Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K191106

    Validate with FDA (Live)

    Date Cleared
    2019-12-13

    (232 days)

    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K123043, K103604, K040183, K053174, K002734, K110645

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The C50 and C80 patient monitors are intended to be used for monitoring, displaying, reviewing, alarming and storing of multiple physiological parameters as following: ECG (3-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), temperature (Temp), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG), cardiac output (C.O.) for single patient. The C80 can also monitor the ICG (impedance cardiography) and BIS (bispectral index).

    All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:

    · The ICG monitoring is applicable to the adult patients of 122229cm in height and 30159Kg (67~341 pounds) in weight only;

    · NIBP measurement continual mode is not applicable to neonates;

    · Anesthetic depth (BIS) of the multi-parameter patient monitor is not intended for neonatal patients;

    The monitors are to be used in general healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. The monitors are not intended for home use.

    Device Description

    The C50 and C80 patient monitors are intended to be used for monitoring, displaying, reviewing, alarming and storing of multiple physiological parameters as following: ECG (3-lead, 5-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), temperature (Temp), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG), cardiac output (C.O.) for single patient. The C80 can also monitor the ICG (impedance cardiography) and BIS (bispectral index).

    All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
    • The ICG monitoring is applicable to the adult patients of 122229cm in height and 30159Kg (67~341 pounds) in weight only;
    • NIBP measurement continual mode is not applicable to neonates;
    • Anesthetic depth (BIS) of the multi-parameter patient monitor is not intended for neonatal patients;
    Those monitors provide patient monitoring capabilities by using corresponding accessories.
    The multi-parameter Patient monitor, model: C50 and C80 have same design principle and technical characteristics:
    But the software in the host and modules and components are different.
    The differences between C50 and C80 are ICG, BIS and size of monitor.

    AI/ML Overview

    The provided document describes the acceptance criteria and a study proving the C50 and C80 Multi-parameter Patient Monitor meets these criteria.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally established through adherence to recognized international standards and clinical performance metrics. The document details a comparison with a predicate device (K170876, Passport Series Patient Monitors). While explicit "acceptance criteria" for each parameter are not always separately listed with a single performance value from the new device, the "Comparison" column in the table below infers if the new device meets or exceeds the predicate's performance or if sufficient justification is provided for differences.

