Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K041943

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-10-14

    (87 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K992914, K003705, K011382, K020847, K031451

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    UVB Phototherapy for psoriasis, vitiligo, atopic dermatitis, and leukoderma

    Device Description

    The XTRAC XL2 Excimer Laser System Model AL8000 is a complete self-contained compact UVB laser light source, which utilizes a XeCl gas mixture to generate an operator selected dose and target-specific ultraviolet light at monochromatic operater services ann. The laser system consists of a touch-screen display, an advanced fiberoptic cable attached to a handpiece, and a foot-switch to initiate exposure. The laser is enclosed in a protective interlocked housing. The unit is designed to operate on standard AC power available from wall outlets and can accommodate US, European and other nominal supply voltages and operating frequencies.

    AI/ML Overview

    The provided 510(k) summary for the PhotoMedex XTRAC XL2 Excimer Laser System, Model AL8000, does not contain the type of acceptance criteria and study information typically associated with AI/ML device evaluations. This document is for an excimer laser system for dermatological conditions and its clearance is based on substantial equivalence to previously cleared predicate devices, rather than new clinical performance data demonstrating device performance against specific acceptance criteria.

    Therefore, many of the requested sections below cannot be filled from the provided text, as the document focuses on regulatory clearance for a hardware device, not an AI/ML algorithm.

    Here's an attempt to answer the questions based only on the provided text, indicating where information is not available:


    Acceptance Criteria and Device Performance Study (XTRAC XL2 Excimer Laser System)

    This 510(k) submission for the XTRAC XL2 Excimer Laser System, Model AL8000, evaluates its substantial equivalence to previously cleared excimer laser systems for dermatological phototherapy. The submission does not present new clinical data against specific performance acceptance criteria for the device itself, but rather relies on the established safety and effectiveness of its predicates.

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The provided document is for a medical device (laser system), not an AI/ML diagnostic or prognostic tool. Therefore, standard AI/ML performance metrics (sensitivity, specificity, AUC, etc.) and associated acceptance criteria are not applicable or mentioned. The "acceptance criteria" here relate to meeting safety standards and demonstrating equivalence to predicate devices, which is largely qualitative in the provided summary.

    Acceptance Criterion (Implicit from 510(k) process)Reported Device Performance (based on substantial equivalence)
    Safety and Effectiveness Equivalence: No new issues of safety or efficiency compared to predicates.The device performs in an identical manner as the identified predicates, and differences (size, weight, ergonomics, enhanced operating parameters, maintenance improvements) do not affect safety or intended use.
    Intended Use Equivalence: Identical intended use.Intended use (targeted UVB phototherapy for psoriasis, vitiligo, atopic dermatitis, and leukoderma) is identical to predicate devices.
    **Technological Characteristics Equivalence (or
    differences do not raise new questions of safety/effectiveness):** Monochromatic (308nm) UVB light generation and therapeutic mechanism.The device is a "complete self-contained compact UVB laser light source, which utilizes a XeCl gas mixture to generate an operator selected dose and target-specific ultraviolet light at monochromatic wavelength (308nm)." This mechanism is identical to the predicates.
    Compliance with Standards: Meets relevant electrical safety, performance, and risk management standards.Complies with current applicable international EN 60601 series of standards, 21 CFR Part 1040.10 & 1040.11 Performance Standards for Light-Emitting Products, UL 2601-1/UL60601-1, and EN ISO 1497 (risk management).

    2. Sample size used for the test set and the data provenance

    No new clinical test set data was used or presented for this 510(k) submission, as the clearance is based on substantial equivalence. The document explicitly states: "PhotoMedex believes duplicative clinical data is not required as a condition of granting market clearance for the XTRAC XL2 Excimer Laser System Model AL8000."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. No new clinical test set was used.

    4. Adjudication method for the test set

    Not applicable. No new clinical test set was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an excimer laser system, not an AI/ML diagnostic tool, and no MRMC study was conducted or presented.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is an excimer laser system, not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable, as no new clinical data was collected or analyzed for ground truth in this submission. The "ground truth" for the device's efficacy is based on the previously cleared predicate devices and the established understanding of UVB phototherapy for the indicated conditions.

    8. The sample size for the training set

    Not applicable. This document is for a hardware device clearance via substantial equivalence, not an AI/ML training set.

    9. How the ground truth for the training set was established

    Not applicable. This document is for a hardware device clearance via substantial equivalence, not an AI/ML training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K021762

    Validate with FDA (Live)

    Device Name
    BCLEAR
    Manufacturer
    Date Cleared
    2002-08-23

    (86 days)

    Product Code
    Regulation Number
    878.4630
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K011382, K992914, K003705

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BClear™, Targeted PhotoClearing™ System, is a medical ultraviolet lamp and delivery device intended for the treatment of leukoderma. The BClear, Targeted PhotoClearing System, is also indicated for the treatment of psoriasis, vitiligo, atopic dermatitis (eczema), and seborrheic dermatitis. The BClear, Targeted PhotoClearing System, is intended for use on all skin types (I - VI).

