Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K161978

    Validate with FDA (Live)

    Device Name
    AirFit N20
    Manufacturer
    Date Cleared
    2016-11-17

    (122 days)

    Product Code
    Regulation Number
    868.5905
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K031047, K032433, K132013, K132887

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AirFit N20 channels airflow non-invasively to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bilevel device.

    The AirFit N20 is:
    • to be used by patients weighing more than 66 lb (30 kg) for whom positive airway pressure has been prescribed
    • intended for single patient re-use in the home environment and multi-patient re-use in the hospital / institutional environment.

    Device Description

    The modified ResMed AirFit N20 is a non-invasive vented respiratory mask that provides a silicone air seal around the patient's nose and upper lip. The mask is held in place with adjustable head straps. Air flow from a positive air pressure (PAP) source is directed to the patient's airway non-invasively. It connects to a conventional PAP device air delivery hose via a standard 22mm swivel.

    The modified AirFit N20 comprises 4 subassemblies: headgear, frame, cushion and elbow/short tube. The exhaust port is incorporated into the elbow/short tube assembly. For home use, the mask may be cleaned in warm soapy water.

    The AirFit N20 is a prescription device supplied non-sterile.

    AI/ML Overview

    This document describes the regulatory submission for the ResMed AirFit N20, a non-invasive vented respiratory mask. The submission, K161978, is a modification of a previously cleared predicate device, AirFit N20 (K153673), aiming to demonstrate substantial equivalence.

    Based on the provided information, the following can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Functional Performance:
    CO2 rebreathingDemonstrated to be in accordance with ISO 17510:2015. (Implied: Acceptable CO2 levels, comparable to predicate)Verified to meet predetermined acceptance criteria and substantially equivalent to predicate.
    Total mask flowDemonstrated to be in accordance with ISO 17510:2015. (Implied: Flow characteristics comparable to predicate)Verified to meet predetermined acceptance criteria and substantially equivalent to predicate.
    Flow resistanceDemonstrated to be in accordance with ISO 17510:2015. (Implied: Resistance characteristics comparable to predicate) -----------------------------------------------------------Verified to meet predetermined acceptance criteria and substantially equivalent to predicate.
    Through impedanceDemonstrated to be in accordance with ISO 17510:2015. (Implied: Impedance characteristics comparable to predicate) -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Verified to meet predetermined acceptance criteria and substantially equivalent to predicate.
    Operating pressure rangeSimilar to predicateSimilar performance to predicate.
    Pressure flowSimilar to predicateSimilar performance to predicate.
    Flow impedance characteristicsSimilar to predicateSimilar performance to predicate.
    PAP device settings ("Pillows, Mirage or Swift")Operate on the same ResMed flow generator settings as predicate ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Operates on the same settings as predicate.
    Mechanical Integrity:
    Home cleaning for single patient reuseVerified under simulated normal use and reasonable abuse scenarios. (Implied: Maintains integrity and function after home cleaning)Verification confirmed it met predetermined acceptance criteria.
    Reprocessing for multi-patient reuseVerification under simulated normal use and reasonable abuse scenarios, including cleaning efficacy, disinfection efficacy, post-disinfection/sterilization performance, and residual toxicity evaluation. (Implied: Maintains integrity and function, remains safe for multi-patient reuse after reprocessing)Verification confirmed it met predetermined acceptance criteria.
    TransportationVerified under simulated normal use and reasonable abuse scenarios. (Implied: Withstands transportation without damage or loss of function)Verification confirmed it met predetermined acceptance criteria.
    StorageVerified under simulated normal use and reasonable abuse scenarios. (Implied: Maintains integrity and function during storage)Verification confirmed it met predetermined acceptance criteria.
    Biocompatibility:In accordance with ISO 10993-1, ISO 10993-5, ISO 10993-10 and ISO 10993-12 for modified components using new materials with patient exposure classifications of permanent external communicating device (tissue) and/or permanent skin contact. (Implied: No unacceptable biological risks)Evaluation conducted, relying on predication for components with previously cleared materials.
    Material Specifications:Use of polymeric construction materials for pneumatic and structural components and foam padded fabrics for the head strap. ISO 5356-1 compliant 22mm diameter swivel for connection to PAP delivery hose.Similar to predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state sample sizes for specific tests (e.g., number of masks tested for CO2 rebreathing, or number of cleaning/reprocessing cycles). It mentions "testing to simulated normal use and reasonable abuse scenarios," suggesting laboratory-based testing.

    • Data Provenance: The testing was conducted by ResMed Ltd, an Australian company, for a device intended for international markets (indicated by FDA submission). The data is retrospective in the sense that it's bench testing and verification/validation conducted by the manufacturer, not a prospective clinical trial. There is no mention of data from human subjects in the testing phase beyond what is implicitly covered by biocompatibility standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. This is a medical device submission primarily relying on bench testing and engineering verification/validation against international standards (e.g., ISO 17510:2015, ISO 10993 series) and comparison to a predicate device, rather than human expert-established ground truth for a diagnostic output. The "ground truth" here is the adherence to established engineering specifications and safety standards.