    FeaturePredicate Device (K170876) PerformanceSubject Device (C50 & C80) PerformanceComparison to Acceptance Criteria (Predicate) & Justification
    Monitor SizePassport 17m: 400x370x193mm; Passport 12m: 297x336x187mmC80: Approx. 344x291x165mm; C50: Approx. 291.7x250x187mmThe monitor size of the subject device is smaller than the predicate device, but it "won't affect the safety and effectiveness of the subject device." Implied acceptance: smaller size is acceptable if safety/effectiveness are not compromised.
    Integrated display & touch screenColor TFT LCD (17-inch, 1280x1024 pixels for 17m; 12-inch, 800x600 pixels for 12m)Color TFT LCD (C80: 12.1-inch, 800x600 pixels; C50: 10.4-inch, 800x600 pixels)Screen size is smaller, but resolution for C80/C50 is the same as Passport 12m. "Same. The screen of Subject device is smaller than predicate device. Complying with IEC 60601-1 and IEC 62366-1 also indicates the clinical use is safe and effective." Implied acceptance: smaller screen size is acceptable due to compliance with relevant IEC standards for safety and usability.
    Power supplyTwo rechargeable Lithium-ion batteries or AC for Passport 17m; One rechargeable Lithium-ion battery or AC for Passport 12mC50 and C80: Powered either by built-in battery or external AC."Same, both powered by battery and AC." Implied acceptance: functional equivalence.
    Battery11.1V, 4500mAh11.1V, 2200mAh/4400mAh"The battery capacity is different. The batteries of C50/C80 have complied with IEC 62133. C50 and C80 have conformed to IEC 60601-1." Implied acceptance: different capacity is acceptable as long as safety (IEC 62133, IEC 60601-1) is maintained.
    ECG3-lead, 5-lead and 12-lead selectable, heart rate (HR)3-lead, 5-lead and 12-lead selectable, heart rate (HR)"Same." Implied acceptance: functional equivalence.
    ECG (Arrhythmia Analysis)Asystole, VFib/VTac, Vtac, Vent. Brady, Extreme Tachy, Extreme Brady, PVC, Couplet, Bigeminy, Trigeminy, Ron T, Run PVCs, PVCs, Tachy, Brady, Missed Beats, Vent. Rhythm, PNP, PNC, Multif. PVC, Nonsus. Vtac, Pause, Irr. Rhythm, AFib (24 types)Asystole, ventricular fibrillation, R ON T, VT >2, Couplet, PVC, Bigeminy, Trigeminy, Brady (Bradycardia), PNC (Pacer Not Capture), PNP (Pacer Not Pace), Missed Beats, IHB (Irregular Heart Beat), VTAC (Ventricular Tachycardia), Tachy (Tachycardia), PVC Too High, Extreme Tachycardia, Extreme Bradycardia, Ventricular Rhythm, Heart Pause (20 types)"C50/C80 has 20 types of arrhythmias in total, 18 of which is same as the predicate device. The other two (underline) is supported by compliance with EC57." Implied acceptance: fewer types are acceptable as long as key arrhythmia detections are present and supported by EC57 compliance.
    RespirationMethod: Trans-thoracic impedance. Range: adult:0-120 rpm; pediatrics:0-150rpm; neonate:0-150rpm. Accuracy: 7 to 150rpm: ± 2rpm or ± 2%, whichever is greater. 0 to 6rpm: not specified.Method: Trans-thoracic impedance. Range: adult:0-120 rpm; pediatrics:0-150rpm; neonate:0-150rpm. Accuracy: 7 to 150rpm: ± 2rpm or ± 2%, whichever is greater. 0 to 6rpm: not specified."Same." Implied acceptance: functional and performance equivalence.
    Pulse oxygen saturation (SpO2)Method: red and infrared light method. Masimo SpO2: Range:1100%. Accuracy: No motion: 70-100% ±2%(adult/pediatric), ±3%(neonate); Motion: 70-100% ±3%. Nellcor SpO2: Range:0100%. Accuracy: 70-100% ±2%(adult/pediatric), ±3%(neonate). Mindray SpO2: Range:0~100%. Accuracy: 70-100% ±2%(adult/pediatric), ±3%(neonate).Method: same. Masimo SpO2: same range & accuracy. Nellcor SpO2: same range & accuracy. Comen SpO2: Range: 0%-100%. Accuracy: 70-100% ±2%(adult/pediatric, non-motion), ±3%(neonate, non-motion)."Same." (for Masimo and Nellcor modules). For their own Comen SpO2 module, it shows equivalent accuracy. Implied acceptance: performance equivalence with predicate's different OEM modules and new Comen module meets similar accuracy.
    Pulse rate (PR)From Mindray SpO2 (Range 20-254 bpm, Acc ±3 bpm), Masimo SpO2 (Range 25-240 bpm, Acc ±3 bpm no motion, ±5 bpm motion), Nellcor SpO2 (Range 20-300 bpm, Acc ±3 bpm for 20-250 bpm), IBP sensor (Range 25-350 bpm, Acc ±1 bpm or ±1%).From Masimo SpO2 (Range 25-240 bpm, Acc ±3 bpm no motion, ±5 bpm motion), Nellcor SpO2 (Range 20-300 bpm, Acc ±3 bpm for 20-250 bpm), Comen SpO2 (Range 20-254 bpm, Acc ±2 bpm), IBP sensor (Range 25-350 bpm, Acc ±1 bpm or ±1%), NIBP sensor (Range 40-240 bpm, Acc ±3 bpm or ±3%)."The PR from Comen SpO2 is more accuracy than the PR form Mindray SpO2. The PR form Masimo SpO2, Nellcor SpO2 and IBP sensor of C50 and C80 are the same with the predicate device. In addition, C50 and C80 have one more PR data source (NIBP) than predicate device. C50 and C80 have complied with ISO 80601-2-61." Acceptance criteria met or exceeded; added NIBP source (improvement). Adherence to ISO 80601-2-61.
