Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K234063

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-09-13

    (266 days)

    Product Code
    Regulation Number
    866.3960
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    DEN140019

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    T2Candida® 1.1 Panel and T2Dx® Instrument is a qualitative T2 Magnetic Resonance (T2MR®) assy for the direct detection of Candida species in K₂EDTA human whole blood specimens from patients with symptoms of, or medical conditions predisposing the patient to, invasive fungal infections. The T2Candida 1.1 Panel identifies five species of Candida and categorizes them into the following three species groups:

      1. Candida albicans and/or Candida tropicalis
      1. Candida parapsilosis
    • Candida glabrata and/or Candida krusei 3.

    The T2Candida 1.1 Panel does not distinguish between C. albicans and C. tropicalis. The T2Candida 1.1 Panel does not distinguish between C. glabrata and C. krusei.

    The T2Candida 1.1 Panel is indicated for the presumptive diagnosis of candidemia. The T2Candida 1.1 Panel is performed independent of blood culture. Concomitant blood cultures are necessary to recover organisms for susceptibility testing or further identification.

    The T2Candida positive and negative External Controls (T2Candida QCheck Positive Kit and the T2Dx QCheck Negative Kit) are intended to be used as quality control samples with the T2Candida 1.1 Panel when run on the T2Dx Instrument. These controls are not intended for use with other assays or systems.

    Device Description

    The T2Candida 1.1 Panel and T2Dx Instrument is comprised of the T2Candida 1.1 Panel performed on the T2Dx Instrument. The T2Candida 1.1 Panel is a qualitative molecular diagnostic assay that employs a whole blood compatible PCR amplification followed by T2 magnetic resonance (T2MR) detection. The T2Candida 1.1 Panel is performed on the T2Dx Instrument which executes all steps after specimen loading. A K₂EDTA whole blood specimen is loaded onto the T2Candida 1.1 Sample Inlet, which is then placed on the T2Candida 1.1 Cartridge along with the T2Candida 1.1 Reagent Tray. The Reagent Tray contains the internal control, amplification reagent, enzyme and the probe-coupled superparamagnetic particles for each Candida target. Two milliliters of the blood specimen is transferred to the T2Dx Instrument where lysis of the red blood cells, concentration and lysis of the Candida cells and amplification of the Candida DNA takes place. Amplification products are detected by T2MR detection using species-specific probes which are attached to the superparamagnetic particles. The assay identifies Candida albicans and/or Candida tropicalis, Candida parapsilosis, and Candida glabrata and/or Candida krusei. The test does not distinguish between C. albicans and C. tropicalis. The test does not distinguish between C. glabrata and C. krusei

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the T2Candida 1.1 Panel, aimed at amending labeling to include pediatric patients. The information focuses on analytical and clinical performance to demonstrate substantial equivalence to a previously cleared device.

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    The document primarily focuses on demonstrating substantial equivalence by relying on previously obtained performance data and additional tests for pediatric populations and cross-reactivity. Explicit acceptance criteria for clinical performance are not directly stated in percentages (e.g., minimum sensitivity/specificity), but the summary indicates "acceptable performance" was demonstrated. The analytical acceptance criteria for cross-reactivity are defined for the new tests.

    Table of Acceptance Criteria and Reported Device Performance

    CategoryAcceptance Criteria (Implied/Stated)Reported Device Performance
    Clinical Performance (Pediatric)Acceptable performance for detecting Candida albicans, Candida parapsilosis, Candida glabrata, and Candida krusei infection in pediatric patients. (Implied: performance comparable to adult studies and sufficient for clinical utility).Sensitivity (PPA): Ranged from 50-100% in pediatric studies. Specificity (NPA): Ranged from 97-99% in pediatric studies. (Note: These ranges are from external peer-reviewed publications used to support the submission, and low prevalence of positive blood cultures (1.2%) was observed in these studies, which can impact PPA/NPA interpretations).
    Analytical Cross-ReactivityCross-reactivity defined as an increase in T2 signal above the established cutoff for the Candida detection channel when tested at clinically relevant concentrations, requiring both amplification with Candida primers and detection with capture probes. (Acceptance: No cross-reactivity at clinically relevant concentrations).Of 5 organisms tested at 10^6 CFU/mL, 2 (S. agalactiae, H. influenzae) showed some cross-reactivity initially. Retesting at "clinically relevant concentrations" (100-1000 CFU/mL): S. agalactiae: No cross-reactivity observed at 1000 CFU/mL, 100 CFU/mL, or 33 CFU/mL. H. influenzae: No cross-reactivity observed at 1000 CFU/mL or 100 CFU/mL. (One instance of 1/3 positive at 100 CFU/mL was observed but not deemed cross-reactive after additional replicates). N. meningitidis, S. mitis, L. monocytogenes: No cross-reactivity at 10^6 CFU/mL.
    Internal ControlInternal Control (IC) must be valid for the test to be considered acceptable. (Implicit: Pass rate for IC under various conditions).Valid for all cross-reactivity tests (3/3 or 6/6 depending on replicates).

