Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K094009

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-06-14

    (168 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K993944, K032149

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Halt 2000GI™ Electrosurgical Radiofrequency Ablation System is indicated for use in percutaneous, laparoscopic, and intraoperative coagulation and ablation of soft tissue.

    Device Description

    The Halt 2000GI™ Electrosurgical Radiofrequency Ablation System includes the following system components:
    • Halt 2000GI Radiofrequency Generator (RF Generator): that provides RF energy to the RF Probe
    • Disposable RF Probe (Tulip™, TU 1000), a hand piece with a trocar-pointed shaft and 7 deployable RF needle electrodes.
    • RF Probe Extension Cable that connects the RF Probe to the RF Generator.
    • Dispersive Electrode Pads (TSP 115) Set that provides the return path for the RF energy applied by the RF Probe. (Use only Dispersive Electrode Pads provided by Halt Medical, Inc.)
    • Dispersive Electrode Pad Set Extension Cable (TSPCBL) that connects the Dispersive Electrode Pad Set to the RF Generator.
    • Power Cord that is a medical grade power cord providing AC power to the RF Generator.
    • Foot Pedal that is a pneumatic foot pedal with tubing used to turn RF energy on and off.

    The Halt 2000GI™ Electrosurgical Radiofrequency Ablation System is designed to deliver up to 200 W of RF power at 460 kHz in three operational modes: Temperature Control. Manual Control and Coagulation Mode. A touch screen with a graphical user interface (GUI) enables selection of operational parameters such as the mode of operation, the ablation time, the target temperature, and the power delivery level. With the RF probe placed in the tissue to be ablated and its electrodes deployed, RF power can be turned on. The system parameters are continuously monitored and displayed on the RF Generator. If the measured parameters are outside the acceptable limits, the RF energy delivery automatically stops and a message appears on the graphical user interface. RF energy during an ablation or coagulation can also be stopped at any time by the user by pressing the foot pedal.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Halt 2000GI™ Electrosurgical Radiofrequency Ablation System. This document focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a study report as would be found for a novel device. Therefore, much of the requested information, particularly regarding specific performance metrics, sample sizes for test and training sets, expert qualifications, and adjudication methods, is not explicitly available in the provided text.

    However, I can extract and infer some information based on the content.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Inferred from testing types)Reported Device Performance
    Electrical Safety Standards: EN/IEC 60601-1 (General medical electrical equipment safety) EN/IEC 60601-1-2 (Electromagnetic compatibility) EN/IEC 60601-2-2 (High-frequency surgical equipment) ANSI/AAMI HF-18 (Electrosurgical devices safety)The system passed all testing for applicable safety requirements.
    Mechanical Validation Testing:The system passed all testing.
    Software Validation Testing:The system passed all testing.
    Ablation Capability: (Performs as intended and per specifications)Animal and bench testing successfully demonstrated that the Halt RF Ablation System performs as intended and per specifications. The ablation capability was confirmed, providing a reproducible, discretely demarcated zone of tissue necrosis surrounded by normal tissue perfused with blood.
    Biocompatibility: (Safe, suitable, and appropriate for intended use, compliant with ISO 10993-1, ISO 10993-5, ISO 10993-10)A series of biocompatibility testing demonstrated that the device materials are safe, suitable, and appropriate for their intended use and in compliance with the specified ISO standards.

    2. Sample size used for the test set and the data provenance

    The document does not specify a distinct "test set" in the context of clinical or image-based studies. The testing mentioned includes:

    • Electrical, mechanical, and software validation.
    • Animal testing.
    • Bench testing.
    • Biocompatibility testing.

    The size of these "test sets" (e.g., number of animals, number of bench tests) is not provided. Data provenance (country of origin, retrospective/prospective) is also not specified for these types of engineering and pre-clinical tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided. The ground truth for the performance testing appears to be based on the outcomes of engineering tests (e.g., meeting electrical safety standards) and observations from animal and bench studies (e.g., "reproducible, discretely demarcated zone of tissue necrosis"). There is no mention of expert consensus for establishing ground truth in the context of a diagnostic or interpretive device.

    4. Adjudication method for the test set

    Not applicable to the types of testing described. Adjudication methods are typically relevant for human-in-the-loop studies or clinical trials involving subjective assessment, which are not detailed here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not conducted. This device is an electrosurgical ablation system, not a diagnostic imaging device utilizing AI that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This device does not involve an "algorithm only" component in the sense of AI or image interpretation. Its performance is related to its ability to deliver RF energy and achieve tissue ablation, which was evaluated through animal and bench testing, as well as engineering validation.

    7. The type of ground truth used

    The ground truth used for performance validation appears to be:

    • Engineering specifications and regulatory standards for electrical, mechanical, and software validation.
    • Histopathological or visual assessment of tissue necrosis in animal and bench testing.
    • Biocompatibility standards (ISO 10993 series) for material safety.

    8. The sample size for the training set

    Not applicable. This document describes an electrosurgical device, not a machine learning model, so there is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    Not applicable, as no training set is mentioned for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1