510(k) Data Aggregation

    K Number
    K033876

    Validate with FDA (Live)

    Date Cleared
    2004-03-11

    (87 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K030952,K012989,K032947

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dolphin 2000™ Oximetry Sensors are indicated for use in continuous monitoring of arterial oxygen saturation and pulse rate.

    Device Description

    The Dolphin 2000™ Oximetry Sensors are fully compatible disposable and re-usable replacement sensors for use with major brands of pulse oximeter monitors. They are Nellcor and BCI Compatible Sensors.

    The disposable Dolphin 2000 Oximetry Sensors are constructed in a similar manner to predicate devices. The emitter and detector diodes are embedded in a laminate of tapes and development an adhesive bandage that is connected to the cable assombly. "The consortient by wrapping it around a finger backing that allows the sensor to be upplied to the End (Colphin 2000 Oximetry Sensors are of toe (measurement site). Tour sizes of aloposes, infant and neonatal application available, which are indicated for doc for duall, positive, mon-sterile for single patient use.

    The re-usable Dolphin 2000 Finger Clip Oximetry Sensor is an adult-sized clothespin-The re-usable Dolphil 2005 Finger Olip Oximony Consor consists of the style clip that is placed on the one on anyon one on anyon maintained in mild emitter and detector components mounted in opposing clip halves, mainter and consected emitter and delection components mounted in opposing only his house the optiolection in complession by a spring mings. The montain contact with the patient's finger. Clear components within contoured paids that the read to pass through the finger for the measurements.

    The Re-usable Y sensor is for use on the ear, finger, hand, or neonatal foot and held in The Re-usable 1 Sensor is for use on the sensor can also be used on the adult ear with the place with a disposable bandage. The concer or enounted in a sealed pouch (same ear clip accessory." The emitter and other cructed in a Y shape. All Dolphin sensors are provided non-sterile.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Dolphin 2000™ Pulse Oximetry Y Sensor, structured according to your requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The given text does not explicitly state quantitative acceptance criteria for SpO2 accuracy that Dolphin Medical Inc. aims to meet. Instead, it states that "The labeled accuracy of the Dolphin 2000 sensors is equivalent to those of the predicate devices." For the purpose of this analysis, we will infer the de-facto acceptance criteria based on standard pulse oximeter testing practices and the reported performance.

    Acceptance Criteria (Inferred from standard practice)Reported Device Performance (Dolphin 2000 SpO2 Accuracy)
    ARMS (Accuracy Root Mean Square) for SpO2 values within the range of 70-100% must be within a clinically acceptable limit (typically ≤ 3% for pulse oximeters).Data was analyzed to determine the ARMS for each probe. Specific ARMS values are not provided, but the study was conducted to demonstrate "scientific accuracy" by "statistically comparing Dolphin 2000 SpO2 values to functional SaO2 values."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. The text mentions "Volunteers participated in breathing down protocols," indicating human subjects.
    • Data Provenance:
      • Country of Origin: USA (VA Hospital of Wisconsin - Milwaukee).
      • Retrospective or Prospective: Prospective, as it involved "breathe-down protocols" with "volunteers" specifically for the validation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not explicitly stated. The study was overseen by "Dr. Phillip Clifford, MD." It is common for a physician or a team of physicians to oversee such clinical studies and approve the ground truth, but the exact number establishing the ground truth is not specified.
    • Qualifications of Experts: Dr. Phillip Clifford, MD. His specific specialization (e.g., pulmonologist, anesthesiologist) is not mentioned, but an MD degree suggests appropriate medical expertise for overseeing oxygen saturation studies.

    4. Adjudication Method for the Test Set

    Not specified. The clinical testing involved comparing SpO2 values to functional SaO2 values, implying a direct comparison to a gold standard, rather than a consensus-based adjudication of the device's output.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly done. This device is a sensor, not an AI-assisted diagnostic tool that would typically involve human readers interpreting output. The study focused on the standalone accuracy of the sensor itself.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance study was done. The "Clinical Testing" section describes a study where "Dolphin 2000 SpO2 values" were statistically compared to "functional SaO2 values," which is a direct measurement of the sensor's accuracy in a human clinical setting without human interpretation of the sensor's raw output.

    7. The Type of Ground Truth Used

    The ground truth used was functional SaO2 values. This is typically measured using arterial blood gas analysis (CO-oximetry), which is considered the gold standard for measuring arterial oxygen saturation.

