510(k) Data Aggregation

    K Number
    K190051

    Validate with FDA (Live)

    Device Name
    TD Clip
    Date Cleared
    2019-06-07

    (147 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    DEN170090, K103704, K111066

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TD Clip is intended to be used with MATRx or MATRx plus Titration Trays to form a mandibular repositioning device intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in patients 18 years and older. The device is intended to be fitted with assistance from a healthcare professional and used during sleep for a total of less than 30 nights.

    Device Description

    The TD Clip is a temporary oral appliance. The patient uses the temporary oral appliance to alleviate mild to moderate obstructive sleep apnea and/or snoring while waiting for a custom oral appliance to be manufactured. The dentist fabricates temporary titration trays consisting of upper and lower dental trays. The trays are custom fit to the patient using polyvinyl siloxane impression material. The healthcare provider may use the clip assembly to fasten the titration trays at the target therapeutic position to form the TD Clip temporary oral appliance. Patients may use the temporary oral appliance as treatment while waiting for their custom oral appliance. Adjustment of the position requires the setting of a new TD Clip by the healthcare professional.

    AI/ML Overview

    The acceptance criteria and study proving the device meets them are described below for the TD Clip.

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance Criteria (Implied)Reported Device Performance
    Intraoral Material PerformanceNo presence or increased risk of breakage that could lead to unacceptable harm over the maximal study length (30 nights of use).Evaluated under simulated aging, including environmental conditions and representative worst-case loading schemes. The device met the implied acceptance criteria. (Specific quantitative results or detailed findings are not provided in this summary, but the conclusion is that the device is safe and effective).
    TD Clip PerformanceMaintained the set protrusive positioning over 30 nights of use.Evaluated under repeated use and exposed to normal sleep forces over 30 nights of use, including static and cyclic loading under worst-case use. The device met the implied acceptance criteria. (Specific quantitative results or detailed findings are not provided in this summary, but the conclusion is that the device is safe and effective).
    Tray Removal Simulation100% of users were able to remove the Titration Trays from their mouth quickly and without causing injury to themselves or the Titration Trays.Evaluated to ensure users could remove the trays quickly and without injury. The device met the implied acceptance criteria. (Specific quantitative results or detailed findings are not provided in this summary, but the conclusion is that the device is safe and effective).
    Biocompatibility TestingMaterials in contact with mucosal membrane adhere to ISO 10993-1 and associated FDA guidance.All materials in contact with the mucosal membrane were assessed and tested in accordance with ISO 10993-1 and the associated FDA guidance. The device met the implied acceptance criteria.
    Risk ManagementNo unacceptable risks identified.During risk management and performance testing, no new risks were identified. The risk management concluded that the TD Clip had no unacceptable risks.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not specified for any of the performance tests. The document only mentions "mechanical and environmental use conditions" and "repeated use and exposed to normal sleep forces over 30 nights of use".
    • Data Provenance: The tests described are "Bench testing" and "simulated mechanical and environmental use conditions," indicating these were laboratory/engineering tests rather than human clinical trials. Thus, there is no country of origin or retrospective/prospective distinction for data provenance in the traditional sense; the data is from an engineering test setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    • Not applicable. The ground truth for these performance tests would be based on engineering specifications and direct measurements against those specifications, rather than expert interpretation of medical data. The evaluation was for mechanical and material performance.

    4. Adjudication Method for the Test Set:

    • Not applicable. Adjudication methods like 2+1 or 3+1 are used for clinical studies with human evaluators to establish ground truth for ambiguous cases. These were bench tests against predetermined engineering criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size:

    • No MRMC comparative effectiveness study was done. Clinical testing was explicitly stated as "not performed as the device has no technological differences from the predicate."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. The TD Clip is a physical medical device (an intraoral appliance) and not an algorithm or AI software. Therefore, the concept of "standalone performance" for an algorithm doesn't apply here. The performance tests focused on the physical integrity and function of the device itself.

    7. The Type of Ground Truth Used:

    • The ground truth for the performance testing was based on engineering specifications and predefined performance limits (e.g., no breakage, maintenance of protrusive positioning, quick and injury-free removal, adherence to biocompatibility standards). This is a form of objective, empirical measurement and adherence to established material and mechanical standards.

    8. The Sample Size for the Training Set:

    • Not applicable. As a physical medical device undergoing bench testing, there is no "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no training set.