Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K190101

    Validate with FDA (Live)

    Device Name
    UltraFlex IAB
    Date Cleared
    2019-06-28

    (157 days)

    Product Code
    Regulation Number
    870.3535
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K010330 Arrow RediGuard® 9 Fr. 50cc Universal Intra-Aortic Balloon Catheter, K000729 Arrow Ultra 8® 8

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    UltraFlex™ IAB with Intra-Aortic Balloon Pump as a control system is indicated for use in any of the following conditions:

    1. Acute Coronary Syndrome

    2. Cardiac and Non-Cardiac Surgery

    3. Complications of Heart Failure

    Device Description

    The UltraFlex IAB consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning. A computerized control system, also known as Intra-Aortic Balloon Pump (IABP) regulates the inflation and deflation of the balloon.

    The UltraFlex IAB catheter consists of an inner lumen, an outer lumen, and an inflatable balloon. The outer lumen is comprised of an inflatable balloon connected to the distal tip of the catheter shaft and to the IAB catheter tip outer surface. The inner lumen is comprised of a luer adapter connected to the proximal end of the inner lumen and to the IAB catheter tip inner surface.

    AI/ML Overview

    The provided text is a 510(k) summary for the UltraFlex™ IAB, an intra-aortic balloon catheter. This document describes the device, its indications for use, and a comparison to a predicate device. It also mentions various bench tests conducted to demonstrate substantial equivalence, but it does not contain information about a study proving the device meets acceptance criteria using a test set with expert adjudicated ground truth, multi-reader multi-case studies, or information about training sets and their ground truth establishment.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria by extracting information from the given text.

    The closest relevant information extracted is:

    Acceptance Criteria Mentioned (but specific values are not provided):

    • The results of the verification tests met the specified acceptance criteria and performed similar to the predicate device.

    Study Type (Bench Tests, not clinical or image-based AI performance studies):
    The document states that technological differences have been evaluated through bench tests. The following tests were performed:

    • Catheter Insertion Test
    • Aneurysm Test
    • Durability Test
    • Catheter Tip to Balloon Bond Tensile per ISO 10555-1
    • Outer Lumen to Balloon Bond Tensile per ISO 10555-1
    • Sheath and Dilator Surface Visual Inspection per ISO 11070
    • Sheath and Dilator Tensile Testing per ISO 11070
    • Balloon Volume Test
    • Catheter Rate Limit Test
    • Kink Resistance
    • Sheath and Dilator Dimensional Analysis

    Data Provenance/Sample Size (Not Applicable in the provided text for a clinical test set):
    The document does not describe a clinical test set, nor does it specify sample sizes for the bench tests.

    Experts, Adjudication, MRMC, Standalone Performance, Ground Truth Type, Training Set Details (Not Applicable/Not Present in the provided text):
    The 510(k) summary focuses on demonstrating substantial equivalence based on bench testing of physical and functional characteristics of the device compared to a predicate device, not on diagnostic performance or AI algorithm validation using image data, which would typically involve expert readers, ground truth establishment, and MRMC studies. Therefore, the requested details about expert qualifications, adjudication methods, MRMC studies, standalone performance, training sets, and ground truth establishment methods for training sets are not present in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190117

    Validate with FDA (Live)

    Device Name
    Fiberoptix IAB
    Date Cleared
    2019-06-13

    (140 days)

    Product Code
    Regulation Number
    870.3535
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K010330

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FIBEROPTIX® IAB with the Intra-Aortic Balloon Pump as a control system is indicated for use in any of the following conditions:

    1. Acute Coronary Syndrome

    2. Cardiac and Non-Cardiac Surgery

    3. Complications of Heart Failure

    Device Description

    The FIBEROPTIX IAB consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning. A computerized control system, also known as the Intra-Aortic Balloon Pump (IABP) is utilized to requlate the inflation and deflation of the balloon.
    The FIBEROPTIX IAB consists of an inner lumen, an outer lumen, and an inflatable balloon. The outer lumen is comprised of an inflatable balloon connected to the distal tip of the catheter shaft and to the IAB catheter tip outer surface. The inner lumen is comprised of a luer adapter connected to the proximal end of the inner lumen and to the IAB catheter tip inner surface.

    The FIBEROPTIX IAB has a fiber optic pressure sensor which acts as a pressure transducer embedded in the catheter tip.

    AI/ML Overview

    The provided text describes the 510(k) submission for the FIBEROPTIX IAB, focusing on its substantial equivalence to a predicate device. The information details various tests performed to demonstrate this equivalence.

    Here's an analysis of the provided information concerning acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The provided text lists several design verification tests performed. While it explicitly states that the results "met the specified acceptance criteria and performed similar to the predicate device," it does not provide a specific table detailing the acceptance criteria values for each test or the exact performance metrics observed.

    Acceptance Criteria (Not Explicitly Stated - Inferred)Reported Device Performance
    Pass/Meet specified criteria for:Met specified acceptance criteria
    - Balloon Volume TestPerformed similar to the predicate device
    - Aneurysm TestMet specified acceptance criteria
    - Durability TestPerformed similar to the predicate device
    - Catheter Insertion TestMet specified acceptance criteria
    - Catheter Tip to Balloon Bond Tensile per ISO 10555-1Performed similar to the predicate device
    - Outer Lumen to Balloon Bond Tensile per ISO 10555-1Met specified acceptance criteria
    - Catheter Rate Limit TestPerformed similar to the predicate device
    - Sheath and Dilator Surface Visual Inspection per ISO 11070Met specified acceptance criteria
    - Sheath and Dilator Tensile Testing per ISO 11070Performed similar to the predicate device
    - Sheath and Dilator Dimensional AnalysisMet specified acceptance criteria

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The text mentions "bench tests" for design verification. However, it does not specify the sample size used for any of these tests. It also does not provide any information regarding data provenance (e.g., country of origin, retrospective or prospective nature), as these are bench tests and not human clinical studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This type of information is usually relevant for studies involving subjective human interpretation (e.g., image analysis). For the described bench tests, no experts were used to establish ground truth in this manner. The ground truth for these engineering tests would be established by objective measurements against predefined engineering specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Adjudication methods like 2+1 or 3+1 are typically used in clinical studies where multiple human readers interpret data, and discrepancies need to be resolved. Since the described tests are bench tests, no such adjudication method was applicable or employed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The provided information focuses on the substantial equivalence of an intra-aortic balloon catheter, a physical medical device, through a series of engineering bench tests. It does not describe an AI-powered device or an MRMC comparative effectiveness study involving human readers or AI assistance. Therefore, there is no mention of effect size for human readers with or without AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Similar to the point above, the FIBEROPTIX IAB is a physical medical device, not an algorithm. Therefore, a standalone algorithm-only performance study was not conducted or is not applicable in this context.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the bench tests described, the ground truth would be established by engineering specifications and objective measurements. For example:

    • Dimensional Analysis: Ground truth is the specified dimensions (e.g., diameter, length).
    • Tensile Testing: Ground truth is the specified material strength or bond strength.
    • Volume Test: Ground truth is the specified balloon inflation volume.

    There is no mention of expert consensus, pathology, or outcomes data being used to establish ground truth for these specific tests.

    8. The sample size for the training set

    The provided text only discusses pre-market submission via a 510(k) pathway, focusing on substantial equivalence to a predicate device through bench testing. It does not mention any training set size, as it does not describe the development or evaluation of a machine learning or AI algorithm.

    9. How the ground truth for the training set was established

    As there is no mention of a training set (because it's not an AI/ML device), this information is not applicable and therefore not provided in the text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1