    Non-invasive blood pressure (NIBP)Method: Oscillometry. Range: Adult (systolic 25-290mmHg, diastolic 10-250mmHg), pediatric (systolic 25-240mmHg, diastolic 10-200mmHg), neonate (systolic 25-140mmHg, diastolic 10-115mmHg). Accuracy: Max mean error: ±5 mmHg; Max standard deviation: 8 mmHg.Method: same. Range: Adult (systolic 40-270mmHg, diastolic 10-215mmHg), pediatric (systolic 40-200mmHg, diastolic 10-150mmHg), neonate (systolic 40-135mmHg, diastolic 10-100mmHg). Accuracy: 0-300mmHg: ±3mmHg."The NIBP measure range for adult/pediatric/neonate in C50 and C80 is smaller than that in predicate device. The C50 and C80 is more accuracy than predicate device. C50 and C80 have conformed to IEC 80601-2-30." Acceptance criteria met or exceeded in accuracy, and justification for smaller range implicitly accepted by meeting IEC 80601-2-30.
    Temperature (Temp)Method: Thermal resistance. Range: 0 ~ 50°C. Accuracy: ±0.1°C.Method: Thermal resistance. Range: 0 ~ 50°C. Accuracy: ±0.2°C."The predicate device is more accuracy than C50 and C80. C50 and C80 have complied with ISO 80601-2-56." Implied acceptance: slightly lower accuracy is acceptable as long as it complies with ISO 80601-2-56.
    Carbon dioxide (CO2)Method: Infrared absorption. Masimo CO2: Sidestream 0-99mmHg, AwRR 0-120rpm. Microstream 0-99mmHg, AwRR 0-150rpm. Accuracy: Sidestream 0-40mmHg ±2mmHg, 41-76mmHg ±5%, 77-99mmHg ±10%, AwRR ±2rpm. Microstream 0-38mmHg ±2mmHg, 39-99mmHg ±5% +0.08%, AwRR 0-70rpm ±1rpm, 71-120rpm ±2rpm, 121-150rpm ±3rpm.Method: same. Masimo CO2: Sidestream 0-190mmHg, 0-25% (760mmHg), AwRR 0-150rpm. Mainstream 0-190mmHg, 0-25% (760mmHg), AwRR 0-150rpm. Respironics CO2: Sidestream 0-150mmHg, 0-19.7%, AwRR 0, 2-150rpm. Mainstream 0-150mmHg, 0-19.7%, AwRR 0, 2-150rpm. Accuracy (Masimo): ±(2.25mmHg +reading x 4%), AwRR ±1rpm. Accuracy (Respironics): 0-40mmHg ±2mmHg, etc., AwRR ±1rpm."The C50 and C80 are more accuracy and measured wider than predicate device. For C50 and C80, Masimo CO2 module (K123043 and K103604) and Respironics CO2 module (K040183 and K053174) have been cleared. Both modules have complied with ISO 80601-2-55." Acceptance criteria met or exceeded; wider range and higher accuracy; existing modules previously cleared and comply with ISO 80601-2-55.
    Invasive blood pressure (IBP)Method: Direct invasive measurement. Range: -50 to 300 mmHg. Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor).Method: Direct invasive measurement. Range: -50 to 300 mmHg. Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor)."Same." Implied acceptance: functional and performance equivalence.
    Cardiac output (C.O.)Method: Thermodilution method. Range: 0.1 to 20 L/min. Accuracy: ±5% or ±0.1 L/min, whichever is greater.Method: Thermodilution method. Range: 0.1 to 20 L/min. Accuracy: ±5% or ±0.1 L/min, whichever is greater."Same." Implied acceptance: functional and performance equivalence.
    Anesthetic gas (AG)Method: Infrared absorption. Range (CO2, N2O, Hal, Enf, Iso, Sev, O2, AwRR). Accuracy (CO2 ±0.3% ABS, N2O ±(8%REL+2%ABS), Other AG 8%REL).Method: same. Range (CO2, N2O, Hal, Enf, Iso, Sev, O2, AwRR). Accuracy (CO2 0-15%: ±(0.2kPa+readingx2%), N2O ±(2 kPa+readingx2%), Hal/Enf/Iso 0-8%: ±(0.15%+readingx5%), Sev 0-10%: ±(0.15%+readingx5%), Des 0-22%: ±(0.15%+readingx5%), O2 ±(1%+readingx2%), Masimo AG AwRR ±1rpm)."For CO2/Enf/Hal/Iso/Sev/Des, The measurement range of C50 and C80 is smaller than the predicate device. For awRR, the measurement range of C80 is better than the predicate device. The C50 and C80 are more accuracy than predicate device. The measurement range of C50 and C80 is enough for most environments; it won't affect the safety and effectiveness. The C50 and C80 supports two AG modules: MASIMO ISA AX+ Sidestream module and Masimo IRMA AX+ Mainstream module, both of them have been cleared in K103604." Acceptance criteria: different ranges are acceptable with justification of sufficient range for most environments and high accuracy. Utilizes previously cleared and compliant modules.
    BIS (C80 only)Range: BIS, BIS L, BIS R: 0-100; SQI, SQI L, SQI R: 0-100%; EMG, EMG L, EMG R: 0-100 Db; SR, SR L, SR R: 0-100%; SEF, SEF L, SEF R: 0.5-30.0 Hz; TP, TP L, TP R: 40-100 Db; BC, BC L, BC R: 0-30; sBIS L, sBIS R: 0-10.0; Semg L, Semg R: 0-10.0; ASYM: 0-100%. Accuracy: 1% for BIS, SQI, EMG, ESR (implied).Range and Accuracy: BIS: same; accuracy: 1%. SQI: same; accuracy: 1%. EMG: same; accuracy: 1%. ESR: 0-100%; accuracy: 1%."The underline parameters are what C50 and C80 doesn't have. The BIS module and sensor have been cleared by FDA and its 510k numbers are K040183 and K002734." Acceptance criteria for essential parameters are met by using previously cleared FDA modules.
    ICG (C80 only)Method: Indirect impedance cardiograph measurement. Range: SV: 5-250 ml; HR: 44-2m; C.O. 1.4-15 L/min. Accuracy: SV: Not specified; HR: ±2 bpm; C.O. Not specified.Method: Indirect impedance cardiograph measurement. Range: HR: 40-250bpm; SV: 0-250mL; C.O.:0-30L/min; TFC: 5-150 /KΩ; SVR:0-3500 dyn・s ・ cm-5. Accuracy: SV: Not specified; HR: ±2 bpm; C.O. Not specified."The underline parameter is what C50 and C80 doesn't have. The ICG electrode cable and sensors are all cleared by FDA, 510(k) number of which is K110645." Acceptance criteria for essential parameters are met by using previously cleared FDA components.

