Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DORA Disposable A.V. Fistula Needle Sets (Safety Needle Series) are single-use sterile medical devices intended to be used as vein puncture for hemodialysis treatment. Protective Shield aids in the prevention of accidental needlesticks.

    The DORA Disposable A.V. Fistula Needle Sets are single-use sterile medical devices intended to be used as vein puncture for hemodialysis treatment.

    Device Description

    The proposed device, DORA Disposable A.V. Fistula Needle Sets, is a non-implantable blood access device, which mainly consists of an adaptor, flexible tube and sharp needle. It is available in two types, 1) Needle sets with safety feature and 2) Needle sets without safety feature. Both the two types of proposed device are provided sterile and are for single use only.

    The two types of proposed devices are offered in difference configurations with options that include needle gauge, needle length, wing types (fixed wing or rotatable wing).

    The proposed device and its package are designed to be provided in Gamma sterilization. The package could maintain the sterility of the device for three years.

    The Protective Shield for injury prevention requires physical action by the clinician to activate and is designed to cover the cannula after treatment. Correct use of this safety feature will eliminate accidental needlestick injuries.

    The proposed device is used in conjunction with following hemodialysis blood tubing sets during clincial use.

    AI/ML Overview

    The provided document, a 510(k) Summary for the DORA Disposable A.V. Fistula Needle Sets, details the premarket notification of intent to market the device and its substantial equivalence to a legally marketed predicate device. This type of submission focuses on demonstrating that a new device is as safe and effective as an already approved device, rather than proving novel effectiveness through extensive clinical trials.

    As such, the document does not contain the information requested regarding acceptance criteria and a study proving a device meets those criteria in the context of:

    • A table of acceptance criteria and reported device performance for an AI/ML or diagnostic device. This document is for a physical medical device (AV fistula needle sets).
    • Sample sizes for a test set (for AI/ML validation). The testing described is for physical characteristics and biocompatibility, not for a 'test set' in the AI/ML sense.
    • Data provenance for a test set (for AI/ML validation). Not applicable to a physical medical device.
    • Number of experts and their qualifications for establishing ground truth. Not applicable.
    • Adjudication method for a test set. Not applicable.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness study. Not applicable.
    • Standalone (algorithm only) performance. Not applicable.
    • Type of ground truth used (expert consensus, pathology, outcomes data). Not applicable.
    • Sample size for training set. Not applicable.
    • How ground truth for training set was established. Not applicable.

    Instead, the document focuses on non-clinical testing and substantial equivalence to a predicate device.

    Here's what the document does provide regarding acceptance criteria and performance:

    1. Acceptance Criteria (via Standards Compliance and Performance Testing) and Reported Device Performance:

    The acceptance criteria for this device are demonstrated through adherence to various international and national standards, as well as the successful completion of specific non-clinical performance tests. The reported performance is that the device "met all design specifications" and "is similar as that of the predicate device" or "acceptable."

    Acceptance Criteria CategorySpecific Criterion/StandardReported Device Performance
    Material/Component StandardsISO/FDIS 9626:2016 (Stainless steel needle tubing)Complies with ISO/FDIS 9626:2016 (Needle Performance)
    Packaging IntegrityASTM F88/F88M - 15 (Seal Strength)Complies
    ASTM F1929 – 15 (Seal Leaks by Dye Penetration)Complies
    Biological Safety (Biocompatibility)USP <85> (Bacterial Endotoxin Limit)Complies
    ISO 10993-3:2014 (Genotoxicity, carcinogenicity, reproductive toxicity)Complies with ISO 10993 series standards; tests performed (Cytotoxicity;Sensitization; Irritation sensitivity; Systemic toxicity;Pyrogen; Hemolysis; Thromboresistance; Partial thromboplastin time;Complement activation; Bacterial reverse mutation;Mammalian chromosome aberration; Mouse bone marrow micronucleus;Activated clotting time of whole blood; Platelet adhesion;Muscle implantation; Subchronic systemic toxicity)
    ISO 10993-4:2002 A1:2006 (Interactions with blood)Complies
    ISO 10993-5:2009 (Vitro Cytotoxicity)Complies
    ISO 10993-6:2007 (Local effects after implantation)Complies
    ISO 10993-10:2010 (Irritation and skin sensitization)Complies
    ISO 10993-11:2006 (Systemic toxicity)Complies
    ASTM F756-13 (Hemolytic Properties)Complies
    Functional/Mechanical PerformanceISO 594-1:1986 (Luer Taper - General)Complies (Female Conical Fitting Testing)
    ISO 594-2:1998 (Luer Taper - Lock Fittings)Complies (Female Conical Fitting Testing)
    ISO 10555-1:2013 (Intravascular catheters – General requirements)Complies
    Shipping/Packaging PerformanceASTM D4169-16 (Performance Testing Of Shipping Containers)Complies
    Specific Device PerformanceSimulated Use Needlestick Prevention TestingTest results demonstrated similarity to predicate device
    Needle Performance TestingTest results demonstrated similarity to predicate device
    Mechanical Testing (Wing torque, final lock, needle pushback, mechanical hemolysis)Test results demonstrated similarity to predicate device
    Tensile Strength Testing (Tube to wing pull, tube to joint, needle to cover pull, cannula to hub)Test results demonstrated similarity to predicate device
    Tubing Kinking TestTest results demonstrated similarity to predicate device
    Leakage Testing (Liquid leakage, air leakage)Test results demonstrated similarity to predicate device
    Clamp Stop TestingTest results demonstrated similarity to predicate device
    Flow Rate Testing (Also listed as an AV Fistula Needle Set Performance Test)Test results demonstrated similarity to predicate device
    Particulate Contamination Testing (Also listed as an AV Fistula Needle Set Performance Test)Test results demonstrated similarity to predicate device
    Sterility AssuranceSAL: 10-6 (Sterility Assurance Level) via Gamma SterilizationAchieved 10-6 SAL

