510(k) Data Aggregation

    K Number
    K093327

    Validate with FDA (Live)

    Date Cleared
    2010-05-10

    (196 days)

    Product Code
    Regulation Number
    870.2360
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K030509

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INTCO Tab Electrodes are designed for use with general electrocardiographic procedures when ECG monitoring is deemed necessary by a physician. Such procedures include ECG surveillance and diagnosis recording. INTCO Tab Electrodes are non-sterile and are to only be used on intact (uninjured) skin.

    Device Description

    INTCO Tab electrodes are composed of PET tape, Ag/AgCl or carbon conductive layer, and a conductive gel. These are configured as ten electrodes applied to a siliconized card, with 10 cards per pouch

    AI/ML Overview

    The provided 510(k) summary describes the Intco Tab Electrode, a disposable ECG electrode. This device does not involve AI or algorithms, so the majority of the requested information about AI model performance, training sets, and expert adjudication is not applicable.

    Here's an analysis based on the available information:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard Compliance)Reported Device Performance
    ANSI/AAMI EC12:2000/(R)2005Conforms to the standard
    Biocompatibility (ISO 10993-1)Non-irritating
    Non-cytotoxic
    Non-sensitizing
    Shelf life24 months (accelerated aging tests performed)

    Study Details

    The submission focuses on demonstrating substantial equivalence to a predicate device (Leonard Lang GmbH Skintact ECG Tab Electrode K030509) by showing compliance with recognized consensus standards and comparable physical/technical characteristics.

    1. Sample size used for the test set and the data provenance: Not explicitly stated as a clinical study with a "test set" in the context of an AI model. The performance evaluation is based on conformance to industry standards and biocompatibility testing. For the standards testing, specific sample sizes for tests within ANSI/AAMI EC12:2000/(R)2005 are not detailed, but standard conformance implies appropriate sampling for those tests. The data provenance is generally from laboratory testing performed by the manufacturer.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This is not a study requiring expert-established ground truth for an AI model. Biocompatibility testing results are laboratory-derived, and standard conformance is assessed against defined technical parameters.

    3. Adjudication method for the test set: Not applicable. There is no expert adjudication process described.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an electrode, not an AI diagnostic tool.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device does not have an algorithm.

    6. The type of ground truth used:

      • Standard Conformance: The "ground truth" is defined by the technical specifications and test methods outlined in the ANSI/AAMI EC12:2000/(R)2005 standard for disposable ECG electrodes. The device performance is measured against these established parameters.
      • Biocompatibility: The "ground truth" is established by the accepted scientific methodologies for assessing irritation, cytotoxicity, and sensitization as defined in ISO 10993-1.
    7. The sample size for the training set: Not applicable. This device does not utilize a training set as it is not an AI/machine learning device.

    8. How the ground truth for the training set was established: Not applicable.

    Summary of Device Evaluation:

    The INTCO Tab Electrode's acceptance criteria and proven performance are primarily based on:

    • Conformance to the ANSI/AAMI EC12:2000/(R)2005 consensus standard, which specifies requirements for characteristics such as impedance, DC offset, combined offset and offset instability, and input noise performance for disposable ECG electrodes. The submission states the device "have been tested and conform to recognized consensus standard."
    • Biocompatibility testing according to ISO 10993-1, which demonstrated the conductive gel to be non-irritating, non-cytotoxic, and non-sensitizing.
    • Accelerated aging tests to substantiate a 24-month shelf life.

    The submission aims to demonstrate substantial equivalence to a legally marketed predicate device by showing that the INTCO Tab Electrode meets these established performance and safety benchmarks.