    Study that Proves the Device Meets Acceptance Criteria

    The essential studies to prove the device meets acceptance criteria are divided into Non-Clinical Performance Data and Clinical Studies.

    Non-Clinical Performance Data:

    • Software Verification and Validation Testing: Conducted in accordance with FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Risk analysis was performed, and the software was deemed a "major" level of concern.
    • Electrical safety and Electromagnetic Compatibility (EMC): Testing conducted according to:
      • ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012 (IEC 60601-1:2005, MOD)
      • IEC 60601-1-2 Edition 4: 2014-02
    • Bench Testing: Functional and system-level tests were performed, showing the devices meet specifications and perform equivalently to the predicate.
    • Biocompatibility Testing: Performed for patient-contacting components (ECG cable, SpO2 probes, temperature probes) that were not previously cleared. Cytotoxicity, sensitization, and irritation testing were conducted based on ISO 10993-1 and FDA's 2016 biocompatibility guidance.
    • Conformance with Applicable Standards: The device demonstrated compliance with numerous recognized consensus standards, in addition to those listed above for specific modules (e.g., ISO 80601-2-55 for Respiratory Gas Monitors, ISO 80601-2-56 for Thermometers, ISO 80601-2-61 for Pulse Oximeter Equipment, ANSI AAMI EC57:2012 for Cardiac Rhythm and ST-Segment measurement algorithms).

    Clinical Studies:

    1. NIBP Clinical Study - for adults and children

    • Device Parameter: Non-invasive Blood Pressure (NIBP)
    • Acceptance Criteria/Standard: Complies with ISO 81060-2 Second Edition 2013-05-01 "Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type."
    • Sample Size: 25 patients (19 adults, 6 children).
      • Demographics: 10 men, 15 women. 6 patients aged 3-12, 2 aged 12-40, 10 aged 40-60, 7 above 60.
    • Data Provenance: Retrospective or Prospective not explicitly stated, but implies prospective data collection from "The Second Affiliated Hospital of Guangzhou Medical University and The First Affiliated Hospital of Guangzhou Traditional Chinese Medical University" from 2012 to 2013 (China).
    • Ground Truth: Implicitly referent method (e.g., auscultation by trained observers) as required by ISO 81060-2.
    • Adjudication Method: Not specified, but standard practice for NIBP clinical validation involves multiple observers for reference measurements.
    • Results: "The results are accurate and reliable, and the repeated measurement consistency is in good condition, within the measurement range. Moreover, No adverse events and side effects were found in clinical trials."

