "worst-case single and multiple devices," and simulations in "10 cm increments." This suggests a comprehensive set of simulated scenarios rather than a traditional physical sample size.
    • Data Provenance: Not explicitly stated as "country of origin" or "retrospective/prospective" in the same way clinical data is. The data is generated through computational modeling and simulation using MED Institute's FDA-qualified Medical Device Development Tool (MDDT) and the Duke virtual human anatomy. This is a form of prospective simulation data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable in the traditional sense for this type of non-clinical, simulation-based study. The "ground truth" for the RF-induced heating simulations is derived from the established physics and engineering principles embedded in the FDA-qualified MDDT and the Duke virtual human anatomy model. The expertise lies in the development and validation of these computational tools and the interpretation of the simulation results by experts in MR safety and medical device engineering at MED Institute and the submitting company. The document does not specify the number or qualifications of individual experts validating the computational model, but implies that the MDDT itself is "FDA-qualified," indicating a level of expert review and agreement on its methodology.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like "2+1" or "3+1" are typically used for consensus building among human expert readers for clinical studies, especially when establishing ground truth from image interpretation. This study is based on physical property testing and computational simulations, not human interpretation of clinical data.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not a study involving human readers or AI-assisted diagnostic performance. It focuses on the physical safety of implants in an MR environment.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The RF-induced heating assessment involved "Computational modeling and simulation (CM&S) ... using MED Institute's FDA-qualified Medical Device Development Tool (MDDT) and in a clinically relevant position within the Duke virtual human anatomy." This is a standalone algorithm/model-based assessment without a human-in-the-loop for the performance evaluation itself. Human experts design the simulations, configure the models, and interpret the results, but the "performance" (temperature rise, SAR calculations) is computed by the algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests:

    • Magnetically induced displacement force, torque, and image artifacts: The "ground truth" is based on the physical properties of the materials and device designs, measured or calculated according to established ASTM standards (F2052-21, F2213-17, F2119-07).
    • RF-induced heating: The "ground truth" for the simulations is derived from the established electromagnetic physics and thermal dynamics principles implemented in the FDA-qualified Medical Device Development Tool (MDDT) and applied to the Duke virtual human anatomy model. The MDDT's qualification process by the FDA implicitly establishes the reliability of its results as a form of "ground truth" for simulation-based assessments.

    8. The sample size for the training set

    Not applicable. This is a non-clinical study for MR safety assessment, not a machine learning model requiring a training set in the typical sense. The "training" for the MDDT is its initial validation and qualification against known physical phenomena and experimental data, which is a separate process from this submission.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for a machine learning model in this context. The "ground truth" for qualifying the simulation tool (MDDT) would have been established through extensive validation against experimental measurements and recognized physical theories.

    Ask a Question

    Ask a specific question about this device

    K Number
    K042573

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-10-18

    (27 days)

    Product Code
    Regulation Number
    872.4880
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K980760, K971297, K033483

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KLS Martin Drill Free® MMF Screws is intended for use in maxillomandibular fixation to provide stabilization of fractures of the maxilla, mandible, or both.

    Device Description

    The KLS-Martin Drill Free® MMF Screw provides temporary occlusal and fracture stabilization. These screws may be applied prior to or after exposure of the fracture.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the KLS Martin Drill Free® MMF Screw.

    Unfortunately, the provided text does not contain any information about acceptance criteria or a study proving the device meets said criteria. The document is a 510(k) summary and the FDA's clearance letter for the KLS Martin Drill Free® MMF Screw. It focuses on demonstrating substantial equivalence to predicate devices rather than providing performance data from a specific study against predefined acceptance criteria.

    Therefore, I cannot populate most of the requested sections. Here's what I can extract from the provided text based on the limited information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Not specified in documentNot specified in document

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample size for test set: Not specified.
    • Data provenance: Not specified.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Not applicable as no specific study data or ground truth establishment is described in the provided document.

    4. Adjudication Method for the Test Set

    • Not applicable as no specific study data or adjudication is described in the provided document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No, there is no mention of an MRMC study.
    • Effect size of human reader improvement with AI vs. without AI assistance: Not applicable, as this is a medical screw, not an AI-assisted diagnostic device.

    6. Standalone (Algorithm Only) Performance Study

    • Was a standalone study done? No, there is no mention of a standalone performance study. This device is a physical medical implant, not an algorithm.

    7. Type of Ground Truth Used

    • Not applicable as no specific study data or ground truth is described in the provided document.

    8. Sample Size for the Training Set

    • Not applicable as no "training set" is relevant for this type of medical device submission.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable as no "training set" is relevant for this type of medical device submission.

    Summary of Document's Content:

    The provided documents describe the submission of a 510(k) premarket notification for the KLS Martin Drill Free® MMF Screw. The core of the submission is to demonstrate substantial equivalence to existing predicate devices based on:

    • Identical intended use: For maxillomandibular fixation to stabilize fractures of the maxilla, mandible, or both.
    • Similar application: Compared to other drill-free and anchorage systems.
    • Key difference: The new device is "drill-free" (self-tapping), while its primary predicate required a pilot hole.

    The FDA's clearance letter confirms their determination of substantial equivalence, allowing the device to be marketed. This type of submission relies on comparisons to already cleared devices rather than a de novo study with acceptance criteria for a novel device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1