Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K130740

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-07-23

    (126 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K001775

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Zoe Medical Nightingale Monitoring System is indicated for use in adult & pediatric patient populations.

    The Zoe Medical Nightingale Monitoring System facilitates the monitoring of:

    • ECG -
    • Impedance respiration -
    • Non-Invasive blood pressure -
    • Invasive blood pressure -
    • Body temperature "
    • Functional arterial oxygen saturation (SpO2) -
    • -End-tidal & inspired CO2

    The Zoe Medical Nightingale Monitoring System is a prescription device intended to be used by healthcare professionals in all areas of a healthcare facility.

    Device Description

    The Zoe Medical Nightingale Monitoring System (NMS) facilitates the monitoring of patient physiological parameters both at the bedside on the Nightingale PPM3 monitor and remotely on the Nightingale MPC central station. The physiological parameters monitored by the NMS include: ECG. impedance respiration, non-invasive blood pressure, invasive blood pressure, body temperature, functional arterial oxygen saturation (SpO2), and end-tidal & inspired CO2 (capnography). The Nightingale Monitoring System as described in this submission is a new device, and will be marketed as the next generation product superseding the original and previously cleared Zoe Medical NMS (K001775).

    The part numbers for the four Nightingale PPM3 variants and their associated configurations are given below.

    179-0003 NIGHTINGALE PPM3 (with NIBP, SPO2, ECG, and Temp)

    179-1000 NIGHTINGALE PPM3 WITH IBP (with NIBP, SPO2, ECG, Temp, & IBP)

    179-1001 NIGHTINGALE PPM3 WITH ETCO2 (with NIBP, SPO2, ECG, Temp, & Microstream ETCO2)

    179-1002 NIGHTINGALE PPM3 (with NIBP, SPO2, ECG, Temp, IBP, & Microstream ETCO2)

    The Nightingale MPC central station (171-0001) will be sold as an option available to all customers based on their needs.

    AI/ML Overview

    The provided document describes the Zoe Medical Nightingale Monitoring System, a patient physiological monitor. Here's an analysis of its acceptance criteria and the studies performed:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission focuses on establishing substantial equivalence to a predicate device (Spacelabs Healthcare Vital Signs Monitoring System, K090556) rather than detailing specific numerical acceptance criteria for each physiological parameter. The acceptance criteria are broadly defined by compliance with applicable standards and demonstration that the device's performance meets its predetermined specifications.

    CategoryAcceptance Criteria (Implied)Reported Device Performance and Compliance
    SoftwareDesigned and developed according to Zoe Medical processes; verified and validated.Software was designed and developed in accordance with Zoe Medical software development processes, and was verified and validated. Test results indicated that the Nightingale Monitoring System complies with its predetermined specification.
    Electrical SafetyCompliance with specified electrical safety standards (e.g., AAMI ES 60601-1:2005, IEC 60601-2-XX series).Tested for patient safety in accordance with AAMI ES 60601-1:2005, ANSI/AAMI SP10:2002 / A1:2003, IEC 60601-1-4:2000, IEC 60601-1-8:2006, IEC 60601-2-27:2011, IEC 80601-2-30:2009, IEC 60601-2-34:2011, ISO 80601-2-55:2011, ISO 80601-2-56:2009, ISO 80601-2-61:2011, IEC 62366:2007, IEC 62304:2006. Test results indicated compliance with predetermined specification.
    Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2:2007.Tested for EMC in accordance with IEC 60601-1-2:2007. Test results indicated that the Nightingale Monitoring System complies with its predetermined specification.
    Bench PerformanceCompliance with Zoe Medical internal requirements and procedures; functional, environmental, and shipping/transportation performance.Tested in accordance with Zoe Medical internal requirements and procedures. Test results indicated that the device complies with the predetermined requirements, including performance, functional, environmental, and shipping/transportation testing.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a separate "test set" in the context of device performance data beyond the compliance testing. The performance and safety validations were primarily based on engineering and laboratory testing against established standards and internal requirements. Therefore:

    • Sample Size: Not applicable in terms of a clinical dataset for a test set. The samples would be the physical devices and components undergoing the various safety and performance tests.
    • Data Provenance: The data comes from internal testing conducted by Zoe Medical, Inc., as part of their design and development processes. It is retrospective in the sense that the testing was completed before submission for clearance. The country of origin of this internal testing data is likely the USA (Topsfield, MA, where the company is located).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. The device's performance is assessed against recognized industry standards and internal specifications, not against a "ground truth" established by experts in the typical sense of a clinical diagnostic study. The "ground truth" for electrical safety would be the requirements of the IEC 60601 series, and for EMC, the IEC 60601-1-2 standard. Internal engineering experts and certified testing laboratories would be involved in verifying compliance.