    2. NIBP Clinical Study - for neonate and infants

    • Device Parameter: Non-invasive Blood Pressure (NIBP)
    • Acceptance Criteria/Standard: Complies with ISO 81060-2 Second Edition 2013-05-01 "Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type."
    • Sample Size: 20 subjects.
      • Demographics: 10 males, 10 females. 13 subjects <29 days, 4 subjects 29 days to 1 year, 3 subjects 1 to 3 years old.
      • Data Points: 10 groups of data collected for each subject, totaling 200 groups.
    • Data Provenance: Retrospective or Prospective not explicitly stated, but implies prospective data collection from "The First Affiliated Hospital of Guangxi Medical University and The First Affiliated Hospital of Xinxiang Medical University" from 2012 to 2013 (China).
    • Ground Truth: Implicitly referent method (e.g., auscultation by trained observers) as required by ISO 81060-2.
    • Adjudication Method: Not specified.
    • Results: "The results are accurate and reliable, and the repeated measurement consistency is in good condition, within the measurement range. Moreover, No adverse events and side effects were found in clinical trials."

    3. SpO2 Clinical Study

    • Device Parameter: Oxygen saturation (SpO2)
    • Acceptance Criteria/Standard: Complies with FDA Certification Guidance Document of Photoelectric Oximeter and ISO 80601-2-61 "Medical Electrical Equipment - Part 2-61: Particular Requirements For Basic Safety And Essential Performance Of Pulse Oximeter Equipment."
    • Sample Size: 30 subjects (24 adult subjects, 6 neonates).
      • Demographics (Adults): 24 subjects (7 males, 17 females) aged 24-44 years old (20 yellows, 4 blacks).
      • Demographics (Neonates): 6 neonates (5 males, 1 female) aged 1-24 days.
    • Data Provenance: Implies prospective data collection from "the Affiliated Hospital of Guilin Medical University" from January 20, 2016 to June 10, 2016 (China).
    • Ground Truth: Implicitly a controlled desaturation study where arterial blood gas samples are taken and analyzed for SaO2, which is compared to the pulse oximeter readings.
    • Adjudication Method: Not specified.
    • Results: "Per the clinical test analysis result, there is no adverse events, the oxygen saturation module and probe are safe and effective for the intended use, as well as demonstrating accurate pulse measurement."

    Additional Requested Information:

    • 2. Sample size for the test set and the data provenance:

      • NIBP (Adult/Children): 25 patients from China (2012-2013).
      • NIBP (Neonate/Infants): 20 subjects (200 data groups) from China (2012-2013).
      • SpO2: 30 subjects (24 adults, 6 neonates) from China (Jan-June 2016).
      • All clinical data appears to be prospective as it describes clinical trials "conducted in" specific hospitals during specific periods.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not specify the number or qualifications of experts (e.g., radiologists) for establishing ground truth in clinical studies. For NIBP validation against ISO 81060-2, ground truth typically involves at least two trained observers performing auscultatory measurements. For SpO2, ground truth is established by laboratory CO-oximetry measurements of arterial blood, usually managed by laboratory technicians in a controlled setting, not interpretation by "experts" in the sense of image analysis.
    • 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • The document does not specify the adjudication method for any of the clinical or non-clinical tests. For NIBP, ISO 81060-2 dictates specific statistical analysis methods (e.g., Bland-Altman) for comparing device readings to reference measurements, often requiring two or more observers.
    • 5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This device is a multi-parameter patient monitor measuring physiological parameters (ECG, SpO2, NIBP, Temp, etc.). It does not describe an AI-powered diagnostic imaging device or an AI-assistance feature for human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • The device itself is a standalone patient monitor that provides automated measurements and alarms. The performance of its modules (e.g., NIBP, SpO2 algorithms) was evaluated independently against reference standards (clinical validation) and through bench testing. So, yes, standalone performance of the underlying algorithms and hardware was demonstrated through the clinical and non-clinical tests.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • NIBP: Likely comparative measurements against a reference standard method (e.g., auscultatory measurements by trained observers) as specified by ISO 81060-2.
      • SpO2: Involves laboratory arterial blood gas analysis (CO-oximetry) during controlled desaturation studies, comparing the device's SpO2 readings to the directly measured SaO2 values.
      • Other Parameters (ECG, Temp, IBP, CO2, AG, BIS, ICG): Performance is primarily verified through bench testing against calibrated simulators and reference equipment and adherence to respective ISO/IEC standards.
    • 8. The sample size for the training set:

