Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K133912

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-05-14

    (142 days)

    Product Code
    Regulation Number
    884.6180
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K080473, K120136, K030490, K133707, K991279

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ORIGIO® Sequential Fert™ is for the fertilization of oocytes in vitro.
    ORIGIO® Sequential Cleav™ is for the culture of embryos until the 2-8 cell stage. ORIGIO® Sequential Cleav™ can also be used for embryo transfer at day 2 or 3.

    Device Description

    ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ are aseptically filtered, non viscous solutions, light pink or colorless solutions, which are ready to use by professionals within assisted reproduction.
    ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ are contained in 10 mL or 60 mL transparent polyethylene terephthalate glycol (PETG) bottles with high density polyethylene (HDPE) closures, available in card board boxes of 1 x 10 mL and 1 x 60 mL bottles. The bottles and boxes are individually labeled. The boxes also contain instruction for use provided as package insert.

    AI/ML Overview

    This document describes the ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ medical devices. The submission focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets acceptance criteria from a standalone clinical study.

    Therefore, a table of "acceptance criteria" and "reported device performance" specifically for the device's diagnostic performance (like sensitivity/specificity) is not provided in the typical sense of a diagnostic medical device. Instead, the "acceptance criteria" are implied by the comparison to predicate devices' specifications and functional properties through various laboratory tests.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally expressed as ranges or thresholds for various physicochemical properties and biological performance in comparison to predicate devices and recognized ART media. The reported device performance is stated to meet these specifications.

    CharacteristicAcceptance Criteria (ORIGIO® Sequential Fert™)Reported Device Performance (ORIGIO® Sequential Fert™)Acceptance Criteria (ORIGIO® Sequential Cleav™)Reported Device Performance (ORIGIO® Sequential Cleav™)
    pH7.3-7.5Meets specification (implied)7.2-7.4Meets specification (implied)
    Osmolality (mOsm/kg)277-293Meets specification (implied)272-288Meets specification (implied)
    Endotoxin (EU/mL)<0.15Meets specification (implied, better than predicate)<0.15Meets specification (implied, better than predicate)
    Aseptically filteredXMeets specification (implied)XMeets specification (implied)
    1-cell MEA (Mouse Embryo Assay)≥80%≥80%≥80%≥80%
    SterilitySterileMeets specification (implied)SterileMeets specification (implied)
    HSA ConcentrationStable through shelf lifeMeets specification (implied)Stable through shelf lifeMeets specification (implied)
    BiocompatibilityCytotoxicity, sensitization, and irritation tests demonstrate biocompatibility (for ORIGIO® Sequential Cleav™ only)Biocompatible (for short duration contact with mucosal tissues)N/AN/A
    Shelf Life36 weeksValidated to 36 weeks36 weeksValidated to 36 weeks

    Note: The document does not provide specific numerical "reported device performance" results for pH, osmolality, endotoxin, or sterility, but states that these parameters were tested and met specifications or were comparable to predicate devices. For MEA, the specific percentage is given as >=80% in the table, implying this was maintained.

    2. Sample size used for the test set and the data provenance

    The document does not describe a clinical "test set" in the context of diagnostic device performance (e.g., patient data for sensitivity/specificity). Instead, the "testing" refers primarily to:

    • Physicochemical analyses: pH, osmolality, endotoxin, sterility performed on batches of the manufactured media. The sample size for these manufacturing release tests is not specified but would typically follow internal quality control procedures.
    • Mouse Embryo Assay (MEA): This is a biological test performed on mouse embryos to assess the media's ability to support embryo development. The sample size for MEA is stated as achieving "≥80%," indicating a pass/fail criterion rather than a detailed study sample size. The provenance of these mouse embryos or the exact number tested is not detailed.
    • Stability Studies: Conducted to determine shelf life, testing pH, osmolality, endotoxin, HSA concentration, MEA, and sterility over time. The sample size (number of batches, number of samples per batch/time point) for these studies is not specified.
    • Biocompatibility Testing: Performed for ORIGIO® Sequential Cleav™. This involves standardized in vitro and in vivo tests (cytotoxicity, sensitization, irritation). The sample size for these specific tests is not provided, but generally involves a predetermined number of test samples.

    All data described appears to be retrospective (part of product development and qualification) and is generated by ORIGIO a/s (Denmark) or contracted laboratories.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    N/A. This product is an in vitro fertilization (IVF) medium, not a diagnostic device that requires expert ground truth for interpretation of discrete outputs. The "ground truth" for its performance is assessed through its physical and chemical properties and its ability to support embryo development (MEA), which are objective laboratory measurements, not subjective expert interpretations.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    N/A. As this is not a diagnostic device or a study involving human interpretation, an adjudication method for a "test set" is not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A. No MRMC study was conducted. This device is an IVF medium, not an AI or imaging device involving human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    N/A. There is no algorithm or AI component to this device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this product's performance is established by:

    • Physicochemical standards: pH, osmolality, endotoxin levels are measured against predefined acceptable ranges.
    • Biological performance standards: The Mouse Embryo Assay (MEA) serves as a biological ground truth, where successful development of mouse embryos (≥80% 1-cell MEA) indicates the medium's suitability.
    • Comparison to predicate devices: The "ground truth" for substantial equivalence is met by demonstrating that the new devices have comparable technological characteristics and performance to legally marketed predicate devices.

    8. The sample size for the training set

    N/A. This is not a machine learning or AI device, so there is no training set in that context. The "training" in product development refers to formulation optimization and initial testing, but no specific "training set" of data is mentioned.

    9. How the ground truth for the training set was established

    N/A. See above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1