Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K151367

    Validate with FDA (Live)

    Date Cleared
    2016-01-27

    (251 days)

    Product Code
    Regulation Number
    884.6180
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K023353

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sydney IVF Hyaluronidase is intended for use to facilitate removal of the cumulus cells surrounding oocytes in assisted reproduction technology (ART) procedures.

    Device Description

    Sydney IVF Hyaluronidase is an enzymatic preparation containing 80 IU/mL Hyaluronidase (of ovine origin) in a Bicarbonate buffered media supplemented with 5 mg/mL Human Serum Albumin (HSA) and 10 ug/mL Gentamicin.

    Sydney IVF Hyaluronidase is provided sterile (aseptically filtered) and is packaged in glass vials with Fluorotec coated rubber stoppers held in place with a tamper evident seal. It is available in a 5 pack carton box with each vial containing 1 mL of Sydney IVF Hvaluronidase.

    AI/ML Overview

    The provided text describes the Sydney IVF Hyaluronidase device and its substantial equivalence to a predicate device, focusing on its formulation, shelf-life, and performance. However, it does not contain the kind of detailed information typically found in a study demonstrating the device meets specific acceptance criteria with reported device performance against those criteria in a tabular format, nor does it provide information about the methodology of such a study (sample sizes, expert involvement, adjudication, etc.).

    The document is a 510(k) summary for regulatory clearance, which primarily focuses on demonstrating substantial equivalence to a previously cleared device. It mentions "performance data" and "stability & shelf life" but doesn't present a formal study report with acceptance criteria and results in the requested format.

    Therefore, I cannot fulfill your request for:

    • A table of acceptance criteria and the reported device performance.
    • Sample sizes used for the test set and data provenance.
    • Number of experts and their qualifications for ground truth.
    • Adjudication method for the test set.
    • MRMC comparative effectiveness study effect size.
    • Standalone performance.
    • Type of ground truth used (beyond "MEA data").
    • Sample size for the training set.
    • How ground truth for the training set was established.

    However, I can extract the relevant information known about the device's performance based on the document:

    The document states that the Sydney IVF Hyaluronidase (the proposed device) and the predicate device (K023353 - Sydney IVF Hyaluronidase) share similar performance specifications and that the proposed device has been validated in stability studies.

    Here's what can be pieced together from your request and the provided text:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria CategoryPredicate Device Specification (K023353)Proposed Device Performance (Sydney IVF Hyaluronidase)
    Osmolality285-295 mOsm/kgSame (285-295 mOsm/kg)
    Endotoxin< 0.40 EU/mLSame (< 0.40 EU/mL)
    SterilitySterile (aseptically filtered)Same (aseptically filtered)
    Embryo Toxicity (MEA)1-cell MEA (96hrs) with ≥75% of control that develop to blastocyst2-cell MEA (72hrs) with ≥80% of control that develop to blastocyst
    Shelf-Life6 months at -20°CValidated to 20 weeks at 2-8°C
    Stability Tests(Implied by predicate and proposed device sharing "performance specifications")pH, osmolality, endotoxin, MEA, sterility, concentrations of proline, pyruvate, ammonia, and enzymatic activity of Hyaluronidase

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size: Not specified for any specific "test set." The document refers to "MEA data" and "stability studies" without providing sample sizes.
    • Data Provenance: The device is from William A. Cook Australia Pty Ltd. The "MEA data" and "stability studies" would presumably be conducted by them. The document mentions the ovine derived product has been "licensed and sold globally since 2007," and its performance is supported by "experience" during this time, suggesting real-world usage data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not specified. The "MEA data" itself serves as a performance metric, but the number or qualifications of experts involved in running or interpreting these assays are not mentioned.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not specified.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This device is a reproductive media and supplement (Hyaluronidase Solution), not an AI or imaging device with "human readers." Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • This is not an algorithm. The relevant "standalone performance" refers to the intrinsic performance of the chemical/biological solution, which is assessed through the MEA and stability studies.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The primary "ground truth" or performance indicator for embryo toxicity is the Mouse Embryo Assay (MEA), which assesses the development of mouse embryos to the blastocyst stage. This is a laboratory-based, biological assay. Stability tests also involve objective measurements of chemical and physical properties (pH, osmolality, concentrations, enzymatic activity).

    8. The sample size for the training set:

    • This concept is not applicable as this is not an AI or machine learning device that requires a "training set."

    9. How the ground truth for the training set was established:

    • Not applicable (see point 8).

    In summary, the document addresses regulatory clearance based on substantial equivalence, and while it references performance data (like MEA and stability studies), it does not present a detailed study report with the specific methodological details requested for evaluating AI/diagnostic devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1