      • The document describes pre-market notification (510k) for a patient monitor, not an AI/machine learning device that typically involves a training set. The clinical studies performed are for validation of the device's performance against human subjects, not for training internal algorithms. Therefore, there is no explicit "training set" mentioned in the context of machine learning.
    • 9. How the ground truth for the training set was established:

      • As there is no explicit mention of an AI/machine learning training set, this question is not applicable. The ground truth for validation was established by reference standard methods in clinical trials and bench testing as described in point 7.
    Ask a Question

    Ask a specific question about this device

    K Number
    K052145

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-08-23

    (15 days)

    Product Code
    Regulation Number
    882.1400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K011534, K011534, K994330, K002734, K001980

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Datex-Ohmeda S/5™ BIS Module, E-BIS and accessories are intended to be used with Datex-Ohmeda modular multiparameter monitors for monitoring neurophysiological status of hospitalized patients.

    The Datex-Ohmeda S/5TM BIS Module, E-BIS and accessories are indicated for monitoring the state of the brain by data acquisition of electroencephalograph (EEG) signals of all hospitalized patients. The Bispectral index (BIS), a processed EEG variable, may be used as an aid in monitoring the effects of certain anesthetic agents. The device is indicated for use by qualified medical personnel only.

    Device Description

    The Datex-Ohmeda S/5™ BIS Module, E-BIS and accessories is a single-width plug-in parameter module for a Datex-Ohmeda modular monitoring system. The S/5 E-BIS module, M-BIS can be used with the following Datex-Ohmeda modular monitors: S/5 Anesthesia Monitor (AM), S/5 Compact Anesthesia Monitor (CAM), S/5 Critical Care Monitor (CCM) and S/5 Compact Critical Care Monitor (CCCM) with main software L-(C)ANE02(A)..00 or L-(C)ICU02(A)..00 or newer version. The Datex-Ohmeda S/5™ BIS Module, E-BIS and accessories is a parameter module for monitoring the state of the brain by data acquisition of EEG signals of all hospitalized patients. The BIS, a processed EEG variable, may be used as an aid in monitoring the effects of certain anesthetic agents. The Datex-Ohmeda S/5TM BIS Module, E-BIS and its predicate M-BIS (K013389) are used in conjunction with Aspect Medical Systems, Digital Signal Converter-Expanded Performance, DSC-XP (K011534), Patient Interface Cable; PIC (K011534), and BIS Sensor Plus (K994330), BIS Sensor XP (K002734) or BIS Sensor Pediatric (K001980). The raw EEG signal can be displayed from one of the two monitored channels. The waveform size, color and sweep speed can be adjusted.

    Calculated parameters are:

    • Bispectral Index,BIS(Range=0-100),continuous processed EEG parameter correlating to the patient's level of hypnosis, where 100=awake and 0=comatose.
    • Suppression Ratio,SR,(Range=0-100%), the percentage of epochs in the past 63 seconds in which the EEG signal is considered suppressed.
    • . Electromyograph, EMG, the absolute power (in decibels) in the frequency range 70-110 Hz
    • Signal Quality Index, SQI (Range:0-100%), the percentage of good epochs in the last 60sec. that could be used to calculate the Bispectral Index and spectral variables.

    All the calculated parameters can be selected on the display, and trended.

    Alarms for E-BIS are taken care of by the host monitor and follow the user interface for alarms in Datex-Ohmeda $/5 patient monitors. There are auditory and visual alarms and user adjustable limits for E-BIS. The default is OFF, because it doesn't provide information to be used for treatment or therapy. The BIS Engine at BIS module dictates error messages displayed at Datex-Ohmeda's host monitor's message fields and service page. These error messages are related to the BIS measurement, but follow Datex-Ohmeda user interface rules,

    AI/ML Overview

    The provided text describes a Premarket Notification 510(k) Summary for the Datex-Ohmeda S/5™ BIS Module, E-BIS and accessories. However, it does not contain specific acceptance criteria or performance study details (e.g., accuracy, sensitivity, specificity, or results from clinical trials) in the format requested. The document primarily focuses on establishing substantial equivalence to a predicate device.

    Here's an analysis based on the provided text:

    • No detailed acceptance criteria or numerical performance metrics are explicitly stated in the document for the new device itself. The filing aims to demonstrate substantial equivalence to a predicate device (Datex-Ohmeda M-BIS Module (K013389)) based on similar technology, intended use, indications for use, and safety/effectiveness profiles.