    2. Sample sizes used for the test set and the data provenance:

    • Clinical Performance (Pediatric):
      • Sample Size: A total of 246 pediatric samples were evaluated across two peer-reviewed publications.
      • Data Provenance: The data came from existing studies (peer-reviewed publications) where the T2Candida 1.1 Panel was utilized. The document does not specify the country of origin, but generally, such studies supporting FDA submissions would often include data from the US or other regions with comparable clinical practices. The studies were retrospective in the sense that they were "existing studies" identified and utilized for this submission, although the original data collection within those studies might have been prospective.
    • Analytical Cross-Reactivity:
      • Sample Size:
        • Initial testing: 3 replicates per organism at 10^6 CFU/mL.
        • For organisms showing initial cross-reactivity: Additional 6 replicates (from two additional sample preparations) at 10^6 CFU/mL, and 3 replicates at lower concentrations (1000 CFU/mL, 100 CFU/mL, 33 CFU/mL).
      • Data Provenance: This appears to be prospective laboratory testing conducted specifically for this submission, as it's described as "Additional cross-reactivity testing was performed in this submission."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Clinical Performance (Pediatric): The ground truth for this segment of the study was primarily established by blood culture results. The document does not mention the use of experts (e.g., radiologists) for establishing this ground truth, as it's a molecular diagnostic device measuring specific microbial presence. Blood culture is a laboratory-based gold standard for candidemia diagnosis.
    • Analytical Cross-Reactivity: Ground truth for this was based on known spiked concentrations of the organisms and the inherent characteristics of the T2Candida 1.1 Panel's detection mechanism (T2 signal cutoff). No external human experts are mentioned for ground truth establishment here.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Clinical Performance (Pediatric): Not applicable. The ground truth was based on blood cultures.
    • Analytical Cross-Reactivity: A form of adjudication was applied for cross-reactivity. If an organism demonstrated cross-reactivity at 10^6 CFU/mL, it was "further evaluated with additional replicates from two additional sample preparations." Furthermore, the rule for confirming cross-reactivity was: "Results were not considered cross-reactive if only one replicate demonstrated cross-reactivity." This implies a majority rule or consistency requirement rather than a specific expert consensus adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a molecular diagnostic test for detecting Candida species directly from blood, not an imaging-based AI diagnostic. Therefore, a multi-reader multi-case study involving human readers and AI assistance is not relevant to this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, this is effectively a standalone device. The T2Candida 1.1 Panel (and T2Dx Instrument) runs the assay and provides results (positive/negative for specific Candida groups, or invalid). Its performance is evaluated intrinsically through its ability to detect the target organisms (clinical sensitivity/specificity) and avoid false positives/negatives (analytical cross-reactivity). While a clinician interprets the results, the device's diagnostic output itself (e.g., "Candida albicans/tropicalis detected") is generated by the algorithm/system without human intervention in the diagnostic process of reading the T2MR signals.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Clinical Performance (Pediatric): Primarily blood culture results.
    • Analytical Cross-Reactivity: Known concentrations of spiked organisms in blood, with the "ground truth" for cross-reactivity being the absence or presence of specific amplification and detection events as defined by the assay's cutoffs.

    8. The sample size for the training set:

    • The document does not explicitly mention a "training set" in the context of machine learning, as this is a molecular diagnostic device with a defined mechanism (T2MR technology, PCR amplification) rather than a machine learning algorithm that learns from data. Its "training" is inherent in its design and optimization during development, validated by analytical and clinical studies. No specific sample size for "training" is provided in the submission summary.

    9. How the ground truth for the training set was established:

    • As above, "training set" and its associated ground truth establishment methods (e.g., expert labels for images) are not applicable in the typical AI/ML sense for this device. The development process for such molecular diagnostics involves extensive analytical characterization (e.g., limit of detection, inclusivity, exclusivity, precision studies) to define performance parameters and establish expected results, which serves a similar function to providing "ground truth" for the device's operational parameters.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1