Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K981722

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-06-03

    (19 days)

    Product Code
    Regulation Number
    862.1155
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K022818, K021203, K981177

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The test is a rapid immunoassay for the qualitative detection of hCG in urine as an aid in the early detection of pregnancy. The test is intended for overthe-counter home use.
    The Quidel Home Pregnancy Test is intended for the qualitative detection of hCG in urine for the early detection of pregnancy. The test is for over-the-counter home use.

    Device Description

    The test, a sandwich-format, lateral-flow immunoassay, employs monoclonal and polyclonal antibodies. If the sample contains hCG, a pink Test Line along with a blue Control Line is visible in the large Result Window to indicate a positive result. If hCG is not present in the sample, only the blue Control Line is visible in the large Result Window to indicate a negative result.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Quidel Home Pregnancy Test, based on the provided text:

    Acceptance Criteria and Device Performance

    The document doesn't explicitly state quantitative acceptance criteria in a table format. However, it outlines the areas where the device's performance was evaluated and deemed acceptable. The reported device performance is described generally as "excellent" or "similar to other commercially available products."

    Acceptance CriteriaReported Device Performance
    Substantial Equivalence: Similar features and intended use to existing marketed tests.The test was shown to be similar to other commercially distributed tests in terms of features and intended use.
    Precision: Intra- and inter-assay consistency.The test was shown to have excellent intra- and inter-assay precision.
    Lot-to-Lot Consistency: Reproducible manufacturing.Lot-to-lot consistency analyses showed the test to be reproducibly manufacturable.
    Interference: Not affected by common drugs, chemicals, and biologicals.Common drugs, chemicals, and biologicals were shown not to interfere with the test's performance.
    Consumer Usability: Lay users can correctly perform and interpret the test.A consumer study was conducted in order to show that lay users could read the package insert, perform the test, and obtain results similar to results obtained by clinic personnel.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • The document does not specify the exact sample sizes for any of the test sets (precision, lot-to-lot consistency, interference, or consumer study).
      • Data Provenance: Not explicitly stated, though a consumer study would imply a prospective, real-world setting. The other studies (precision, lot-to-lot, interference) are typically conducted in a laboratory setting. There is no mention of country of origin for the data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • For the consumer study, the "ground truth" was established by clinic personnel whose results were compared to those of lay users. The number of clinic personnel and their specific qualifications (e.g., nurses, lab technicians) are not specified.
      • For the other performance studies (precision, consistency, interference), the ground truth would be based on established analytical methods and reference standards for hCG detection, likely performed by trained laboratory professionals, but the document does not specify the number or qualifications of these experts.
    3. Adjudication method for the test set:

      • The document does not describe any specific adjudication method (e.g., 2+1, 3+1). For the consumer study, it implies direct comparison between lay user results and clinic personnel results rather than an adjudication process between conflicting interpretations.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This device is a rapid immunoassay for home use, not an AI-assisted diagnostic device, so this type of study is irrelevant.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, in a sense, the intrinsic performance characteristics such as precision, lot-to-lot consistency, and interference studies represent the standalone performance of the assay itself, demonstrating its ability to detect hCG in urine samples. The device is a "standalone" test kit that provides a visible result without external analytical equipment or further human interpretation beyond reading the lines.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Analytical Ground Truth for Performance Studies: For precision, lot-to-lot, and interference studies, the ground truth would be based on known concentrations of hCG in prepared samples or reference materials, confirmed by established laboratory methods.
      • Clinical/User Ground Truth for Consumer Study: For the consumer study, the ground truth for correct test interpretation was established by the results obtained by clinic personnel. The ultimate "ground truth" for pregnancy itself is physiological (presence of hCG), but the study focused on whether users could correctly identify the presence/absence of hCG based on the test lines.
    7. The sample size for the training set:

      • The document does not mention a training set. This is because the device is a chemical immunoassay, not a machine learning or AI-based system that requires training data.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1