Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K020716

    Validate with FDA (Live)

    Date Cleared
    2002-05-06

    (62 days)

    Product Code
    Regulation Number
    862.3650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K003809, K003809/A002

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rapid Opiates Test Card II is an imunochromatorgraphy based one step in vitro test. It is designed for qualitative determination of Opiates and its metabolites in human urine specimens. The presence of Morphine in human urine above a cutoff level of 2,000ng/ml can be detected.

    The test kit is used to obtain a visual, qualitative result and is intended for professional use. It is not intended for over the counter sale to lay persons.

    Device Description

    The Rapid Drug tests are based on the principle of specific immunochemical reaction between antibodies and antigens to analyze particular compounds in human urine specimen. The assay relies on the competition for antibody binding between drug conjugate and free drug which may be present in the urine specimen being tested.

    When drug is present in the urine specimen, it competes with drug conjugate for the limited amount of antibody-dye conjugate. When the amount of drug is equal or more than the cutoff level, it will prevent the binding of drug conjugate to the antibody. As a result, a positive urine specimen will not show a colored band on the test line zone, indicating a positive result, while the presence of a colored band indicates a negative result.

    AI/ML Overview
    1. A table of acceptance criteria and the reported device performance

      The provided text does not explicitly state formal acceptance criteria with specific numerical targets. However, the performance section indicates that the device's accuracy, precision, and other characteristics were evaluated, and the overall conclusion is that it is "substantially equivalent" to a legally marketed predicate device.

      Performance Characteristic EvaluatedReported Device Performance (Implied Acceptance)
      Sensitivity Accuracy"comparison study of clinical urine specimens" showed the device to be "substantially equivalent to the Applied Biotech SureStep ™Drug Screen Test Morphine II."
      Stability - SpecimenEvaluated (details not provided)
      Stability - ProductEvaluated (details not provided)
      PrecisionEvaluated (details not provided)
      ReproducibilityEvaluated (details not provided)
      SpecificityEvaluated (details not provided)
      InterferenceEvaluated (details not provided)
      Overall Effectiveness"demonstrate Rapid Opiates Test Strip II and Rapid Test Card II to be substantially equivalent to the Applied Biotech SureStep ™Drug Screen Test Morphine II"
    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

      The text states: "Both urine control specimen and clinical urine specimen were tested to evaluate the safety and effectiveness of Rapid Opiates Test Strip II and Rapid Opiates Test Card II."

      • Sample Size: Not specified.
      • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). It only mentions "clinical urine specimens."
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

      This information is not provided in the given text. The method for establishing ground truth is not detailed beyond the use of "urine control specimen and clinical urine specimen."

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

      This information is not provided in the given text.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      No, an MRMC comparative effectiveness study involving human readers or AI assistance was not mentioned. This device is an in-vitro diagnostic test, not an AI-assisted diagnostic tool for interpretation by human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

      Yes, the performance characteristics provided are for the device (Rapid Opiates Test Strip II and Rapid Test Card II) operating as a standalone diagnostic test, without human interpretation beyond reading the visual result. The "Indications for Use Statement" specifies it is "designed for qualitative determination... The test kit is used to obtain a visual, qualitative result and is intended for professional use."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      The text mentions the use of "urine control specimen and clinical urine specimen." For drug tests, the ground truth is typically established through a reference method (e.g., Gas Chromatography-Mass Spectrometry (GC/MS) or other highly accurate laboratory methods) that quantifies the presence and concentration of the drug in the urine. While not explicitly stated as GC/MS, the "comparison study of clinical urine specimens" and the determination of "sensitivity accuracy" imply a highly reliable reference method was used to establish the true presence or absence of opiates.

    8. The sample size for the training set

      This device is an immunochromatography-based test, not an AI or machine learning algorithm that requires a "training set" in the conventional sense. Therefore, the concept of a training set size does not apply. The development would involve analytical verification and validation using various concentrations of analytes and interferents, not a data-driven training set like for AI.

    9. How the ground truth for the training set was established

      As explained in point 8, this device does not utilize a training set in the AI/machine learning context. The "ground truth" for the development and analytical validation of such a test would be established through precisely prepared controls (known concentrations of opiates and other substances) and clinical samples verified by a gold-standard laboratory method.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1