510(k) Data Aggregation

    K Number
    K072567

    Validate with FDA (Live)

    Date Cleared
    2007-10-30

    (48 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K063826, K031825, K051460

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV.

    Device Description

    The Symbia-E is a gamma camera used to view and analyze images of the human body and the distribution of administered radionuclides. This system is designed for whole-body oncology, neurology, cardiology and general diagnostic examinations. The Symbia-E is intended to be utilized by appropriately trained health care professionals to image and measure the distribution of radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body.

    The Symbia-E system which is the subject of this application is substantially equivalent to the commercially available e.cam gamma camera (K963983) and the commercially available Siemens Enhanced Imaging System (K041166).

    The changes incorporated into Symbia E include modifications made to address obsolescence issues, cost reduction efforts and to change the industrial design of the system. Additionally, the system software has been updated (Syngo MI Applications 2007B).

    AI/ML Overview

    The provided text is a 510(k) summary for the Symbia-E gamma camera. It describes the device, its intended use, and claims substantial equivalence to previously marketed devices. However, this document does not contain any information about acceptance criteria, device performance studies, sample sizes, ground truth establishment, expert qualifications, or MRMC studies.

    Therefore, I cannot fulfill your request for a table of acceptance criteria and reported device performance or details about a study demonstrating the device meets acceptance criteria, as this information is not present in the provided text.

    Based on the information provided, here's what can be inferred:

    • Device Name: Symbia-E
    • Device Type: Gamma camera (Emission Computed Tomography System)
    • Purpose: To view and analyze images of the human body and the distribution of administered radionuclides for various diagnostic examinations (oncology, neurology, cardiology, general diagnostics).
    • Equivalence Claim: Substantially equivalent to previously cleared devices (e.g., e.cam gamma camera, Siemens Enhanced Imaging System, Syngo MI Applications 2007A, c.cam, c.cam-AC).
    • Changes in Symbia-E: Modifications for obsolescence, cost reduction, industrial design change, and updated system software (Syngo MI Applications 2007B).
    • Indications for Use: To detect or image the distribution of radionuclides in the body or organ, using planar imaging, whole body imaging, and tomographic imaging for isotopes with energies up to 588keV.
    • Regulatory Outcome: FDA found the device substantially equivalent to legally marketed predicate devices.

    To provide the requested information, you would need a different type of document, such as a test report, clinical study summary, or a more detailed section of the 510(k) submission that discusses performance data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062450

    Validate with FDA (Live)

    Date Cleared
    2006-10-20

    (59 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K031825, K053062

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    D-SPECT is an emission computed tomography system intended to detect the location and distribution of gamma ray radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. The device includes display equipment, patient and equipment supports, component parts, and accessories. D-SPECT is primarily intended for cardiac applications. D-SPECT supports radionuclides within the energy range of 40 -170 keV.

    Device Description

    Spectrum Dynamics' D-Spect™ Cardiac Scanner System is a SPECT device, which is designed to perform myocardial perfusion imaging. The device is comprised of a detector head, gantry, and patient chair. Device operation is controlled from an acquisition station console. The system is supported by use of data-transfer accessories (RFID tags), which are attached to the patient's wrist and to the syringe containing the radio-pharmaceutical agent, for patient and syringe positive identification. The cardiac gamma camera is designed such that there are no external moving parts that surround the patient. Detector boards rotate within the closed detector head. The special scanning geometry and detector technology, enable shorter scan times.

    AI/ML Overview

    The provided document is a 510(k) summary for the D-SPECT System, an emission computed tomography system for cardiac applications. However, it does not contain specific acceptance criteria or an explicit study report detailing the device's performance against such criteria for the purpose of demonstrating substantial equivalence beyond general safety and performance standards.

    The document primarily focuses on establishing substantial equivalence to predicate devices (Virgo and Cardiarc) based on similar intended use and technological characteristics, and by demonstrating compliance with recognized general performance and safety standards.

    Therefore, the following information cannot be fully provided based on the given text:

    1. A table of acceptance criteria and the reported device performance: No specific clinical acceptance criteria (e.g., sensitivity, specificity, accuracy for a particular clinical task) are mentioned, nor are quantitative performance metrics reported against such criteria. The document states that performance tests demonstrate the device can be safely and effectively used, but not how it quantitatively meets specific performance benchmarks.

    2. Sample size used for the test set and the data provenance: Only "Sample images from three clinical cases" are mentioned as clinical performance data. This is an extremely small sample and likely not a formal test set for statistical evaluation. Data provenance (country of origin, retrospective/prospective) is not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned.

    4. Adjudication method for the test set: Not mentioned.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: An MRMC comparative effectiveness study was not done. The device is a standalone imaging system, not an AI-assisted interpretation tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The document describes the D-SPECT system as producing images through computer reconstruction of data, and these images are then presumably interpreted by human readers. The clinical performance data presented (Sample images from three clinical cases) suggests a qualitative assessment rather than a quantitative standalone performance evaluation with metrics like sensitivity/specificity against a ground truth. The focus is on the device's ability to "produce cross-sectional images" and "detect the location and distribution of gamma ray radionuclides."

    7. The type of ground truth used: Not explicitly stated for the "three clinical cases." Given the nature of a SPECT device, ground truth for myocardial perfusion imaging would typically involve a combination of patient symptoms, additional diagnostic tests (e.g., angiography, stress echocardiography), and clinical follow-up. However, this is not detailed in the document.

    8. The sample size for the training set: Not mentioned. The document does not describe the use of machine learning that would typically involve a training set. The device is described as operating on "the same basic principle and the same basic technology as other SPECT devices."

    9. How the ground truth for the training set was established: Not mentioned, as no training set for a machine learning algorithm is described.

    Summary of what can be extracted from the document:

    • Device Name: D-Spect™ Cardiac Scanner System
    • Intended Use/Indication for Use: "An emission computed tomography system intended to detect the location and distribution of gamma ray radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. Primarily intended for cardiac applications. Supports radionuclides within the energy range of 40 -170 keV."
    • Study Design (Clinical Data): "Sample images from three clinical cases using the D-SPECT device." This is an observational presentation, not a formal clinical study with performance metrics.
    • Non-Clinical Test Data: The device underwent "extensive safety testing, performance testing, and validation" including:
      • EMC and Safety testing (IEC/EN-60601-1 and IEC/EN 60601-1-2)
      • Testing according to relevant sections of NEMA NU 1 standard (Performance Measurements of Scintillation Cameras)
      • Software validation testing.
    • Conclusion from Tests: "The performance tests demonstrate that D-SPECT device may be safely and effectively used myocardial perfusion imaging. The software validation and performance validation tests for scintillation cameras (NEMA NU 1), as well as the clinical images provided, demonstrate that the D-SPECT device meets its design and performance specifications and is substantially equivalent to the predicate devices."

    In essence, the 510(k) summary provided here relies on compliance with established safety and performance standards (like NEMA NU 1 for scintillation cameras) and qualitative clinical image review, rather than a detailed clinical study with quantitative acceptance criteria for diagnostic performance.