510(k) Data Aggregation

    K Number
    K033416

    Validate with FDA (Live)

    Date Cleared
    2003-11-05

    (9 days)

    Product Code
    Regulation Number
    870.4210
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K840002, K000776, K013013, K010737, K891687

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These cannulae are intended for use in perfusion of the ascending aorta during short term (6 hours or less) cardiopulmonary bypass.

    Device Description

    The Select 3D™ and Select CAP™ Arterial Cannulae both have clear flexible, thin w.ill wire-wound PVC bodies with angled, beveled tips. The proximal end of each cannula includes a 3/8" (0.95 cm) vented or non-vented connector with a peel cap. The vented connector allows air to be vented from the cannula before connection to the perfusion line.

    The Select 3D™ Arterial Cannula tip has three integrated flutes which help diffuse and disperse blood flow. The cannula body features a tip orientation line indicate direction of the cannula tip during cannulation. Overall cannula length is 11.5" (29.2 cm). It is available in 22 and 24 Fr sizes.

    The Select CAP™ Arterial Cannula tips have an integrated pressure monitoring port. The pressure monitoring port on the tip of the cannulae provides the real time capability of accurately measuring central arterial pressure within the aorta. Overall cannula length is 12" (30.5 cm). It is available in 18, 20, 22, and 24 Fr. sizes.

    The devices may include a Carmeda® BioActive Surface.

    AI/ML Overview

    This submission describes device modifications, specifically adding a Carmeda® BioActive Surface coating to existing Select 3D™ and Select CAP™ Arterial Cannulae. Therefore, the "study" is focused on demonstrating that the coated devices maintain the performance characteristics of the uncoated predicate devices and that the coating itself is safe and effective.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategoryReported Device Performance (Summary)
    Functional PerformancePerformance characteristics of the materials (after Carmeda® coating) were established through in vitro visual and functional testing.
    Coating Integrity/CoverageCoverage, bioactivity, and leach testing were performed on Carmeda® coated devices. (Specific quantitative acceptance criteria are not provided in the document).
    Substantial EquivalenceDemonstrated through design, test results, and indications for use, confirming equivalence to predicate devices.

    Missing Information: The document does not explicitly present a table of quantitative acceptance criteria (e.g., minimum flow rates, burst pressures, specific leach limits) that were met. Instead, it refers to "functional testing," "coverage, bioactivity, and leach testing" as having been performed and the results supporting substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: Not explicitly stated. The description mentions "in vitro visual and functional testing" and "coverage, bioactivity, and leach testing" performed on "Carmeda® coated devices," implying a sample of such devices, but the exact number is not provided.
    • Data Provenance: The testing was "in vitro," meaning laboratory-based testing. There is no mention of country of origin for the data or whether it was retrospective or prospective in the context of human subjects.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This is not applicable as the study did not involve human subjects or expert assessment for ground truth establishment. The testing described is entirely in vitro (laboratory-based chemical and physical performance testing).

    4. Adjudication Method for the Test Set:

    This is not applicable as the study did not involve human subjects, diagnostic assessments, or a need for adjudication among experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. This submission is for a medical device (arterial cannulae), not an AI/imaging diagnostic device. Therefore, no MRMC study, human readers, or AI assistance is relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable as this is not an AI/algorithm-based device.

    7. The Type of Ground Truth Used:

    The "ground truth" for this device's performance is based on:

    • In vitro measurements and observations: This includes visual inspections, functional tests (e.g., flow characteristics, mechanical integrity), and chemical analyses (e.g., coverage, bioactivity, leach testing) conducted in a laboratory setting.
    • Comparison to predicate devices: The "ground truth" for performance is effectively defined by the established performance characteristics of the prior, uncoated, predicate devices, which the new coated devices must match or exceed.

    8. The Sample Size for the Training Set:

    This is not applicable as this is not an AI/algorithm-based device, so there is no "training set."

    9. How the Ground Truth for the Training Set was Established:

    This is not applicable as this is not an AI/algorithm-based device, so there is no "training set" or ground truth establishment for it.