    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K922058, K992636, K021422

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For central monitoring of adult, pediatric, and neonatal patients; and where the clinician decides to monitor cardiac arrhythmia of adult, pediatric, and neonatal patients and/or ST segment of adult patients to gain information for treatment, to monitor adequacy of treatment, or to exclude causes of symptoms.

    Device Description

    The Philips Drug Calculator Software served by the Philips IntelliVue application server software (M2387A Application Server Software). The modification is primarily a software based change that incorporates the Rule of Six algorithm for IV drug computations in 100 mL IV bags.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Philips Drug Calculator Software. However, the documentation focuses primarily on the regulatory aspects of substantial equivalence and does not contain the specific details required to answer all questions about acceptance criteria and a dedicated study proving device performance against those criteria.

    Here's what can be extracted and what information is missing:


    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly list acceptance criteria or specific reported device performance in a quantitative manner. It states:

    • "Verification testing activities were conducted to establish the performance and reliability characteristics of the new device with respect to the predicate."
    • "Pass/Fail criteria were based on the specifications cleared for the predicate devices and test results showed substantial equivalence."

    This indicates that internal specifications and the performance of predicate devices served as benchmarks, but the actual criteria and performance metrics are not detailed.


    Missing Information for 2-9:

    The provided text focuses on the 510(k) summary for the Philips Drug Calculator Software, emphasizing its substantial equivalence to previously cleared devices. It states that "Verification testing activities were conducted to establish the performance and reliability characteristics of the new device with respect to the predicate," and that "Pass/Fail criteria were based on the specifications cleared for the predicate devices and test results showed substantial equivalence."

    However, the document does not provide the detailed information requested regarding sample sizes, data provenance, expert involvement, adjudication methods, details of comparative effectiveness studies (MRMC), standalone performance, types of ground truth, or training set specifics. The document is a regulatory submission summary, not a detailed study report.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Reference Devices :

    K942843, K992674, K003038, K021422

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters patients.

    Intended use: For monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in health care facilities.

    EASI 12-lead ECG is only for use on adult and pediatric patients.

    ST Segment monitoring is restricted to adult patients only.

    The transcutaneous gas measurement (tcp02 / tcpCO2) is restricted to neonatal patients only.

    Bispectral Index (BIS) monitoring is for use in monitoring the state of the brain by data acquisition of EEG signals in the intensive care unit, operating room, and for clinical research. The Bispectral Index, a processed EEG variable, may be used as an aid in monitoring the effects of certain anesthetic agents*.

    Device Description

    Philips Medical Systems, M1275B Component Compact Monitor, Release A.04, with M1021A SvO2 Module, M1027A EEG Module, and M1034A BIS Module. The modification is updated software and the ability to support the Philips M1021A SvO2 Module, M1027A EEG Module, and M1034A BIS Module as well as the M2385A Application Server.

    AI/ML Overview

    The provided 510(k) summary focuses on the Philips Medical Systems, M1275B Component Compact Monitor. However, the document does not contain explicit acceptance criteria tables or detailed study results demonstrating the device's performance against specific metrics.

    Instead, the submission states that:

    • "Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the new device with respect to the predicate."
    • "Testing involved system level tests, performance tests, and safety testing from hazard analysis."
    • "Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence."
    • "The results demonstrate that Component Compact Monitor meets all reliability requirements and performance claims."

    This indicates that internal testing was conducted to ensure the device met the established specifications of its predicate devices, but the specific numerical acceptance criteria and reported performance values are not disclosed in this summary.

    Given the information provided, it's not possible to populate the requested table comprehensively or answer all specific questions about a detailed performance study with quantitative results, as the provided text only refers to the existence of such testing and its conclusion of substantial equivalence.

    However, based on the general statements and the context of a 510(k) submission, we can infer some information:

    1. Acceptance Criteria and Reported Device Performance: This information is not explicitly provided in a table format. The document states that "Pass/Fail criteria were based on the specifications cleared for the predicate device." This implies that the device was expected to perform at least as well as the predicate devices across all measured parameters (e.g., accuracy of gas analysis, arrhythmia detection, blood pressure measurement, ECG performance). The "reported device performance" is simply stated as meeting "all reliability requirements and performance claims" and demonstrating "substantial equivalence" to the predicate.

    2. Sample Size and Data Provenance: Not specified in the document. The testing described is internal verification and validation, likely conducted by the manufacturer.

    3. Number of Experts and Qualifications: Not specified. This type of expert review is typically not required for the technical performance testing of a multi-parameter monitor unless a specific component (e.g., arrhythmia detection algorithm) involved subjective interpretation, which is not detailed here.

    4. Adjudication Method: Not applicable. The testing described is objective performance and safety testing against predefined specifications, not subjective reviews requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not mentioned and typically not applicable for this type of medical device (a multi-parameter monitor). MRMC studies are common for diagnostic imaging algorithms to assess human-AI interaction.

    6. Standalone Performance: Yes, the described "system level tests, performance tests, and safety testing" refer to the standalone performance of the device without human interaction being part of the performance metric itself.

    7. Type of Ground Truth: The ground truth for this type of device would be objective, established reference standards for physiological measurements (e.g., calibrated gas mixtures for gas analyzers, known signal generators for ECG and arrhythmia detection, calibrated blood pressure cuffs for NBP, etc.). The document doesn't explicitly state the specific ground truth methodologies.

    8. Sample Size for Training Set: Not applicable. This device is a physiological monitor, not an AI/ML-based diagnostic algorithm that learns from a training set in the way a deep learning model would. The "software update" mentioned is more likely a traditional software update rather than a new AI model.

    9. How Ground Truth for Training Set was Established: Not applicable as there is no specific "training set" for an AI/ML model described.

    In summary of what can be extracted from the provided text:

    While detailed quantitative data is absent, the submission asserts that the Philips M1275B Component Compact Monitor underwent verification, validation, and testing activities. These tests demonstrated that its performance, functionality, and reliability were substantially equivalent to its predicate devices, meeting all established specifications and performance claims. The "Pass/Fail criteria were based on the specifications cleared for the predicate device." This constitutes the implicit acceptance criteria.