    2. Sample sizes used for the test set and the data provenance:

    • Sample Sizes: Not explicitly stated with specific numbers (e.g., "n=X needles were tested"). The document generally states "Non clinical tests were conducted" and "The test results demonstrated...". For a 510(k) submission for a physical device, detailed sample sizes for each specific test are typically found in the full technical documentation, not summarized in the 510(k) summary as it is here.
    • Data Provenance: The tests were conducted internally by Bain Medical Equipment (Guangzhou) Co., Ltd. or by test labs they contracted. The document implicitly indicates the tests originate from China (Guangzhou) as that's the manufacturer's location. The tests are "non-clinical" and likely conducted in laboratory settings, not on patient data or retrospective/prospective human studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this is a physical medical device. "Ground truth" in this context refers to compliance with established engineering and biocompatibility standards, not expert human interpretation for a diagnostic or AI device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a MRMC study was not done. This device is a physical medical instrument, not an AI/ML diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No, not applicable.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    • For physical devices, the "ground truth" is adherence to predefined engineering specifications, internationally recognized standards (ISO, ASTM, USP), and established safety and performance benchmarks (e.g., sterilization effectiveness, material strength, biocompatibility profiles). It's a truth based on physical and chemical properties and their interaction, rather than clinical outcomes or diagnostic interpretations.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device that uses a "training set."

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K013634

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-07-09

    (246 days)

    Product Code
    Regulation Number
    876.5820
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K972206, K001465

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BIOTEQ® 3 in 1 Hemodialysis Blood Tubing Packs ( BT-102DA · BT-102DB · BT-102DC · BT-102DD) are sets of single-use disposable components intended to provide extracorporeal access to blood of patients suffering from end stage renal disease during Hemodialysis therapy. Each set will include an arterial/venous bloodline set, and any combination of two or three component of the following items: An intravenous administration set for (use in administration of intravenous fluids to a dialysis set in conjunction with hemodialysis procedures.) , AV fistula needle set to (apply on the access site of patient vessel to draw the blood flow adequate to pass through the dialyer and reinfusion of dialysed blood back to patient vessel via the A.V. Fistula Needle Set during hemodialysis procedures.) , and/or A transducer protector for (use as protective devices for pressure monitors and to help protect the sterility of the fluid pathway).

    Device Description

    The BIOTEQ® 3 in 1 Hemodialysis Blood Tubing Packs (BT-102DA 、BT-102DB、BT-102DC、BT-102DD) are sets of single-use disposable components intended to provide extracorporeal access to blood of patients suffering from end stage renal disease during Hemodialysis therapy. Each set will include an arterial/venous bloodline set, and any combination of two or three component of the following items: An intravenous administration set for use in administration of intravenous fluids to a dialysis set in conjunction with hemodialysis procedures. , AV fistula needle set to apply on the access site of patient vessel to draw the blood flow adequate to pass through the dialyer and reinfusion of dialysed blood back to patient vessel via the A.V. Fistula Needle Set during hemodialysis procedures. , and/or A transducer protector for use as protective devices for pressure monitors and to help protect the sterility of the fluid pathway. BIOTEQ® 3 in 1 Hemodialysis Blood Tubing Packs includes 4 models (BT-102DA 、BT-102DB、BT-102DC、BT-102DD)

    AI/ML Overview

    This document describes a 510(k) premarket notification for a medical device and, as such, focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with detailed acceptance criteria and reported device performance in the way requested for AI/diagnostic devices. The information provided heavily emphasizes physical, chemical, biological, and sterilization specifications rather than clinical performance metrics often associated with "device performance" in the context of diagnostic or AI-driven tools.

    Therefore, the requested table of acceptance criteria and reported device performance, and several other points (2, 3, 4, 5, 6, 7, 8, 9), cannot be directly extracted or are not applicable from the provided text in the way they would be for a typical AI/diagnostic device study. The text primarily focuses on conformity to standards and substantial equivalence.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Applicable Standards/Specifications)Reported Device Performance (as stated in the document)
    Physical SpecificationConforms to applicable standards
    Chemical SpecificationConforms to applicable standards
    Biological SpecificationConforms to applicable standards (including ISO 10993 series, USP Pyrogenic standards)
    Sterilization SpecificationConforms to applicable standards (including ISO 11607-1, ISO 11135)
    Substantial Equivalence to Predicate DeviceBench testing demonstrates that differences in technological characteristics do not raise new questions of safety or effectiveness, making it substantially equivalent to Nipro Blood Tubing Set for Hemodialysis with Transducer Protectors and Priming set (K972206) & (K001465).

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
    The document mentions "bench testing," but it does not specify the sample size used for these tests. Data provenance (country of origin, retrospective/prospective) is not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. This device is a blood tubing pack, and its evaluation relies on engineering, material, and sterility testing against established standards, not expert interpretation of outputs like an AI diagnostic tool.

    4. Adjudication method for the test set
    Not applicable. The evaluation is based on conformance to engineering and safety standards, not a diagnostic interpretation process requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is not an AI or diagnostic device that involves human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
    Not applicable, as this is not an AI algorithm. The performance evaluation focuses on the physical and biological characteristics of the blood tubing pack itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    The "ground truth" for this device is based on established international and national standards for medical device safety, material compatibility, and sterilization (e.g., ISO 10993 series for biological evaluation, ISO 11607-1 for packaging, ISO 11135 for sterilization, USP Pyrogenic standards).

    8. The sample size for the training set
    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established
    Not applicable. This is not an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1