    8. The Sample Size for the Training Set

    Not applicable. The Dolphin 2000 Oximetry Sensor is a hardware device (sensor) and not an AI/Machine Learning algorithm that requires a training set. Its "performance" is based on its physical design and electro-optical properties.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as this device does not use a training set in the context of AI/ML. Its ground truth for performance validation (as described in point 7) was established through clinical measurements against a gold standard.

    Ask a Question

    Ask a specific question about this device

    K Number
    K030952

    Validate with FDA (Live)

    Date Cleared
    2003-05-19

    (54 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K012989, K863784, K991823

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dolphin 2000 Oximetry Sensors are indicated for use in continuous monitoring of arterial oxygen saturation and pulse rate.

    Device Description

    The Dolphin 2000 Oximetry Sensors are fully compatible disposable and re-usable replacement sensors for use with Nellcor pulse oximeter monitors. They represent a design change to the Dolphin 2000 BCI Compatible Sensors.

    The disposable Dolphin 2000 Oximetry Sensors are constructed in a similar manner to predicate devices. The emitter and detector diodes are embedded in a laminate of tapes that is connected to the cable assembly. The sensors have an adhesive bandage backing that allows the sensor to be applied to the patient by wrapping it around a finger or toe (measurement site). Four sizes of disposable Dolphin 2000 Oximetry Sensors are available, which are indicated for use for adult, pediatric, infant and neonatal application sites. The Dolphin 2000 disposable sensors are provided non-sterile for single patient use.

    The re-usable Dolphin 2000 Finger Clip Oximetry Sensor is an adult-sized clothespinstyle clip that is placed on the end of a finger. The finger clip sensor consists of the emitter and detector components mounted in opposing clip halves, maintained in mild compression by a spring hinge. The molded outer components house the optoelectric components within contoured pads that maintain contact with the patient's finger. Clear windows within these pads permit the optical energy to pass through the finger for the measurements. The Dolphin 2000 re-usable sensors are provided non-sterile.

    AI/ML Overview

    The Dolphin 2000 Oximetry Sensors were validated in clinical testing to demonstrate their accuracy in measuring arterial oxygen saturation (SpO2).

    1. Table of Acceptance Criteria and Reported Device Performance:

    ParameterAcceptance Criteria (Clinical Validation)Reported Device Performance (Clinical Validation)
    ARMS for Reusable, Adult Disposable, and Pediatric/Infant Disposable probes< 2.0% in the range of 70-100% SaO2< 2.0% ARMS in the range of 70-100% SaO2
    ARMS for Neonatal Disposable probes< 3.0% in the range of 70-100% SaO2< 3.0% ARMS in the range of 70-100% SaO2

    2. Sample Size and Data Provenance:

    • Test Set Sample Size: Not explicitly stated how many volunteers participated, but the study implies a group of "Volunteers."
    • Data Provenance: Prospective (breathe-down protocol in a clinical setting), conducted at the VA Hospital of Wisconsin - Milwaukee, USA.

    3. Number of Experts and Qualifications:

    • Number of Experts: One expert is explicitly named: Dr. Phillip Clifford, MD.
    • Qualifications: "Dr. Phillip Clifford, MD." implies a medical doctor, likely with expertise in physiology or critical care relevant to oximetry. Specific years of experience or specialization are not detailed.

    4. Adjudication Method:

    • Not applicable as the ground truth (functional SaO2) was established by direct measurement from arterial blood gas analysis, not expert consensus or review.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The study focuses on the standalone accuracy of the device against a physiological ground truth.

    6. Standalone Performance:

    • Yes, a standalone study was performed. The device's SpO2 values were statistically compared to functional SaO2 values obtained from volunteers during a breathe-down protocol.

    7. Type of Ground Truth Used:

    • Physiological Ground Truth: Functional SaO2 values, likely obtained through co-oximetry via arterial blood gas analysis, are implied as the comparison standard ("statistically comparing Dolphin 2000 SpO2 values to functional SaO2 values").

    8. Sample Size for the Training Set:

    • The document does not explicitly mention a separate "training set" or its size. The described "Clinical Testing" focuses on validation. It is common for such medical devices to be developed and refined using internal data, but details of that process are not provided in this 510(k) summary.

    9. How the Ground Truth for the Training Set Was Established:

    • As a separate training set is not explicitly mentioned, the method for establishing its ground truth is not detailed in this document.