    • The document implies that the device meets existing standards by referencing various IEC, EN, CAN/CSA, UL, and AAMI standards, along with FDA guidance documents. These standards likely contain performance requirements, but the specific acceptance criteria and the device's reported performance against them are not enumerated in this summary.

    Therefore, I cannot populate the table or answer all of your specific questions as the information is not present in the provided text.

    Here's what can be extracted and inferred:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Inferred from Predicate Equivalence)Reported Device Performance
    Safety and Effectiveness: Equivalent to predicate Datex-Ohmeda M-BIS Module (K013389)"have the same safety and effectiveness" as predicate M-BIS (K013389)
    Functionality: Calculation of Bispectral Index (BIS), Suppression Ratio (SR), Electromyograph (EMG), Signal Quality Index (SQI) within defined ranges.The device calculates and displays these parameters with the same operating principle as the predicate. Ranges are described (e.g., BIS 0-100, SR 0-100%, SQI 0-100%).
    Standards Compliance: Adherence to various international and national standards for medical electrical equipment, including safety, EMC, and software.The device "has been thoroughly tested through validation and verification of specifications" against a long list of standards (e.g., IEC 60601-1, AAMI ES1-1993, IEC 60601-2-26:2002, ISO 14971, etc.).
    Intended Use/Indications: Same as the predicate device."identical intended use and indications for use"
    Technological Characteristics: Identical fundamental scientific technology, electronic measurement board, BIS Engine, operating principle, accessories, user interface, and parameter specifications."identical fundamental scientific technology," "identical electronic measurement board," "same BIS Engine," "use the same operating principle," "identical accessories," "have the same user interface...and alarms," "the Customer and parameter specifications are the same."

    2. Sample size used for the test set and the data provenance
    The document does not mention a specific test set, its sample size, or data provenance (e.g., country of origin, retrospective/prospective status) for a performance study. Nonclinical testing is mentioned in terms of validation and verification against standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    This information is not provided. No specific ground truth establishment for a test set is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    This information is not provided.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    This type of study is not mentioned. The device is an EEG measurement module with a BIS index, and the context doesn't suggest an AI-driven interpretive system that would typically involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    The document doesn't explicitly describe a standalone performance study with specific metrics. It states the device calculates parameters like BIS, SR, EMG, and SQI. The performance is implied to be equivalent to the predicate, which also operates in a "standalone" or "algorithm only" fashion in terms of producing these processed EEG variables.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    The concept of "ground truth" in the context of this device (an EEG measurement module with a BIS index) would typically relate to the accuracy of its physiological measurements against a reference standard or validated algorithms. However, this level of detail regarding specific ground truth for performance evaluation is not present in the summary. The "ground truth" used for training (if applicable) is also not specified.

    8. The sample size for the training set
    The document does not mention any training set or its sample size. This device appears to be based on established signal processing algorithms for EEG, rather than a machine learning model that would typically require a "training set" in the modern sense. The "BIS Engine" is mentioned as being the same as the predicate, suggesting a stable and established algorithm.

    9. How the ground truth for the training set was established
    This information is not provided, as no training set is mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013389

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-01-10

    (90 days)

    Product Code
    Regulation Number
    882.1400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K994330, K002734, K001980

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Datex-Ohmeda BIS module, M-BIS and accessories are indicated for monitoring the state of the brain by data acquisition of electroencephalograph (EEG) signals of all hospitalized patients. The Bispectral index (BIS), a processed EEG variable, may be used as an aid in monitoring the effects of certain anesthetic agents. The device is indicated for use by qualified medical personnel only.

