510(k) Data Aggregation

    K Number
    K242572

    Validate with FDA (Live)

    Date Cleared
    2024-12-20

    (113 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K002452

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TopCEM-Try in Veneer Try-in Gel is used to evaluate the shade and the matching of the restoration prior to permanent cementation.

    Device Description

    TopCEM-Try in Veneer Try-in Gel is primarily made of polyethylene glycol, inorganic fillers and pigments. It is used to indicate the color of the Light Cure Veneer Cement before use, so as to select the right color to achieve the best aesthetic effect. The polyethylene glycol and inorganic filler in the Veneer Try-in Gel component are hydrophilic and can be rinsed off with a water gun nozzle after the color test operation. The principal organic components of TopCEM-Try in Veneer Try-in Gel are mixtures of polyethylene glycol, and the inorganic fillers loading is about 50% by volume.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a dental product, TopCEM-Try in Veneer Try-in Gel. It concerns the substantial equivalence determination for this device, which means proving it is as safe and effective as a legally marketed predicate device.

    The document does not describe an AI medical device or a study involving AI, machine learning, or human readers. Therefore, many of the requested details about acceptance criteria, test sets, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, and standalone performance are not applicable to this document.

    However, I can extract the information that is provided regarding the device's performance based on the general format and content of the filing:


    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Stated as "Pass/fail criteria" or "internal standards")Reported Device Performance (Conclusion)
    Meet internal standards for AppearanceSatisfactory
    Meet internal standards for ShadeSatisfactory
    Meet internal standards for LiquiditySatisfactory
    Meet internal standards for ConsistenceSatisfactory
    Conform to HuGe internal inspection standard for Physical PropertiesSubstantially equivalent (to predicate)
    Biological evaluation according to ISO 10993 standards and FDA guidancePassed Cytotoxicity, Sensitization, and Irritation tests

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document refers to "Bench testing" performed on the subject device and the primary predicate device. It also states that biological evaluation (biocompatibility tests) was performed.

    • Sample size: Not specified.
    • Data provenance: Not specified, but the tests were performed by Rizhao HuGe Biomaterials Company, Ltd. in China. The nature of these tests (bench testing, biocompatibility) implies they are prospective tests conducted on manufactured samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. The "ground truth" for the physical and chemical properties as well as biocompatibility tests would be the established scientific and regulatory standards (e.g., ISO 10993 for biocompatibility, and the company's internal standards for physical properties). These do not typically involve human expert consensus in the same way an AI diagnostic device's ground truth would.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing a consensus ground truth in studies involving human interpretation or subjective assessments, which is not the case for the type of bench testing and biocompatibility reported here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI medical device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For physical and chemical properties: "HuGe internal inspection standard" for parameters like appearance, shade, liquidity, and consistence.
    • For biocompatibility: International Standard ISO 10993-1 and FDA guidance for biological evaluation of medical devices.

    8. The sample size for the training set

    Not applicable. There is no training set mentioned as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set mentioned as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153231

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-07-06

    (240 days)

    Product Code
    Regulation Number
    872.3275
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K001823, K002452, K972594

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    G-Cem LinkForce

      1. Cementation of all types of all ceramic, resin and metal-based inlays, onlays, crowns and bridges.
      1. Cementation of metal, ceramic, fiber posts, and cast post and cores.
      1. Cementation of all ceramic and composite veneers (up to 2 teeth)
      1. Permanent cementation of crowns and bridges on implant abutments.

    GC DUAL CURE ACTIVATOR

    1. Cementation of indirect restorations.
    2. Bonding of core build-ups made of light-curing composite or core build-up materials to tooth structure.
    3. Bonding of resin cements and core build-up materials and self-cure composites to tooth structure.

    TRY-IN PASTE
    To evaluate the shade and the fitting of the restoration prior to permanent cementation.

    MULTI PRIMER

      1. Priming of the adherend surfaces of indirect restorations made of glass ceramics, oxide ceramics, hybrid ceramics, composite and metal before luting procedures.
      1. Intra-oral and extra-oral repair of fractured indirect restorations made of ceramics, hybrid ceramics and composites.
    Device Description

    G-CEM LinkForce is a dual-cured resin cement. The components consists of Paste A and B, which are filled in a one-body syringe. Both pastes are automixed with a mixing tip. G-CEM LinkForce is available in 4 shades: Translucent, A2, Opaque, Bleach.

    GC DUAL CURE ACTIVATOR is a primer to a tooth abutment. It must be used with G-Premio BOND (510(k) No. K143140).

    TRY-IN PASTE is a hue-conformity verification material. The product is placed into a dental restoration and set on an abutment tooth to verify hue and conformity. The product has a same color as G-CEM LinkForce.

