Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K234107

    Validate with FDA (Live)

    Date Cleared
    2024-04-23

    (119 days)

    Product Code
    Regulation Number
    876.1725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K022023

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Solar anorectal manometry catheter is a non-sterile, single patient use catheter for use on patient requiring anorectal manometry testing.

    Device Description

    The Solar™ Anorectal Manometry Catheter is a high-resolution anorectal manometry (HRAM) catheter. It is a diagnostic device used to assess the function of anal and rectum muscles. The device consists of 10 air-charged sensing balloons which can sense pressures during anorectal manometry, along with a distal balloon used for the assessment of recto anal inhibitory reflex (RAIR). The catheter is disposable and non-sterile. Through manual insertion in the device assesses pressures within the anorectal anatomy as well as assessing sensation and reflexes. This product is non-sterile, single patient use catheter intended for use with a reusable Solar™ Anometry Charger on patients requiring lower gastrointestinal pressure measurements.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria for an AI/ML device or a study proving that such a device meets acceptance criteria. The document is an FDA 510(k) clearance letter for a Solar™ Anorectal Manometry Catheter, which is a medical device for measuring gastrointestinal motility, not an AI/ML product.

    Therefore, I cannot extract the requested information regarding acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth for an AI/ML device from this document.

    The document focuses on:

    • The regulatory clearance of the catheter.
    • Comparisons of the new catheter’s design and technology to a predicate device.
    • Non-clinical bench performance testing (visual, dimensional, functional, destructive, biocompatibility, bioburden, transit simulation) for the catheter and its charger. These tests aim to demonstrate substantial equivalence to the predicate device, not the performance of an AI/ML algorithm.
    • Usability assessment according to EN IEC 62366-1 for the device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1