    Device Description

    The BClear, Targeted PhotoClearing System, is an ultraviolet light source and energy delivery system that provides targeted energy to the treatment site while avoiding unnecessary exposure to non-affected tissue. The light source is contained within a protective console. The complete system also includes a handpiece connected to the console via an umbilical. Timing and dosing parameters and an interface to other system features are controlled from a display panel on the console. The delivery system allows UV-B light to pass through the handpiece to selectively treat skin and nail lesions without exposure to the healthy skin.

    AI/ML Overview

    The provided document is a 510(k) summary for the BClear, Targeted PhotoClearing System. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed performance study data against specific acceptance criteria. Therefore, much of the requested information regarding acceptance criteria and performance study details is explicitly not applicable or not provided in this specific document.

    Here's an analysis based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    CriterionAcceptance CriteriaReported Device Performance
    SafetyNot explicitly stated, but implied to be equivalent to predicate devices."The technological specifications of the BClear, Targeted PhotoClearing System, are the same or very similar to those of the claimed predicate devices." No adverse events or safety concerns are raised.
    EffectivenessNot explicitly stated, but implied to be equivalent to predicate devices for the specified indications."The BClear, Targeted PhotoClearing System, has the same indications for use for which the claimed predicates have been cleared." No specific performance metrics (e.g., clearance rates, time to response) are provided.
    Indications for Use"Intended for the treatment of leukoderma. ... also indicated for the treatment of psoriasis, vitiligo, atopic dermatitis (eczema), and seborrheic dermatitis. ... intended for use on all skin types (I - VI)."The document states in its "Substantial Equivalence" section: "The BClear, Targeted PhotoClearing System, has the same indications for use for which the claimed predicates have been cleared." This implies it meets the same intended use as the predicate devices.
    Technological CharacteristicsNot explicitly stated, but the expectation is for the BClear system to be similar or identical to the predicate devices."The technological specifications of the BClear, Targeted PhotoClearing System, are the same or very similar to those of the claimed predicate devices."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not applicable. The document explicitly states: "Performance Data: None. The technological specifications of the BClear, Targeted PhotoClearing System, are the same or very similar to those of the claimed predicate devices. The BClear, Targeted PhotoClearing System, has the same indications for use for which the claimed predicates have been cleared. Therefore, performance data is not required."
    • Data Provenance: Not applicable, as no new performance data was generated for this 510(k) submission. The decision is based on substantial equivalence to previously cleared predicate devices.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not applicable, as no new performance study with a test set was conducted.

    4. Adjudication Method for the Test Set

    • Not applicable, as no new performance study with a test set was conducted.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No. The document explicitly states "Performance Data: None."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done

    • No. This device is a physical medical ultraviolet lamp and delivery system, not an AI algorithm. Therefore, "standalone algorithm performance" is not relevant. The substantial equivalence argument is based on the device's physical and functional similarity to predicate devices.

    7. The Type of Ground Truth Used

    • Not applicable, as no new performance study was conducted. The "ground truth" for the device's safety and effectiveness relies on the prior clearance and established use of the predicate devices for their indicated conditions. The document mentions "Currently published clinical data supports UVB treatment of hypopigmented or depigmented skin (leukoderma)," which indirectly points to existing clinical evidence for the treatment modality, but not for this specific device's new performance evaluation.

    8. The Sample Size for the Training Set

    • Not applicable, as this device is not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable, as this device is not an AI/ML algorithm that requires a training set.

    Summary of the Study Proving Acceptance Criteria:

    The "study" in this case is not a clinical trial or performance study of the BClear device itself. Instead, the document describes a substantial equivalence claim based on a comparison to legally marketed predicate devices:

    • Predicate Devices: Lumenis Inc., BClear, Targeted PhotoClearing System (K011197, K020591); PhotoMedex, XTRACTM Excimer Laser Phototherapy System, AL7000 (K011382, K992914, K003705).
    • Argument: The applicant argues that the BClear, Targeted PhotoClearing System, is substantially equivalent because:
      • Its technological specifications are the same or very similar to those of the claimed predicate devices. (This implies that the underlying technology for delivering UV-B light to treat dermatological conditions is consistent with established and cleared devices).
      • It has the same indications for use as the cleared predicate devices (treatment of leukoderma, psoriasis, vitiligo, atopic dermatitis (eczema), and seborrheic dermatitis, for all skin types I-VI).
    • Basis for No Performance Data: Because of this perceived substantial equivalence in both technology and indications for use, the submission explicitly states that "performance data is not required." The assumption is that since the device is essentially the same as (or very similar to) a device already on the market and cleared for the same purposes, its performance will be comparable.

    Therefore, the "study" proving the device meets acceptance criteria is fundamentally a comparative analysis of device characteristics and intended use against already cleared predicate devices, rather than a de novo performance evaluation.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1