    4. Adjudication Method for the Test Set

    This is not applicable as there is no mention of a "test set" in the context of human interpretation or clinical data requiring adjudication. Compliance with standards is typically a pass/fail outcome based on measurements and observations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The submission relies on demonstrating substantial equivalence to a predicate device and compliance with relevant safety and performance standards for a physiological monitor, not on assessing the improvement of human readers with this AI-assisted technology. This device is a monitor, not an AI diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable. The Zoe Medical Nightingale Monitoring System is a physiological patient monitor, a hardware device with embedded software. Its performance is inherent to the device itself and its sensors, not a standalone algorithm in the AI sense. Its intended use involves healthcare professionals continuously monitoring the patient via the device.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance and safety testing is based on:

    • Pre-determined specifications: Internal requirements established by Zoe Medical.
    • International and National Standards: Compliance with recognized standards such as AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-2-XX series, ISO 80601-2-XX series, IEC 62366, and IEC 62304. These standards define acceptable levels of safety and performance for physiological monitors.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physiological monitor, not an AI/ML algorithm that requires a "training set" in the machine learning context. Its functionality is based on established engineering principles and signal processing, not on learning from a large dataset.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable, as there is no "training set."

    Ask a Question

    Ask a specific question about this device

    K Number
    K040012

    Validate with FDA (Live)

    Device Name
    WELL@HOME SYSTEM
    Manufacturer
    Date Cleared
    2004-04-02

    (88 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K952979,K001775,K024194

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Well@Home System is indicated for use for patients who need home-based health care from professional clinicians. The system provides benefits for patients who need periodic remote monitoring of their vital signs (such as noninvasive blood pressure, oxygen saturation, pulse rate, temperature, respiration rate, weight, blood glucose level, heart rate, or ECG) or self-reported symptoms, who need reminders to take prescribed medications, or who need training about how to manage their illness.

    Device Description

    The Well@Home System consists of the home-based "Well@Home Monitor" and the office-based "Well@Home Clinician Application". Using the Well@Home System, clinicians can monitor theised at home or in another remote health care facility. The use model is "daily checkups", rather them continuous monitoring.

    The Well@Home Monitor collects vital signs and symptom information on a periodic basis via a patientfriendly user interface. The user interface guides the patient through a schedule of activities using voice promots, instructional diagrams and large, easy-to-read buttons. The scheduled of doutinities can'nclude reminders to take medications or to take vital signs measurements. The schedule for these activities can be customized by the clinician according to the patient's medical condition. The user interface also provides opportunities for the patient to report symbons at any time, and to review training ontent related to managing the patient's illness.

    The Well@Home monitor measures noninvasive blood pressure, oxygen saturation, pulse rate, temperation | President rate, heart rate, and ECG. The Well@Home Monitor is designed to interface to Other Medical Devices (OMD's) for the purpose of gathering physiological data from those devices. The interface is through the monitor's three OMD serial ports. These serial ports are electrically isoated from each other and from the monitor's own physiological front-end. The list of supported OMD's is as follows:

    • . Fairbanks Digital Scale (private labeled and device listed as a Zoe Medical accessory)
    • Lifescan One Touch Ultra Blood Glucose Meter (or compatible BGM from Lifescan) .

    The OMD interface mechanism is designed to be expandable. Future OMD's may include a Spirometer or a PT/INR measurement device.

    The Well@Home Monitor transmits the gathered information to a clinician over the patient's existing telephone line. This information is received and stored by the Well(@Home Clinican Application.

    The Well@Home Clinician Application is essentially a patient data management and record keeping program that includes an interface to the Well@Home Monitor. The Well@Home Clinician Application stores information that is gathered from multiple Well@Home Monitors and allows a clinician to review a given patient's information. It also allows the clinician to make adjustments to a patient's schedule of activities as needed. These adjustments are sent back to the patient's Well@Home Monitor over the same telephone link.

    The Well@Home Clinician Application software was developed by Patient Care Technologies (PtCT) of Atlanta, Georgia, and is designed to run on standard Windows-based PC's. These PC's are typically installed at the office of a home care agency, and are referred to in this 510(k) as "Agency Servers". The Agency Servers may run other information management software developed by PtCT that is not part of the Well@Home System. The Well@Home System is product-branded to tie into PtCT's information management software product line.

    Because of the remote-monitoring and spot check nature of the Well@Home System, there are no realtime "alarms" as in a hospital-style monitor. Rather, the clinician defines "alerts" that are to be generated (e.g., when a vital sign parameter exceeds set limits or when a particular symptom is reported). These alerts are part of the information package that is sent from the Well@Home Monitor to the Well@Home Clinician Application. The alerts are then visually highlighted when the clinician reviews the patient's information.

