510(k) Data Aggregation

    K Number
    K030457

    Validate with FDA (Live)

    Device Name
    REX, VERSION 3.0
    Manufacturer
    Date Cleared
    2003-04-08

    (56 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K021099, K012290

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    REX™ 3.0 is a software package intended for viewing and manipulating DICOM-compliant medical images acquired from CT and MR scanners. REX™ 3.0 can be used for real-time image viewing, image manipulation, 3D volume rendering, virtual endoscopy, and issuance of reports.

    Device Description

    REX™ 3.0 is a tool for 3D (three dimensional) and 2D (two dimensional) viewing and manipulation of DICOM compliant CT and MR images. The proposed software provides real-time image viewing, image manipulation, 3D volume rendering, virtual endoscopy, and issuance of reports.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification Summary for the REX™ 3.0 PACS / Image Processing Software. It focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a specific study proving the device meets them in the way modern AI/medical device submissions typically do.

    Based on the provided text, here's a breakdown of the information requested, with "N/A" where the information is not available in the document:


    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes the device's performance through a comparison to predicate devices rather than against pre-defined, quantitative acceptance criteria. The "acceptance criteria" here are implicitly that the REX™ 3.0 software performs all specified functions in line with software requirements and safety standards, and is substantially equivalent to predicate devices.

    Feature/CriterionREX™ 3.0 Reported Performance (Implicit Acceptance)
    DICOM ConformanceConforms to DICOM Version 3.0.
    Functional RequirementsPerforms all input functions, output functions, and all required actions according to the functional requirements specified in the Software Requirements Specification. Validation testing confirmed this.
    Non-Clinical Performance (Safety/Hazards)Potential hazards identified in Hazard Analysis are controlled by design controls, protective measures, and user warnings. No new potential safety risks identified.
    Intended UsePerforms in accordance with its intended use (viewing and manipulating DICOM-compliant medical images from CT and MR scanners, real-time image viewing, image manipulation, 3D volume rendering, virtual endoscopy, and issuance of reports).
    Equivalence to REX™ 1.0Substantially equivalent to REX™ 1.0 with the addition of MR image analysis functions and a dual-monitor setup (one for image viewing, one for report viewing).
    Equivalence to Rapidia® V 2.0Substantially equivalent in common features and specifications.
    Image SourcesSupports CT and MR images (enhancement over REX™ 1.0 which only supported CT).
    Operating SystemOperates on Windows 2000. (Note: Not on Windows XP or NT, unlike Rapidia® V 2.0).
    Multi Planar ReformattingYes (enhancement over REX™ 1.0 which did not have this).
    Other Features (GUI, Patient Demographics,Yes (comparable to predicate devices for these listed features: GUI, Platform, PC, Patient Demographics, Networking (TCP/IP), DICOM 3.0 compliant, PNG (Lossless) image compression, Annotations, 3D Volume rendering, Still/Window/Level/Zoom/Pan/Flip for image review, 2D measurements (length, area), DICOM 3.0 image input, PNG (lossless snapshots) image output, Standard monitor use, Patient and Study Browser, Measure Image Intensity Values (ROI), Standalone software, Virtual Endoscopy (instant access to lesions, real-time display, internal/external viewing of hollow structures), Local Image Storage, True Color, User Login, Preset Window and Level, Image Conversion (for browser viewing), Trained Physician users, Volume Rendering algorithms, Reporting, Off-the-shelf hardware, Windows 2000 OS, DICOM compatible).
    Image Communication, Image Processing etc.)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document refers to "Validation testing" but does not specify a separate "test set" with a defined sample size for clinical or image-based performance evaluation. The "test set" is implicitly the DICOM-compliant images used during software validation, but no details are provided about their origin, number, or whether they were retrospective or prospective.

    • Sample Size: N/A (Not specified as a distinct clinical test set with a quantifiable size)
    • Data Provenance: N/A (Not specified)

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    N/A. The submission does not describe a process for establishing ground truth by expert consensus for a test set, as it is a PACS/image processing software focused on viewing and manipulation, not diagnostic interpretation or algorithm-based detection needing labeled ground truth in the context of an AI device. The validation is focused on software functionality and compliance.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    N/A. Since no specific test set with expert-established ground truth is described, no adjudication method is mentioned.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A. This is not an AI-assisted diagnostic device. It is a PACS/image processing software. Therefore, an MRMC study concerning AI assistance is not relevant or described.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device itself is described as "Standalone" software (meaning it's not embedded within a larger system). However, a "standalone algorithm performance" study related to an AI diagnostic function is not applicable here as it is not an AI diagnostic algorithm. The safety statement explicitly mentions: "Clinician interactive review/editing of data integral to use," indicating human-in-the-loop is part of its intended operational model.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    N/A. Ground truth in the context of diagnostic accuracy is not discussed because this is an image viewing and manipulation software, not a diagnostic algorithm. The "truth" for its proper functioning is adherence to DICOM standards and its own software requirements specification.