    4. Adjudication Method for the Test Set

    Not applicable for this type of submission. The evaluation is based on objective measurements against pre-defined engineering and performance criteria and adherence to recognized standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This document explicitly states: "Clinical data was not relied upon to demonstrate Substantial Equivalence to the predicate device. Bench testing demonstrates that the modified AirFit N20 mask device performs in an equivalent manner and is as safe and as effective as the previously cleared predicate AirFit N20 device (K153673)." Therefore, an MRMC study is not relevant to this submission.

    6. Standalone Performance (Algorithm Only without Human-in-the-loop Performance)

    Yes, in the sense that the device's functional and mechanical performance was evaluated in isolation ("bench testing") against pre-defined specifications and standards. This is not an AI algorithm, but a physical medical device; its "standalone performance" refers to its intrinsic characteristics and operation.

    7. Type of Ground Truth Used

    The "ground truth" for this device, as described in the submission, is based on:

    • International Standards: Primarily ISO 17510:2015 (Medical devices – Sleep apnoea breathing therapy Masks and application accessories) and ISO 10993 series for biocompatibility.
    • Engineering Specifications: Predetermined acceptance criteria derived from the device's design requirements and performance characteristics of the legally marketed predicate device (K153673).
    • Predicate Device Performance: The original AirFit N20 (K153673) serves as a benchmark for "similar performance."

    8. Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As noted above, there is no AI/ML training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K042403

    Validate with FDA (Live)

    Device Name
    MIRAGE SWIFT
    Manufacturer
    Date Cleared
    2004-10-08

    (35 days)

    Product Code
    Regulation Number
    868.5905
    Reference & Predicate Devices
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mirage Swift is intended for Multiple Patient Reuse by adult patients (>30 Kg) prescribed continuous positive airway pressure or bilevel therapy for use in home, hospitals or clinics.

    Device Description

    The Mirage Swift is designed for adult patients for the delivery of non-invasive ventilatory support using continuous positive airway pressure or bi-level therapy. It is intended for multiple patient re-use and is minimally obtrusive to the user providing a high level of comfort, ease-ofuse and seal.

    AI/ML Overview

    The provided document is a 510(k) summary for the ResMed Mirage Swift Nasal Mask, seeking clearance for "Multiple Patient Reuse" after initially being cleared as a single patient reuse device.

    Based on the provided text, there is no specific performance study with acceptance criteria and a detailed report of device performance as one might expect for a new AI/software device. This document is for a physical medical device (a CPAP mask) and focuses on demonstrating substantial equivalence to predicate devices for an expanded indication for use (multiple patient reuse), not for establishing performance metrics of a novel diagnostic or therapeutic algorithm.

    Therefore, many of the requested points regarding acceptance criteria, sample sizes for test/training sets, expert adjudication, MRMC studies, or standalone algorithm performance are not applicable or discoverable from this particular 510(k) summary.

    Here's an attempt to answer the questions based only on the provided text:


    1. Table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria or a direct performance report in the context of a new diagnostic/therapeutic device. Instead, the "acceptance criterion" is implicitly demonstrating substantial equivalence to predicate devices that are already cleared for multiple patient reuse. The reported "performance" is that this substantial equivalence has been met.

    Acceptance Criterion (Implicit)Reported Device Performance
    Substantial equivalence to predicate masks (Mirage Activa and Mirage Full Face Series 2) for Multiple Patient Reuse, particularly concerning cleaning/disinfection.The Mirage Swift mask is substantially equivalent to the previously cleared predicate masks and can be relabeled for multiple-patient, multiple-use.
    (Performance data and rationale are provided to demonstrate this equivalence, but no specific metrics are reported in this summary.)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. This is not a study assessing diagnostic performance against a test set. The "performance data" mentioned would likely pertain to cleaning, disinfection, and material compatibility relevant to multiple patient reuse, rather than a clinical outcome study with a "test set" of patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. "Ground truth" in the context of expert consensus for a diagnostic test is not relevant here. The "ground truth" for multiple patient reuse would be established through testing cleaning efficacy, material degradation after disinfection cycles, and biocompatibility, typically performed in a lab setting rather than by clinical experts establishing a diagnosis.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There is no "test set" in the context of expert adjudication for this type of submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document is about a physical CPAP mask, not an AI or software-assisted diagnostic device. Therefore, no MRMC study or AI improvement metrics are relevant or present.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The concept of "ground truth" as it applies to diagnostic algorithms is not relevant here. For the "Multiple Patient Reuse" claim, the "ground truth" would be objective evidence from laboratory testing demonstrating that the device can be effectively cleaned and disinfected according to a validated protocol without material degradation or compromising safety/performance, making it suitable for subsequent patient use. The document states that "Performance data and rationale are provided" to support this.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable. As above, this is not an algorithm requiring a "training set" or "ground truth" in that context.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1