    AI/ML Overview

    The provided text describes the Zoe Medical Well@Home System, which is a noninvasive blood pressure measurement system for remote patient monitoring. However, it does not include detailed acceptance criteria or a specific study proving the device meets those criteria in the way a modern medical device submission typically would.

    Instead, the document focuses on demonstrating substantial equivalence to existing predicate devices through:

    1. Technological Characteristics Comparison: Showing that the Well@Home System's design, functions, and principles of operation are similar to previously cleared devices.
    2. Performance Testing: Stating that internal engineering tests were conducted to verify performance against functional requirements and to show substantial equivalence.

    Given this, I will extract the information available and note where specific details (like numerical acceptance criteria, sample sizes for test sets, expert qualifications, etc.) are absent.


    Acceptance Criteria and Device Performance (as inferred from the document)

    Acceptance Criterion (Inferred)Reported Device Performance
    Overall System Functionality (compared to HANC Network)The Well@Home System performed according to its functional requirements and passed all internal engineering tests, verifying performance in all functions similar to the HANC Network (scheduling activities, providing information, collecting vital sign/symptom data, transferring data, modifying schedules). The differences are described as "minor" and do not diminish the claim of substantial equivalence.
    Vital Signs Measurement (compared to Nightingale PPM)The Well@Home Monitor's performance for noninvasive blood pressure, oxygen saturation, pulse rate, temperature, respiration rate, heart rate, and ECG was shown to be "substantially equivalent" to the Nightingale PPM based on bench tests using simulators. This equivalence is attributed to using the same signal processing front-end hardware and software algorithms as the cleared Nightingale PPM.
    Blood Glucose Measurement (compared to Lifescan One Touch Ultra)Internal engineering tests confirmed that the measurement results provided by the Lifescan One Touch Ultra Blood Glucose Meter (BGM) were correctly collected and displayed by the Well@Home System when interfaced. The BGM's performance is stated to be "no different" whether connected to Well@Home or functioning as a standalone device.

    Detailed Study Information (Based on Available Text):

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Sample Size: Not specified. The document states "internal engineering tests" and "bench tests using simulators" were conducted, but no numerical sample sizes for these tests are provided for either human subjects or simulated data points.
      • Data Provenance: Not specified. The tests were "internal engineering tests," implying they were conducted by the manufacturer, Zoe Medical, but details on location or whether they involved human subjects are absent.
      • Retrospective or Prospective: Not specified. The nature of "internal engineering tests" and "bench tests using simulators" suggests they were likely prospective tests conducted on the device hardware and software.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable/Not specified. The "ground truth" for the vital signs measurements appears to be established by the predicate devices themselves (Nightingale PPM and One Touch Ultra) and, for overall system functionality, by functional requirements. There is no mention of independent experts establishing ground truth for a test set in the context of clinical accuracy or interpretation that would require such expertise. "Bench tests using simulators" imply comparison against known simulator outputs.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable/Not specified. The document does not describe a clinical study design that would involve expert adjudication of results. The testing described is primarily technical and functional equivalence vs. predicate devices or simulators.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is not an AI-assisted diagnostic tool for human readers. It is a remote monitoring system that collects and transmits vital signs and symptom data. Therefore, an MRMC study or AI-assistance effect size is not relevant to this submission.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, implicitly. The "internal engineering tests" and "bench tests using simulators" for vital signs measurement evaluate the device's capability to measure data and display it correctly. For the blood glucose meter, the BGM's performance is explicitly stated as "no different when it is connected to the Well@Home Monitor as when it functions as a standalone device," indicating a standalone performance assessment of the BGM component. The Well@Home System functions as an "information transfer technology" for the BGM.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • For vital signs measurement (Nightingale PPM comparison): The ground truth was presumably the known accurate outputs from physiological simulators or the validated measurements from the predicate Nightingale PPM itself.
      • For blood glucose measurement (One Touch Ultra comparison): The ground truth was the measurements provided by the standalone, predicate Lifescan One Touch Ultra Blood Glucose Meter.
      • For overall system functionality (HANC Network comparison): The ground truth was likely defined by functional requirements derived from the predicate HANC Network's capabilities.
    7. The sample size for the training set:

      • Not applicable/Not specified. This document is for a medical device submission focused on substantial equivalence through hardware and software design, and functional testing. It does not describe a machine learning or AI model that would require a "training set" in the conventional sense. The software algorithms for vital signs measurement are stated to be "the same software algorithms" as the predicate Nightingale PPM, implying they were developed and validated previously, not newly trained for this specific application.
    8. How the ground truth for the training set was established:

      • Not applicable. As there is no described training set for an AI/ML model, this question is not relevant to the provided information.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1