510(k) Data Aggregation

    K Number
    K141529

    Validate with FDA (Live)

    Date Cleared
    2014-12-03

    (177 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K062420, K021256, K022764, K023402, K043387

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Alviguide™ Blue+ Interventional Cardiology Guiding Catheter is intended for use in the intravascular introduction of interventional/diagnostic devices into the coronary vascular system.

    The Alguide™ Endovascular Guiding Catheter is intended for use in the intravascular introduction of interventional/ diagnostic devices into the peripheral vascular system.

    Device Description

    The Alviguide™ Blue+ and Alguide™ Guide Catheters are sterile, non-pyrogenic, single lumen catheters with a tri-layer catheter body construction consisting of a polymer inner layer and a radiopaque outer layer surrounding a stainless steel wire braid, and a soft tip at the distal end. The catheters are for single-use only.

    The Alviguide™ Blue+ and Alguide™ Guide Catheters are provided in a variety of distal shape configurations and are available in Fr Sizes 5, 6, 7 and 8 and lengths of 45, 60, 65 and 80cm (Alguide™), and 100cm (Alviguide™ Blue+ and Alguide™).

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for two guiding catheters and lists the testing performed to demonstrate substantial equivalence to predicate devices. However, it does not contain information about the acceptance criteria for a device's performance in a study, nor does it detail a study that proves the device meets specific acceptance criteria in the context of an AI/algorithm-based medical device.

    The document focuses on:

    • Device Description: Alviguide™ Blue+ Interventional Cardiology Guiding Catheter and Alguide™ Endovascular Guiding Catheter.
    • Intended Use: Intravascular introduction of interventional/diagnostic devices into the coronary/peripheral vascular system.
    • Substantial Equivalence: Comparison to predicate devices based on similar indications for use, design, materials, and nonclinical (bench) and biocompatibility testing.

    The "testing in support of substantial equivalence determination" section lists various bench tests and biocompatibility tests, but these are focused on the physical and biological properties of the catheter itself, not on the performance of a diagnostic algorithm or AI system. There is no mention of "acceptance criteria" for a study in the sense of accuracy, sensitivity, specificity, or performance thresholds for an AI.

    Therefore, I cannot fulfill the request to provide a table of acceptance criteria and reported device performance or details about an AI-driven study from the provided text. The document pertains to a physical medical device (catheter), not an AI/software-as-a-medical-device (SaMD).

    Ask a Question

    Ask a specific question about this device

    K Number
    K091188

    Validate with FDA (Live)

    Date Cleared
    2010-02-26

    (309 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K023402, K022764, K972484, K061601, K020028

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ASAHI ZenyteEX Guiding Catheter is intended for use in coronary or peripheral vascular applications and is designed to provide a pathway through which medical instruments, such as balloon catheters, guide wires or other therapeutic devices may be introduced. This device is not intended for use in the cerebral vasculature.

    Device Description

    The ASAHI ZenyteEX Guiding Catheter is intended for use in coronary or peripheral vascular applications and is designed to provide a pathway through which medical instruments, such as balloon catheters, guide wires or other therapeutic devices may be introduced. The Catheter consists of a tube, which is to be inserted into vasculature, a proximal hub/connector, and strain relief/protector for the joint portion of the first 2 sections. The tube body section consists of soft tip, soft tube and shaft. The tube body itself is constructed of a polyamide elastomer. The shaft is reinforced by stainless steel braid, and it consists of two layers. The distal portion of the tube is made soft in order to easily bend.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called the ASAHI ZenyteEX Guiding Catheter. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a study with specific acceptance criteria and performance metrics in the way that would typically be done for a novel AI/software device.

    Therefore, much of the requested information (such as sample sizes for test/training sets, number of experts, adjudication methods, MRMC studies, standalone performance for an algorithm, and ground truth establishment for a training set) is not applicable to this type of device submission as described in the provided document.

    However, I can extract the relevant information regarding acceptance criteria and the type of study performed to show the device meets these criteria.


    Acceptance Criteria and Device Performance for ASAHI ZenyteEX Guiding Catheter

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/Tests PerformedReported Device Performance
    BiocompatibilityCompatibility with patient contact (materials)"All components that come in direct contact with the patient have a long history of use in medical devices and are proven to be biocompatible for use in the vasculature. Furthermore, this submission contains biocompatibility testing conducted on the subject device." (Implies successful completion.)
    Technical/DimensionalDimensional specifications equivalence to predicate devices"The dimensional specifications are equivalent to those listed for the currently cleared predicate devices."
    Material CompositionUse of same materials as predicate devices"The ASAHI ZenyteEX Guiding Catheter is made of the same materials that have been used in other predicate devices that are labeled for the similar indications."
    Functional PerformanceMechanical and Functional Bench Testing"This 510(k) notice also includes mechanical and functional bench testing that demonstrates that the ASAHI ZenyteEX Guiding Catheter performs as intended." (Implies successful performance.)
    Substantial EquivalenceEquivalence to specified predicate devices for same indication for use"The ASAHI ZenyteEX Guiding Catheter characteristics are substantially equivalent to the specified predicate devices and other currently marketed devices for the same indication for use."
    Intended UseAbility to provide a pathway for medical instruments"Bench testing demonstrates that the device functions as intended." (Supports the intended use.)