    Device Description

    The Datex-Ohmeda BIS module, M-BIS is a single-width plug-in parameter module for a Datex-Ohmeda modular monitoring system. The Datex-Ohmeda BIS module, M-BIS can be used with the following Datex-Ohmeda modular monitors: S/5 Anesthesia Monitor (AM), S/5 Compact Anesthesia Monitor (CAM), S/5 Critical Care Monitor (CCM) and S/5 Compact Critical Care Monitor (CCCM) with main software L-ANE02(A)... 00 or L-ICU02(A)... 00 or newer version. The Datex-Ohmeda BIS module M-BIS, and accessories is a parameter module for monitoring the state of the brain by data acquisition of EEG signals of all hospitalized patients. The BIS, a processed EEG variable, may be used as an aid in monitoring the effects of certain anesthetic agents. The accessories: The Datex-Ohmeda BIS module, M-BIS is used in conjunction with Aspect Medical Systems, Digital Signal Converter-Expanded Performance, DSC-XP (KOI 1534), Patient Interface Cable; PIC (KOI 1534), and BIS Sensor Plus (K994330), BIS Sensor XP (K002734) or BIS Sensor Pediatric (K001980). The raw EEG signal can be displayed from one of the two monitored channels. The waveform size, color and sweep speed can be adjusted. Calculated parameters are: Bispectral Index,BIS(Range=0-100),continuous processed EEG parameter correlating to the patient's level of hypnosis, where 100=awake and 0=comatose. Suppression Ratio.SR.(Range=0-100%), the percentage of evochs in the past 63 seconds in which the EEG signal is considered suppressed. Electromyograph, EMG, the absolute power (in decibels) in the frequency range 70-110 Hz Signal Quality Index, SQI (Range:0-100%), the percentage of good epochs in the last 60sec. that could be used to calculate the Bispectral Index and spectral variables. All the calculated parameters can be selected on the display, and trended. Alarms for BIS are taken care of by the host monitor and follow the user interface for alarms in Datex-Ohmeda S/5 patient monitors. There are auditory and visual alarms and user adjustable limits for BIS. The default is OFF, because it doesn't provide information to be used for treatment or therapy. The BIS Engine at BIS module dictates error messages displayed at Datex-Ohmeda's host monitor's message fields and service page. These error messages are related to the BIS measurement, but follow Datex-Ohmeda user interface rules.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Datex-Ohmeda S/5 BIS Module, M-BIS, and accessories. This submission focuses on establishing substantial equivalence to a predicate device (Aspect Medical Systems Inc. BIS Engine K011534) rather than presenting a de novo study with specific acceptance criteria and performance data for a new, unique device.

    Therefore, the document does not contain a table of acceptance criteria or a study proving the device meets those criteria in the way a clinical trial for a novel device would.

    Instead, the submission demonstrates equivalence by stating:

    • Identical fundamental technology: The BIS engine technology (hardware and software) is identical to the predicate.
    • Same basic functions, operating principles, and signal processing: These are described as being the same as the predicate.
    • Identical parameters: Raw EEG, processed EEG variables (BIS Index, Suppression Ratio, Electromyograph, Signal Quality Index), and electrode impedance information are supplied by the same BIS Engine as the predicate.
    • Identical user features: Multiple BIS smoothing rates, update rates, and various filters are available and identical to the predicate.
    • Identical accessories: Uses the same DSC-XP, PIC Plus, and BIS Sensors as the predicate.
    • Compliance with safety standards: The device complies with numerous electrical safety, electromagnetic compatibility, mechanical, environmental, and software validation standards, as listed.

    Based on the provided text, the following information is NOT available and would typically be found in a study proving acceptance criteria for a novel device:

    1. A table of acceptance criteria and the reported device performance: This document doesn't define specific performance metrics with acceptance thresholds (e.g., accuracy, sensitivity, specificity, or correlation coefficients with specific targets) for the M-BIS module, nor does it report the results of a study against such criteria. Its entire premise is that the core technology is identical to a previously cleared device.
    2. Sample size used for the test set and the data provenance: Not applicable as no new clinical performance study is described for the M-BIS module's core function. The core technology is considered identical to the predicate.
    3. Number of experts used to establish the ground truth for the test set and qualifications of those experts: Not applicable.
    4. Adjudication method for the test set: Not applicable.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a measurement module for physiological signals, not an AI diagnostic tool involving human readers.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The document states the "BIS engine technology (Hardware and Software) are identical compared to the predicate," implying that the core algorithm performance is already established through the predicate. It doesn't describe new standalone testing for this specific submission's purpose.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable to this type of submission. The ground truth for the predicate device would have been established previously, likely through correlation with clinical observations of anesthetic depth.
    8. The sample size for the training set: Not applicable, as no new algorithm development or training is described. The algorithm is stated to be identical to the predicate.
    9. How the ground truth for the training set was established: Not applicable.

    In summary:

    This 510(k) submission focuses on demonstrating substantial equivalence to a predicate device by asserting that the underlying technology, functions, accessories, and core algorithms are identical. It relies on the pre-existing clearance and validation of the predicate device (Aspect Medical Systems Inc. BIS Engine K011534) for its safety and effectiveness claims. Therefore, the document does not detail a new study with explicit acceptance criteria and corresponding performance data for the M-BIS module itself, as would be expected for an entirely novel device or a de novo submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1