    MULTI PRIMER is a priming agent designed to treat the adherend surface of glass ceramic, oxide ceramic, hybrid ceramic, composite and metal restorations to promote a strong bond to resin-based materials as resin-based cements in case of luting procedures or direct composites in case of repairs.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the G-Cem LinkForce, GC Dual Cure Activator, Try-In Paste, and Multi Primer devices, based on the provided text.

    Note: The provided document is a 510(k) premarket notification and primarily focuses on demonstrating substantial equivalence to predicate devices rather than reporting on a standalone clinical study for the new device. Therefore, much of the requested information regarding detailed clinical study design (sample size, ground truth, experts, MRMC studies, training set details) is not present. The studies mentioned are primarily in vitro performance bench tests against ISO standards.


    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily defined by ISO standards and internal company specifications. The document states that the applicant device conforms to these required specifications.

    Table of Acceptance Criteria and Reported Device Performance (Summary from Tables 5.2, 5.3, 5.4, 5.5)

    Device GroupPropertyStandard / Test MethodAcceptance Criteria (Requirements)Reported Device Performance (Implied)
    G-CEM LinkForceFilm thicknessISO 4049: 2009 5.2.2No greater than 50 µmMeets criteria (implied)
    Working timeISO 4049: 2009 5.2.4No detectable change in its homogeneityMeets criteria (implied)
    Setting timeISO 4049: 2009 5.2.6Not more than 10 minMeets criteria (implied)
    Depth of cure (Opaque shade)ISO 4049: 2009 5.2.8Not less than 0.5 mmMeets criteria (implied)
    Depth of cure (Other shade)ISO 4049: 2009 5.2.8Not less than 1.5 mmMeets criteria (implied)
    Flexural strengthISO 4049: 2009 5.2.9Greater than 50 MPaMeets criteria (implied)
    Water sorptionISO 4049: 2009 5.2.10Less than 40 µg/mmMeets criteria (implied)
    SolubilityISO 4049: 2009 5.2.10Less than 7.5 µg/mmMeets criteria (implied)
    RadiopacityISO 4049: 2009 5.5Greater than the same thickness of aluminumMeets criteria (implied)
    GC DUAL CURE ACTIVATOR (with G-Premio BOND)Adhesion to Bovine DentinISO 11405: 2015 5.1.4 (Shear bond strength)Greater than 5MPaMeets criteria (implied)
    Adhesion to Bovine EnamelISO 11405: 2015 5.1.4 (Shear bond strength)Greater than 9MPaMeets criteria (implied)
    TRY-IN PASTEFilm thicknessCompany SpecificationNo greater than 50 µmMeets criteria (implied)
    MULTI PRIMERTensile bond strength to PorcelainCompany SpecificationGreater than 10MPaMeets criteria (implied)
    Tensile bond strength to ZirconiaCompany SpecificationGreater than 10MPaMeets criteria (implied)
    Tensile bond strength to Au-Ag-Pd alloyCompany SpecificationGreater than 10MPaMeets criteria (implied)

    Biocompatibility for all devices (G-CEM LinkForce, GC DUAL CURE ACTIVATOR, TRY-IN PASTE, MULTI PRIMER):

    • Cytotoxicity (ISO 10993-5): No cytotoxic effect.
    • Sensitization (ISO 10993-10): 0% sensitization rate, non-irritant to buccal tissues.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes performance bench tests according to ISO standards, not a clinical test set with human subjects.

    • Sample Size: Not explicitly stated for each specific test, but ISO standards typically involve a defined number of specimens for mechanical and physical property testing.
    • Data Provenance: In-vitro laboratory testing, likely conducted by the manufacturer, GC America Inc. Country of origin not specified, but the submission is to the US FDA. The data is retrospective in the sense that it's reported on already conducted tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The ground truth for performance bench tests is the objective measurement of physical and mechanical properties against pre-defined engineering and material standards (ISO standards, company specifications). There's no "expert ground truth consensus" involved in this type of testing.


    4. Adjudication Method for the Test Set

    Not applicable. This type of performance testing involves direct physical measurements, not interpretation or adjudication by experts.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This submission is for dental cements and primers, which are materials, not AI-powered diagnostic or assistive devices that would involve human readers or AI assistance.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. These are physical dental materials, not algorithms.


    7. The Type of Ground Truth Used

    The ground truth used for these performance bench tests are:

    • Established ISO standards for dental materials (ISO 4049:2009, ISO 11405:2015, ISO 10993-5, ISO 10993-10). These standards define the methodology and acceptable ranges for various physical, mechanical, and biological properties.
    • Company Specifications for properties like film thickness (Try-in Paste) and tensile bond strength to various restorative materials (Multi Primer).

    8. The Sample Size for the Training Set

    Not applicable. These are dental materials tested against standards, not AI algorithms requiring a training set.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI algorithm.