    8. The sample size for the training set

    N/A. This is not an AI/machine learning device that relies on a "training set" in the context of deep learning models. The software performs deterministic image processing and viewing functions.

    9. How the ground truth for the training set was established

    N/A. Not applicable, as there is no training set mentioned or implied for an AI/ML model.


    Summary of the Study and Device Performance:

    The "study" described in K030457 is primarily a software validation and verification process to ensure the REX™ 3.0 software conforms to its design specifications, DICOM standards (Version 3.0), and relevant regulations. It is a non-clinical performance data assessment rather than a clinical trial or performance study involving patient data in a diagnostic context.

    The primary method to "prove" the device meets acceptance criteria (which are largely functional and safety-based for this type of software) is through:

    1. Conformance to DICOM 3.0: Stated directly in the "Non-Clinical Performance Data" section.
    2. Validation and Verification Process: PointDx followed established procedures for software development, validation, and verification which confirm that REX™ 3.0 "performs all input functions, output functions, and all required actions according to the functional requirements specified in the Software Requirements Specification."
    3. Hazard Analysis: Potential hazards were identified and controlled through design, protective measures, and user warnings, concluding that REX™ 3.0 "does not result in any new potential safety risks."
    4. Substantial Equivalence Comparison: A detailed tabular comparison against predicate devices (REX™ 1.0 and Rapidia® V 2.0) highlights that REX™ 3.0 has similar features and functionalities, with improvements such as MR image support and multi-planar reformatting compared to REX™ 1.0, and overall equivalence in common features to Rapidia® V 2.0. This comparison implicitly serves as evidence that the device meets the "acceptance criteria" of being similar in performance and safety to already cleared devices.

    In essence, the submission relies on software engineering best practices and regulatory compliance to demonstrate that the REX™ 3.0 software functions as intended and is safe, rather than a clinical study measuring diagnostic performance or accuracy against ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K021913

    Validate with FDA (Live)

    Device Name
    CALSCOR SOFTWARE
    Manufacturer
    Date Cleared
    2002-09-06

    (87 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K012290

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CalScoR Software is intended to be used with the Rapidia software by a trained physician for the review and analysis of CT images as an aid in cardiac analysis.

    Device Description

    CalScoR is a Windows-compatible software program that runs with the previously cleared Rapidia Software (K012290). The Rapidia Software is a fast, practical, and accurate tool for 3D (three dimensional) and 2D (two dimensional) viewing and manipulation of CT and MRI images using the most advanced graphics-rendering technology.

    CalScoR can specify the location of calcium in images, calculate the calcium score based on Agaston's method display the percentile of the score with the graph, and issue reports.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the 3DMed Co., Ltd. CalScoR software, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Scores obtained from CalScoR and predicate device (AccuScore) are similar. Linear regression slope = 1, intercept = 0.Linear regression demonstrated the scores were similar, with a calculated slope of 1 and intercept of 0.
    Scores obtained from CalScoR and predicate device (AccuScore) are essentially identical.F-test resulted in an F-value of 13742853.39, showing the two scores are identical within 0.0001 (0.01%) error level.
    CalScoR performs all functions according to the functional requirements specified in the Software Requirements Specifications.Testing demonstrated that the CalScoR performs all functions according to the functional requirements.

    Study Information

    1. Sample size used for the test set and the data provenance: Not explicitly stated. The document mentions "Tests were performed to compare the accuracy," but does not specify the number of cases or the origin of the data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as the comparison was against a predicate device's output, not expert-established ground truth.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable, as the comparison was against a predicate device's output.
    4. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done: No, an MRMC study was not done. The study focused on comparing the algorithm's output to a predicate device's output.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Yes, the testing compared the CalScoR software's output directly to the predicate device's output, indicating a standalone algorithm comparison.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for this study was the output of a legally marketed predicate device, AccuScore/Accuview Software (K990241), which also calculates calcium scores based on Agaston's method.
    7. The sample size for the training set: Not applicable. The document does not describe a machine learning model that would require a distinct training set. The device's algorithm for calculating calcium score is stated to be based on Agaston's method.
    8. How the ground truth for the training set was established: Not applicable.

    Overall, the study focused on demonstrating substantial equivalence to a predicate device by comparing the outputs of the two software programs using statistical methods (linear regression and F-test), rather than directly validating against clinical outcomes or a human expert standard.