    2. Sample Sizes Used for the Test Set and Data Provenance

    Not applicable. This is a physical medical device (guiding catheter), not a software or AI device requiring a "test set" of data. The "testing" refers to physical and material testing. The document does not specify sample sizes (e.g., number of catheters tested) for the mechanical, functional, or biocompatibility bench tests. Data provenance is also not applicable in the context of data for an AI/software device; the provenance here would be the manufacturing site (Japan).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. As this is a physical device, there is no "ground truth" in the context of expert review of data. Performance is evaluated via engineering, material, and biocompatibility testing.

    4. Adjudication Method for the Test Set

    Not applicable. There is no "adjudication method" as would be used for diagnostic output from an AI/software device. Performance is determined by meeting pre-defined physical and chemical specifications.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical medical device, not an AI/software diagnostic aid.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    Not applicable in the context of expert consensus, pathology, or outcomes data. For a physical device like this, the "ground truth" for performance is established through adherence to engineering specifications, material properties, and regulatory standards demonstrated via bench testing.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI/software device that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. This is a physical medical device, not an AI/software device that requires a training set.


    Ask a Question

    Ask a specific question about this device

    K Number
    K023402

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-11-06

    (27 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K022764

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medtronic® Launcher™ Guiding Catheter is designed to provide a pathway through which therapeutic devices are introduced. The Launcher catheter is intended to be used in the coronary or peripheral vascular system.

    Device Description

    The Medtronic Launcher Guide Catheter is constructed with an inner liner, stainless steel braid, outer shaft jacket, sleeve, marker band and a soft distal tip. The inner lumen of the Launcher Guide Catheter has a thin lubricious coating.

    AI/ML Overview

    The provided document describes a Special 510(k) Notification for the Medtronic 8F Launcher Guiding Catheter. This type of submission is for modifications to a currently cleared device, demonstrating substantial equivalence to a predicate device. As such, the study focuses on functional and material equivalence rather than clinical performance against specific disease detection criteria.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are based on demonstrating substantial equivalence to the predicate device, the Medtronic 7F Launcher Guide Catheter (K022764), in terms of materials, manufacturing processes, design, and performance in in-vitro tests.

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    BiocompatibilityCompliance with ISO 10993 for external-communicating devices, circulating blood, limited exposure (<24 hours), including: Acute Intracutaneous Reactivity, Acute Systemic Toxicity, Cytotoxicity, Hemolysis, Sensitization."All materials used for the Medtronic® Launcher Guiding Catheter passed the following tests: Acute Intracutaneous Reactivity, Acute Systemic Toxicity, Cytotoxicity, Hemolysis, Sensitization."
    Functional PerformanceMeeting applicable specifications for: hub to shaft, shaft, distal tip to shaft tensile, torque response, lumen lubricity, radiopacity, pressure shaft burst and hub leak, tip compression (softness)."Results verified that the Medtronic® Launcher Guiding Catheter meets all of the applicable specifications and considered an appropriate device for the intended use."
    Technological EquivalenceManufacturing processes, materials, and design are substantially equivalent to the 7F Launcher Guide Catheter."The manufacturing processes, materials and design are substantially equivalent to the 7F Launcher Guide Catheter."
    Intended UseConsistent with the predicate device: providing a pathway for therapeutic devices in the coronary or peripheral vascular system."The Medtronic Launcher Guide Catheter is designed to provide a pathway through which therapeutic devices are introduced. The guiding catheter is intended for used in the coronary or peripheral vascular system." (Matches predicate's use)

    Study Details

    Given this is a 510(k) notification for a guiding catheter, the "study" is primarily an in-vitro testing and comparison assessment rather than a clinical trial with human subjects for diagnostic accuracy.

    1. Sample size used for the test set and the data provenance:

      • The document does not specify a numerical sample size for the in-vitro functional and biocompatibility tests. It refers to "tests" being performed.
      • Data Provenance: The tests were conducted internally by Medtronic AVE, Inc. (the submitter). The provenance is therefore considered internal company testing, and by nature, it is prospective with respect to the testing itself (i.e., the tests were performed specifically for this submission). There's no indication of country of origin for the data other than the submitter's location in Danvers, MA, USA.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable in the context of this type of device and submission. "Ground truth" in this scenario would refer to the established performance specifications for medical devices of this type (e.g., ISO standards, internal Medtronic specifications). These specifications are typically set by engineering and scientific teams within the company and regulatory bodies, not by "experts" in the sense of clinical reviewers for diagnostic accuracy.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. Adjudication methods like 2+1 are used in clinical studies when there are differing opinions from human readers on a diagnostic task. For in-vitro functional and biocompatibility testing, adherence to a specification is typically a pass/fail outcome, not subject to human adjudication of interpretation.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is a physical medical instrument (a catheter), not an AI-powered diagnostic or assistive tool. MRMC studies are specific to evaluating diagnostic systems, especially those involving AI or human interpretation.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. As mentioned, this is a physical medical device, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for the performance claims relies on established engineering specifications, material standards (e.g., ISO 10993 for biocompatibility), and manufacturing quality control parameters. For the claim of substantial equivalence, the "ground truth" is the proven performance of the predicate device.
    7. The sample size for the training set:

      • Not applicable. This device does not involve a "training set" in the machine learning sense. The device's design and manufacturing processes are likely derived from years of experience and iterative development, but not a formally defined "training set" for an algorithm.
    8. How the ground truth for the training set was established:

      • Not applicable. As there is no training set, there is no ground truth establishment for it.