Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K140200

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-07-18

    (172 days)

    Product Code
    Regulation Number
    882.1350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K022914

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    E-Shield Multi-paired Subdermal Needle Electrodes are intended for use with recording and monitoring equipment for the purpose of recording of biopotential signals. Examples include: Electromyography (EMG), Electroencephagraph (EEG) and Nerve potential signals. The electrodes are sterile and for single patient use only.

    Device Description

    E-Shield Multi-paired Subdermal Needle Electrodes are single use, disposable, monopolar, non-pyrogenic and sterilized stainless steel subdermal needle electrodes. Electrodes are applied in the study of biopotentials such as electroencephalograph (EEG), electromyography (EMG). Electrodes are invasive as they are placed subcutaneously or in contact with nerve or muscle tissue. The subdermal needle electrodes can be used to record variety of types of neurological and neurophysiological evoked potentials, including but not limited to electroencephalogram (EEG), electromyogram (EMG) in hospital and clinical setting. It is highly recommended to be used of intra-operative neuromonitoring (IOM). The electrodes consist of a stainless steel needle with a lead wire attached and the lead wires terminate in a safety connector. The electrodes are used under the supervision of a physician. Each sterilized PET box contains 2 sets of 5-paired subdermal needle electrodes and a safety shield attached to each set. The safety shield is used to cover the needle sharps and the multi-paired lead wires are designed in group for easy organizing. Grouped wire cable can be separated into 5 paired lead wires to reach different recording sites of a patient over desired length. For instance, it can be used for the recording in one limb muscle groups.

    AI/ML Overview

    The manufacturer, NeuroInvent Inc., established acceptance criteria and conducted non-clinical testing to demonstrate that "E-Shield Multi-paired Subdermal Needle Electrodes" meet these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Testing ItemAcceptance Criteria (Standard Applied)Reported Device Performance
    BiocompatibilityISO 10993-5:2009 (Tests for in vitro cytotoxicity) ISO 10993-10:2010 (Tests for irritation and skin sensitization) ISO 10993-11:2009 (Tests for systemic toxicity) ISO 10993-12:2012 (Sample preparation and reference material) ASTM F756-08 (Standard Practice For Assessment Hemolytic Properties Of Materials) USP 35-NF30:2012, <151> Pyrogen Test (USP Rabbit Test) ASTM F750-87 (Reapproved 2012) (Standard Practice For Evaluating Material Extracts By Systemic Injection In The Mouse)All test results demonstrate the device meets the requirements of its pre-defined acceptance criteria and intended uses.
    SterilizationISO 10993-7:2008 (Ethylene Oxide sterilization residuals) ISO 11135-1:2007 (Ethylene oxide - Requirements for development, validation and routine control of a sterilization process for medical devices) ISO 11135-2:2008 (Ethylene oxide -- Guidance on the application of ISO 11135-1) ISO 11737-1:2006 (Sterilization of Medical Devices - Microbiological Method - Part I: Determination of a population of microorganisms on products) ISO 11737-2:2009 (Sterilization of medical devices - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process)All test results demonstrate the device meets the requirements of its pre-defined acceptance criteria and intended uses.
    Shelf LifeASTM F1980-07 (Standard Guide for Accelerated Aging of Sterile Medical Device Packages) ASTM F88/F88M-09 (Standard Test Method for Seal Strength of Flexible Barrier Materials) ASTM F1140-13 (Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages for Medical Applications) ASTM F1929-12 (Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration) ASTM D4332:2001 (Standard practice conditioning containers, packages or packaging components for testing) ASTM F1608 (Standard Test Method for Microbial Ranking of Porous Packaging Materials (Exposure Chamber Method)) ASTM D4169-09 (Standard Practice For Performance Testing Of Shipping Containers And Systems) ISO 11607-1 (Packaging For Terminally Sterilized Medical Devices - Part 1: Requirements For Materials, Sterile Barrier Systems And Packaging Systems) ISO 11607-2 (Packaging For Terminally Sterilized Medical Devices - Part 2: Validation Requirements For Forming, Sealing And Assembly Processes) ASTM F2096-11 (Standard Test Method For Detecting Gross Leaks In Packaging By Internal Pressurization (Bubble Test)) The International Safe Transit Association (ISTA) Procedure 1AAll test results demonstrate the device meets the requirements of its pre-defined acceptance criteria and intended uses.
    Electromagnetic Compatibility & Electrical SafetyIEC 60601-1 (Medical electrical equipment Part 1: General requirements for basic safety and essential performance)All test results demonstrate the device meets the requirements of its pre-defined acceptance criteria and intended uses.
    PerformanceISO 9626 First Edition 1991-09-01, Amendment 1 2001-06-01 (Stainless Steel Needle Tubing For The Manufacture Of Medical Devices) ASTM B193-87(1992) (Standard Test Method for Resistivity of Electrical Conductor Materials) IEC 60028 (1925) (International standard of resistance for copper) GB/T 3048.2-2007 (TEST method for electrical properties of electric cables and wire-part2: Test of electrical resistivity of metallic materials)All test results demonstrate the device meets the requirements of its pre-defined acceptance criteria and intended uses.

    2. Sample size used for the test set and the data provenance:

    The document does not specify the exact sample sizes for each non-clinical test (e.g., how many electrodes were tested for biocompatibility, sterility, etc.). It only lists the standards to which the tests were conducted.

    The data provenance is from non-clinical testing performed by NeuroInvent Inc. There is no mention of patient data or specific geographic origin for the test materials beyond the product being manufactured by a company based in Mountain View, CA, with contact information in Taiwan.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable as the study involved non-clinical, bench testing rather than studies requiring expert interpretation of patient data to establish ground truth. The acceptance criteria were based on recognized international and national standards for medical devices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable as the study involved non-clinical, bench testing. Adjudication methods are typically employed in clinical studies or studies involving human readers to resolve discrepancies in interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC comparative effectiveness study was done. The document explicitly states: "No clinical test data was used to support the decision of safety and effectiveness." The device is a physical medical electrode, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable. The device is a physical medical electrode, not a standalone algorithm. The non-clinical testing evaluated the physical and material properties of the electrodes.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the non-clinical testing was established by adherence to established international and national standards and regulations for medical devices (e.g., ISO, ASTM, USP, IEC, GB/T). The tests performed according to these standards (e.g., cytotoxicity tests, pyrogen tests, seal strength tests, electrical resistivity tests) provide the "ground truth" for the device's physical and material properties.

    8. The sample size for the training set:

    This information is not applicable. There was no training set, as the study involved non-clinical, bench testing of the physical device to meet predefined standards, not machine learning model training.

    9. How the ground truth for the training set was established:

    This information is not applicable as there was no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1